For the administration of fluids from a container into the patient's vascular system through a vascular access device.
Device Story
Intravascular (IV) extension sets and accessories facilitate fluid administration from a container to a patient's vascular system. Components include PVC or polyethylene-lined PVC tubing, notch clamps, luer locks, and caps. Accessories include anti-siphon valves (prevent free flow/backflow), back check valves (prevent backflow), and 1.2 µm filters (particulate/air removal). Devices are single-use, sterile, and non-pyrogenic. Used in clinical settings by healthcare providers to deliver solutions, drugs, antibiotics, or lipids. They connect to vascular access devices, potentially using syringes. The system ensures controlled fluid delivery and reduces risks associated with infusion, such as air embolism or backflow.
Clinical Evidence
Bench testing only. No clinical data. Performance testing included ISO 80369-7 luer testing, tensile strength, leak testing, counter flow, blocking performance, opening pressure, clamp activation/shut-off, and filter testing. Biocompatibility was evaluated per ISO 10993-1, including cytotoxicity, sensitization, irritation, systemic toxicity, hemolysis, and extractable/leachable profiles. Sterility validated per ANSI/AAMI/ISO 11137-2 (gamma radiation). Microbial ingress and USP 788 particulate matter testing were also performed.
Technological Characteristics
Single-use IV extension sets and accessories. Materials: PVC, polyethylene-lined PVC, polymethyl methacrylate (acrylic), silicone, polyethersulfone, polytetrafluoroethylene, and high-density polyethylene. Components: tubing, notch clamps, luer locks, anti-siphon/back check valves, and 0.2/1.2 µm filters. Sterilization: Gamma radiation (ANSI/AAMI/ISO 11137-2). Connectivity: Standard luer lock interfaces. No software or electronic components.
Indications for Use
Indicated for the administration of fluids from a container into the patient's vascular system through a vascular access device.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The FDA text logo is in blue and includes the acronym "FDA" followed by "U.S. FOOD & DRUG ADMINISTRATION" in a stacked format.
July 20, 2020
Baxter Healthcare Corporation Gary Chumbimune Associate Director, Regulatory Affairs 32650 N. Wilson Road Round Lake, Illinois 60073
Re: K192366
Trade/Device Name: Intravascular Extension Sets and Accessories Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: Class II Product Code: FPA, FPB Dated: June 16, 2020 Received: June 19, 2020
Dear Gary Chumbimune:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for Payal Patel Acting Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices. and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K192366
#### Device Name
Intravascular Extension Sets and Accessories
Indications for Use (Describe)
For the administration of fluids from a container into the patient's vascular system through a vascular access device.
Type of Use (Select one or both, as applicable)
| <span style="font-family: DejaVu Sans, sans-serif">✔</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: DejaVu Sans, sans-serif">▢</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
|----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font is sans-serif and has a slightly italicized appearance. The word is the logo for the healthcare company Baxter International.
## 510(k) Summary-K192366
July 20, 2020
### OWNER:
Baxter Healthcare Corporation One Baxter Parkway Deerfield, Illinois 60015
# CONTACT PERSON:
Gary Chumbimune Associate Director, Regulatory Affairs 32650 N. Wilson Road Round Lake, IL 60073 Telephone: (224) 270 3312 Fax: (224) 270 4119
# IDENTIFICATION OF THE DEVICE:
Trade/Device Name: Intravascular Extension Sets and Accessories Classification Panel: 80 General Hospital Regulation Number: 21 CFR 880.5440 Regulation Name : Set, Administration, Intravascular Regulatory Class: Class II Product Code: FPA, FPB
| Code # | Device Description |
|------------|-------------------------------------------------------|
| 2N335<br>4 | Micro-Volume Extension Set, 79" (200 cm), Vol 0.47 mL |
| 1: Non-Vented Cap |
|---------------------|
| 2: Female Luer Lock |
| 3: Notch Clamp |
| 4: Tubing |
| 5: Male Luer Lock |
| 6: Male Luer Cap |
### Table 1. Proposed IV Extension Set Configurations
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "Baxter" in a bold, sans-serif font. The word is in blue and appears to be a logo. The letters are evenly spaced and the overall design is simple and modern. The image is clear and well-defined.
| Code # | Device Description | |
|------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|
| 2N338<br>0 | Micro-Volume Extension Set, 59" (150 cm), Vol 0.40 mL<br>Image: [Micro-Volume Extension Set, 59" (150 cm), Vol 0.40 mL] | 1: Non-Vented Cap<br>2: Female Luer<br>Lock<br>3: Notch Clamp<br>4: Tubing<br>5: Male Luer Lock<br>6: Male Luer Cap |
| 2N335<br>8 | Micro-Volume Extension Set, 59" (150 cm), Vol 0.40 mL<br>Image: [Micro-Volume Extension Set, 59" (150 cm), Vol 0.40 mL] | 1: Non-Vented Cap<br>2: Female Luer<br>Lock<br>3: Tubing<br>4: Male Luer Lock<br>5: Male Luer Cap |
| 2N335<br>7 | Micro-Volume Extension Set with 0.2 µm Filter, 79" (200 cm), Vol 0.88 mL<br>Image: [Micro-Volume Extension Set with 0.2 µm Filter, 79" (200 cm), Vol 0.88 mL] | 1: Non-Vented Cap<br>2: Female Luer<br>Lock<br>3, 6: Tubing<br>4: 0.2µm Filter<br>5: Notch Clamp<br>7: Male Luer Lock<br>8: Male Luer Cap |
| 2N338<br>2 | Micro-Volume Extension Set with 0.2 µm Filter, 59" (150 cm), Vol 0.75 mL<br>Image: [Micro-Volume Extension Set with 0.2 µm Filter, 59" (150 cm), Vol 0.75 mL] | 1: Non-Vented Cap<br>2: Female Luer<br>Lock<br>3, 6: Tubing<br>4: 0.2µm Filter<br>5: Notch Clamp<br>7: Male Luer Lock<br>8: Male Luer Cap |
| 2N338<br>6 | Micro-Volume Catheter Extension Set with 0.2 µm Filter, 10" (25 cm), Vol 0.44 mL<br>Image: [Micro-Volume Catheter Extension Set with 0.2 µm Filter, 10" (25 cm), Vol 0.44 mL] | 1: Non-Vented Cap<br>2: Female Luer<br>Lock<br>3, 6: Tubing<br>4: 0.2µm Filter<br>5: Notch Clamp<br>7: Male Luer Lock<br>8: Male Luer Cap |
| 2N335<br>5 | Mini-Volume Extension Set, 59" (150 cm), Vol 0.91 mL<br>Image: [Mini-Volume Extension Set, 59" (150 cm), Vol 0.91 mL] | 1: Non-Vented Cap<br>2: Female Luer<br>Lock<br>3: Notch Clamp<br>4: Tubing |
# Table 1. Proposed IV Extension Set Configurations
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The word is slightly italicized, giving it a sense of movement. The font is sans-serif and the letters are evenly spaced.
Image /page/5/Figure/1 description: The image shows a table with different medical devices and their descriptions. The first device is a Mini-Volume Extension Set, 79" (200 cm), Vol 1.1 mL, with code 2N335 6. The second device is a Mini-Volume Extension Set, 118" (300 cm), Vol 1.4 mL, with code 2N335 9. The third device is a Polyethylene (PE) Lined Micro-Volume Extension Set, 59" (150 cm), Vol 0.40 mL, with code 2N336 0.
#### Table 1. Proposed IV Extension Set Configurations
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font is sans-serif and has a slightly italicized appearance. The word is the logo for the healthcare company Baxter International.
| Code # | Device Description | |
|--------|------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|
| 2N3364 | Anti-Siphon Valve, Vol 0.24 mL<br>Image: Anti-Siphon Valve | 1: Non-Vented Cap to Female Luer<br>2: Anti-Siphon Valve<br>3: Touch Contamination Protector |
| 2N3365 | Back Check Valve, Vol 0.28 mL<br>Image: Back Check Valve | 1: Non-Vented Cap to Female Luer<br>2: Back Check Valve<br>3: Touch Contamination Protector |
| 2N3368 | Air-Eliminating 1.2 μm Solution Filter, Vol 0.80 mL<br>Image: Air-Eliminating 1.2 μm Solution Filter | 1: Non-Vented Cap to Female Luer<br>2: 1.2 μm Filter<br>3: Touch Contamination Protector |
### Table 2. Proposed IV Set Accessories
# PREDICATE DEVICE:
### Table 3. Predicate De vice
| Device | Company | Predicate 510(k) | Clearance Date |
|-----------------------------------------------------------------------------------------------|----------------------------------|------------------|------------------|
| Clearlink Luer<br>Activated Valve,<br>Clearlink SystemNon-<br>DEHP Catheter<br>Extension Sets | Baxter Healthcare<br>Corporation | K112893 | October 18, 2011 |
## REASON FOR SUBMISSION:
The basis for this premarket notification is the intend to market Intravascular (IV) Extension Sets and Accessories. The proposed devices in this submission are single-use, disposable devices, intended for the administration of fluids from a container into the patient's vascular system through a vascular access device.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font is sans-serif and the letters are slightly italicized. The word is likely a logo or brand name.
### DESCRIPTION OF THE DEVICE:
The proposed devices consist of IV Extension sets and IV set accessories. They are single use disposable devices intended for the administration of fluids from a container into the patient's vascular system. They are non-pyrogenic, sterile devices that can be used with or without a syringe.
The extension sets consist of PVC tubing or polyethylene lined PVC tubing, a notch clamp, female luer, non-vented cap, male luer, and filter vented cap. They are used to administer solutions, drugs, antibiotics, lipids to the patient.
The accessories consists of an anti-siphon valve, back check valve, and 1.2 µm Filter. They attach to the proposed sets to add a specific feature to facilitate the administration of fluid when used with a syringe. The anti-siphon valve reduces the risk of free flow from the syringe and backflow into the primary infusion line. The back check valve prevents backflow into the primary infusion line. The 1.2 um Filter prevents particulate matter and eliminates air bubbles.
#### INDICATIONS FOR USE:
For the administration of fluids from a container into the patient's vascular system through a vascular access device.
# TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE:
The proposed devices are substantially equivalent to the predicate device, previously cleared under 510(k) premarket notification K112893 on October 18, 2011. The intended use and function of the proposed devices are equivalent to the predicate device.
Table 4 is a device comparison table outlining the differences between the predicate and proposed devices.
| Features | Predicate Device<br>Cleared under<br>K112893 | Proposed Devices | Assessment of Differences |
|--------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | For use with a vascular<br>access device for the<br>administration of drugs<br>and solutions. The<br>Clearlink Luer<br>Activated Valve is an<br>in-line injection site. | For the administration<br>of fluids from a<br>container into the<br>patient's vascular<br>system through a<br>vascular access device. | Same<br>Minor rewording of the Intended<br>Use statement has been made to<br>better align with 21 CFR<br>880.5440 and for the purpose of |
| | Predicate Device<br>Cleared under<br>K112893 | Proposed Devices | Assessment of Differences |
| Features | which can be<br>connected to the<br>standard male Luer<br>adapters (e.g., syringes<br>or sets) for continuous<br>or intermittent fluid<br>administration or the<br>withdrawal of fluid. | | streamlining the information<br>provided to the user.<br>The general purpose of the<br>device and its function remain<br>unchanged.<br>The minor rewording of the<br>Intended Use statement does not<br>raise different questions of safety<br>and effectiveness. |
| Indication for<br>Use | For use with a vascular<br>access device for the<br>administration of drugs<br>and solutions. The<br>Clearlink Luer<br>Activated Valve is an<br>in-line injection site,<br>which can be<br>connected to the<br>standard male Luer<br>adapters (e.g., syringes<br>or sets) for continuous<br>or intermittent fluid<br>administration or the<br>withdrawal of fluid. | For the administration<br>of fluids from a<br>container into the<br>patient's vascular<br>system through a<br>vascular access device. | Same<br>Minor rewording of the<br>Indications for Use statement has<br>been made to better align with 21<br>CFR 880.5440 and for the<br>purpose of streamlining the<br>information provided to the user.<br>This minor modification does not<br>alter the disease or condition the<br>device will diagnose, treat,<br>prevent, cure/mitigate, or the<br>patient population for which the<br>device is intended to be used. In<br>addition, the minor rewording<br>does not reflect a different<br>anatomical site from which a<br>disease state or population may<br>be inferred.<br>The minor rewording of the<br>Intended Use statement does not<br>raise different questions of safety<br>and effectiveness. |
| Sterile | Yes | Same | N/A |
| Non-<br>Pyrogenic | Yes | Same | N/A |
| Single Use | Yes | Same | N/A |
| Fluid path Components/Materials | | | |
| Anti-Siphon<br>Valve | Not Applicable | Polymethyl<br>methacrylate (Acrylic)<br>[Female and Male Luer]<br>Silicone (Membrane) | The predicate device does not<br>have an anti-siphon check valve.<br>Design control activities have<br>been conducted and have<br>confirmed that the different<br>technological characteristics of<br>the proposed devices do not raise |
| Features | Predicate Device<br>Cleared under<br>K112893 | Proposed Devices | Assessment of Differences |
| Back Check<br>Valve | Not Applicable | (2N3364)<br>Polymethyl<br>methacrylate (Acrylic)<br>[Female and Male Luer]<br>Silicone (Membrane)<br>(2N3365) | different questions of safety and<br>effectiveness<br>The predicate device does not<br>have a check valve. Design<br>control activities have been<br>conducted and have confirmed<br>that the different technological<br>characteristics of the proposed<br>devices do not raise different<br>questions of safety and<br>effectiveness |
| 0.2 μm Filter | Not Applicable | Polymethyl<br>methacrylate (Acrylic)<br>[Housing]<br>Polyvinylidene fluoride<br>(Air Vent Membrane)<br>Polyethersulfone<br>(Solution Membrane)<br>(2N3382, 2N3357,<br>2N3386) | The predicate device does not<br>have a 0.2 μm Filter. These<br>materials have been used in<br>another Baxter cleared device<br>(cleared in K113227, 12/08/11)<br>with the same/similar intended<br>use and with the same type and<br>duration of contact. Design<br>control activities have been<br>conducted and have confirmed<br>that the different technological<br>characteristics of the proposed<br>devices do not raise different<br>questions of safety and<br>effectiveness. |
| Air-<br>Eliminating<br>1.2 μm<br>Solution Filter | Not Applicable | Polymethyl<br>methacrylate (Acrylic)<br>[Housing]<br>Polytetrafluoroethylene<br>(Air Vent Membrane)<br>Polyethersulfone<br>(Solution Membrane)<br>(2N3368) | The predicate device does not<br>have a 1.2 μm Filter. Design<br>control activities have been<br>conducted and have confirmed<br>that the different technological<br>characteristics of the proposed<br>devices do not raise different<br>questions of safety and<br>effectiveness. |
| Tubing | Polyvinyl Chloride<br>(2N8374) | Same<br>(2N3358, 2N3380,<br>2N3354, 2N3382,<br>2N3357, 2N3355,<br>2N3356, 2N3359,<br>2N3386) | N/A |
| Features | Predicate Device<br>Cleared under<br>K112893 | Proposed Devices | Assessment of Differences |
| Tri-Layer<br>Tubing | Not Applicable | Low-Density<br>Polyethylene<br><br>Ethylene vinyl acetate<br><br>Polyvinyl chloride<br><br>(2N3360, 2N3361,<br>2N3362) | The predicate device does not<br>have the same type of material<br>formulation. Design control<br>activities have been conducted<br>and have confirmed that the<br>different technological<br>characteristics of the proposed<br>devices do not raise different<br>questions of safety and<br>effectiveness. |
| Female Luer<br>Lock | Copolyester<br>(2N8374) | Polymethyl<br>methacrylate (Acrylic)<br><br>(2N3358, 2N3380,<br>2N3354, 2N3382,<br>2N3357, 2N3355,<br>2N3356, 2N3359,<br>2N3360, 2N3361,<br>2N3362, 2N3386) | The predicate device does not<br>have the same type of material<br>formulation. Design control<br>activities have been conducted<br>and have confirmed that the<br>different technological<br>characteristics of the proposed<br>devices do not raise different<br>questions of safety and<br>effectiveness. |
| Male Luer<br>Lock | Acrylonitrile<br>Butadiene Styrene<br>(2N8374) | Same<br>(2N3358, 2N3380,<br>2N3354, 2N3382,<br>2N3357, 2N3355,<br>2N3356, 2N3359,<br>2N3360, 2N3361,<br>2N3362, 2N3386) | N/A |
| Male Luer<br>Cap | Polypropylene<br>(2N8374) | High Density<br>Polyethylene (Cap)<br><br>Hydrophobic Filter,<br>Acrylic (W/Non-Woven<br>Nylon Substrate) (Filter<br>Membrane)<br><br>(2N3358, 2N3380,<br>2N3354, 2N3382,<br>2N3357, 2N3355,<br>2N3356, 2N3359,<br>2N3360, 2N3361,<br>2N3362, 2N3386) | The predicate device does not<br>have the same type of material.<br>Design control activities have<br>been conducted and have<br>confirmed that the different<br>technological characteristics of<br>the proposed devices do not raise<br>different questions of safety and<br>effectiveness. |
Table 4. Device Comparison
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font is sans-serif and appears to be slightly italicized. The word is the main focus of the image and is positioned horizontally.
Table 4. Device Comparison
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font appears to be sans-serif. The word is slightly italicized, giving it a sense of movement.
# Table 4. Device Comparison
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The word is slightly blurred, giving it a sense of depth. The font is sans-serif and appears to be a company logo.
# Table 4. Device Comparison
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font appears to be sans-serif. The word is slightly blurred, suggesting that the image may have been taken from a distance or with a slightly out-of-focus lens. The background is white.
# DISCUSSION OF NONCLINICAL TESTS:
Baxter Healthcare Corporation conducts risk analyses and design verification tests based on the result of these analyses. All test results meet their acceptance criteria and support that the proposed devices are appropriately designed for their intended use.
# Performance Data:
The following bench tests (Table 5) were conducted to evaluate the functional performance of the proposed devices:
| Test | Acceptance Criteria |
|---------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ISO 80369-7 Luer Tests on male Luer Lock Connector | BS EN ISO 80369-7:2016, Clause 6.1.2,<br>BS EN ISO 80369-7:2016, Clause 6.1.3,<br>ISO 80369-7:2016, Clause 6.2,<br>ISO 80369-7:2016, Clause 6.3,<br>ISO 80369-20:2015 AnnexE,<br>ISO 80369-7:2016, Clause 6.1.2,<br>ISO 80369-7:2016, Clause 6.1.3,<br>ISO 80369-7:2016, Clause 6.4,<br>ISO 80369-7:2016, Clause 6.5,<br>ISO 80369-7:2016, Clause 6.6,<br>ISO 80369-7: 2016, Clause 5,<br>ISO 594-1:1986, Clause 3,<br>ISO 594-2:1998, Clause 3 (as applicable) |
| ISO 80369-7 Luer Tests on female Luer Lock<br>Connector | BS EN ISO 80369-7:2016, Clause 6.1.2,<br>BS EN ISO 80369-7:2016, Clause 6.1.3,<br>ISO 80369-7:2016, Clause 6.2,<br>ISO 80369-7:2016 Clause 6.3,<br>ISO 80369-20:2015 AnnexE,<br>ISO 80369-7:2016, Clause 6.1.2,<br>ISO 80369- 7:2016, Clause 6.1.3,<br>ISO 80369-7:2016, Clause 6.4,<br>ISO 803697:2016, Clause 6.5.<br>ISO 80369-7:2016, Clause 6.6,<br>ISO 80369-7:2016, Clause 5 (as applicable),<br>ISO 594-1:1986, Clause 3,<br>ISO 594-2:1998, Clause 3 |
| Tensile Strength Test | BS EN ISO 8536-9, Clause 5.3,<br>BS EN ISO 8536-10, Clause 4.3,<br>BS EN ISO 8536-11, Clause 5.3,<br>BS EN ISO 8536-12, Clause 6.2, |
| Leak Test | BS EN ISO 8536-9: 2015, Section A.4,<br>BS EN ISO 8536-11:2015, Section A.4,<br>ISO 8536-10: 2015, section A.4 |
| Counter Flow Test | ISO 8536-12:2007+A1:2013, Clause A.4. |
# Table 5. Performance Data
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font appears to be sans-serif. The word is slightly italicized, giving it a sense of movement. The background is plain white, which makes the blue text stand out.
| Test | Acceptance Criteria |
|------------------------------------------|----------------------------------------------------------------------|
| Blocking Performance Test | ISO 8536-12:2007+A1:2013, Clause A.6. |
| Opening Pressure Test | ISO 8536-12:2007+A1:2013, Clause A.7.1<br>and per Baxter Test Method |
| Notch Clamp Activation Force Test | Per Baxter Test Method |
| Notch Clamp Shut-Off Test | ISO 8536-14:2015 |
| Filter Tests | EN ISO 8536-11, Clause 5.1<br>and per Baxter Test Method |
| Non-DEHP Claim Verification (<0.1% DEHP) | Per Baxter Test Method (as tested in<br>K161808) |
### Table 5. Performance Data
All tests met the acceptance criteria.
### Biocompatibility:
Biocompatibility assessments were conducted based on ISO-10993-1, Biological Evaluation of Medical Devices for prolonged duration, external communicating device, indirect blood path and FDA-2013-D-0350 Guidance for Industry & FDA Staff, Use of ISO 10993-1, "Biological evaluation and medical devices - Part 1: Evaluation and testing within a risk management process, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing",as recommended in the Intravascular Administration Sets guidance, "Guidance for Industry and FDA Staff: Intravascular Administration Sets Premarket Notification Submissions [510(k)]". Biocompatibility assessments were conducted on a worst case/representative final, finished device for all materials of the proposed devices. The following tests were conducted as part of the biocompatibility testing for the proposed devices' fluid path:
- Cytotoxicity ISO 10993-5
- Sensitization ISO 10993-10
- Intracutaneous (Irritation) Reactivity ISO 10993-10
- Systemic Toxicity (acute dose) ISO 10993-11
- Toxicological Assessment of Extractable Profile ISO 10993-17
- Materials Mediated Pyrogen ISO 10993-11
- Hemolysis ISO 10993-4
- Extractable Assessment ISO 10993-18
{13}------------------------------------------------
Image /page/13/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font appears to be sans-serif. The word is slightly italicized, giving it a sense of movement. The background is plain white, which makes the blue text stand out.
- . 14 Day Systemic Repeat Dose Toxicity Study ISO 10993-11
Based upon the results of this prolonged duration, external communicating, indirect blood path testing, the proposed devices have been shown to be biocompatible and appropriate for their intended use.
## Sterility:
The proposed devices are sterilized with gamma radiation. The products are in the bioburden (sub) category "General Sets Labeled "Sterile" and "General Small Devices Labeled 'Sterile''. The Minimum Sterilizing Dose (MSD) required to provide a 106 Sterility Assurance Level (SAL) for this (sub) category was established and validated at the manufacturing facility as described in ANSI/AAMI/ISO 11137-2, Sterilization of health care products-Radiation-Part 2: Establishing the sterilization dose. The dose setting method used includes, but is not limited to. Method 1 or VDmax. Generally, the MSDs are between 14.2 - 25.0 kGy. The continued validity of the MSD for this (sub) category is confirmed via periodic dose audit studies. In addition. routine periodic presteriilzation bioburden testing is performed for each (sub) category.
## Shelf-Life :
Baxter has performed aging testing to support a shelf-life claim of 3 (three) years.
# Microbial Ingress Testing:
Baxter has conducted testing on all potential points of microbial entry into the sterile fluid pathway of the proposed devices subject to this premarket notification. The Luer Connector Sites were tested following Baxter's test method. All test results meet their acceptance criteria and support that the proposed devices are appropriately designed for their intended use.
## Particulate Matter Testing
Baxter has performed particulate matter testing per <USP 788> on the proposed devices subject to this premarket notification. All test results meet their acceptance criteria and support that the proposed devices are appropriately designed for their intended use.
# CONCLUSION:
The non-clinical data demonstrate that the subject devices are substantially equivalent and perform comparably to the predicate device that is legally marketed for the same intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.