Omnipod® Insulin Management System, Omnipod DASH® Insulin Management System
Device Facts
| Record ID | K211575 |
|---|---|
| Device Name | Omnipod® Insulin Management System, Omnipod DASH® Insulin Management System |
| Applicant | Insulet Corporation |
| Product Code | LZG · General Hospital |
| Decision Date | Aug 13, 2021 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 880.5725 |
| Device Class | Class 2 |
| Attributes | Therapeutic, Pediatric |
Indications for Use
The Omnipod® Insulin Management System is intended for subcutaneous delivery of insulin at set and variable rates for the management of diabetes mellitus in persons requiring insulin and for the quantitative measurement of glucose in fresh whole capillary blood (in vitro) from the finger. The glucose measurements should not be used for the diagnosis or screening for diabetes. The PDM glucose meter is intended for single patient use and should not be shared. Abbott FreeStyle® test strips are used with the built-in FreeStyle meter for the quantitative measurement of blood glucose in fresh whole capillary blood from the finger, upper arm, and palm. Abbott Freestyle Control Solutions are used to verify that the meter and test strips are working together properly and that the test is performed correctly. The Omnipod DASH™ Insulin Management System is intended for subcutaneous delivery of insulin at set and variable rates for the management of diabetes mellitus in persons requiring insulin. Additionally, the Omnipod DASH System is interoperable with a compatible blood glucose meter to receive and display glucose measurements.
Device Story
System manages insulin therapy via disposable body-worn infusion pump (Pod) and wireless remote controller (Personal Diabetes Manager, PDM). Pod delivers subcutaneous insulin at set/variable rates; PDM controls basal/bolus delivery. PDM includes bolus calculator using carbohydrates, blood glucose, correction factor, insulin-to-carbohydrate ratio, target glucose, and Insulin on Board (IoB). Omnipod PDM features integrated glucose meter; DASH PDM uses Bluetooth Low Energy for interoperability with compatible meters. Users wear Pod on arm, back, abdomen, or thigh for up to 3 days. Healthcare providers use system to manage diabetes; patients operate PDM to adjust therapy. Output includes insulin delivery, glucose readings, and therapy history logs. Device benefits patients by automating insulin delivery and simplifying dosage calculations based on clinical parameters.
Clinical Evidence
Bench testing only. In-use stability and leachables testing conducted to validate compatibility of Lyumjev U100 insulin with the infusion systems. Safety assurance cases provided per FDA guidance for infusion pumps.
Technological Characteristics
Pod housing: Polycarbonate; Adhesive: Medical Grade Acrylic with polyester backing; Fluid path: FEP, silicone, stainless steel, polypropylene, PVC, cyclic olefin copolymer, EPDM, polyisoprene. Pumping: Step drive mechanism via microprocessor-controlled leadscrew. Communication: 433.92MHz RF (Omnipod) or Bluetooth Low Energy (DASH). Sterilization: Ethylene Oxide (SAL 10^-6).
Indications for Use
Indicated for subcutaneous insulin delivery at set/variable rates for management of diabetes mellitus in insulin-requiring patients. Omnipod system includes integrated glucose monitoring for capillary blood. Omnipod DASH system is interoperable with compatible glucose meters.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
Predicate Devices
- Omnipod Insulin Management System and Omnipod DASH Insulin Management System (K192659)