Remunity 2.0 Pump for Remodulin (treprostinil) Injection
K241736 · Deka Research and Development Corp. · FRN · Jan 16, 2025 · General Hospital
Device Facts
Record ID
K241736
Device Name
Remunity 2.0 Pump for Remodulin (treprostinil) Injection
Applicant
Deka Research and Development Corp.
Product Code
FRN · General Hospital
Decision Date
Jan 16, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5725
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Remunity® 2.0 Pump for Remodulin® (treprostinil) Injection (the Remunity 2.0 System) is intended for continuous subcutaneous delivery of Remodulin (treprostinil) Injection for use in patients, ages 17 and older.
Device Story
Wearable infusion pump for continuous subcutaneous delivery of Remodulin (treprostinil) in PAH patients; consists of durable pump assembly, disposable 3mL user-filled cassette, and touchscreen remote interface. Pump utilizes micro-dosing mechanism with acoustic volume sensor (AVS) feedback for delivery accuracy monitoring. System operates via rechargeable lithium-ion battery; provides visual, audible, and vibratory user feedback. Used by patients in ambulatory settings; prescription-only. Pump infuses drug based on individualized programmed rates (8–225 µL/hr). Healthcare providers program pump; patients manage daily use and cassette changes (up to 72-hour wear time). System improves patient quality of life by enabling wearable, ambulatory drug delivery; AVS feedback ensures precise dosing and occlusion detection, enhancing safety compared to manual delivery methods.
Clinical Evidence
No clinical data or animal studies were obtained. Substantial equivalence supported by bench performance testing (delivery accuracy, occlusion performance, fault insertion, drug stability, battery performance, durability) and human factors validation.
Indicated for continuous subcutaneous delivery of Remodulin (treprostinil) Injection in patients aged 17 and older with pulmonary arterial hypertension (PAH).
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
Predicate Devices
Unity Subcutaneous Delivery System for Remodulin® (K190182)
K250357 — RemunityPRO Pump for Remodulin® (treprostinil) Injection · Deka Research & Development Corp. · Mar 11, 2025
K191313 — Unity Subcutaneous Infusion System for Remodulin · DEKA Research & Development · Feb 21, 2020
K240256 — Remunity System · Deka Research and Development · Jun 12, 2024
K202690 — Remunity Pump for Remodulin (treprostinil) Injection · Deka Research and Development · Dec 30, 2020
K243354 — Remunity System · Deka Research & Development Corp. · Nov 27, 2024
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 16, 2025
DEKA Research and Development Corp. Paul Smolenski Regulatory Affairs 340 Commercial Street Manchester, New Hampshire 03101
Re: K241736
Trade/Device Name: Remunity® 2.0 Pump for Remodulin® (treprostinil) Injection Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: Class II Product Code: FRN Dated: December 18, 2024 Received: December 18, 2024
Dear Paul Smolenski:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
# Rebecca Dorsey -S
For Jake Lindstrom, Ph.D. Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
510(k) Number (if known) K241736
Device Name
Remunity® 2.0 Pump for Remodulin® (treprostinil) Injection
Indications for Use (Describe)
The Remunity® 2.0 Pump for Remodulin® (treprostinil) Injection (the Remunity 2.0 System) is intended for continuous subcutaneous delivery of Remodulin (treprostinil) Injection for use in patients, ages 17 and older.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# K241736 510(k) Summary
This 510(k) Summary is prepared in accordance with the requirements of 21 CFR Part 807.92.
# Submitter Information
| 510(k) Sponsor | DEKA Research and Development Corp<br>340 Commercial Street<br>Manchester, NH 03101 |
|-------------------|------------------------------------------------------------------------------------------------------------------------------|
| Contact Person(s) | Paul Smolenski (primary)<br>Lauren Blake (secondary) |
| | Regulatory Affairs<br>Phone: (603) 669-5139<br>Fax: (603) 624-0573<br>psmolenski@dekaresearch.com<br>lblake@dekaresearch.com |
Date Prepared January 10, 2025
## Proposed Device
| Common/Usual Name: | Infusion Pump |
|-------------------------|------------------------------------------------------------|
| Trade/Proprietary Name: | Remunity® 2.0 Pump for Remodulin® (treprostinil) Injection |
| Classification Name: | Infusion pump |
| Device Classification: | 880.5725 |
| Product Code: | FRN |
| Class: | II |
| Device Panel: | General Hospital |
# Predicate Device
Unity Subcutaneous Delivery System for Remodulin®, K190182
The predicate device has not been subject to a design-related recall.
### Reference Devices
The cassette and pump hardware of the subject system are identical to those of the reference device, the DEKA ACE Pump System cleared on March 13, 2024 under K233952.
{5}------------------------------------------------
The remote interface hardware of the subject system is identical to that of the reference device, the Remunity System which was previously cleared under premarket application K240256 on June 12, 2024.
# Device Description
The Remunity 2.0 System is a wearable infusion pump designed to deliver Remodulin® (treprostinil) subcutaneously for the treatment of pulmonary arterial hypertension (PAH). Remodulin® was FDA-approved under NDA 021272. The system consists of several components: a wearable pump assembly, a remote interface, and accessories (e.g., rechargeable batteries, battery charger, charging cable, power adapter). A commercially available subcutaneous infusion set is connected to the pump assembly via a standard luer connector for the delivery of Remodulin® from the system to the patient. The Remunity 2.0 System is prescription use only.
The pump assembly is composed of a durable pump and a disposable, single-use cassette with a user-filled drug reservoir. The pump infuses Remodulin® subcutaneously into the patient based on an individualized programmed rate. Each disposable cassette may be used for up to 72 hours after attachment to the pump. The subject system utilizes a micro-dosing pump mechanism supplemented with acoustic volume sensor feedback to ensure delivery accuracy.
# Indications for Use
The Remunity® 2.0 Pump for Remodulin® (treprostinil) Injection (the Remunity 2.0 System) is intended for continuous subcutaneous delivery of Remodulin (treprostinil) Injection for use in patients, ages 17 and older.
{6}------------------------------------------------
# Substantial Equivalence Discussion
### Intended Use Comparison
The tables below includes a matrix of the intended use between the subject device and the predicate device.
| Characteristic | Predicate Device<br>(K190182) | Subject Device | Equivalence |
|------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device<br>Classification<br>Regulation and<br>Product Code | Class II Infusion Pump<br>21 CFR 880.5725<br>Product Code: FRN | Class II Infusion Pump<br>21 CFR 880.5725<br>Product Code: FRN | No change |
| Indications for<br>Use | The Unity<br>Subcutaneous Delivery<br>System for<br>Remodulin® (the Unity<br>System) is intended for<br>continuous<br>subcutaneous delivery<br>of Remodulin<br>(treprostinil) injection<br>for use in adults<br>(greater than 22 years<br>of age). | The Remunity® 2.0<br>Pump for Remodulin®<br>(treprostinil) Injection<br>(the Remunity 2.0<br>System) is intended for<br>continuous<br>subcutaneous delivery<br>of Remodulin<br>(treprostinil) Injection<br>for use in patients, ages<br>17 and older. | Equivalent<br>Both devices are labeled<br>specifically for use with<br>Remodulin. The subject system<br>has a widened indicated patient<br>population age range in order to<br>align with the indications for<br>Remodulin. |
| Prescription<br>Use or Over the<br>Counter (OTC) | Prescription | Prescription | No change |
| Intended<br>Patient<br>Population | Greater than 22 years<br>of age | Age 17 years and older | Equivalent<br>The subject system has a<br>widened indicated patient<br>population age range, in<br>alignment with the Remodulin<br>indications. |
| Patient<br>Environment | On-body wearable<br>ambulatory pump | On-body wearable<br>ambulatory pump | No change |
The intended use remains unchanged in the subject device with respect to the predicate device (K190182). The indications for use have been updated expand the indicated patient population, in alignment with Remodulin drug labeling. This update has been verified through bench performance testing and validated through human factors testing.
{7}------------------------------------------------
### Technological Characteristic Comparison
The table below compares the updates made to the subject device as compared to the predicate device and includes an assessment of these updates in the context of safety and effectiveness.
| Characteristic | Predicate Device<br>(K190182) | Subject Device | Equivalence |
|--------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Delivery Method | Microprocessor controlled<br>micro-dosing pump<br>mechanism supplemented<br>with acoustic volume<br>sensor (AVS) feedback<br>for monitoring delivery<br>accuracy | Microprocessor controlled<br>micro-dosing pump<br>mechanism supplemented<br>with acoustic volume<br>sensor (AVS) feedback for<br>monitoring delivery<br>accuracy | No change |
| Delivery type | Subcutaneous infusion | Subcutaneous infusion | No change |
| Dimensions | 6 cm x 6 cm x 2 cm<br>(2.4 in x 2.4 in x 0.4 in) | 6 cm x 6 cm x 2 cm<br>(2.4 in x 2.4 in x 0.4 in) | No change |
| Weight | 50 g (1.76 oz.) | 50 g (1.76 oz.) | No change |
| Basal Delivery<br>Rate Range | 16 µL/hr – 225 µL/hr,<br>with increments of 1<br>µL/hr | 8 µL/hr – 225 µL/hr, with<br>increments of 1 µL/hr | Equivalent<br>The basal delivery rate<br>range for the subject<br>system is wider than<br>that of the predicate,<br>but remains in<br>alignment with the<br>Remodulin prescribing<br>information.<br>Bench testing,<br>including delivery<br>accuracy and occlusion<br>detection, demonstrates<br>equivalent performance<br>for the expanded flow<br>rate range. No<br>additional risks. |
| Basal Accuracy | +6% | +6% | No change |
| Characteristic | Predicate Device<br>(K190182) | Subject Device | Equivalence |
| Bolus Volume<br>after Occlusion<br>Release | < 40 µL at all rates | < 8 µL at all rates | Equivalent<br>The subject system<br>demonstrates improved<br>post-occlusion bolus<br>volume performance<br>compared to the<br>predicate device. No<br>additional risks. |
| Time to<br>Occlusion Alarm | Maximum time to<br>occlusion alarm:<br>• ≥ 100 µL/hr: 12<br>minutes<br>• < 100 µL/hr: 8 hours | Maximum time to<br>occlusion alarm:<br>• ≥ 100 µL/hr: 12<br>minutes<br>• < 100 µL/hr: 8 hours | No change |
| Material<br>Biocompatibility | Fluid contacting:<br>Polycarbonate (PC),<br>Bromobutyl, SEBS,<br>Polyurethane<br>Patient contacting: PC,<br>Acrylic, Polyurethane,<br>ABS, Aluminum, PC-ABS | Fluid contacting:<br>Polycarbonate (PC), COP,<br>Bromobutyl, SEBS,<br>Polyurethane<br>Patient contacting: PC,<br>Acrylic, Polyurethane,<br>ABS, Aluminum | Equivalent<br>The materials of the<br>subject device meet the<br>same biocompatibility<br>standards as the<br>predicate device. No<br>additional risks. |
| Cassette Shelf<br>Life | 2 years | 2 years | No change |
| Service Life | 3 years | 3 years | No change |
| Applicable Safety<br>Standards | • IEC 60601-1<br>• IEC 60601-1-2<br>• IEC 60601-1-8<br>• IEC 60601-1-11<br>• IEC 60601-2-24<br>• ISO 11137-1<br>(Sterilized via<br>Gamma Radiation)<br>• ISO 10993-1<br>• ISO 14971 | • IEC 60601-1<br>• IEC 60601-1-2<br>• IEC 60601-1-8<br>• IEC 60601-1-11<br>• IEC 60601-2-24<br>• ISO 11137-1<br>(Sterilized via Gamma<br>Radiation)<br>• ISO 10993-1<br>• ISO 14971<br>• IEC 60601-4-2<br>• ANSI C63.27<br>• AIM 7351731 | Equivalent<br>The subject device<br>complies with<br>additional safety<br>standards. No<br>additional risks. |
| Power Source | Rechargeable Lithium Ion<br>Battery | Rechargeable Lithium Ion<br>Battery | No change |
| Characteristic | Predicate Device<br>(K190182) | Subject Device | Equivalence |
| Pump Storage<br>Conditions | Temperature: -25 °C to 70 °C (-13 °F to 158 °F) Non-condensing humidity: up to 90% Pressure: 50 kPa to 106 kPa | Temperature: -25 °C to 70 °C (-13 °F to 158 °F) Non-condensing humidity: up to 90% Pressure: 50 kPa to 106 kPa | No change |
| Operating<br>Conditions | Temperature: 5 °C to 40 °C (41 °F to 104 °F) Non-condensing humidity: up to 90% Pressure: 70 kPa to 106 kPa | Temperature: 5 °C to 40 °C (41 °F to 104 °F) Non-condensing humidity: up to 90% Pressure: 70 kPa to 106 kPa | No change |
| Remote Interface | Push-Button Device | Touchscreen Device | Equivalent<br>The subject Remote<br>Interface has identical<br>hardware and<br>functionality to the<br>reference device<br>(K240256). No<br>additional risks. |
| System User<br>Feedback | Visual, audible, vibratory | Visual, audible, vibratory | No change |
| Battery Operating<br>Time | 72 hours | 72 hours | No change |
| Infusion Set | Medtronic Quick-set Infusion Set: 23 in (58 cm), MMT-392, MMT-393 [K991759] Medtronic Silhouette Infusion Set: 23 in (58 cm), MMT-373 [K162812] Smiths Medical Cleo 90 Infusion Set: 24 in (61 cm), 21-7230-24, 21-7220-24 [K042172] | Medtronic Quick-set Infusion Set: 23 in (58 cm), MMT-392, MMT-393 [K991759] Medtronic Silhouette Infusion Set: 23 in (58 cm), MMT-373 [K162812] Smiths Medical Cleo 90 Infusion Set: 24 in (61 cm), 21-7230-24, 21-7220-24 [K042172] | No change |
| Characteristic | Predicate Device<br>(K190182) | Subject Device | Equivalence |
| Cassette | Remodulin Unity<br>Cassettes, 3 mL, user-filled | Unity 2.0 Cassettes, 3 mL, user-filled | Equivalent<br>The subject Cassette is<br>identical to the<br>reference device<br>Cassette (K233952).<br>The updated design<br>removes the filling aid<br>present in the predicate<br>Cassette. Workflows<br>associated with<br>removal of the filling<br>aid were validated<br>through human factors<br>testing. No additional<br>risks. |
| Priming Method | Manually primed by user<br>before Cassette<br>attachment to the Pump | Primed by Pump after<br>Cassette attachment to the<br>Pump | Equivalent<br>Human factors and<br>bench testing<br>demonstrates that the<br>updated prime method<br>is safe and effective.<br>No additional risks. |
{8}------------------------------------------------
### K241736 Page 5 of 8
{9}------------------------------------------------
### K241736 Page 6 of 8
{10}------------------------------------------------
### Performance Data
Performance testing was conducted in order to verify and validate the subject device remote interface for its intended use, and to establish substantial equivalence to the respective predicate device remote interface in terms of safety and effectiveness. The FDA guidance document, Infusion Pumps Total Product Life Cycle, issued December 2, 2014 was followed.
| Device<br>Performance | The following functional areas were evaluated:<br>Delivery accuracy Occlusion performance Fault Insertion Drug stability and particulate Incidental delivery performance Remodulin exposure Battery performance Durability Shipping and Storage |
|-----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
{11}------------------------------------------------
| Software and<br>Cybersecurity | The subject device is considered to require an “Enhanced Documentation Level” per<br>the 2023 FDA guidance document.<br><br>Software and Cybersecurity testing was conducted per ANSI AAMI IEC<br>62304:2006/A1:2016 and the following FDA guidance documents:<br>Content of Premarket Submissions for Device Software Functions, issued June 14, 2023 Off-The-Shelf Software Use in Medical Devices, issued September 27, 2019 Content of Premarket Submissions for Management of Cybersecurity in Medical<br>Devices, issued September 27, 2023 Cybersecurity in Medical Devices: Quality System Considerations and Content<br>of Premarket Submissions, issued April 8, 2020 Cybersecurity in Medical Devices: Refuse to Accept Policy for Cyber Devices<br>and Related Systems Under Section 524B of the FD&C Act, issued March 30, 2023 |
|-------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Human<br>Factors | The human factors validation followed the FDA guidance document, Applying<br>Human Factors and Usability Engineering to Medical Devices, issued February 3,<br>2016, and the following standards:<br>IEC 60601-1-6:2020 ANSI AAMI IEC 62366-1:2015+A1:2020 |
| Electrical<br>Safety | Per IEC 60601-1:2020 |
| EMC | Per IEC 60601-1-2:2020, AIM 7351731 |
| Alarms | Per IEC 60601-1-8:2020 |
### Animal and Clinical Studies
No animal study or clinical trial data was obtained in support of this premarket submission.
### Design Control
The subject system was specified and developed by DEKA. DEKA complies with the FDA Quality System Regulation as specified in 21 CFR Part 820, as well as to ISO 13485.
## Conclusion
The differences between the predicate and the subject device do not raise different questions of safety or effectiveness. The Remunity 2.0 System is substantially equivalent to the Unity Subcutaneous Delivery System for Remodulin® cleared under K190182.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.