AUTOPRIME
Device Facts
| Record ID | K964283 |
|---|---|
| Device Name | AUTOPRIME |
| Applicant | Arbor Technologies, Inc. |
| Product Code | FPB · General Hospital |
| Decision Date | Jan 22, 1997 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 880.5440 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Indications for Use
The Arbor Medical AutoPrime™ I.V. is intended to filter intravenous solutions. Per FDA Safety Alert: Hazards of Precipitation Associated with Parental Nutrition, issued April 18, 1994, an IV filter is indicated for use when infusing either central or peripheral parenteral nutrition admixtures. Both pore sizes of filters are being offered, a 0.22um air eliminating filter for nonlipid containing admixtures, and a 1.2um air eliminating filter for lipid containing admixtures.
Device Story
AutoPrime™ I.V. is an infusion line filter and administration set; used to filter intravenous solutions during parenteral nutrition delivery. Device incorporates air-eliminating filters to remove particulates and air from infusion lines. 0.22um filter variant targets nonlipid admixtures; 1.2um filter variant targets lipid admixtures. Operated by clinicians in clinical settings. Device ensures patient safety by preventing infusion of precipitates or air into the bloodstream.
Clinical Evidence
Bench testing only; no clinical data provided. Equivalence established via comparison of materials and performance specifications to predicate devices.
Technological Characteristics
Infusion line filter and administration set. Materials: polyethersulfone main membranes (0.22um or 1.2um), 0.02um PTFE vent filters, modified acrylic housings. Air-eliminating design. No electronic or software components.
Indications for Use
Indicated for patients requiring intravenous therapy, specifically for the filtration of central or peripheral parenteral nutrition admixtures. 0.22um filter for nonlipid containing admixtures; 1.2um filter for lipid containing admixtures.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
Predicate Devices
- 3M Health Care AVI™ Administration Set (0.22um filter)
- Block Medical Verifuse® Administration Set (1.2um filter)
Related Devices
- K030385 — IV ADD-ON FILTER ADMINISTRATION SET WITH AND WITHOUT Y CONNECT OR NEEDLELESS ACCESS CONNECTOR · Kawasumi Laboratories America, Inc. · Apr 15, 2003
- K974661 — 0.2 U BACTERIAL FILTER · Sims Deltec, Inc. · Feb 11, 1998
- K964850 — SOLUTION ADMINISTRATION SET WITH 0.22 MICRON HIGH PRESSURE EXTENDED LIFE FILTER · Baxter Healthcare Corp · Feb 25, 1997
- K993379 — PALL SUPOR AEF FILTER · Pall Corp. · Dec 16, 1999
- K202672 — Precision Delivery Infusion Set · Quest Medical, Inc. · Jan 21, 2021