K955514 · Dura-Vac Products · MMK · Aug 1, 1996 · General Hospital
Device Facts
Record ID
K955514
Device Name
SHARPS SAFE 1.5 & SHARPS SAFE 2.0
Applicant
Dura-Vac Products
Product Code
MMK · General Hospital
Decision Date
Aug 1, 1996
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5570
Device Class
Class 2
Indications for Use
Sharps Safes are designed to be used by Paramedics or Home Care Providers who need a small, portable and rugged sharps container which can be used in remote locations to dispose of contaminated needles.
Device Story
Sharps Safe 1.5 and 2.0 are portable, rugged sharps containers; constructed of plastic tubes with secure end caps; bottom cap contains hard plastic disc for puncture resistance; top cap features slit-flap opening for needle insertion; includes vinyl cover for moisture-tight seal. Used by paramedics and home care providers in remote locations for safe disposal of contaminated needles. Device addresses limitations of existing containers regarding size, leakage, shatter resistance, and visibility. Provides secure storage for sharps during transport.
Clinical Evidence
No clinical data. Bench testing only, including impact resistance (8-foot drop test), puncture resistance, wall thickness measurements, pull tests (45 lb), capacity measurements, and leakproof testing.
Technological Characteristics
Plastic tube construction (1.5-inch and 2.0-inch diameters). Features include puncture-resistant bottom disc, slit-flap top entry, and vinyl moisture-tight cover. Mechanical performance validated via impact, puncture, and pull testing. No electronic or software components.
Indications for Use
Indicated for use by paramedics and home care providers for the disposal of contaminated needles in remote or portable settings.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
K133031 — SHARPSOLOGY 2 · Industrial Water Solutions, Inc. · Nov 15, 2013
Submission Summary (Full Text)
{0}
DURA-VAC
AUG - I 1996
4464 McGRATH STREET #114 • VENTURA, CA 93003 • (805) 658-7768
REVISION OF 510K SUMMARY
(as required by 807.92 (c))
May 20, 1996
K955514
Submitted by:
Neal Shults (owner)
4464 McGrath Street #114
Ventura, CA 93003
(805) 658-7768
(805) 658-8353 (FAX)
November 27, 1995
Trade name - Sharps Safe 1.5
Sharps Safe 2.0
Common name - Sharps Container
Classification name - Sharps Container 80MMK 880.5570
Equivalent device being legally marketed:
Trade name - Sharps A Gator
Manufactured by: Devon Industries, Inc.
510K Approval Number: K943892
Description:
The device is comprised of a 1.5 inch diameter plastic tube (model 1.5), or a 2.0 diameter plastic tube (model 2.0). Securely attached to each end is a plastic cap. Secured inside the cap attached to the bottom end is a hard plastic disc designed to prevent needle penetration. The cap on the top end has a ¾ inch hole in the center. Radiating from the hole are slits cut in the plastic which form flaps through which a sharps can be inserted. A vinyl cover is provided for the top to provide a moisture tight seal.
Intended Use:
Sharps Safes are designed to be used by Paramedics or Home Care Providers who need a small, portable and rugged sharps container which can be used in remote locations to dispose of contaminated needles.
Technological Characteristics:
Predicate devices which we surveyed all had some shortcomings. They were either too large, too small, wouldn't contain the sharps if mishandled, blood could leak out, would shatter if struck, couldn't be attached to the outside of a drug box, or the contents couldn't be seen.
All of these failings were addressed when designing the Sharps Safe and eliminated.
{1}
COPY
K955s14
1
DURA-VAC
4464 McGRATH STREET #114 • VENTURA, CA 93003 • (805) 658-7768
Non Clinical Tests:
The summary data derived from mechanical testing are as follows:
1) Impact Resistance: No needle ejection when dropped from eight (8) feet onto concrete. No damage to container.
2) Puncture Resistance:
1.5 - 5.12 lb. average for sides.
6.56 lb. average for bottom.
2.0 - 5.80 lb average for sides.
6.99 lb. average for bottom.
3) Wall Thickness:
1.5 - .072 inches average for sides.
.1174 inches average for bottom.
2.0 - .078 inches average for sides.
.1125 inches average for bottom.
4) Sharps Closure - Pull Test - 45 lb. for 24 hours.
5) Capacity:
1.5 - 15.8 cu. in./6.75 oz.
2.0 - 28.2 cu. in./12.30 oz.
6) Leakproof Test: The bottom is leakproof. The top is also leakproof when the vinyl cover is in place.
7) Disposal: Sharps Safes contain no heavy metals and can be incinerated.
8) Attachment Strap Test: A Sharps Safe will support 45 pounds when secured to a surface by the attachment straps.
{2}
K950519
4404 McGRATH STREET #114 • VENTURA, CA 93003 • (805) 658-7768
DURA-VAC
Clinical Tests:
Not required.
Conclusions:
It is very clear that there is a real need for a sharps container that is designed specifically for Paramedics and Home Health Care Professionals.
All of the existing devices on the market have shortcomings which make them unsuitable for Paramedic use. Sharps Safes were designed to fill this void.
The nonclinical tests provided us with the design and materials to be used, and the clinical tests refined the parameters of the materials. These tests also revealed that a better system was needed to secure the container to the drug box. It also became apparent that two sizes would be beneficial (2 inch and 1.5 inch).
The combined talents of sixty (60) fire department Paramedics were used to develop Sharps Safes.
We feel that they meet the need.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.