ASANTE CONSET AND ASANTE COMFORT SUBCUTANEOUS INFUSION SETS FOR USE WITH ASANTE PEARL INFUSION PUMPS

K120872 · Unomedical A/S Infusion Devices · FPA · Apr 20, 2012 · General Hospital

Device Facts

Record IDK120872
Device NameASANTE CONSET AND ASANTE COMFORT SUBCUTANEOUS INFUSION SETS FOR USE WITH ASANTE PEARL INFUSION PUMPS
ApplicantUnomedical A/S Infusion Devices
Product CodeFPA · General Hospital
Decision DateApr 20, 2012
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 880.5440
Device ClassClass 2
AttributesTherapeutic

Indications for Use

These infusion sets are indicated for the subcutaneous infusion of medication from an external Infusion Pump. These infusion sets are intended to be used only with the Asante Pearl™ Insulin Infusion Pump.

Device Story

Asante Comfort™ and Asante Conset™ are sterile, single-use subcutaneous infusion sets designed for use with the Asante Pearl™ Insulin Infusion Pump. The devices function as a fluid path between the external pump and the patient's subcutaneous tissue. The Conset™ variant includes an integrated spring-powered catheter insertion device. The primary modification from predicate devices is the replacement of the standard female luer connector with a proprietary connector/adapter specifically compatible with the Asante Pearl™ pump. This adapter is bonded to the tubing using the same adhesive as the predicate. The devices are used in clinical or home settings by patients or caregivers to deliver medication. The healthcare provider or patient attaches the connector to the pump per manufacturer instructions. The device facilitates continuous medication delivery, benefiting patients requiring subcutaneous infusion therapy.

Clinical Evidence

Bench testing only. Testing included flow (occlusion) tests (minimum 40 mL/min for Comfort, 80 mL/min for Conset at 1 bar), leak testing (no leaks at 0.56 bar for 30 seconds), and pull testing (minimum 15 N for tubing to adapter bond).

Technological Characteristics

Sterile, non-pyrogenic, single-use subcutaneous infusion sets. Features proprietary connector/adapter for Asante Pearl™ pump. Materials, fluid contact surfaces, and sterilization processes are identical to predicate devices. Mechanical connection via adhesive bonding.

Indications for Use

Indicated for subcutaneous infusion of medication from an external infusion pump. Intended for use with the Asante Pearl Insulin Infusion Pump.

Regulatory Classification

Identification

An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.

Special Controls

*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K120872 #### 8.0 Revised 510(K) SUMMARY Date Prepared: April 10, 2012 APR 2 0 2012 ### 8.1 SUMMARY OF SAFETY AND EFFECTIVENESS Submitted By: John M. Lindskog President Unomedical A/S Infusion Devices Aaholmvei 1-3. Osted DK-4320 lejre, Denmark Telephone: + 45 48 16 70 00 8.2 asante conset™ and asante comfort™ Trade/Proprietary Name: Subcutaneous Infusion Sets for use with Asante - Subcutaneous Infusion Set 8.3 Common/Usual Name - Classification Name Intravascular Administration Set 8.4 #### 8.5 Predicate Devices The asante comfort™ infusion sets for use with the Asante Pearl™ Pump are substantially equivalent to the current Unomedical Comfort™(K972135) and asante conset™ infusion sets for use with the Asante Pearl™ Pump are substantially equivalent to the Unomedical Inset™ (K032854) infusion sets. Pearl™ infusion Pumps #### 8.6 General Description - 8.6.1 Asante comfort™ The asante comfort™ Subcutaneous Infusion Sets for use with the Asante Pearl™ pump and the currently marketed Comfort™ Subcutaneous Infusion Sets are both sterile, non-pyrogenic, single use subcutaneous infusion sets. For the connection to the infusion set, the current Comfort™ sets have a standard female luer connector compatible with many commercially available infusion devices while the new asante comfort™ sets have the unique connector/adapter only compatible with the Asante Insulin Infusion Pump cleared as part of the Asante Pearl™ Diabetes Management System 510(k), file number K100567. This connector/adapter is provided to Unomedical as a finished, tested component by Asante Solutions, the manufacturer of the Asante Pump, and is bonded,us ing the same adhesive, to the pump end of a standard Comfort™ infusion set in lieu of the female luer by Unomedical. The remainder of the set, including all fluid contact materials, sterile packaging, and the manufacturing and sterilization process are unchanged. The method of use of the devices is the same as for the unmodified devices including attaching the connector end of the set to the pump per the pump manufacturers' instructions and the use of the pump with the Comfort™ Subcutaneous Infusion Sets were included as part of the Asante 510(k) file. {1}------------------------------------------------ ### 8.6.2 Asante conset™ The asante conset™ Subcutaneous Infusion Sets for use with the Asante Pearl™ pump and the current Inset™ Subcutaneous infusion Sets are both sterile, nonpyrogenic, single use subcutaneous infusion sets with an integrated springpowered catheter insertion device. For the connection to the infusion set, the current Inset™ infusion sets have a standard luer connector compatible with commercially available infusion devices while the new Asante Conset™ sets have a unique connector/adapter compatible with the Asante Pearl™ Infusion Pump cleared as part of the Asante Pearl™ Diabetes Management System 510(k), file number K100567. This part is provided to Unomedical as a finished, tested component by Asante Solutions, the manufacturer of the Asante Pump, and is bonded, using the same adhesive, to the pump end of a standard Inset™ infusion set in lieu of the female luer by Unomedical. The remainder of the set, including all fluid contact materials, sterile packaging, and the manufacturing and sterilization process are unchanged. The method of use of the devices is the same as for the unmodified devices including attaching the connector end of the set to the pump per the pump manufacturers' instructions and the use of the pump with the Inset™ Subcutaneous Infusion Sets were included as part of the Asante 510(k) file. - 8.7 Intended Use 8.7.1 Indication for Use These infusion sets are indicated for the subcutaneous infusion of medication from an external Infusion Pump. 8.7.2 Intended use These infusion sets are intended to be used only with the Asante Pearl™ Insulin Infusion Pump. #### 8.8 Technological Characteristics The asante comfort™ and asante conset™ Subcutaneous Infusion Sets have all of the same technological characteristics as the Unomedical Comfort™ and Unomedical Inset™ Infusion Sets except that a proprietary connector, uniquely compatible with the Asante Pearl™ Insulin Infusion Pump has been substituted for the female luer connector on the pump end of the set. #### 8.9 Performance Data The following verification testing was performed on the modified sets: ### 8.9.1 Flow (Occlusion) test | 8.9.1.1 | asante comfort™ - minimum 40 mL/min at 1 bar pressure | |-----------------|-------------------------------------------------------| | 8.9.1.2 | asante conset™ - minimum 80 mL/min at 1 bar pressure | | 8.9.2 Leak test | | | 8.9.2.1 | No Leaks at 0.56 bar pressure for 30 seconds | | 8.9.3 Pull test | | ## 8.9.3 Pull test Tubing to Asante Adapter - min 15 N 8.9.3.1 ### 8.10 Conclusion Unomedical A/S assessed the sterilization cycle, material biocompatibility, and connector function, flow, leak and pull test results to conclude that the new product lines are substantially equivalent to products currently legally marketed in the USA. {2}------------------------------------------------ # DEPARTMENT OF HEALTH & HUMAN SERVICES. Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services USA. The seal features a stylized eagle with its wings spread, symbolizing protection and care. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle, indicating the department's name and national affiliation. The seal is presented in black and white. ### Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Mr. John M. Lindskog President Unomedical a/s Infusion Devices Aaholmvej 1-3, Osted 4320 Lejre Denmark APR 2 0 2012 Re: K120872 Trade/Device Name: Asante Conset™ and Asante Comfort™ Subcutaneous Infusion Set Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: II Product Code: FPA Dated: March 9, 2012 Received: March 22, 2012 Dear Mr. Lindskog: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ # Page 2- Mr. Lindskog - Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem /default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Bag Anthony D. Watson, B.S., M.S., M.B.A. Director > Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # INDICATIONS FOR USE 510(k) Number: Device Name: Asante Consetm and Asante Comfortm Subcutaneous Infusion Sets "for" the Asante Pearl Insulin Infusion Pump Indications for Use: These infusion sets are indicated for the subcutaneous infusion of medication from an external Infusion Pump. Prescription Use (Per 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) 4/19/12 (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices 510(k) Number: 120872
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...