K061179 · C.R. Bard, Inc. · LJS · Aug 9, 2006 · General Hospital
Device Facts
Record ID
K061179
Device Name
POWERHICKMAN LONG-TERM INTRAVASCULAR CATHETER
Applicant
C.R. Bard, Inc.
Product Code
LJS · General Hospital
Decision Date
Aug 9, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5970
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The PowerHickman™ catheter is intended for short or long term access to the central venous system for intravenous therapy and blood sampling.
Device Story
PowerHickman™ is an open-ended, long-term central venous catheter; available in 8 Fr single lumen and 9.5 Fr dual lumen configurations. Constructed from purple-tinted ChronoFlex polyurethane; designed for differentiation from other power-injectable catheters. Used in clinical settings for central venous access; enables I.V. therapy, blood sampling, power injection of contrast media, and central venous pressure monitoring. Operated by healthcare professionals. Device provides reliable vascular access; facilitates diagnostic imaging via power injection and hemodynamic monitoring; benefits patients requiring frequent or prolonged intravenous therapy.
Clinical Evidence
Bench testing only. No clinical data provided. Performance testing conducted per ISO 10555-1, ISO 10555-3, ISO 594-2, ASTM F640-79, and IEC 60601-2-34 to validate mechanical integrity, radiopacity, biocompatibility (ISO 10993), and pressure monitoring performance.
Technological Characteristics
Materials: ChronoFlex polyurethane (shaft and junction/bifurcation). Configuration: 8 Fr single lumen, 9.5 Fr dual lumen. Features: Purple colorant for identification, open-ended tip. Standards: ISO 10555-1/3, ISO 594-2, ASTM F640-79, IEC 60601-2-34, ANSI/AAMI BP22. Sterilization: Ethylene Oxide (AAMI/ANSI/ISO 11135).
Indications for Use
Indicated for short or long term access to the central venous system for administration of I.V. fluids, blood products, drugs, parenteral nutrition solutions, blood withdrawal, power injection of contrast media (max 5ml/sec), and central venous pressure monitoring (lumen 20 gauge or larger recommended).
Regulatory Classification
Identification
A percutaneous, implanted, long-term intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings, such as luer hubs, and accessories that facilitate the placement of the device. The device allows for repeated access to the vascular system for long-term use of 30 days or more, and it is intended for administration of fluids, medications, and nutrients; the sampling of blood; and monitoring blood pressure and temperature. The device may be constructed of metal, rubber, plastic, composite materials, or any combination of these materials and may be of single or multiple lumen design.
Special Controls
*Classification.* Class II (special controls) Guidance Document: “Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters.”
K061289 — ARROW PRESSURE INJECTABLE PICC · Arrow International, Inc. · Aug 24, 2006
Submission Summary (Full Text)
{0}------------------------------------------------
K061179
PowerHickman™ Catheter Traditional 510(k)
### Section 5 PowerHickman Central Venous Catheter 510(k) Summary of Safety and Effectiveness 21 CFR 807.92(a)
#### 5.1 General Information
| Submitter Name: | Bard Access Systems, Inc. (BAS)<br>[Wholly owned Subsidiary of C. R. Bard, Inc.] |
|----------------------------------|----------------------------------------------------------------------------------|
| Address: | 5425 West Amelia Earhart Drive<br>Salt Lake City, Utah 84116 |
| Telephone Number: | (801) 595-0700 ext. 5541 |
| Fax Number: | (801) 595-5425 |
| Contact Person: | Michaela Rivkowich |
| Date of Preparation: | June 30, 2006 |
| Registration Number: | 1720496 |
| Additional Registration Numbers: | |
| C. R. Bard: | 2212754 |
#### 5.2 Subject Device Information
| Device Name: | PowerHickman™ Central Venous Catheter |
|-----------------------|------------------------------------------------------------------------------------------------------------|
| Trade Name: | PowerHickman™ |
| Common/Usual Name: | Central Venous Catheter |
| Classification Name: | Long Term Intravascular Catheter<br>21 CFR 880.5970, Class II<br>80 LJS - Long Term Intravascular Catheter |
| Classification Panel: | General Hospital |
#### 5.3 Predicate Device Information
| Device Name: | PowerLine™ Central Venous Catheter |
|-----------------------|---------------------------------------------------------------------------------------------------------------------------------|
| Trade Name: | PowerLine™ |
| Common/Usual Name: | Central Venous Catheter |
| Classification Name: | Long Term Intravascular Catheter<br>21 CFR 880.5970, Class II<br>80 LJS – Long Term Intravascular Catheter |
| Classification Panel: | General Hospital |
| 510(k) Clearance: | K050185, concurrence date May 26, 2005<br>K051417, concurrence date June 30, 2005<br>K051991, concurrence date October 20, 2005 |
#### 5.4 Intended Use
The PowerHickman™ catheter is intended for short or long term access to the central venous system for intravenous therapy and blood sampling.
#### ર્ડ Indications for Use
The PowerHickman™ catheter is indicated for short or long term access to the central venous system. The PowerHickman™ catheter is designed for the administration of I.V. fluids, blood products, drugs, parenteral nutrition solutions, as well as blood withdrawal, power injection of
Image /page/0/Picture/13 description: The image shows a sequence of numbers. The numbers are "000008". There are five zeros followed by the number eight.
AUG - 9 2006
{1}------------------------------------------------
PowerHickman™ Catheter Traditional 510(k)
contrast media and allows for central venous pressure monitoring. The maximum recommended infusion rate is 5ml/sec for power injection of contrast media. For central venous pressure monitoring, it is recommended that catheter lumen of 20 gauge or larger be used.
#### 5.6 Device Description
The PowerHickman™ catheters are open-ended, long-term central venous catheters. The catheters are made of polyurethane material to which a purple colorant has been added. Purple colorants were added to the catheter materials to provide the catheter with an appearance that allows the end user to differentiate Bard's power injectable catheters from other manufacturers' power injectable catheters. The PowerHickman™ catheters are available in 8 Fr single lumen and 9.5 Fr dual lumen catheter configurations. The catheters are packaged in intermediate and microintroducer kits that incorporate kit components designed to meet the needs of the respective placer.
#### 5.7 Technological Comparison to Predicate Device
The technological characteristics of the PowerHickman™ catheters are substantially equivalent to those of the predicate PowerLine™ catheters in terms of intended use, application, user population, basic design, performance and labeling.
#### ર. જ 510(k) Substantial Equivalence Decision Tree
New device is compared to Marketed Device? Yes.
Does the new device have the same indication statement and intended use as the predicate? Yes. The intended use and indications for use are a combination of the indications from the predicate PowerLine catheters 510(k) K050185, K051417 and K051991.
Does the new device have the same technological characteristics, e.g. design, materials, etc.? Not in all regards. The principles of operation and basic design are the same as the predicate PowerLine catheter. The main differences are new catheter sizes and the use of ChronoFlex polyurethane material for catheter shaft and junction/bifurcation.
### Could the new characteristics affect safety or effectiveness?
Yes. The design changes may affect safety or effectiveness of the device.
Do the new characteristics raise new types of safety or effectiveness questions? No. Safety and effectiveness questions are the same as for the predicate devices.
Do accepted scientific methods exist for assessing effects of the new characteristics? Yes. Testing was performed to assess the new characteristics and was based on the following applicable standards and FDA guidance document:
- Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term જુન Intravascular Catheters, dated 3/16/95
- ું. BS/EN/ISO 10555-1: 1997, Sterile, single-use intravascular catheters, Part 1. General requirements
- ISO 10555-1:1995, Sterile, single-use intravascular catheters, Part 1. General requirements, ું મું Amendment 1:1999
- 4 BS/EN/ISO 10555-3:1997, Sterile, single-use intravascular catheters, Part 3. Central venous catheters
Image /page/1/Picture/19 description: The image shows the number 000009 in a bold, sans-serif font. The numbers are black against a white background. The numbers are evenly spaced and aligned horizontally.
{2}------------------------------------------------
- ISO 594-2: 1998, Conical fittings with a 6% (Luer) taper for syringes, needles and certain な other medical equipment - Part 2: Lock Fittings
- ું. ASTM F640-79 (reapproved 2000), Standard Test Methods for Radiopacity of Plastics for Medical Use
- AAMI/ANSI/ISO 11135:1994, Medical Devices Validation and Routine Control of Ethylenc ું. Oxide Sterilization
- AAMI/ANSI/ISO 10993-1:2003, Biological Evaluation of Medical Devices Part 1: Evaluation ું. and Testing, and the FDA Modified ISO 10993 Test Profile
- ું. IEC 60601-2-34: 2000-10. Medical electrical equipment - Particular requirements for the safety, including essential performance, of invasive blood pressure monitoring equipment
- AAMI TIR9: 1992, Evaluation of Clinical Systems for Invasive Blood Pressure Monitoring ું.
- な ANSI/AAMI BP22: 1994, Blood Pressure Transducers
### Are performance data available to assess effects of new characteristics?
Yes. Bench testing was based on the above referenced guidance document and standards.
### Performance data demonstrate equivalence?
Yes. The PowerHickman™ Central Venous Catheters met performance criteria of the safety and effectiveness tests performed and, based on FDA's decision tree, are substantially equivalent to the predicate PowerLine™ Central Venous Catheters, K050185, concurrence date May 26, 2005, K051417, concurrence date June 30, 2005, and K051991, concurrence date October 20, 2005
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes entwined around it. The caduceus is positioned to the right of the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES . USA", which is arranged in a circular fashion around the symbol. The logo is black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
## AUG - 9 2006
C.R. Bard, Incorporated Ms. Michaela Rivkowich Senior Regulatory Affairs Specialist Bard Access Systems, Incorporated 5425 West Amelia Earhart Drive Salt Lake City, Utah 84116
Re: K061179
Trade/Device Name: 8 Fr Single Lumen and 9.5 Fr Dual Lumen PowerHickman™ Central Venous Catheters Regulation Number: 880.5970
Regulation Name: Percutaneous, Implanted, Long-Term Intravascular Catheter Regulatory Class: II Product Code: LJS Dated: June 30, 2006
Received: July 3, 2006
Dear Mr. Rivkowich:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include bonnois pro for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 -Ms. Rivkowich
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits vour device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Syndite y. Michie Davis
Chu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
### Section 4 Indications for Use
510(k) Number (if known): _K061179
# Device Name: 8 Fr Single Lumen and 9.5 Fr Dual Lumen PowerHickman™ Central Venous Catheters
Indications for Use:
The PowerHickman catheter is indicated for short or long term access to the central The I ower Hickman catheter is designed for the administration of I.V. fluids, blood products, drugs, parenteral nutrition solutions, as well as blood withdrawal, power injection of contrast media and allows for central venous pressure monitoring. power injoction of commended infusion rate is 5ml/sec for power injection of contrast I he maximum rocommended iniser monitoring, it is recommended that catheter lumen of 20 gauge or larger be used.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Anthony D. n
Anesthesiology, General Hospital, on Control. Dental Dev
Number: K 461179
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.