K012189 · Angiplast Private , Ltd. · LHI · Mar 5, 2002 · General Hospital
Device Facts
Record ID
K012189
Device Name
ANGELTOUCH VENTED IV ADMINISTRATION SET
Applicant
Angiplast Private , Ltd.
Product Code
LHI · General Hospital
Decision Date
Mar 5, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
To administer parenteral fluids / medication into the patient's intravascular system.
Device Story
Single-use sterile infusion set designed to deliver parenteral fluids or medications from a container to a patient's intravascular system. Device components include a protective cap, closure-piercing device, air vent filter assembly, drip chamber, fluid filter, tubing, flow regulator, injection site, and male fitting. Operated by healthcare professionals in clinical settings. The device facilitates controlled fluid delivery; flow is managed via a manual flow regulator. Benefits include reliable, sterile fluid administration. No electronic or software components.
Clinical Evidence
Bench testing only. The device complies with ISO 8536-4 requirements for intravascular administration sets.
Technological Characteristics
Materials: Polypropylene (PP), Acrylonitrile Butadiene Styrene (ABS), Polyvinyl Chloride (PVC) with DEHP plasticizer, High-Density Polyethylene (HDPE), Low-Density Polyethylene (LDPE), and Latex. Mechanical flow regulation. Sterile, single-use. No energy source, software, or connectivity.
Indications for Use
Indicated for the administration of parenteral fluids or medication into the patient's vascular system.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
Predicate Devices
PRIMARY I.V. SET, Convertible Pin, 100 Inch with Backcheck Valve and 2 CLAVE® ports Piggyback MICRODRIP® with OPTION-LOK® (LifeShield®)
Continu-Flo® Solution Set (Interlink® System)
Submission Summary (Full Text)
{0}------------------------------------------------
| Angiplast Pvt. Ltd., Ahmedabad, India | Issue No.: 1 |
|---------------------------------------|-----------------------|
| Device Master File - 510(k) | Date: January 1, 2001 |
| L.V. ADMINISTRATION SET | Revision No.: 2 |
| | Date: January 1, 2002 |
K012/89
5 2002 mar
# 15.0 PREMARKET NOTIFICATION [ 510(K) ] SUMMARY
| ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | 1 | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>-------<br>-------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>--------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>.<br>1<br><br>AND OF LE LE LE BERESH AND A<br><br>CARD L. CARD COLUM |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| .<br>LAND BUILD COMMENT<br>----------<br>. | Associate<br>we all consumer of the distribution of control and consider of can and any and<br>Comments of Concess of Children of Children a<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | AND AND AN<br>I assessment the distributions | .<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ |
| | | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>A career commends of<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ |
{1}------------------------------------------------
| Angiplast Pvt. Ltd., Ahmedabad, India | Issue No.: 1 |
|---------------------------------------|-----------------------|
| | Date: January 1, 2001 |
| Device Master File - 510(k) | Revision No.: 2 |
| | Date: January 1, 2002 |
| I.V. ADMINISTRATION SET | |
- Name, Address and Phone numbers of Manufacturer 15.1
5 2002 MAR
- Name of Submitter .
Image /page/1/Figure/4 description: The image shows a signature block. The name "Mr. Ashit B. Shah" is printed on the left side of the block, with the title "Director" printed below it. On the right side of the block, the word "Signature" is printed below a signature line, and a signature is written above the line.
- Registered address and address of manufacturing and sterilization plant .
| ANGIPLAST PRIVATE LIMITED | |
|----------------------------------|-------------------------------|
| 4803, Phase-IV, G.I.D.C. Estate, | |
| Vatva, Ahmedabad 382 445 | |
| India | |
| Phone No.: | +91 79 584 0661 / 584 1967 |
| Fax No.: | +91 79 584 1009 |
| E-mail: | medical.appl@gems.vsnl.net.in |
- Date: 18-6-2001 ●
Details of legally marketed device to which we claim equivalence 15.2
| 1. Name & Description | PRIMARY I.V. SET, Convertible Pin, 100 Inch with Backcheck Valve and 2 CLAVE® ports Piggyback MICRODRIP® with OPTION-LOK® |
|--------------------------------|---------------------------------------------------------------------------------------------------------------------------|
| Trade Name | LifeShield® |
| Name & Address of Manufacturer | Abbott Laboratories, North Chicago, IL 60064, USA |
| <br>with the first to the first and the charges and any and any and any and a | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | .<br>---------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br> | and and season and the direction of the climit of the climit of the climit of the climit of the climit of the climit of the climit of the climit of the climit of the client o<br><br>-------------------<br>Concession Colline Collins of<br>------------------------------------------------------------------------------------------------------------------------<br>.<br><br>------- |
|-------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| I | CARDER CONSULTION CONSULTION CONSULTION CONTRACT COLLEGION CONTRACT CONTRACT COLLEGION CONTRACT COLLECTION CONTRACT COLLEGION CONTRACT COLLECTION CONTRACT COLLECTION CONTRACT<br>LANDRY & MOND ANNUAL A<br> | -------------------------------------------------------------------------------- | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ |
| | | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>1 | .<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>. In come a second a some a more a<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ |
{2}------------------------------------------------
| Angiplast Pvt. Ltd., Ahmedabad, India | Issue No.: 1 |
|---------------------------------------|----------------------|
| Device Master File - 510(k) | Revision No.: 2 |
| I.V. ADMINISTRATION SET | Date: January 1, 200 |
| | Date: January 1, 200 |
| 2. | Name & Description | : Continu-Flo® Solution Set |
|----|-----------------------------------|-------------------------------------------------------------|
| | Trade Name | : Interlink® System |
| | Name & Address of<br>Manufacturer | : Baxter Healtchcare Corporation<br>Deerfield, IL 60015 USA |
#### Description of the Device 15:3
- The product is Single Use Sterile Infusion Set for Medical Use. .
The infusion set consists of the components as illustrated in figure given below.
Image /page/2/Figure/5 description: This image shows a diagram of an intravenous (IV) infusion set and a table describing the components. The diagram shows the different parts of the IV set, labeled with numbers 1 through 10. The table lists four components: a protective cap, a closure-piercing device, an air vent filter assembly, and a drip chamber. The table also includes the intended use and material of each component, such as polypropylene (PP), acrylonitrile butadiene styrene (ABS), and polyvinyl chloride (PVC).
| ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>-------<br>A LINE START LEBELL MODEL & MILLER<br>1 | .<br>---------------<br> | <br>.<br>-----------<br>LEBEL & MIN-RESEAR<br>--------- | ----<br>-<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>LAND AN |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br><br>.<br><br>. | ------------------------------------------------------------------------------------------------------------<br>Concession Collection Corporation Concession Corporation Compressional Concession Compressional Concessional Concession Compressional Concessional Concessional Concession Com<br>Commended College | ------------------<br>I<br><br>--------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br><br>------- | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>and the control and call call and also and all a collection and and any and<br>Million in all come of the charges of the driver and<br>------------------- |
| | | <br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br> | Company of Children College Court of Children Company of Children<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ |
{3}------------------------------------------------
| Angiplast Pvt. Ltd., Ahmedabad, India | | | Issue No.: 1<br>Date: January 1, 2001 | |
|---------------------------------------|----------------|-----------------------------------------------------------------------------------|------------------------------------------|--|
| Device Master File - 510(k) | | | Revision No.: 2<br>Date: January 1, 2002 | |
| I.V. ADMINISTRATION SET | | | | |
| 5 | Fluid filter | To filter parenteral solution | HDPE | |
| 6 | Tubing | To transfer the parenteral solution from bottle /<br>plastic container to patient | PVC using<br>DEHP<br>plasticizer | |
| 7 | Flow regulator | To control flow of infusion of parenteral solution | PP, HDPE | |
| 7 Flow regulator | To control flow of infusion of parenteral solution | PP, HDP |
|------------------------------------------|-----------------------------------------------------------------------------------------|---------|
| 8 Injection site | To administer extra medicament during infusion | Latex |
| 9 Male fitting (Adaptor) | To connect with female fitting used to infuse<br>parenteral solution | HDPE |
| 10 Protective cap of the male<br>fitting | To protect end of male fitting and to maintain<br>sterility of interior of infusion set | LDPE |
### Intended use of the Medical Device: 15:4
8 9
10
To administer parenteral fluids / medication into the patient's intravascular system.
- Summary of the technical characteristics of our device compared to legally marketed ાર રિંદર devices (two American brands)
- Summary of comparative technical characteristics 1.
From the comparison it is observed that our product is found to be substantially equivalent to Two similar American Products. The two American brands as well as our brand comply with the requirements of ISO 8536-4 in all critical requirements.
| -----------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br> | N. I. S. S. S. S. L. L. L. M. L. M. BELL MORE S. C.<br>American Article Article Article Antiques Article An<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>I MILLER BOOK EAR<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | <br>month of the charge of the charges of the company of the comments of the comments of the comments of the comments of the comments of the comments of the comments of the comme<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>1.000<br>AN A SENSION LINE I NO END SERVED MANAGE | A BREACH AND AND AND AND AND AND<br>.<br>------<br>.<br>A + A == + 4 == = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = =<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br> |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| And Advertising and<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>--------<br>Alle All Alle All All All All All All All All All All All All All All All All All All All All All All All All All All All All All All All All All All All All All All All All | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>Company of Concession and Concession of the consideration<br> |
| | | --------------------------------------------------------------------------------------------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>.<br>.<br>------- | I ER MELER BORLER<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>.<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird-like figure with three wing-like shapes.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR 5 2002
Mr. Ashit B. Shah Director Angiplast Private Limited 4803, Phase IV, G. I. D. C., Vatva Ahmedabad, INDIA
Re: K012189
Trade/Device Name: Angeltouch Vented IV Adminstration Set Regulation Number: 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: Il Product Code: FPA Dated: January 11, 2002 Received: January 15, 2002
Dear Mr. Shah:
We have reviewed your Section 510(k) premarket notification of intent to market the device We nuro roviewed your becamined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device mendments, or to devices that have been reclassified in accordance with the provisions of Amendinone, or to act roometic Act (Act) that do not require approval of a premarket the rederal I vou, Drag, and Ocamons , therefore, market the device, subject to the general approvisions of the Act. The general controls provisions of the Act include controls providents of a registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting (1 wr t), it may over in the Code of Federal Regulations, Title 21, Parts 800 to 898. In your device ear or roundsh further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
### Page 2 - Mr. Shah
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. or the For of all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4618 . Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and alle Net Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Timothy A. Ulatowski Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
| Angiplast Pvt. Ltd.; Ahmedabad, India | Issue No.: 1 |
|---------------------------------------|-------------------------|
| Device Master File-510(k) | Date: January 1, 2001 |
| I.V. ADMINISTRATION SET | Revision No.: 3 |
| | Date: February 20, 2002 |
# 12.0 STATEMENT OF INDICATIONS FOR USE
To administer parenticral fluids / medication into the patient's vascular system.
:: ….
The sterile I.V. Administration Set is indicated for use to administrater parenteral fluids medication into the paticits vascular system.
Patricio Cucente
(Division Sign-Off) (Division Sign-On)
Division of Dental, Infection Control, Division of a Hospital Devices 510(k) Number -----------10
| | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Company of the status and the status of the status of the status of the status of the status of the status of the status of the status of the status of the status of the stat | | |
| | | |
| -----------------------------------------------------------------------------------<br>Company of American Andrews of Children<br>contrasses of confire defense count<br>1 | .<br>.<br>------<br><br>.<br>Comprehibition of the works and would be works of the count of the comment of the comments of the comments of the contribution of the contribution of the contribution of the | |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.