Precision Delivery Infusion Set

K202672 · Quest Medical, Inc. · FPA · Jan 21, 2021 · General Hospital

Device Facts

Record IDK202672
Device NamePrecision Delivery Infusion Set
ApplicantQuest Medical, Inc.
Product CodeFPA · General Hospital
Decision DateJan 21, 2021
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5440
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Precision Delivery Infusion Sets are intended to administer or aspirate fluids. These devices may be used with any patient population with consideration given to the procedure being performed and fluids being infused.

Device Story

Precision Delivery Infusion Sets are single-use, disposable extension sets for fluid administration or aspiration. Components include tubing, filters, clamps, needleless connectors, and luers. Operated by clinicians in clinical settings, the device functions via gravity-fed fluid delivery. The device acts as a conduit between an IV source and a patient's intravascular catheter. The inclusion of a fluid filter allows for particulate removal. The device benefits patients by providing a sterile, reliable pathway for fluid therapy or blood/fluid aspiration. No complex electronics or software are involved.

Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing, including priming volume, backpressure leak, flow rate, simulated use, tubing bond strength, clamping, kink resistance, aspiration, ISO 8536-4 chemical/particulate/filter efficiency, ISO 80369-7 luer compliance, USP <788> particulate, biocompatibility (cytotoxicity, sensitization, irritation, pyrogenicity, hemocompatibility, systemic toxicity), and sterilization validation.

Technological Characteristics

Materials: Non-DEHP PVC tubing, HDPE clamp/cap, Acrylic/PTFE/PES filter, Polycarbonate/Silicone connector, ABS/Co-Polyester luers. Principle: Gravity-fed fluid conduit. Dimensions: 7"-62" length. Sterilization: ISO 11135:2014. Connectivity: None (mechanical).

Indications for Use

Indicated for administration or aspiration of fluids in any patient population, contingent upon the specific procedure and fluids involved.

Regulatory Classification

Identification

An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.

Special Controls

*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. January 21, 2021 Quest Medical, Inc. Tosan Eweka Regulatory Affairs Supervisor One Allentown Parkway Allen, Texas 75002 Re: K202672 Trade/Device Name: Precision Delivery Infusion Sets Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: Class II Product Code: FPA Dated: December 17, 2020 Received: December 18, 2020 Dear Tosan Eweka: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part {1}------------------------------------------------ 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, CAPT Alan M. Stevens Director (acting) DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K202672 Device Name Precision Delivery Infusion Sets #### Indications for Use (Describe) The Precision Delivery Infusion Sets are intended to administer or aspirate fluids. These devices may be used with any patient population with consideration given to the procedure being performed and fluids being infused. Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter-Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW." The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. 116 {3}------------------------------------------------ ## GENERAL INFORMATION [807.92(a)(1)] #### Applicant: Quest Medical, Inc. One Allentown Parkway Allen, TX 75002-4211 USA Phone: 972-332-6338 Fax: 972-390-2881 #### Contact Person: Tosan Eweka Regulatory Affairs Supervisor Quest Medical, Inc. One Allentown Parkway Allen, TX 75002-4211 USA Phone: 972-332-6338 972-390-2881 Fax: Email: teweka@questmedical.com #### Date Prepared: September 17, 2020 ## DEVICE INFORMATION [807.92(a)(2)] #### Trade Name: Precision Delivery Infusion Sets #### Generic/Common Name: Intravascular Administration Set #### Classification: Class II per 21CFR§880.5440 # Product Code: FPA {4}------------------------------------------------ ## PREDICATE DEVICE(S) [807.92(a)(3)] Extension Sets with BD Q-Syte Luer Access Split Septum (K142527) ## DEVICE DESCRIPTION [807.92(a)(4)] Ouest Medical's Precision Delivery Infusion Sets are single use, disposable, extension sets used to deliver fluids to a patient. The Precision Delivery Sets consist of various configurations, which includes tubing, filter, clamp, needleless connector and luers. ### INDICATIONS FOR USE [807.92(a)(5)] The Precision Delivery Infusion Sets are intended to be used to administer or aspirate fluids. These devices may be used with any patient population with consideration given to the procedure being performed and fluids being infused. #### TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE [807.92(a)(6)] The Precision Delivery Infusion Set has the same intended use, indications for use, and principles of operation and similar design features and technological characteristics as the predicate device. The differences in the technological characteristics between the Precision Delivery Infusion Set and the predicate device do not raise different questions of safety or effectiveness. Thus, the Precision Delivery Infusion Set is substantially equivalent to the predicate device. An analysis of the differences in technological characteristics is provided in the table below. | Element of<br>Comparison | Predicate Device –<br>Extension Sets with BD<br>Q-Syte Luer Access<br>Split Septum<br>(K142527) | Subject Device | Analysis of<br>Differences | |--------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Indications for Use | Extension Sets with an<br>attached BD Q-Syte<br>Luer Access Split<br>Septum are intended to<br>be used with<br>intravascular catheters to<br>aspirate blood or<br>administer fluids,<br>including medications<br>and blood or blood<br>products. These devices<br>may be used with any<br>patient population with<br>consideration given to<br>the procedure being<br>performed and fluids<br>being infused. | The Precision Delivery<br>Infusion Sets are<br>intended to be used to<br>administer or aspirate<br>fluids. These devices<br>may be used with any<br>patient population with<br>consideration given to<br>the procedure being<br>performed and fluids<br>being infused. | The main difference<br>is the indications for<br>use of the predicate is<br>broader than the<br>indications for use of<br>the subject device.<br>The predicate is<br>indicated for<br>aspiration and<br>administration of<br>blood, blood products<br>and fluids while the<br>subject device is<br>indicated for<br>administration or<br>aspiration of fluids.<br>The indications for<br>the subject device<br>falls within the<br>indications for the<br>predicate device. The<br>difference in the | | Element of<br>Comparison | Predicate Device -<br>Extension Sets with BD<br>Q-Syte Luer Access<br>Split Septum<br>(K142527) | Subject Device | Analysis of<br>Differences | | | | | indications for use<br>therefore does not<br>alter the intended use<br>of the predicate<br>device and the subject<br>device, which is to<br>provide a conduit for<br>the administration of<br>fluids. | | Design<br>Features/Components | Tubing with Slide<br>clamp, Needleless<br>connector, Luers and<br>Caps. Extension Sets are<br>provided in single<br>tubing, bifurcated tubing<br>or trifurcated tubing<br>configurations with<br>lengths ranging from 5"-<br>21". | Tubing with Slide<br>clamp, Needleless<br>connector, Luers, Caps<br>and Filter. Extension<br>Sets are provided in<br>single tubing and<br>trifurcated tubing<br>configurations with<br>lengths ranging from 7"-<br>62". | The predicate device<br>and subject device are<br>both offered in single<br>and trifurcated tubing<br>configurations. The<br>predicate device<br>additionally has a<br>bifurcated tubing<br>configuration. The<br>subject device is not<br>offered in bifurcated<br>models. The predicate<br>device and subject<br>device are comprised<br>of the same<br>component types with<br>the exception of a<br>filter component<br>included in the<br>subject device. The<br>fluid filter in the<br>subject device is<br>intended to filter<br>particulates. Fluid<br>filter efficiency test<br>was conducted on the<br>subject device and all<br>acceptance criteria<br>was met. The<br>minimum/maximum<br>length configuration<br>of the subject device<br>is shorter/longer than<br>the predicate device.<br>The results from<br>performance testing<br>conducted on the<br>subject device<br>demonstrates that the | | Element of<br>Comparison | Predicate Device -<br>Extension Sets with BD<br>Q-Syte Luer Access<br>Split Septum<br>(K142527) | Subject Device | Analysis of<br>Differences | | | | | difference in design<br>features and<br>components between<br>the subject device and<br>predicate device do<br>not raise different<br>questions of safety<br>and effectiveness. The<br>subject device met all<br>performance<br>specifications<br>necessary to fulfil its<br>intended use. | | Mode of Fluid<br>Delivery | Gravity | Same | No difference | | Device Materials | Tubing<br>• Non-DEHP<br>PVC<br>Slide clamp<br>• Unknown<br>Luer Cap<br>• Unknown<br>Needleless Connector<br>• Unknown<br>Luers<br>• Unknown | Tubing<br>• Non-DEHP<br>PVC<br>Slide clamp<br>• HDPE<br>Luer Cap<br>• HDPE<br>Filter<br>• Acrylic/PTFE/PES<br>Needleless Connector<br>• Polycarbonate/Silicone<br>Luers<br>• ABS/Co-Polyester | The subject device<br>tubing and predicate<br>device tubing are<br>made from Non-DEHP PVC. The<br>materials used on<br>other components of<br>the predicate device is<br>unknown. Results<br>from biocompatibility<br>testing conducted on<br>the subject device<br>demonstrate that<br>differences (if any) in<br>the materials used in<br>the subject device and<br>predicate device do<br>not raise different<br>questions of safety<br>and effectiveness. | | Patient Contact<br>Category/Duration | Externally<br>Communicating, Blood<br>Path Indirect, Prolonged<br>contact | Same | No difference | | Sterilization | SAL 10-6 | Same | No difference | {5}------------------------------------------------ {6}------------------------------------------------ {7}------------------------------------------------ ## PERFORMANCE DATA [807.92(b)] All necessary testing was conducted on the Precision Delivery Infusion Sets to support a determination of substantial equivalence to the predicate device. ## Nonclinical Testing Summary [807.92(b)(1)] The bench and nonclinical tests conducted on the Precision Delivery Infusion Sets in this submission included: - Priming Volume - Backpressure Leak ● - Flow Rate ● - Simulated Maximum Use ● - o Tubing Bond Strength - Multiple Clamping - Prolonged Clamping ● - o Tubing Kink Resistance - Aspiration Test ● - ISO 8536-4 Chemical and Particulate Tests ● - ISO 8536-4 Filter Efficiency Test - ISO 80369-7:2016 Luer compliance - USP <788> Particulate Test - Biocompatibility Tests (Cytotoxicity, Sensitization, Irritation, Pyrogenicity, o Hemocompatibility, Acute Systemic Toxicity, Sub-Chronic Toxicity) - Sterilization validation per ISO 11135:2014 ● - Bacterial Endotoxin Tests ● Microbial ingress data for the needleless connector component was referenced to the previously cleared K002689. # Clinical Testing Summary [807.92(b)(2)] Not applicable. Clinical testing was not performed to support this 510(k) submission. # CONCLUSIONS [807.92(b)(3)] The collective results of the performance testing demonstrate that the Precision Delivery Infusion Sets meets the established specifications necessary for consistent performance during its intended use. In addition, the collective performance testing demonstrate that the Precision Delivery Infusion Sets does not raise different questions of safety or effectiveness when compared to the predicate device. The results from the performance testing support the conclusion that the Precision Delivery Infusion Sets is substantially equivalent to the predicate device. ## SUMMARY The Precision Delivery Infusion Sets is substantially equivalent to the predicate device.
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%