Remunity System

K240256 · Deka Research and Development · QJY · Jun 12, 2024 · General Hospital

Device Facts

Record IDK240256
Device NameRemunity System
ApplicantDeka Research and Development
Product CodeQJY · General Hospital
Decision DateJun 12, 2024
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5725
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Remunity® Pump for Remodulin® (treprostinil) Injection (the Remunity System) is intended for continuous subcutaneous delivery of Remodulin (treprostinil) Injection for use in adults (greater than 22 years old).

Device Story

Wearable infusion pump for continuous subcutaneous delivery of Remodulin (treprostinil) to treat pulmonary arterial hypertension (PAH). System comprises a reusable pump, disposable single-use cassette (pharmacy or user-filled), and a touchscreen remote controller. Pump utilizes micro-dosing mechanism with acoustic volume sensor (AVS) feedback for delivery accuracy. Used in professional healthcare facilities and home environments; operated by patients or caregivers. Healthcare providers program delivery rates; pump provides visual, audible, and vibratory feedback. Touchscreen remote facilitates user interaction and monitoring. Benefits include precise, ambulatory drug delivery for chronic PAH management.

Clinical Evidence

No clinical or animal study data provided. Substantial equivalence is supported by bench testing, including functional performance, electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), alarm testing (IEC 60601-1-8), and human factors validation.

Technological Characteristics

Wearable micro-dosing pump; AVS feedback for accuracy. Materials: PC, Bromobutyl, SEBS, polyurethane (cassette); ABS, PC, Aluminum (pump). Power: Rechargeable Li-ion battery. Connectivity: Wireless communication between pump and touchscreen remote. Sterilization: Gamma radiation (ISO 11137-1). Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11, IEC 60601-2-24, ISO 10993-1, ISO 14971.

Indications for Use

Indicated for continuous subcutaneous delivery of Remodulin (treprostinil) in adults >22 years old with pulmonary arterial hypertension.

Regulatory Classification

Identification

An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue. June 12, 2024 Deka Research and Development Paul Smolenski Regulatory Affairs 340 Commercial Street Manchester, New Hampshire 03101 Re: K240256 Trade/Device Name: Remunity System Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion pump Regulatory Class: Class II Product Code: QJY Dated: Mav 9, 2024 Received: May 10, 2024 Dear Paul Smolenski: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" {1}------------------------------------------------ (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). {2}------------------------------------------------ Sincerely, # Jake K. Lindstrom -S Jake Lindstrom, Ph.D. Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ #### Indications for Use 510(k) Number (if known) K240256 Device Name Remunity System Indications for Use (Describe) The Remunity® Pump for Remodulin® (treprosimil) Injection (the Remunity System) is intended for continuous subcutaneous delivery of Remodulin (treprostinil) Injection for use in adults (greater than 22 years old). | Type of Use (Select one or both, as applicable) | | |--------------------------------------------------------------------------------|------------------------------------------------------------------------------| | <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the word "DEKA" in a bold, sans-serif font. The letters are all capitalized and black. The "K" is stylized with an arrow shape pointing to the left. ## 510(k) Summary This 510(k) Summary of Safety and Effectiveness information is prepared in accordance with the requirements of 21 CFR Part 807.92. ### Submitter Information | 510(k) Sponsor | DEKA Research & Development<br>340 Commercial Street<br>Manchester, NH 03101 | |-------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact Person(s) | Paul Smolenski (primary), Lauren Blake (secondary)<br>Regulatory Affairs<br>Phone: (603) 669-5139<br>Fax: (603) 624-0573<br>psmolenski@dekaresearch.com, lblake@dekaresearch.com | Date Prepared January 30, 2024 ### Proposed Device(s) | Common/Usual Name: | Infusion Pump | |-------------------------|------------------------------------------------------| | Trade/Proprietary Name: | Remunity Subcutaneous Delivery System for Remodulin® | | Classification Name: | Infusion pump, drug specific, pharmacy-filled | | Device Classification: | 880.5725 | | Product Code: | QJY | | Class: | II | | Device Panel: | General Hospital | ### Predicate Device(s) The predicate device is the Remunity Subcutaneous Delivery System for Remodulin®, which was previously cleared under premarket application K202690 on December 30, 2020. ### Device Description The Remunity Subcutaneous Delivery System for Remodulin® (Remunity System) is a wearable infusion pump designed to deliver Remodulin® for the treatment of pulmonary arterial hypertension (PAH). It is intended for continuous subcutaneous delivery of FDA-approved {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the word "DEKA" in a bold, sans-serif font. The letters are all capitalized and black. The "K" in the word is stylized with an arrow shape in the middle. Remodulin® (treprostinil) (hereinafter referred to as 'Remodulin®' or 'Remodulin® (treprostinil)'), NDA 021272. The Remunity System consists of several components: a wearable pump assembly, a remote interface, a filling and priming aid, and accessories (e.g., rechargeable batteries, battery charger, charging cable, power adapter). A commercially available subcutaneous infusion set is connected to the pump assembly via a standard luer connector for the delivery of Remodulin® from the Remunity System to the patient. The pump assembly is composed of a reusable pump and a disposable single-use cassette with a pharmacy-filled or user-filled drug reservoir, which infuses Remodulin® subcutaneously into the patient based on an individualized programmed rate. Each disposable cassette may be used for up to 72 hours after attachment to the pump. The Remunity System utilizes a micro-dosing pump mechanism supplemented with acoustic volume sensor feedback to ensure delivery accuracy. The device is prescription use only. The modified device utilizes a touchscreen remote controller. ### Indications for Use The Remunity® Pump for Remodulin® (treprostinil) Injection (the Remunity System) is intended for continuous subcutaneous delivery of Remodulin (treprostinil) Injection for use in adults (greater than 22 years old). ### Substantial Equivalence Discussion #### Intended Use Comparison | Characteristic | Predicate Device - Remunity<br>System (K202690) | Subject Device – Remunity<br>System | |---------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Indications for Use | The Remunity Subcutaneous<br>Infusion System (the Remunity<br>System) is intended for<br>continuous subcutaneous<br>delivery of Remodulin®<br>(treprostinil) Injection for use in<br>adults (greater than 22 years of<br>age). | The Remunity® Pump for<br>Remodulin® (treprostinil)<br>Injection (the Remunity System)<br>is intended for continuous<br>subcutaneous delivery of<br>Remodulin (treprostinil)<br>Injection for use in adults<br>(greater than 22 years old). | The tables below includes a matrix of the intended use between the subject device and the predicate device. {6}------------------------------------------------ Image /page/6/Picture/0 description: The image shows the word "DEKA" in a bold, sans-serif font. The letters are all capitalized and black. The letters are closely spaced together, and the overall appearance is modern and industrial. | Characteristic | Predicate Device – Remunity<br>System (K202690) | Subject Device – Remunity<br>System | |--------------------------------------------------|-------------------------------------------------------------------|-------------------------------------| | Prescription Use or<br>Over the Counter<br>(OTC) | Prescription | No change | | Intended Patient<br>Population | Greater than 22 years of age | No change | | Patient Environment | On-body wearable ambulatory<br>pump | No change | | Environment of Use | In professional healthcare<br>facilities and home<br>environments | No change | The intended use remains unchanged in the subject device with respect to the predicate device (K202690). The Indications for Use have been updated to reflect the current branding for the device. This update does not affect the intended patient population or environment of use. #### Technological Characteristic Comparison The table below compares the characteristics of the predicate and subject devices, and includes an assessment of the differences and why these differences do not introduce new or different questions of safety and effectiveness. | Characteristic | Predicate Device –<br>Remunity System<br>(K202690) | Subject Device –<br>Remunity System | Equivalence | |--------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Delivery<br>Method | Microprocessor<br>controlled micro-<br>dosing pump<br>mechanism<br>supplemented with<br>acoustic volume sensor<br>(AVS) feedback for<br>monitoring delivery<br>accuracy | Same | N/A | | Delivery type | Subcutaneous infusion | Same | N/A | | Dimensions | 6 cm x 6 cm x 2 cm<br>(2.4 in x 2.4 in x 0.4 in) | Same | N/A | | Weight | 50 g (1.76 oz.) | Same | N/A | | Basal Delivery<br>Rate Range | 16 μL/hr – 225 μL/hr,<br>with increments of 1<br>μL/hr | Same | N/A | | Basal Accuracy | +6% | Same | N/A | | Characteristic | Predicate Device –<br>Remunity System<br>(K202690) | Subject Device -<br>Remunity System | Equivalence | | Bolus Volume<br>after Occlusion<br>Release | < 40 µL at all rates | Same | N/A | | Time to<br>Occlusion<br>Alarm | Maximum time to<br>occlusion alarm:<br>• ≥ 100 µL/hr: 12 minutes<br>• ≤ 100 µL/hr: 8 hours | Maximum time to<br>occlusion alarm:<br>• ≥ 100 µL/hr: 12 minutes<br>• ≤ 100 µL/hr: 8 hours | N/A | | Pump Service<br>Life | 3 years | Same | N/A | | Ingress<br>Protection | IP58 when connected to<br>the Cassette | Same | N/A | | Applicable<br>Safety Standards | • IEC 60601-1<br>• IEC 60601-1-2<br>• IEC 60601-1-8<br>• IEC 60601-1-11<br>• IEC 60601-2-24<br>• ISO 11137-1<br>(Sterilized via<br>Gamma Radiation)<br>• ISO 10993-1<br>• ISO 14971 | Same | N/A | | Power Source | Rechargeable Lithium<br>Ion Battery | Same | N/A | | Pump Storage<br>Conditions | Temperature: -25 °C to<br>70 °C (-13 °F to 158 °F)<br>Humidity: up to 90%<br>Pressure: 50 kPa to 106 kPa | Same | N/A | | Operating<br>Conditions | Temperature: 5 °C to 40 °C (41 °F to 104 °F)<br>Humidity: up to 90%<br>Pressure: 70 kPa to 106 kPa | Same | N/A | | Characteristic | Predicate Device –<br>Remunity System<br>(K202690) | Subject Device –<br>Remunity System | Equivalence | | Remote<br>Interface | • Push-button device | • Touchscreen device | Equivalent<br>The subject device provides a touchscreen device for use with the pump. The text and format of information displayed to the user, as well as workflows and required user input remains unchanged. No additional risks have been introduced. | | System User<br>Feedback | Visual, audible, vibratory | Same | N/A | {7}------------------------------------------------ Image /page/7/Picture/0 description: The image shows the word "DEKA" in a bold, sans-serif font. The letters are all capitalized and black. The letters are closely spaced together, and the overall impression is one of strength and stability. {8}------------------------------------------------ Image /page/8/Picture/0 description: The image shows the word "DEKA" in a bold, stylized font. The letters are black and have a unique design, with the "E" having three horizontal lines and the "K" resembling an arrow pointing to the left. The "A" has a flat top and a horizontal line through the middle, giving the word a distinctive and modern appearance. The overall design is simple yet eye-catching. {9}------------------------------------------------ Image /page/9/Picture/0 description: The image shows the word "DEKA" in a bold, sans-serif font. The letters are all capitalized and black. The letters are close together, and the overall impression is one of strength and stability. The background is white. | Characteristic | Predicate Device –<br>Remunity System<br>(K202690) | Subject Device –<br>Remunity System | Equivalence | |---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------|-------------| | Common<br>Alarms | Battery Depleted<br>Cassette Depleted<br>Cassette Removed<br>Cassette Problem<br>Occlusion<br>Pump Error<br>Pump Failure<br>Basal Not Started<br>Adjust Pump<br>Delivery Stopped<br>Depletes Soon<br>Excessive Noise<br>Message Timeout<br>No Communication<br>Pairing Failed<br>Pairing Lost<br>Pump Idle<br>Pump Battery Low<br>Remote Battery Low<br>Request Refused<br>Software Version Error<br>Tech<br>Walkaway | Same | N/A | | Battery<br>Operating Time | 72 hours | Same | N/A | {10}------------------------------------------------ Image /page/10/Picture/0 description: The image shows the word "DEKA" in a bold, stylized font. The letters are black and have a unique design, with rounded edges and geometric shapes. The overall appearance is modern and industrial. The letters are all capitalized and evenly spaced. | Characteristic | Predicate Device –<br>Remunity System<br>(K202690) | Subject Device –<br>Remunity System | Equivalence | |--------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------|-------------| | Infusion Set | Medtronic Quick-set<br>Infusion Set [K991759]<br>42" (MMT-390, MMT-391)23" (MMT-392, MMT-393)<br>Medtronic Silhouette<br>Infusion Set [K162812]<br>23" (MMT-373)<br>Smiths Medical Cleo 90<br>Infusion Set [K042172]<br>24" (21-7220-24, 21-7230-24)31" (21-7221-24, 21-7231-24)42" (21-7222-24, 21-7232-24) | Same | N/A | | Compatible<br>Cassette | Remodulin Unity<br>Cassettes, 3 mL | Same | N/A | | Cassette Filling<br>Method | Specialty pharmacyUser-filled | Same | N/A | | Cassette Shelf<br>Life (empty) | 2 years | Same | N/A | | Pharmacy-filled<br>Shelf-Life | 14 days (per <USP<br>797>) | Same | N/A | {11}------------------------------------------------ Image /page/11/Picture/0 description: The image shows the word "DEKA" in a bold, sans-serif font. The letters are all capitalized and black. The "K" is stylized with a sharp, arrow-like shape pointing to the left. The background is white. | Characteristic | Predicate Device –<br>Remunity System<br>(K202690) | Subject Device –<br>Remunity System | Equivalence | |----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------|-------------| | Material<br>Biocompatibility | Cassette fluid path:<br>Polycarbonate (PC),<br>Bromobutyl, SEBS,<br>polyurethane | Same | N/A | | | Pump: ABS, PC,<br>Aluminum | | | | | Cartridge: PC, Acrylic,<br>polyurethane | | | | | Filling Aid: PC-ABS | | | | | Cap (Pharmacy-fill):<br>ABS | | | | | | | | | Pharmacy-fill<br>End-user<br>Packaging | Aseptically filled<br>cassette with female<br>luer lock fluid path<br>closure, placed in<br>plastic clamshell tray<br>and sealed in foil pouch | Same | N/A | | Priming Method | Manually primed by<br>user before Cassette<br>attachment to the Pump | Same | N/A | #### Performance Data Performance testing was conducted in order to verify and validate the subject device remote interface for its intended use, and to establish substantial equivalence to the respective predicate device remote interface in terms of safety and effectiveness. The FDA Guidance Document, Infusion Pumps Total Product Life Cycle, issued December 2, 2014 was followed. | Device<br>Performance | The following functional areas, specific to the remote interface, were evaluated:<br>Touchscreen Remote Reliability Touchscreen Remote Functionality Wireless Communication | |-------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Software and<br>Cybersecurity | The subject device is considered to require an “Enhanced Documentation Level” per the 2023 FDA guidance document. | {12}------------------------------------------------ Image /page/12/Picture/0 description: The image shows the word "DEKA" in a bold, sans-serif font. The letters are all capitalized and black. The letters are stylized with rounded corners and a slightly futuristic look. The "K" is designed with an arrow-like shape pointing to the left. | | Software and Cybersecurity testing was conducted per ANSI AAMI IEC<br>62304:2006/A1:2016 and the following FDA Guidance Documents: | |-------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | • Content of Premarket Submissions for Device Software Functions,<br>issued June 14, 2023 | | | • Off-The-Shelf Software Use in Medical Devices, issued September 27,<br>2019 | | | • Content of Premarket Submissions for Management of Cybersecurity<br>in Medical Devices, issued September 27, 2023 | | | • Cybersecurity in Medical Devices: Quality System Considerations and<br>Content of Premarket Submissions, issued April 8, 2020 | | | • Cybersecurity in Medical Devices: Refuse to Accept Policy for Cyber<br>Devices and Related Systems Under Section 524B of the FD&C Act,<br>issued March 30, 2023 | | Human Factors | The human factors evaluation followed the FDA Guidance Document,<br>Applying Human Factors and Usability Engineering to Medical Devices,<br>issued February 3, 2016, and the following standards: | | | • IEC 60601-1-6:2020 | | | • ANSI AAMI IEC 62366-1:2015+A1:2020 | | Electrical Safety | Per IEC 60601-1:2020 | | EMC | Per IEC 60601-1-2:2020 | | Alarms | Per IEC 60601-1-8:2020 | #### Animal and Clinical Studies No animal study or clinical trial data was obtained in support of this premarket submission. #### Design Control The Remunity System was specified and developed by DEKA. DEKA complies with the FDA Quality System Regulation as specified in 21 CFR 820, as well as to ISO 13485:2016. ### Conclusion The differences between the predicate and the subject device do not raise any new or different questions of safety or effectiveness. The Remunity Subcutaneous Delivery System for Remodulin® is substantially equivalent to the Remunity Subcutaneous Delivery System for Remodulin® cleared under K202690 with respect to the indications for use, target populations, treatment method, and technological characteristics.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...