Steripath Micro Blood Collection System

K200661 · Magnolia Medical Technologies, Inc. · FPA · Oct 8, 2020 · General Hospital

Device Facts

Record IDK200661
Device NameSteripath Micro Blood Collection System
ApplicantMagnolia Medical Technologies, Inc.
Product CodeFPA · General Hospital
Decision DateOct 8, 2020
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 880.5440
Device ClassClass 2

Indications for Use

The Steripath® Micro Blood Collection System is a system to draw blood for in vitro diagnostic testing. The Steripath® Micro Blood Collection System is indicated for use as a blood collection system that diverts and sequesters the initial specimen prior to collection of a subsequent test sample to reduce the frequency of blood culture contamination when contaminants are present in the initial blood sample compared to blood cultures drawn with standard procedure without manual diversion. Additionally, components of the system may be used for infusion following sample collection after disconnection of the Initial Specimen Diversion Device® (ISDD®).

Device Story

Steripath® Micro is a single-use, sterile, mechanical blood collection system; diverts/sequesters initial 0.6mL–0.9mL of blood (potentially contaminated) into a reservoir; subsequent blood sample flows through a secondary pathway into a syringe for culture bottle inoculation. Designed for patients with limited blood draw volumes or difficult venous access. Operated by clinicians in clinical settings. After ISDD® removal, system components can be used for infusion. Benefits include reduced blood culture contamination rates compared to standard procedures.

Clinical Evidence

Bench testing only. Verification and validation activities confirmed performance against predetermined acceptance criteria. Testing included sequestration performance, vacuum performance, diversion volume accuracy, and positive pressure integrity. Biocompatibility testing (ISO 10993-1) included cytotoxicity, sensitization, acute systemic toxicity, pyrogenicity, and hemocompatibility. Packaging and shelf-life (1-year) validated per ASTM standards. Literature (Patton et al.) referenced for diversion volume efficacy.

Technological Characteristics

Single-use, sterile, mechanical device. Materials: medical-grade plastics (polycarbonate, TPE), PVC tubing, medical-grade adhesives, silicone. Dimensions/form factor: ISDD® base assembly with Luer extension and syringe interface. Energy source: None (mechanical). Sterilization: Gamma radiation (ISO 11137-1). Connectivity: None. Software: None.

Indications for Use

Indicated for patients requiring blood collection for in vitro diagnostic testing, specifically those where maximum allowable total blood draw volume is limited or patients with difficult venous access. Used to reduce blood culture contamination. Components may be used for infusion after ISDD® disconnection.

Regulatory Classification

Identification

An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.

Special Controls

*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. October 8, 2020 Magnolia Medical Technologies Inc. Gregory J. Bullington CEO 200 West Mercer Street, Suite 500 Seattle, Washington 98119 Re: K200661 Trade/Device Name: Steripath® Micro Blood Collection System Regulation Number: 21 CFR 862.1675 Regulation Name: Blood Specimen Collection Device Regulatory Class: Class II Product Code: JKA. FPA Dated: September 10, 2020 Received: September 10, 2020 Dear Gregory J. Bullington: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. for Payal Patel Acting Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K200661 Device Name Steripath® Micro Blood Collection System #### Indications for Use (Describe) The Steripath® Micro Blood Collection System is a system to draw blood for in vitro diagnostic testing. The Steripath® Micro Blood Collection System is indicated for use as a blood collection system that diverts and sequesters the initial specimen prior to collection of a subsequent test sample to reduce the frequency of blood culture contamination when contaminants are present in the initial blood sample compared to blood cultures drawn with standard procedure without manual diversion. Additionally, components of the system may be used for infusion following sample collection after disconnection of the Initial Specimen Diversion Device® (ISDD®). X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## K200661 510(k) Summary In accordance with 21 CFR 807.92(c) the following summary information is provided: 510(k) Number: K200661 Date Prepared: September 10, 2020 Submitter: Magnolia Medical Technologies, Inc. 200 West Mercer Street Suite 500 Seattle WA 98119 Registration number: 3009976527 ## Contact Person: Greg Bullington CEO Phone: (888) 617-3420 regulatory@magnolia-medical.com Trade Name: Steripath® Micro Blood Collection System Common Name: Blood collection set Regulation Name: Blood Specimen Collection Device Regulation Number: 862.1675 Regulatory Class: Class II Product Code: JKA and FPA Predicate Device: K192247-Steripath® Gen2 Blood Collection System K192247, Product codes JKA & FPA, Class II Device Description: The Steripath® Micro Blood Collection System diverts and sequesters the initial portion of the blood specimen (potentially contaminated blood) in the diversion reservoir. When diversion is complete, a subsequent blood sample flows through a second pathway within the device. The subsequent blood sample is collected into a syringe that is used to inoculate culture bottles. Upon removal of the Initial Specimen Diversion Device®, ISDD®, components of the system can be used for infusion per the included manufacturer's instructions for use. The subject device incorporates multiple syringe oulet accessory configurations that include and three sizes of syringes that are previously cleared as referenced below. The Steripath® Micro Blood Collection System are available: | Steripath®<br>Micro Kit<br>Model<br>Number | ISDD® | Inlet Accessory | Outlet Accessory | |--------------------------------------------|--------|----------------------------------------------------------------------|--------------------------------------------------------------| | 4005-EN | P00353 | Luer Extension, 7"<br>ICU Medical, Inc.<br>Model B1754-NS<br>K964435 | Syringe, 5ml<br>Becton Dickinson<br>Model 301027<br>K980987 | | 4010-EN | P00353 | Luer Extension, 7"<br>ICU Medical, Inc.<br>Model B1754-NS<br>K964435 | Syringe, 10ml<br>Becton Dickinson<br>Model 301029<br>K980987 | | 4020-EN | P00353 | Luer Extension, 7"<br>ICU Medical, Inc.<br>Model B1754-NS<br>K964435 | Syringe, 20ml<br>Becton Dickinson<br>Model 301031<br>K980987 | Table 1: Steripath® Micro Configurations Available #### Intended Use/Indications for Use: The Steripath® Micro Blood Collection System is a system to draw blood for in vitro diagnostic testing. The Steripath® Gen2 Blood Collection System is indicated for use as a blood collection system that diverts and sequesters the initial specimen prior to collection of a subsequent test sample to requency of blood culture contamination when contaminants are present in the initial blood sample compared to blood cultures drawn with standard procedure without manual diversion. Additionally, components of the system may be used for infusion following sample collection after disconnection of the Initial Specimen Diversion Device® (ISDD®). {4}------------------------------------------------ ## Differences in Intended Use/Indications for Use The Intended Use / Indications for use are the predicate device except that the Micro device removes the venipuncture needles and associated 2 hour limitation on infusion. ## Technology: The Steripath® Micro Blood Collection System is a single use, sterile, mechanical device that initial 0.6mL to 0.9mL of blood from the patient. The Steripath® Micro is designed for patients where maximum allowable total blood draw volume is limited or for patients with difficult venous access. The system consists of an Initial Specimen Diversion Device® (ISDD®) made of injection molded, medical grade plastics. Off-the-Shelf (OTS) components provide interface to the patient vasculature (via a peripheral IV start not provided by Magnolia) and sample collection is achieve by a syringe. Upon removal of the ISDD®, components of the system can be used for infusion per the included manufacturer's instructions for use. Differences between the Steripath® Micro Blood Collection System and the Predicate Steripath® Gen2 Blood Collection System are noted in Table 2 below. | ltem | Steripath® Micro<br>Blood Collection<br>System<br>K200661 | Predicate Device,<br>Steripath® Gen2<br>Blood Collection<br>System (K192247) | Difference between<br>Steripath® Micro<br>Blood Collection<br>System and<br>Predicate Device | |-----------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | FR Number(s) | 862.1675 | 862.1675 | Same | | Product Code | JKA and FPA | JKA and FPA | Same | | Classification Name | Blood Specimen<br>Collection Device | Blood Specimen<br>Collection Device | Same | | Common Name | Blood collection set | Blood collection set | Same | | Regulatory Class | Class II | Class II | Same | | Intended Use | The Steripath® Micro<br>Blood Collection<br>System is a system to<br>draw blood for in vitro<br>diagnostic testing. | The Steripath® Micro<br>Blood Collection<br>System is a system to<br>draw blood for in vitro<br>diagnostic testing. | Same | | Indications for Use | The Steripath® Micro<br>Blood Collection<br>System is indicated for<br>use as a blood<br>collection system that<br>diverts and sequesters<br>the initial specimen<br>prior to collection of a<br>subsequent test<br>sample to reduce the<br>frequency of blood<br>culture contamination<br>when contaminants<br>are present in the<br>initial blood sample<br>compared to blood<br>cultures drawn with<br>standard procedure<br>without manual<br>diversion.<br>Additionally,<br>components of the<br>system may be used<br>for infusion following<br>sample collection after<br>disconnection of the<br>Initial Specimen<br>Diversion Device®. | The Steripath® Gen2<br>Blood Collection<br>System is indicated for<br>use as a blood<br>collection system that<br>diverts and sequesters<br>the initial specimen<br>prior to collection of a<br>subsequent test<br>sample to reduce the<br>frequency of blood<br>culture contamination<br>when contaminants<br>are present in the<br>initial blood sample<br>compared to blood<br>cultures drawn with<br>standard procedure<br>without manual<br>diversion.<br>Additionally,<br>components of the<br>system may be used<br>for infusion following<br>sample collection after<br>disconnection of the<br>Initial Specimen<br>Diversion Device®.<br>Venipuncture needles<br>are indicated for short<br>term infusion (less<br>than 2 hours). | Indications for use<br>changed slightly due<br>to removal of<br>venipuncture needles.<br>The changes raise no<br>new questions of<br>safety or<br>effectiveness. | | Contraindications | None | None | Same | | Prescription Status | Prescription Use Only | Prescription Use Only | Same | | | | | | | Item | Steripath® Micro<br>Blood Collection<br>System | Predicate Device,<br>Steripath® Gen2<br>Blood Collection<br>System (K192247) | Difference between<br>Steripath® Micro<br>Blood Collection<br>System and<br>Predicate Device | | Patient Interface<br>(ISDD® Inlet<br>Accessories) | Luer Extension, 7"<br>ICU Medical, Inc.<br>Model B1754-NS<br>K964435 | Blood Collection Set,<br>21G<br>Becton Dickinson<br>Model 367326<br>(K030573)<br>Blood Collection Set,<br>23G<br>Becton Dickinson<br>Model 367324<br>(K030573)<br>Luer Extension, 9"<br>ICU Medical, Inc.<br>Model B1798-NS<br>(K964435) | Micro adds a 7" Luer<br>extension and<br>removes venipuncture<br>needle patient<br>interfaces. The<br>changes raise no new<br>questions of safety or<br>effectiveness. | | Culture<br>Bottle/<br>Syringe<br>Interface<br>(ISDD® Outlet<br>Accessories) | Syringe, 10ml<br>Becton Dickinson<br>Model 301029<br>(K980987)<br>Syringe, 20ml<br>Becton Dickinson<br>Model 301031<br>(K980987)<br>Syringe, 5ml<br>Becton Dickinson<br>Model 301027<br>K980987 | Transfer Adapter,<br>BD Luer-Lok Access<br>Device<br>Model 364902<br>(K991088)<br>BD Bottle Interface<br>Transfer Adapter,<br>Saf-T Holder Blood<br>Culture Device with<br>Male Luer Adapter<br>Smiths Medical<br>Model 96004 &<br>Model 96000S<br>(K081229)<br>BioMerieux® and<br>ThermoFisher® Bottle<br>interfaces<br>Syringe, 10ml<br>Becton Dickinson<br>Model 301029<br>(K980987)<br>Syringe, 20ml<br>Becton Dickinson<br>Model 301031<br>(K980987) | Micro adds a 5mL<br>syringe and removes<br>Transfer Adapter<br>bottle interfaces. The<br>changes raise no new<br>questions of safety or<br>effectiveness. | | Initial Specimen<br>Diversion Device®<br>(ISDD®) | P00353 Base<br>Assembly, Steripath®<br>Micro | P00133 Base<br>Assembly, Gen2 | Average Micro<br>diversion volume =<br>0.75mL. Average<br>Gen2 diversion<br>volume = 1.75mL.<br>Based on these<br>averages, a 1mL<br>reduction in diversion<br>volume raises no new<br>questions of safety or<br>effectiveness.<br>Literature supporting<br>Micro diversion<br>volume is referenced<br>in Patton et al.1 | | Packaging | Chevron Pouch, 12" x<br>6"<br>TPT-0270 to TPF-<br>0524a | Chevron Pouch, 12" x<br>6"<br>TPT-0270 to TPF-<br>0524a | Same | | Sterilization Method | Gamma Radiation<br>Steris, Ontario CA | Ethylene Oxide<br>Steris, Temecula CA | The Steripath® Micro<br>gamma radiation<br>sterilization process is<br>validated to FDA<br>recognized consensus<br>standards to the same<br>sterility assurance<br>level as the predicate<br>device (10-6).<br>Therefore this change<br>raises no new | | ltem | Steripath® Micro<br>Blood Collection<br>System | Predicate Device,<br>Steripath® Gen2<br>Blood Collection<br>System (K192247) | Difference between<br>Steripath® Micro<br>Blood Collection<br>System and<br>Predicate Device | | | | | questions of safety or<br>effectiveness. | | SAL Level | 10-6 | 10-6 | Same | | Non-pyrogenic | Yes | Yes | Same | | Shelf Life | 1 year | 1 year | Same | | Materials | Medical grade<br>materials (pvc tubing,<br>medical grade<br>adhesives,<br>polycarbonate,<br>silicone, TPE) | Medical grade<br>materials (stainless<br>steel, pvc tubing,<br>medical grade<br>adhesives,<br>polycarbonate) | The Steripath® Micro<br>Blood Collection<br>System has slightly<br>different medical<br>grade materials,<br>(thermoplastics, and<br>elastomers) than<br>Steripath® Gen2.<br>Having completed<br>appropriate<br>biocompatibility testing<br>per FDA recognized<br>consensus standards,<br>the changes in<br>materials of the<br>Steripath® Micro of<br>the raise no new<br>questions of safety or<br>effectiveness. | | Biocompatibility<br>Testing | ISO 10993-1<br>ISO 10993-4<br>ISO 10993-5<br>ISO 10993-10<br>ISO 10993-11 | ISO 10993-1<br>ISO 10993-4<br>ISO 10993-5<br>ISO 10993-10<br>ISO 10993-11 | The Steripath® Gen2<br>Blood Collection<br>System was tested in<br>accordance with FDA<br>recognized consensus<br>biocompatibility<br>standards for short<br>duration, blood<br>contacting devices. | | Transport<br>Environment | ASTM D4169-09<br>distribution cycle 13,<br>assurance level II | ASTM D4169-09<br>distribution cycle 13,<br>assurance level II | Same | {5}------------------------------------------------ {6}------------------------------------------------ ## Table 2: Substantial Equivalence Table ² Patton RG, Schmitt T. (2010) "Innovation for reducing blood culture contamination: initial specimen diversion technique". J. Clin. Microbio; 48:4501-3 #### Summary of Performance Testing: The Steripath® Micro Blood Collection System has been found to its System, Labeling, Controls, Interfaces, Accessory, Functional, Physical, Biological Safety and Packaging required by the risk analysis, all design verification and validation activities were performed by the designated individual(s) and the results demonstrated that the predetermined acceptance criteria were met. It has also been found to FDA consensus, medical device safety and international harmonized standards. Conformity to key medical device safety requirements include: Sterilization – The system is sterilized using validation processes in conformance with ISO 11137-1 "Sterilization of health care products – Radiation Part 1: Requirement, validation, and routine control of a sterilization process for medical devices". Aging/Shelf Life Test – The system is validated to achieve an accelerated 1-year shelf-life. Prior to distribution, Accelerated Aging is performed in conformity with ASTM F1980-16 "Standards Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices" Biological Safety (Biocompatibility Tests) – The system meets the requirements of ISO 10993-1:2018 "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process", for a short duration (<24hrs.), blood path indirect, contacting device. Testing included Cytotoxicity, Sensitization (intracutaneous reactivity), Acute System Toxicity, Pyrogenicity, and Hemocompatibility. Packaging Integrity Testing / Shipping Tests – The system meets the requirements of ASTM D4169-16, "Standard Practice for Performance Testing of Shipping Containers and Systems", Distribution Cycle 13, Assurance Level II. Functional and Performance Testing – The system meets its functional requirements for safe and effective performance as noted below. {7}------------------------------------------------ | Requirement | Description | Verification<br>Test Result | |-----------------------------|------------------------------------------------------------------------------------------------|-----------------------------| | Sequestration | The ISDD® shall sequester the diversion volume. | PASS | | Minimum Vacuum performance. | The ISDD® shall meet its performance requirements under minimum vacuum conditions. | PASS | | Maximum Vacuum performance. | The ISDD® shall meet its performance requirements under maximum vacuum conditions. | PASS | | Diversion Volume. | The ISDD® shall meet the minimum and maximum diversion volume requirements | PASS | | Positive Pressure Maximum | The ISDD® shall remain functionally intact and safe under maximum positive pressure conditions | PASS | Conclusions: Through performance testing the Sterion System has demonstrated substantial equivalence to the predicate device, the Steripath® Gen2 Blood Collection System (K192247)
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...