K202492 · Teleflex Medical · FMI · Mar 22, 2021 · General Hospital
Device Facts
Record ID
K202492
Device Name
EZ-IO Intraosseous Vascular Access System
Applicant
Teleflex Medical
Product Code
FMI · General Hospital
Decision Date
Mar 22, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5570
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
For intraosseous access anytime in which vascular access is difficult to obtain in emergent, urgent or medically necessary cases for up to 24 hours. Insertion sites: ADULTS (≥22 years old) - Proximal humerus - Proximal tibia - · Distal tibia PEDIATRICS (<21 years old) - Proximal humerus - Proximal tibia - · Distal tibia - · Distal femur For patients ≥12 years old, use of the device may be extended for up to 48 hours when alternate intravenous access is not available or reliably established.
Device Story
The EZ-IO Intraosseous Vascular Access System provides vascular access by inserting a needle set (stylet and catheter) into the bone marrow medullary space. The system consists of a reusable cordless battery-powered driver and single-use needle sets (15mm, 25mm, 45mm) made of 304 stainless steel with polycarbonate hubs. Used in pre-hospital, hospital, and acute care settings by clinicians. The clinician identifies anatomical landmarks, attaches the needle set to the driver, and drills through the bone cortex until a tactile 'pop' indicates entry into the medullary space. The driver is detached, the stylet is removed, and the catheter is connected to standard IV tubing via a Luer lock for fluid/medication administration. The device facilitates rapid vascular access when traditional IV access is difficult, potentially life-saving in emergent scenarios.
Clinical Evidence
Bench testing only. MRI compatibility testing was conducted per ASTM F2052, ASTM F2213, ASTM F2182, and ASTM F2119 to evaluate RF-induced heating in 1.5T and 3T scanners, supporting the MR Conditional labeling.
Technological Characteristics
Needle sets: 304 stainless steel cannula/stylet, polycarbonate hubs, 15G (1.8mm) diameter, 15/25/45mm lengths. Reusable cordless driver powered by lithium batteries. Biocompatible materials per ISO 10993-1. Sterilization: Ethylene Oxide. Depth control via visual markers and tactile feedback. MR Conditional status.
Indications for Use
Indicated for adult and pediatric patients requiring vascular access when difficult to obtain in emergent, urgent, or medically necessary cases. Contraindications: fracture in target bone; previous significant orthopedic procedure, prosthetic limb, or joint at site; IO access in targeted bone within past 48 hours; infection at insertion site; excessive tissue/obesity preventing identification of anatomical landmarks.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
Predicate Devices
EZ-IO Intraosseous Vascular Access System (K180395)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
March 22, 2021
Teleflex Medical Rachel Rehl Regulatory Affairs Specialist 3015 Carrington Mill Blyd. Morrisville, North Carolina 27560
Re: K202492
Trade/Device Name: EZ-IO Intraosseous Vascular Access System Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: Class II Product Code: FMI Dated: February 19, 2021 Received: February 23, 2021
Dear Rachel Rehl:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Rumi Young Acting Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K202492
#### Device Name
EZ-IO Intraosseous Vascular Access System
Indications for Use (Describe)
For intraosseous access anytime in which vascular access is difficult to obtain in emergent, urgent or medically necessary cases for up to 24 hours.
Insertion sites:
ADULTS (≥22 years old)
- · Proximal humerus
- Proximal tibia
- · Distal tibia
PEDIATRICS (<21 years old)
- Proximal humerus
- Proximal tibia
- · Distal tibia
- · Distal femur
For patients ≥12 years old, use of the device may be extended for up to 48 hours when alternate intravenous access is not available or reliably established.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------|----------------------------------------------------------------------|
| <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) | <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### K202492 510(k) SUMMARY
### EZ-IO Intraosseous Vascular Access System MR Conditional Safety Status Labeling
### 1. Submitter Information
| Name: | Teleflex Medical |
|-------------------|---------------------------------------------------------|
| Address: | 3015 Carrington Mill Blvd.<br>Morrisville, NC 27560 USA |
| Contact Person: | Rachel Rehl |
| Telephone Number: | (919)433-2588 |
| Email: | rachel.rehl@teleflex.com |
Date Prepared: August 28, 2020
## 2. Device Name
Device Trade Name: EZ-IO Intraosseous Vascular Access System Common Name: Intraosseous Infusion System Classification Name: Needle, Hypodermic, Single Lumen (Class II, FMI, 21 CFR 880.5570)
### 3. Predicate Device
EZ-IO Intraosseous Vascular Access System (K180395)
### 4. Device Description and Clinical Operation
#### Device Description
The EZ-IO Intraosseous Vascular Access System previously cleared with K180395 is designed to allow the user to insert a needle set consisting of a stylet and catheter through the cortex of the bone to a desired depth within the bone marrow to facilitate intraosseous infusion of desired fluids and medications for vascular access. All system materials are biocompatible. Needle sets are single-use and composed of 304 stainless steel with polycarbonate hubs and available in 15 mm (for patients 3-39 kg); 25 mm (for patients 3 kg or over) and 45 mm (for patients 40 kg or over) lengths. Black lines on the needle set serve as depth markers. The reusable cordless driver/drill is powered by lithium batteries with a battery-power indicator light. An extension tubing set accessory, the EZ-Connect, is included with every needle set. The EZ-Connect contains a needleless connector system and Luer lock adapter. An optional dressing, the EZ-Stabilizer, is an accessory to the EZ-IO Vascular Access System. It is designed as a securement device with an adhesive backing that is placed over an EZ-IO Needle to
{4}------------------------------------------------
keep the needle securely anchored to the patient; and is recommended in the Instructions for use (IFU).
## Clinical Operation
Clinicians locate anatomical landmarks and clean the insertion site. Using the cordless driver with needle set attached, the needle set is pressed through the soft tissue to the outer cortex of the bone. At least one depth marker on the cannula must be visible prior to powering the driver to ensure adequate needle length for proper placement within the medullary space. Clinicians then squeeze the driver trigger and apply moderate, steady pressure. The trigger is released when a sudden "give" or "pop" is felt, which indicates entry into the medullary space; the needle set will not always be inserted to the hub. After insertion of the needle set, the driver unit is detached from the needle set, leaving the stylet and cannula firmly seated in the bone. The stylet is then separated and removed from the cannula by turning the stylet hub counterclockwise leaving the catheter with a standard Luer lock hub securely seated in the bone. The cannula Luer lock permits attachment of the provided EZ-Connect, standard syringes or other IV tubing for administration of medications and fluids.
### 5. Indications for Use
For intraosseous access anytime in which vascular access is difficult to obtain in emergent, urgent or medically necessary cases for up to 24 hours.
Insertion sites:
ADULTS (≥22 years old)
- Proximal humerus
- Proximal tibia
- · Distal tibia
# PEDIATRICS (≤21 years old)
- Proximal humerus
- Proximal tibia
- · Distal tibia
- · Distal femur
For patients 2 years old, use of the device may be extended for up to 48 hours when alternate intravenous access is not available or reliably established.
- 6. Technological Characteristics and SubstantialEquivalence
The proposed device is identical to the predicate device described in K180395 in design, materials of construction, functional performance, principles of operation, manufacturing, packaging, sterilization, and shelf life.
{5}------------------------------------------------
The only change is the labeled MR Conditional Safety Status. Table 5-1 below provides a comparison of the proposed and predicate devices.
| | Proposed Device<br>EZ-IO Intraosseous Vascular<br>Access System<br>(MR Conditional Safety<br>Status Labeling) | Predicate Device K180395<br>EZ-IO Intraosseous Vascular<br>Access System | Comparison |
|--------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|
| Indications<br>for Use | For intraosseous access anytime in<br>which vascular access is difficult to<br>obtain in emergent, urgent or<br>medically necessary cases for up to 24<br>hours.<br>Insertion sites:<br>ADULTS (≥22 years old)<br>• Proximal humerus<br>• Proximal tibia<br>• Distal tibia<br><br>PEDIATRICS (≤21 years old)<br>• Proximal humerus<br>• Proximal tibia<br>• Distal tibia<br>• Distal femur<br><br>For patients ≥12 years old, use of the<br>device may be extended for up to 48<br>hours when alternate intravenous<br>access is not available or reliably<br>established. | For intraosseous access anytime in<br>which vascular access is difficult to<br>obtain in emergent, urgent, or<br>medically necessary cases for up to 24<br>hours. Insertion sites:<br>ADULTS (≥22 years old): proximal<br>humerus, proximal tibia, distal tibia<br>PEDIATRICS (≤21 years old): proximal<br>humerus, proximal tibia, distal tibia,<br>distal femur<br><br>Use of the device may be extended for<br>up to 48 hours when alternate<br>intravenous access is not available or<br>reliably established. Insertion sites:<br>ADULTS (≥22 years): proximal<br>humerus, proximal tibia, distal tibia<br>PEDIATRICS (≥12 years through 21<br>years old): proximal humerus,<br>proximal tibia, distal tibia, distal femur | Wording<br>updated for<br>clarity but the<br>indications for<br>use are the<br>same. |
| Contraindicati<br>ons | • Fracture in target bone.<br>• Previous, significant orthopedic<br>procedure at the site,<br>prosthetic limb or joint.<br>• IO access (or attempted IO<br>access) in targeted bone within<br>past 48 hours.<br>• Infection at the area of<br>insertion.<br>• Excessive tissue (severe<br>obesity) and/or absence of<br>adequate anatomical<br>landmarks. | • Fracture in target bone.<br>• Previous, significant orthopedic<br>procedure at the site,<br>prosthetic limb or joint.<br>• IO access (or attempted IO<br>access) in targeted bone within<br>past 48 hours.<br>• Infection at the area of<br>insertion.<br>• Excessive tissue (severe<br>obesity) and/or absence of<br>adequate anatomical<br>landmarks. | Same |
| | Proposed Device<br>EZ-IO Intraosseous Vascular<br>Access System<br>(MR Conditional Safety<br>Status Labeling) | Predicate Device K180395<br>EZ-IO Intraosseous Vascular<br>Access System | Comparison |
| Target<br>Population | Adult and pediatric patients who<br>are in need of vascular access. | Adult and pediatric patients who<br>are in need of vascular access. | Same |
| Where Used | Pre-hospital, In hospital, Acute care | Pre-hospital, In hospital, Acute care | Same |
| Anatomical<br>Sites Used | Proximal Tibia Proximal Humerus Distal Tibia Distal Femur in pediatric<br>population | Proximal Tibia Proximal Humerus Distal Tibia Distal Femur in pediatric<br>population | Same |
| Needle/Cannula<br>Design | Sterile, single use Hubs: polycarbonate and color<br>additive Stylet/catheter: Stainless Steel Faceted tip Standard Luer connection 15 mm;<br>25 mm; 45 mm 15 gauge (0.071", 1.8 mm) Needle Cover: Polypropylene | Sterile, single use Hubs: polycarbonate and color<br>additive Stylet/catheter: Stainless Steel Faceted tip Standard Luer connection 15 mm;<br>25 mm; 45 mm 15 gauge (0.071", 1.8 mm) Needle Cover: Polypropylene | Same |
| Needle Set<br>Guidelines | Available Needle Sets: 15 mm: 3-39 kg 25 mm: 3 kg or over 45 mm: 40 kg or over | Available Needle Sets: 15 mm: 3-39 kg 25 mm: 3 kg or over 45 mm: 40 kg or over | Same |
| Depth Control | Positioning marks at 5 mm and 10<br>cm apart to provide visual reference<br>points<br><br>Tactile feedback for change of<br>pressure | Positioning marks at 5 mm and 10<br>mm apart to provide visual reference<br>points<br><br>Tactile feedback for change of<br>pressure | Same |
| Sterile single<br>use components<br>and accessories | EZ-IO Needle Sets EZ-Connect Extension Set EZ-IO Patient Wristband NeedleVISE 1-port Sharps Block EZ-Stabilizer Dressing | EZ-IO Needle Sets EZ-Connect Extension Set EZ-IO Patient Wristband NeedleVISE 1-port Sharps Block EZ-Stabilizer Dressing | Same |
| Sterility of<br>single use<br>components<br>and accessories | Ethylene Oxide | Ethylene Oxide | Same |
| Shelf life of<br>single use<br>components<br>and accessories | 4 Years | 4 Years | Same |
| | Proposed Device<br>EZ-IO Intraosseous Vascular<br>Access System<br>(MR Conditional Safety<br>Status Labeling) | Predicate Device K180395<br>EZ-IO Intraosseous Vascular<br>Access System | Comparison |
| Biocompatibility | Biocompatible materials used (per<br>ISO 10993-1 prolonged contact<br>duration). | Biocompatible materials used (per<br>ISO 10993-1 prolonged contact<br>duration). | Same |
| MR Safety<br>Status | MR Conditional | MR Unsafe | Update to MR<br>Safety Status<br>as described in<br>submission. |
Table 5-1: Substantial Equivalence Comparison to Predicate
{6}------------------------------------------------
{7}------------------------------------------------
# 7. Non-clinical Testing
Non-clinical testing was conducted as the basis of the MR Conditional safety status claims that are proposed in this 510(k). The testing is listed in Table 5-2 below. Summaries of the testing are provided as well. The test article was the 45 mm needle.
| Testing | Summary |
|--------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| MRI Testing,<br>Exponent, 2020 | This study evaluated the MRI compatibility of the EZ-IO Needle Set in regard to radiofrequency (RF)-induced heating in 1.5 T and 3 T clinical scanners. The evaluation was conducted using methodologies prescribed in ASTM F2052, ASTM F2213, ASTM F2182, and ASTM F2119 as a guide.<br><br>The evaluation concluded that the EZ-IO Needle Set should be labeled as MR Conditional and included labeling guidelines in the report. |
# Table 5-2: Non-Clinical Testing
# 8. Conclusions
The conclusions drawn from the non-clinical tests inform the MRI labeling that is proposed in this 510(k). As there is no change to the device itself, we can conclude that the device is substantially equivalent to the predicate 510(k) K180395.