K252637 · Keter Canada, Inc. · MMK · Oct 14, 2025 · General Hospital
Device Facts
Record ID
K252637
Device Name
Community Containers (Flap and Daisy)
Applicant
Keter Canada, Inc.
Product Code
MMK · General Hospital
Decision Date
Oct 14, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5570
Device Class
Class 2
Indications for Use
The Community Container (Flap and Daisy) are intended to be used for the collection, transportation and disposal of 1 ml and 2 ml hypodermic needles and syringes in health care areas, home care environment and any other area requiring the use of sharps containers for the collection, transportation and disposal of used 1 ml and 2 ml hypodermic needles and syringes.
Device Story
Portable, injection-molded polypropylene sharps collectors; designed for personal storage of small low-volume sharps including blood needles, lancets, and 1-2 ml syringes. Features rectangular shape with conical taper, hinged cap, and temporary closure system; allows reopening for additional sharps storage prior to terminal disposal. Used in healthcare areas and home care environments by patients or healthcare workers. Provides alternative to re-sheathing needles; protects end-user from sharps injury during transport to incineration. Device is single-use, non-sterile, and disposable.
Clinical Evidence
No clinical data submitted. Bench testing only. Performance testing included resistance to penetration (19.04 N minimum) and resistance to spillage by toppling, both meeting ISO 23907-1:2019 requirements.
Technological Characteristics
Material: Polypropylene. Form factor: Rectangular with conical taper, 8.7cm x 4.3cm x 15.1cm. Volume: 325mL. Features: Hinged cap, divider flap, daisy top. Standards: ISO 23907-1:2019 (pocket collector). Manufacturing: Injection molding. Non-sterile, single-use.
Indications for Use
Indicated for collection, transportation, and disposal of 1 ml and 2 ml hypodermic needles and syringes in healthcare and home care environments.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
October 14, 2025
Keter Canada, Inc.
% Trey Thorsen
Regulatory Consultant
The FDA Group
290 Turnpike Road
Suite 200
Westborough, Massachusetts 01581
Re: K252637
Trade/Device Name: Community Containers (Flap and Daisy)
Regulation Number: 21 CFR 880.5570
Regulation Name: Hypodermic Single Lumen Needle
Regulatory Class: Class II
Product Code: MMK
Dated: August 20, 2025
Received: August 20, 2025
Dear Trey Thorsen:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252637 - Trey Thorsen
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K252637 - Trey Thorsen
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Stephen A. Anisko -S
Stephen Anisko
Acting Assistant Director
DHT4C: Division of Infection Control Devices
OHT4: Office of Surgical and Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K252637 | |
| Device Name Community Containers (Flap and Daisy) | |
| Indications for Use (Describe) The Community Container (Flap and Daisy) are intended to be used for the collection, transportation and disposal of 1 ml and 2 ml hypodermic needles and syringes in health care areas, home care environment and any other area requiring the use of sharps containers for the collection, transportation and disposal of used 1 ml and 2 ml hypodermic needles and syringes. | |
| Type of Use (Select one or both, as applicable) ☐ Prescription Use (Part 21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
KETER
510(k) Summary
K252637
October 13, 2025
I. SUBMITTER
Keter Canada, Inc.
205 Market Dr,
Milton, ON L9T 4Z7, Canada
Contact: Lovleen Kaur
Email: Lovleen.kaur@keter.com
Phone: 289-409-9133
II. CORRESPONDANT
The FDA Group
290 Turnpike Road
Suite 200
Westborough, Massachusetts 01581
Contact: Trey Thorsen
Email: athorsen@the-fda-group.com
Phone: 850-450-3932
III. DEVICE
Name of Device: Community Containers (Flap and Daisy)
Common or Usual Name: Container Sharps
Classification Name: Hypodermic single lumen needle 21 CFR 880.5570
Product Code: MMK
Device Class: II
IV. PREDICATE DEVICE
Name of Device: ASP HealthCare Fitpack Product Line cleared via K060037.
Common or Usual Name: Container Sharps
Classification Name: Hypodermic single lumen needle 21 CFR 880.5570
Product Code: MMK
Device Class: II
V. INDICATIONS FOR USE
The Community Containers (Flap and Daisy) are intended to be used for the collection, transportation and disposal of 1 ml and 2 ml hypodermic needles and syringes in health care areas, home care environment and any other area requiring the use of sharps containers for the collection, transportation and disposal of used 1 ml and 2 ml hypodermic needles and syringes.
VI. DEVICE DESCRIPTION
K252637
Page 1 of 5
{5}
KETEN
The Community Containers (Flap and Daisy) are portable molded polypropylene syringe collectors that provide an alternate to re-sheathing a needle with its original protective cover and are designed to safely hold small low volume sharps such as blood needles, lancets, and small syringes.
The Community Containers (Flap and Daisy) are single use, non-sterile disposable, transportable sharps collectors intended to be used for the collection, transportation and disposal of 1 ml and 2 ml hypodermic needles and syringes in health care areas, home care environment and any other area requiring the use of sharps containers for the collection, transportation and disposal of used 1 ml and 2 ml hypodermic needles and syringes.
The Community Containers (Flap and Daisy) are rectangular, with a conical taper and a temporary closure capability, which can be reopened for the storage of additional sharps prior to terminal disposal.
The Community Containers (Flap and Daisy) are approximately 8.7cm (L) x 4.3cm (W) x 15.1cm (H). The cap is opaque plastic with a hinged cap, which can be snapped closed to contain the biohazardous sharps. Each Community Container is individually embossed with a biohazard symbol.
# VII. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
Both the Community Containers (Flap and Daisy) and the predicate device are portable, transportable sharps container, with a temporary closure system which protects the end user prior to incineration.
The subject and predicate device are rectangular with a conical taper and a temporary closure capability, which can be reopened for the storage of additional sharps prior to terminal disposal.
| Element | Subject Device | Predicate Device | Discussion of Differences |
| --- | --- | --- | --- |
| 510(k) Number | K252637 | K060037 | |
| Device Name | Community Containers (Flap and Daisy) | ASP HealthCare Fitpack Product Line | |
| Device Classification | 21 CFR §880.5570: Class II | 21 CFR §880.5570: Class II | No difference. |
| Product Code | MMK – Container Sharps | MMK – Container Sharps | No difference. |
| Device Common Name as Cleared by FDA | Container, Sharps | Container, Sharps | No difference. |
| Intended / Indications for use | The Community Containers (flap and daisy) are intended to be used for the collection, transportation and disposal of 1 ml and 2 ml hypodermic needles and syringes in health care areas, home care environment and any other area requiring the use of sharps containers for the collection, transportation and disposal of used 1 ml and 2 ml hypodermic needles and syringes in health care areas, home care environment and any other area requiring the use of sharps containers for the collection, transportation and disposal of used 1 ml and 2 ml hypodermic needles and syringes. | The Product is intended to be used for the collection, transportation and disposal of 1 ml and 2 ml hypodermic needles and syringes in health care areas, home care environment and any other area requiring the use of sharps containers for the collection, transportation and disposal of used 1 ml and 2 ml hypodermic needles and syringes. | No difference. |
K252637
{6}
KETCA
K252637
Page 3 of 5
| Element | Subject Device | Predicate Device | Discussion of Differences |
| --- | --- | --- | --- |
| 510(k) Number | K252637 | K060037 | |
| Device Name | Community Containers (Flap and Daisy) | ASP HealthCare Fitpack Product Line | |
| | area requiring the use of sharps containers for the collection, transportation and disposal of used I ml and 2 ml hypodermic needles and syringes. | sharps containers for the collection, transportation and disposal of used 1 ml and 2 ml hypodermic needles and syringes. | |
| Device Description | The Community Containers (Flap and Daisy) are compact, injection molded containers complete with integral molded lid.
Specifically designed for personal storage, collection, transportation and disposal of 1ml hypodermic needles and syringes. Ideal for front line health workers, diabetics, in healthcare areas, home care environments and any other area requiring the use of sharps containers. | Compact high quality injection molded container complete with integral molded lid.
Specifically designed for personal storage, collection, transportation and disposal of 1ml hypodermic needles and syringes. Ideal for front line health workers, diabetics, in healthcare areas, home care environments and any other area requiring the use of sharps containers. | No difference. |
| Size (volume) | 325mL nominal | 130mL – 750mL | No difference. The predicate device volume is within the volume range cleared in K060037. |
| Dimension (cm) (L x W x H) | 8.7 x 4.3 x 15.1 | 8.1 x 3.8 x 15.3 | Similar. The subject device’s dimensions are comparable to those of the predicate. The subject device is still considered a pocket sharps container as defined by ISO 23907-1:2019. |
| Single Use | Yes | Yes | No difference. |
| Rx or OTC | OTC | OTC | No difference. |
| Environment of Use | Health care areas, home care environment and any other area requiring the use of sharps containers for the collection, transportation and disposal of used I ml and 2 ml hypodermic needles and syringes. | Health care areas, home care environment and any other area requiring the use of sharps containers for the collection, transportation and disposal of used 1 ml and 2 | No difference. |
{7}
KETEN
| Element | Subject Device | Predicate Device | Discussion of Differences |
| --- | --- | --- | --- |
| 510(k) Number | K252637 | K060037 | |
| Device Name | Community Containers (Flap and Daisy) | ASP HealthCare Fitpack Product Line | |
| | hypodermic needles and syringes. | ml hypodermic needles and syringes. | |
| Application | Specifically designed for personal storage, collection, transportation and disposal of used 1 ml and 2 ml hypodermic needles and syringes. | Specifically designed for personal storage, collection, transportation and disposal of used 1 ml and 2 ml hypodermic needles and syringes. | No difference. |
| Material | Polypropylene | Polypropylene | No difference. |
| Sharps Access | Top opening | Top opening | No difference. |
| Sharps Closure | Hinged closure | Hinged closure | No difference. |
| Configuration | • Divider Flap
• Daisy Top | • Divider Flap
• Daisy Top | No difference. |
| Impact and leak resistance | Yes | Yes | No difference. |
| Needle Penetration Resistance- | Yes | Yes | No difference. |
| Non-Sterile | Yes | Yes | No difference. |
| Color | Multiple colors | Black or Yellow. | Similar. While the predicate comes in only two colors, the subject device comes in multiple colors depending on the customers preference. |
| Method of Manufacture | Injection molding | Injection molding | No difference |
VIII. NON-CLINICAL TESTING SUMMARY
Performance Testing
The Community Containers (Flap and Daisy) were designed using ISO 23907-1:2019 Sharps injury protection — Requirements and test methods-Sharps containers Part 1: Single-use sharps containers, and its size of 0.3L qualifies it as a pocket collector as defined by the standard. It is designed for storage of such sharps as blood needles, lancets, and small syringes prior to terminal disposal. The Community
K252637
{8}
KETEN
Containers (Flap and Daisy) meet the performance requirements for the standard. In addition, it has also been tested by an independent laboratory to meet the requirements for puncture.
| Test | Acceptance Criteria | Result | Comment |
| --- | --- | --- | --- |
| Resistance to Penetration | The force needed to penetrate the test specimens shall be:
• A minimum of 16 N.
• A minimum average of 18 N. | 19.04 N minimum | All samples tested met the acceptance criteria.
The minimum average of all samples complied with the acceptance criterion. |
| Resistance to Spillage by Toppling | There shall be no evidence of a breach of the sample containment area. Five (5) minutes after toppling:
• There shall be no evidence that the function of the container has been compromised.
• The temporary closure must remain intact | No evidence of leakage | All samples tested met the acceptance criteria. |
## Animal Studies
No animal data submitted.
## Clinical Studies
No clinical data submitted.
## IX. CONCLUSIONS
The conclusions drawn for the non-clinical tests demonstrate that the device is as safe, as effective and performs as well as or better than the legally marketed predicate device, K060037.
K252637
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.