Intravascular Extension Set

K210430 · Baxter Healthcare Corporation · FPA · Jun 18, 2021 · General Hospital

Device Facts

Record IDK210430
Device NameIntravascular Extension Set
ApplicantBaxter Healthcare Corporation
Product CodeFPA · General Hospital
Decision DateJun 18, 2021
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5440
Device ClassClass 2
AttributesTherapeutic, Pediatric

Indications for Use

For the administration of fluids from a container into a patient's vascular system through a vascular access device.

Device Story

Single-use, sterile, non-pyrogenic IV extension set; used to administer light-sensitive fluids/drugs from a syringe or primary IV set into a patient's vascular system. Components include non-DEHP PVC tubing, notch clamp, female Luer, and male Luer lock with filter vented cap. Operated by clinicians in clinical settings; connects to existing vascular access devices to extend reach or facilitate syringe-based delivery. Provides light resistance to protect sensitive medications. Benefits include safe, controlled fluid delivery and compatibility with standard Luer-locking systems.

Clinical Evidence

Bench testing only. No clinical data. Performance evaluated via ISO 80369-7 (Luer tests), BS EN ISO 8536-9 (tensile strength, leak testing), USP 788 (particulate matter), and light transmissivity testing. Biocompatibility assessed per ISO 10993-1 (cytotoxicity, sensitization, irritation, systemic toxicity, pyrogenicity, hemolysis). Sterility validated per ANSI/AAMI/ISO 11137-2.

Technological Characteristics

Single-use IV extension set. Materials: non-DEHP PVC tubing, PMMA female Luer, ABS male Luer, HDPE/nylon filter cap. Connectivity: standard Luer lock (ISO 80369-7). Sterilization: gamma radiation (SAL 10^-6). Shelf-life: 3 years. No software or electronic components.

Indications for Use

Indicated for the administration of fluids from a container into the vascular system of patients of all ages (neonatal, pediatric, and adult) via a vascular access device.

Regulatory Classification

Identification

An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.

Special Controls

*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION". June 18, 2021 Baxter Healthcare Corporation Meaghan Bonn Senior Specialist Regulatory Affairs 25212 West Illinois Route 120 Round Lake, Illinois 60073 Re: K210430 Trade/Device Name: Intravascular Extension Set (Light Resistant Extension Set) Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: Class II Product Code: FPA Dated: May 11, 2021 Received: May 21, 2021 #### Dear Meaghan Bonn: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part {1}------------------------------------------------ 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, CAPT Alan M. Stevens Acting Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ #### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration ### Indications for Use 510(k) Number (if known) K210430 Device Name Intravascular (IV) Extension Set (Light Resistant Extension Set) Indications for Use (Describe) For the administration of fluids from a container into a patient's vascular system through a vascular access device. Type of Use (Select one or both, as applicable) × | Prescription Use (Part 21 CFR 801 Subpart D) ] Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the word "Baxter" in a blue, sans-serif font. Below the word "Baxter" is the alphanumeric code "K210430" in a smaller, serif font. The code is left-aligned and positioned directly beneath the word "Baxter". #### Section 5. 510(k) Summarv February 10, 2021 #### OWNER: Baxter Healthcare Corporation One Baxter Parkway Deerfield, Illinois 60015 #### CONTACT PERSON: Meaghan Bonn Senior Specialist, Regulatory Affairs 25212 W. Illinois Route 120 Round Lake, IL 60073 Telephone: (224) 270 6470 ### IDENTIFICATION OF THE DEVICE: Trade/Device Name: Intravascular Extension Set (Light Resistant Extension Set) Classification Panel: 80 General Hospital Regulation Number: 21 CFR 880.5440 Regulation Name: Set, Administration Regulatory Class: Class II Product Code: FPA Image /page/3/Figure/10 description: The image shows a medical device labeled as 'Light Resistant Micro-Volume Extension Set, 59" (150 cm), Vol 0.47 mL' with the code 2N3363. The device consists of a red tube with several components labeled with numbers 1 through 6. These components are identified as 'Non-Vented Cap for Female Luer', 'Female Luer Lock', 'Notch Clamp', 'Light Resistant Microbore Tubing', 'Male Luer Lock', and 'Filter Vented Cap for Male Luer'. # Table 1. Proposed IV Extension Set Configuration {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font appears to be sans-serif. The word is slightly slanted to the right, giving it a dynamic appearance. The background is plain white. ## PREDICATE DEVICE: | Device | Company | Predicate 510(k) | Clearance Date | |-------------------------------------------------|----------------------------------|------------------------|----------------| | Intravascular Extension<br>Sets and Accessories | Baxter Healthcare<br>Corporation | K192366 (Model 2N3380) | July 20, 2020 | #### REASON FOR SUBMISSION: The basis for this premarket notification is the intent to market an Intravascular (IV) Extension Set (Light Resistant Extension Set). The proposed device is a single use disposable device, intended for the administration of fluids from a container into the patient's vascular system through a vascular access device. ### DESCRIPTION OF THE DEVICE: The proposed device consists of an IV Extension Set (Light Resistant Extension Set). It is a single use disposable, non-pyrogenic, sterile device intended for the administration of fluids from a container into the patient's vascular system. This IV extension set can be directly attached to a syringe but also used as an extension set to a primary set that is connected to a syringe (primarily to add length). See Figure 1 and Figure 2 for the clinical use set ups for the proposed device. The Light Resistant Extension Set consists of non-DEHP (<0.1% DEHP) PVC tubing, a notch clamp, female Luer, non-vented cap for female Luer, and filter vented cap for male Luer. It is used to administer fluids from a syringe into the patient's vascular system through a vascular access device, specifically fluids/drugs sensitive to light to the patients of all ages ranges - neonatal, pediatric, and adult. Figure 1. Clinical Use Set Up of Proposed Device when used directly with a Syringe as a Primary Syringe Set Image /page/4/Figure/10 description: The image shows a medical setup with several labeled components. A syringe, labeled as 'Syringe (Container),' is connected via a 'Female Luer Lock' to a tube. The tube has a 'Shut Off' valve in the middle and connects to a 'Male Luer Lock' on the other end, which is attached to a person's arm. {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font is sans-serif and the letters are slightly slanted to the right. The word is centered in the image and takes up most of the space. The background is white. #### Figure 2. Clinical Use Set Up of Proposed Device when used as an Extension to a Primary Syringe Set Image /page/5/Figure/2 description: This image shows a diagram of a medical setup for administering fluids or medication to a patient. The setup includes a syringe connected to a primary syringe set, which is then connected to an extension set. The extension set is connected to a male luer lock that connects to the patient's vascular access device. The diagram also labels a shut-off clamp and female luer locks that connect to the syringe and primary set. # INDICATIONS FOR USE: For the administration of fluids from a container into the patient's vascular system through a vascular access device. # TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE: The proposed device is substantially equivalent to the predicate device (model 2N3380), previously cleared under 510(k) premarket notification K192366 on July 20, 2020. The intended use and function of the proposed device is equivalent to the predicate device. Table 3 is a device comparison table outlining the differences between the predicate and proposed devices. | Features | Predicate Device<br>Cleared under<br>K192366 (2N3380) | Proposed Device | Assessment of Differences | |--------------------|---------------------------------------------------------------------------------------------------------------------------------------|-----------------|---------------------------| | Intended Use | For the administration<br>of fluids from a<br>container into the<br>patient's vascular<br>system through a<br>vascular access device | Same | N/A | | Indication for Use | For the administration<br>of fluids from a<br>container into the<br>patient's vascular<br>system through a<br>vascular access device. | Same | N/A | #### Table 3. Device Comparison {6}------------------------------------------------ Image /page/6/Picture/0 description: The image contains the word "Baxter" in a bold, blue font. The font is sans-serif and slightly italicized. The word is the only element in the image and is centered. | Features | Predicate Device<br>Cleared under<br>K192366 (2N3380) | Proposed Device | Assessment of Differences | |------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|-----------------|---------------------------| | Regulation<br>Number | 21 CFR 880.5440 | Same | N/A | | Product Code | FPA | Same | N/A | | Sterile | Yes | Same | N/A | | Non-Pyrogenic | Yes | Same | N/A | | Single Use | Yes | Same | N/A | | Length | 59" (150cm)<br>(2N3380) | Same | N/A | | Fluid Path Components/Materials | | | | | Tubing | Polyvinyl Chloride<br>(2N3380) | Same | N/A | | Female Luer Lock | Polymethyl<br>methacrylate (Acrylic)<br>(2N3380) | Same | N/A | | Male Luer Lock | Acrylonitrile<br>Butadiene Styrene<br>(2N3380) | Same | N/A | | Filter Vented Cap<br>for Male Luer | High Density<br>Polyethylene (Cap)<br>Hydrophobic Filter,<br>Acrylic (W/Non-<br>Woven Nylon<br>Substrate) (Filter<br>Membrane)<br>(2N3380) | Same | N/A | Table 3. Device Comparison {7}------------------------------------------------ Image /page/7/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font is slightly italicized, giving it a dynamic appearance. The word is the logo for the healthcare company Baxter International. # DISCUSSION OF NONCLINICAL TESTS: Baxter Healthcare Corporation conducts risk analyses and design verification tests based on the result of these analyses. All test results meet their acceptance criteria and support that the proposed device is appropriately designed for its intended use. # Performance Data: The following bench tests (Table 4) were conducted to evaluate the functional performance of the proposed device: | Test | Acceptance Criteria | |------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | ISO 80369-7 Luer Tests on male Luer Lock Connector | ISO 80369-7:2016, Clause 6.1.2 or<br>6.1.3,<br>ISO 80369-7:2016, Clause 6.2,<br>ISO 80369-7:2016, Clause 6.3,<br>ISO 80369-20:2015, Annex E,<br>ISO 80369-7:2016, Clause 6.4,<br>ISO 80369-7:2016, Clause 6.5,<br>ISO 80369-7:2016, Clause 6.6,<br>ISO 80369-7:2016, Clause 5 | | ISO 80369-7 Luer Tests on female Luer Lock Connector | ISO 80369-7:2016, Clause 6.1.2 or<br>6.1.3,<br>ISO 80369-7:2016, Clause 6.2,<br>ISO 80369-7:2016, Clause 6.3,<br>ISO 80369-20:2015, Annex E,<br>ISO 80369-7:2016, Clause 6.4,<br>ISO 80369-7:2016, Clause 6.5,<br>ISO 80369-7:2016, Clause 6.6,<br>ISO 80369-7:2016, Clause 5 | | Tensile Strength Test | BS EN ISO 8536-9, Clause 5.3 | | Leak Test | BS EN ISO 8536-9: 2015, Section A.4 | | Notch Clamp Activation Force Test | Per Baxter Test Method | | Notch Clamp Shut-Off Test | Per Baxter Test Method | | Non-DEHP Claim Verification (< 0.1% DEHP) | Per Baxter Test Method (as tested in K161808) | | Particulate Matter Test | USP 788 | | Light Transmissivity Test | Per Baxter Test Method | # Table 4. Performance Data {8}------------------------------------------------ Image /page/8/Picture/0 description: The image shows the word "Baxter" in a bold, sans-serif font. The word is in blue and appears to be a logo. The letters are closely spaced and have a slight italicized appearance. All tests met the acceptance criteria. #### Biocompatibility: Biocompatibility assessments were conducted based on ISO-10993-1, Biological Evaluation of Medical Devices for prolonged duration, external communicating device, indirect blood path and FDA-2013-D-0350 Guidance for Industry and FDA Staff, "Use of International Standard ISO-10993-1, 'Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process," as recommended in the Intravascular Administration Sets guidance, "Guidance for Industry and FDA Staff: Intravascular Administration Sets Premarket Notification Submissions [510(k)] " Biocompatibility assessments were conducted on a worst case/representative final, finished device for all fluid path materials of the proposed device. The following tests were conducted as part of the biocompatibility assessment for the proposed device: - Cytotoxicity ISO 10993-5 - Sensitization ISO 10993-10 - Intracutaneous (Irritation) Reactivity ISO 10993-10 - Systemic Toxicity (acute and repeat dose) ISO 10993-11 - Materials Mediated Pyrogen ISO 10993-11 - Hemolysis ISO 10993-4 and ASTMF756 Based upon the results of the data supports the ISO 10993-1 categorization of external communicating device, indirect blood path, prolonged contact duration. The proposed device is biocompatible and appropriate for its intended use. ### Sterility: The proposed device is sterilized with gamma radiation. The product is in the bioburden (sub) category "General Sets Labeled Sterile". The Minimum Sterilizing Dose (MSD) required to provide a 106 Sterility Assurance Level (SAL) for this (sub) category was established and validated at the manufacturing facility as described in ANSI/AAMI/ISO 11137-2. "Sterilization of health care products – Radiation-Part 2: Establishing the sterilization dose." The dose setting method used includes, but is not limited to, Method 1 or VDmax. Generally, the MSDs are between 14.2 - 25.0 kGy. The continued validity of the MSD for this (sub) category is confirmed via periodic dose audit studies. In addition, routine periodic pre-sterilization bioburden testing is performed for each (sub) category. {9}------------------------------------------------ Image /page/9/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font is slightly italicized, giving the word a sense of movement. The word is the logo for the healthcare company Baxter International. ### Shelf-Life: Baxter has performed aging testing to support a shelf-life claim of 3 (three) years. ## Microbial Ingress Testing: Baxter has conducted testing on all the potential points of microbial entry into the sterile fluid pathway of the proposed device (per Baxter test method - as tested in K192366). The potential microbial entry points consist of male and female Luer connector sites. All test results meet their acceptance criteria and support that the proposed device is appropriately designed for its intended use. # CONCLUSION: The non-clinical data demonstrate that the subject device is substantially equivalent and performs comparably to the predicate device that is legally marketed for the same intended use.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...