Omnipod Insulin Management System, Omnipod DASH Insulin Management System
K182630 · Insulet Corporation · LZG · Jan 8, 2019 · General Hospital
Device Facts
Record ID
K182630
Device Name
Omnipod Insulin Management System, Omnipod DASH Insulin Management System
Applicant
Insulet Corporation
Product Code
LZG · General Hospital
Decision Date
Jan 8, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5725
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Omnipod Insulin Management System The Omnipod® Insulin Management System is intended for subcutaneous delivery of insulin at set and variable rates for the management of diabetes mellitus in persons requiring insulin and for the quantitative measurement of glucose in fresh whole capillary blood (in vitro) from the finger. The glucose measurements should not be used for the diagnosis or screening for diabetes. The PDM glucose meter is intended for single patient use and should not be shared. Abbott FreeStyle® test strips are used with the built-in FreeStyle meter for the quantitative measurement of blood glucose in fresh whole capillary blood from the finger, upper arm and palm. Abbott Freestyle Control Solutions are used to verify that the meter and test strips are working together properly and that the test is performed correctly. Omnipod DASH Insulin Management System The Omnipod DASH Insulin Management System is intended for subcutaneous delivery of insulin at set and variable rates for the management of diabetes mellitus in persons requiring insulin. Additionally, the Omnipod DASH System is interoperable with a compatible blood glucose meter to receive and display glucose measurements.
Device Story
System manages insulin therapy and blood glucose; comprises disposable insulin infusion pump (Pod) and wireless Personal Diabetes Manager (PDM). Pod affixes to body via adhesive; delivers insulin subcutaneously for up to three days. PDM controls basal/bolus delivery; includes bolus calculator using carbohydrates, blood glucose, correction factor, insulin-to-carbohydrate ratio, target glucose, and Insulin on Board (IoB). PDM maintains history logs and food library. Omnipod PDM features integrated glucose meter (433.92MHz RF communication); Omnipod DASH PDM uses Bluetooth Low Energy to communicate with Pod and compatible glucose meters. Used in home setting by patients; healthcare providers review therapy history. Modification includes updated soft cannula design and compatibility with Admelog U100 insulin.
Clinical Evidence
No clinical trials; evidence based on bench testing and real-world data analysis. Bench testing included drug stability/compatibility with Admelog U100, soft cannula insertion/infusion depth verification, and sterilization validation per AAMI TIR28:2016 and ISO 11737-1. Real-world post-market data analysis of modified cannula design confirmed safety and effectiveness.
Technological Characteristics
System components: disposable Pod (adhesive, soft cannula) and PDM. Communication: 433.92MHz RF (Omnipod) or Bluetooth Low Energy (DASH). Sterilization: Ethylene oxide (ISO 11135). Standards: ISO 10993-1 (biocompatibility), ISO 14971 (risk management). Software: Bolus calculator algorithm.
Indications for Use
Indicated for management of diabetes mellitus in persons requiring insulin. Omnipod system includes integrated blood glucose monitoring for finger-stick capillary blood. Omnipod DASH system is interoperable with compatible blood glucose meters.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
K191679 — Omnipod DASH Insulin Management System with interoperable technology · Insulet Corporation · Sep 20, 2019
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo is composed of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The Department of Health & Human Services logo is a stylized representation of a human figure, while the FDA part includes the acronym "FDA" in a blue square and the words "U.S. FOOD & DRUG ADMINISTRATION" in blue font.
February 11, 2022
Insulet Corporation Julie Perkins Director, Regulatory Affairs/Quality Assurance 100 Nagog Park Acton, Massachusetts 01720
Re: K182630
Trade/Device Name: Omnipod Insulin Management System, Omnipod DASH Insulin Management System Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion pump Regulatory Class: Class II Product Code: LZG, NBW, NDC
Dear Julie Perkins:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated January 8, 2019. Specifically, FDA is updating this SE Letter as an administrative correction.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Courtney Lias, OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices, at courtneyu.lias@fda.hhs.gov.
Sincerely,
Tina Kiang
-s
Tina Kiang, Ph.D. Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image contains the logos of the Department of Health and Human Services (HHS) and the Food and Drug Administration (FDA). The HHS logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue font.
January 8, 2019
Insulet Corporation Julie Perkins Director, Regulatory Affairs/Quality Assurance 100 Nagog Park Acton, Massachusetts 01720
Re: K182630
Trade/Device Name: Omnipod Insulin Management System, Omnipod DASH Insulin Management System Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: Class II Product Code: LZG, NBW, NDC Dated: December 5, 2018 Received: December 6, 2018
Dear Julie Perkins:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{2}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Alan M.
Stevens -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K182630
#### Device Name
Omnipod Insulin Management System and Omnipod DASH Insulin Management System
#### Indications for Use (Describe)
Omnipod Insulin Management System
The Omnipod® Insulin Management System is intended for subcutaneous delivery of insulin at set and variable rates for the management of diabetes mellitus in persons requiring insulin and for the quantitative measurement of glucose in fresh whole capillary blood (in vitro) from the finger.
The glucose measurements should not be used for the diagnosis or screening for diabetes. The PDM glucose meter is intended for single patient use and should not be shared.
Abbott FreeStyle® test strips are used with the built-in FreeStyle meter for the quantitative measurement of blood glucose in fresh whole capillary blood from the finger, upper arm and palm. Abbott Freestyle Control Solutions are used to verify that the meter and test strips are working together properly and that the test is performed correctly.
Omnipod DASH Insulin Management System
The Omnipod DASH Insulin Management System is intended for subcutaneous delivery of insulin at set and variable rates for the management of diabetes mellitus in persons requiring insulin.
Additionally, the Omnipod DASH System is interoperable with a compatible blood glucose meter to receive and display glucose measurements.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------|
| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the Omnipod logo. The logo consists of a circular design made up of many vertical lines, resembling a stylized sun or gear. Below the circular design, the word "omnipod" is written in lowercase letters, and below that, the words "INSULIN MANAGEMENT SYSTEM" are written in smaller, uppercase letters.
# 5.1 510(K) SUMMARY
# 510(k) Summary Complying with 21 CFR 807.92
| Date prepared: | January 8th, 2019 |
|--------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|
| Submitter Name: | Insulet Corporation |
| Submitter Address: | 100 Nagog Park<br>Acton, MA 01720 |
| Contact Person: | Julie Perkins<br>Director of Regulatory Affairs and<br>Quality Assurance |
| Phone: | (978) 600-7951(office) |
| Fax: | (978) 600-0120 |
| Device Trade / Proprietary Name: | Omnipod® Insulin Management System<br>Omnipod DASH <sup>TM</sup> Insulin Management<br>System |
| Device Common Name:<br>Regulation Description:<br>Regulation Medical Specialty:<br>Review Panel: | Pump, Infusion, Insulin<br>Infusion pump<br>General Hospital<br>General Hospital |
| Product Code: | LZG (Infusion Pump)<br>NBW (System, Test, Blood Glucose,<br>Over the Counter)<br>NDC (Calculator, Drug Dose) |
| Submission Type:<br>Regulation Number:<br>Device Class: | Traditional 510(k)<br>880.5725<br>Class II |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Omnipod Insulin Management System. The logo features a circular design made up of many short, vertical lines, resembling a stylized gear or sun. Below the circular design, the word "omnipod" is written in lowercase letters, with a small circle next to the "d". Underneath "omnipod", the words "INSULIN MANAGEMENT SYSTEM" are written in smaller, uppercase letters.
Device predicate:
K162296 Omnipod® Insulin Management System K180045 Omnipod DASH™ Insulin Management System
# Purpose of Submission:
Modification to K16296 Omnipod® Insulin Management System and K180045 Omnipod DASH™ Insulin Management System to address design and labeling changes.
# Device Description:
The subject devices provide for the management of insulin therapy and blood glucose monitoring by patients with diabetes mellitus. They are each comprised of two primary components: the disposable insulin infusion pump (Pod) and an associated wireless remote controller referred to as the Personal Diabetes Manager (PDM). The PDMs incorporate a suggested bolus calculator which aids the user in determining the insulin bolus dosage needed based on carbohydrates ingested, most recent blood glucose reading, programmable correction factor, insulin to carbohydrate ratio, target blood glucose value, and Insulin on Board (IoB).
The Pod is a body-wearable insulin pump that affixes to the user on the back of the arm, the lower back or abdomen, the thigh area, or any site that has a layer of fatty tissue available. It is held in place by an adhesive pad and provides up to three days of insulin before it is removed and replaced with a new Pod. The PDM is a handheld device that controls the Pod. The user interfaces with the device system through the PDM, where they control basal and bolus delivery and various insulin program settings and calculations. The PDM also has a food library to assist with carbohydrate calculations, and it maintains several variables in a history log for the viewer to track their diabetes therapy. The Omnipod Insulin Management System PDM has an integrated blood glucose meter and communicates with the Pod using wirelessly using secure, low power, bi-directional radio frequency (RF) communications at 433.92MHz. The Omnipod DASH Insulin Management System PDM does not have an integrated blood glucose meter, but is interoperable with a compatible blood glucose meter to receive and display glucose measurements. The Omnipod DASH PDM communicates to the Pod and a compatible blood glucose meter using Bluetooth Low Energy.
Both systems are for prescription use only.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the Omnipod logo. The logo consists of a circular design made up of many small, evenly spaced lines, resembling a stylized sun or gear. Below the circular design, the word "omnipod" is written in lowercase letters, with a small circle containing an 'r' next to the 'd', indicating a registered trademark. Underneath "omnipod", the words "INSULIN MANAGEMENT SYSTEM" are written in smaller, uppercase letters.
### Indications for Use:
### Omnipod Insulin Management System
The Omnipod® Insulin Management System is intended for subcutaneous delivery of insulin at set and variable rates for the management of diabetes mellitus in persons requiring insulin and for the quantitative measurement of glucose in fresh whole capillary blood (in vitro) from the finger.
The glucose measurements should not be used for the diagnosis or screening for diabetes. The PDM glucose meter is intended for single patient use and should not be shared.
Abbott FreeStyle® test strips are used with the built-in FreeStyle meter for the quantitative measurement of blood glucose in fresh whole capillary blood from the finger, upper arm and palm. Abbott Freestyle Control Solutions are used to verify that the meter and test strips are working together properly and that the test is performed correctly.
### Omnipod DASH Insulin Management System
The Omnipod DASH Insulin Management System is intended for subcutaneous delivery of insulin at set and variable rates for the management of diabetes mellitus in persons requiring insulin.
Additionally, the Omnipod DASH System is interoperable with a compatible blood glucose meter to receive and display glucose measurements.
The subject devices have the same intended use and indications for use as their predicates.
### Summary of Technological Characteristics Compared to Predicate Device:
The, energy source, principles of operation, and configuration of the Omnipod® Insulin Management System and Omnipod DASH Insulin Management System are not changed from those cleared in K162296 and K180045, respectively. The same modifications are being proposed to both systems as compared to the predicates include:
Hardware: The cannula design was modified. As part of this design change, the cannula depth of insertion and the depth of infusion specifications have been modified. There have been no changes to the material of the cannula.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the Omnipod logo. The logo consists of a circular design made up of many small, vertical lines, resembling a stylized sun or gear. Below the circular design, the word "omnipod" is written in lowercase letters, with a small circle next to the "d". Underneath "omnipod", the words "INSULIN MANAGEMENT SYSTEM" are written in smaller, uppercase letters.
Labeling: Addition of Admelog U100 as an insulin that has been tested for use with the Omnipod Systems.
There have been no changes to the device materials, software, sterilization method, or packaging of the systems that are subject of this 510(k).
### Performance Data and Standards Compliance:
The following performance testing data were provided in support of the substantial equivalence determination.
- Drug Stability and Compatibility; In-use stability and leachables testing was conducted with Admelog U100 insulin to verify and validate that the systems do not adversely affect the insulin.
- Soft Cannula Studies; testing was conducted to verify that the modified soft cannula design met the insertion depth and new insulin infusion depth specifications.
- Real-world data analysis: An analysis of post-market data gathered from devices with the modified soft cannula was conducted to validate the safety and effectiveness of the modified design. The analysis concluded that the modification did not raise any new questions of safety and effectiveness.
- Sterilization; A sterilization product adoption was conducted in accordance with AAMI TIR28:2016 and bioburden testing was conducted in accordance with ISO 11737-1 for the modified soft cannula design.
- Safety Assurance Case: An assurance case for the design modifications was O provided for each system as recommended in the FDA Guidance: Infusion Pumps Total Product Life Cycle.
The stated goal of the Omnipod Insulin Management System safety assurance case is:
> The Omnipod Insulin Management System with blood glucose monitor and dose calculator is acceptably safe for the infusion of U100 insulin that is approved for use in pumps, for use in the home setting by people with diabetes mellitus who require insulin on a daily basis.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Omnipod Insulin Management System. The logo consists of a circular design made up of many small, evenly spaced lines, creating a ring-like shape. Below the circular design, the word "omnipod" is written in lowercase letters, followed by the phrase "INSULIN MANAGEMENT SYSTEM" in smaller, uppercase letters. The text and the circular design are both in a dark blue color.
The stated goal of the Omnipod DASH Insulin Management System safety assurance case is:
> The Omnipod DASH™ Insulin Management System with dose calculator is acceptably safe for the infusion of U100 insulin that is approved for use in pumps, for use in the home setting by people with diabetes mellitus who require insulin on a daily basis.
Additions to the safety assurance cases from the predicate devices' cases included modified soft cannula design and use of the device with Admelog.
- Risk Management; was completed in accordance with ISO 14971:2007-Medical Devices- Application of Risk Management to Medical Devices. Verification activities, as required by the risk analysis, demonstrated that the predetermined acceptance criteria were met and the devices are safe for use.
The Omnipod Insulin Management System and the Omnipod DASH Insulin Management System comply with the following standards as documented in the predicate devices (K162296 and K180045) and in the applicable test reports provided in this 510(k) submission.
- ISO 10993-1 (2009)- 4th Edition Biological Evaluation of Medical Devices- Part ● 1: Evaluation and Testing within a Risk Management Process
- ISO 14971 Second Edition 2007-03-01 Medical Devices- Application of Risk Management to Medical Devices
- ISO 11737-1 (2018) Sterilization of health care products- Microbiological ● methods- Part 1: Determination of the population of microorganisms on products
- ISO 11135 (2014) Sterilization of health care products- Ethylene oxide-● Requirements for development, validation, and routine control of a sterilization process for medical devices
- AAMI TIR28 (2016) Product adoption and process equivalence for ethylene ● oxide sterilization
# Substantial Equivalence Conclusion:
The subject Omnipod® Insulin Management System and Omnipod® DASH Insulin Management System use similar technology, modes of operation, and indications for use as the devices cleared in K162296 and K180045, respectively. Real-world data and performance testing demonstrate that the changes to the subject devices are substantially equivalent to the predicate devices. Therefore, the subject device is substantially equivalent to the predicate devices.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.