OMNIPOD INSULIN MANAGMENT SYSTEM

K122953 · Insulet Corporation · LZG · Dec 7, 2012 · General Hospital

Device Facts

Record IDK122953
Device NameOMNIPOD INSULIN MANAGMENT SYSTEM
ApplicantInsulet Corporation
Product CodeLZG · General Hospital
Decision DateDec 7, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5725
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The OmniPod Insulin Management System is intended for subcutaneous delivery of insulin at set and variable rates for the management of diabetes mellitus in persons requiring insulin and for the quantitative measurement of glucose in fresh whole capillary blood (in vitro) from the finger. The glucose measurements should not be used for the diagnosis of or screening for diabetes. The PDM glucose meter is intended for single-patient use and should not be shared. Abbott FreeStyle® test strips are used with the built-in FreeStyle meter for the quantitative measurement of blood glucose in fresh whole capillary blood from the finger, upper arm and palm. Abbott Freestyle Control Solutions are used to verify that the meter and test strips are working together properly and that the test is performed correctly.

Device Story

System comprises disposable insulin infusion pump (Pod) and wireless Personal Diabetes Manager (PDM) with integrated blood glucose meter. Patient uses PDM to program basal/bolus insulin delivery; PDM communicates wirelessly with Pod. Pod delivers insulin subcutaneously via soft cannula. PDM incorporates Abbott FreeStyle glucose meter for capillary blood glucose testing using test strips. Device used by patients in home/clinic settings. Healthcare providers use PDM data to adjust insulin therapy. Benefits include automated insulin delivery and integrated glucose monitoring to assist in diabetes management.

Clinical Evidence

Bench testing only. No clinical data presented. Testing included usability (IEC 62366, AAMI HE75), risk management (ISO 14971), fluid path integrity, environmental testing (IEC 60601-2-24), cannula depth/angle studies, flow accuracy (basal/bolus +/- 5%), biocompatibility (ISO 10993), insulin compatibility, sterilization, electrical safety (IEC 60601-1), and RF wireless performance.

Technological Characteristics

System includes disposable Pod (insulin reservoir, pump mechanism, cannula) and PDM (controller, glucose meter). Materials biocompatible per ISO 10993. Sterilization via 100% ethylene oxide. Connectivity via RF wireless communication between Pod and PDM. Flow accuracy +/- 5% per IEC 60601-2-24. Software classified as Major level of concern.

Indications for Use

Indicated for persons with diabetes mellitus requiring insulin therapy for subcutaneous insulin delivery and quantitative blood glucose monitoring.

Regulatory Classification

Identification

An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K122953 BEC 0 7 2012 #### Insulet Corporation OmniPod® Insulin Management System 510(k) Premarket Notification Submission #### 7.0 510(k) Summary This 510(k) Summary is being submitted in accordance with the requirements of the Safe Medical Device Act (SMDA) of 1990. The content contained in this 510(k) summary has been provided in conformance with 21 CFR §807.92 9056196 3004464228 #### A. Submitter's Information: Date prepared: September 24, 2012 Name: Insulet Corporation Address: 9 Oak Park Drive Bedford, MA 01730 United States FDA Establishment Owner/Operator Number: Contact Person: Phone: Fax: Manufacturing Site: Michael J. Doyle Director Regulatory & Clinical Affairs (781) 457-5244 (781) 357-4303 Insulet Corporation 9 Oak Park Drive Bedford, MA 01730 United States FDA Establishment Registration Number: #### B. Device Name: Trade/Proprietary Name: Device: Regulation Description: Regulation Medical Specialty: Review Panel: Product Code: Submission Type: Regulation Number: Device Class: Model number (Pod/10-pack): Model number (PDM): Device predicate: OmniPod Insulin Management System Pump. Infusion. Insulin Infusion pump General Hospital General Hospital LZG NBW 510(k) 880.5725 and 862.1345 2 ZXP425 14500-5A (UST400 Series) K111669 - FreeStyle® Glucose Meter incorporated into the OmniPod Insulin Management Systems K042792 - iXL-II Diabetes Management System {1}------------------------------------------------ #### Insulet Corporation OmniPod® Insulin Management System 510(k) Premarket Notification Submission ### C. Device Description/Indications for Use: The OmniPod Insulin Management System provides for the management of insulin therapy and blood glucose monitoring by patients with diabetes mellitus. It is comprised of two primary components: the disposable insulin infusion pump (Pod) and an associated wireless remote controller referred to as the Personal Diabetes Manager (PDM) with an embedded blood glucose meter. ## Indications for use: The OmniPod® Insulin Management System is intended for subcutaneous delivery of insulin at set and variable rates for the management of diabetes mellitus in persons requiring insulin and for the quantitative measurement of glucose in fresh whole capillary blood (in vitro) from the finger. The glucose measurements should not be used for the diagnosis of or screening for diabetes. The PDM glucose meter is intended for single-patient use and should not be shared. Abbott FreeStyle® test strips are used with the built-in FreeStyle meter for the quantitative measurement of blood glucose in fresh whole capillary blood from the finger, upper arm and palm. Abbott Freestyle Control Solutions are used to verify that the meter and test strips are working together properly and that the test is performed correctly. #### D. Technological Characteristics: The OmniPod Insulin Management System has the same technological characteristics and is similar in design and configuration as compared to the predicate device. #### E. Summary of Non-Clinical Test/Performance Testing - Bench Insulet completed the appropriate validation and verification activities required by the Guidance for Industry and FDA Staff - Total Product Life Cycle: Infusion Pump - Premarket Notification [510(k)] Submissions Draft Guidance and other quidance, as applicable. The following performance and safety testing has confirmed the OmniPod Insulin Management System to be substantial equivalent to the predicate device: - . Usability; testing was conducted in accordance with the guidance provided in IEC 62366:2007, Medical devices -- Application of usability engineering to medical devices. AAMI HE75 - Human Factors Engineering - Design of Medical Devices, and the FDA's Medical Device Use-Safety: Incorporating Human Factors Engineering into Risk Management Guidance. Device usability has been fully validated. - . Risk Management; was completed in accordance with ISO 14971:2007 - Medical devices - Application of risk management to medical devices. Verification activities, as required by the risk analysis, demonstrated that the predetermined acceptance criteria were met and the device is safe for use. - Fluid Path Integrity Design; testing was conducted and verified that the fluid path . within the Pod can withstand the required internal pressures. - Environmental Testing, testing consisted of exposing PDMs to various environmental conditions: temperature, humidity, and pressure (atmospheric) per IEC 60601-2-24. The PDM's met the environmental testing requirements and passed all the required functional tests. {2}------------------------------------------------ #### Insulet Corporation OmniPod® Insulin Management System 510(k) Premarket Notification Submission - Soft Cannula Depth and Angle Studies: testing was conducted to verify that the soft . cannula insertion depth and angle of the Pod meets the specification. All Pod samples tested met the requirements. - Priming Volume; testing was performed to verify appropriate insertion mechanism . release and priming of cannula after firing of needle. The Pod met the priming volume requirements after soft cannula deployment as part of the activation sequence. - Rotational Sensor Study; testing was conducted to demonstrate that the rotational . sensor meets its design intent and maintains required electrical contact. Testing confirmed that design intent was met and the rotational sensor is capable of making and maintaining electrical contact in all installed positions. - Reliability and Useful Life; testing was conducted to verify the mechanical and . functional integrity of the common use features of the PDM and ensure that the useful (labeled warranty) life requirement is achieved. Testing verified that the PDM met the reliability and useful life requirements. - Basal Flow Accuracy; testing was performed to confirm the accuracy of basal rate . delivery per IEC 60601-2-24. Verification indicated that an accuracy of +/- 5% at rates of 0.05 U/hr – 30.00 U/hr was met. - Bolus Flow Accuracy; testing was performed to verify bolus delivery accuracy per IEC . 60601-2-24, Pods were tested and confirmed to provide a bolus accuracy of +/- 5% for all set values of 0.05 – 30.00 units. - Basal accuracy at maximum and minimum operational temperature; testing of the . Pod passed the basal flow rates at the minimum and maximum operational temperatures. - Nominal flow basal with vacuum transition at nominal temperature; the proposed . Pod was developed and confirmed to provide a nominal flow basal with vacuum transition at nominal temperature, as defined. - · Power interruptions; the proposed Pod battery connections were tested and determined to be perform as intended. - . Cam finger; the cam finger design has been verified to perform as defined by its requirements. - . Cannula insertion indicator; an indicator has been provided and tested to confirm when the cannula is fully inserted. - . Pod temperature testing; the Pod has been tested and confirmed to operate at a temperature range of 40°F to 98.6°F. - Biocompatibility; an assessment of the OmniPod Insulin Management System was . performed consistent with FDA's Blue Book Memorandum G95-1 "Use of ISO-10993 Biological Evaluation of Medical Devices Part 1: Evaluation and Testing" on the fluid extracts of the Pod fluid path components (Surface device, breached or compromised, permanent wear) and the surface pad and adhesive (Surface device, skin contact, permanent wear) of the Pod base. These body contacting components were tested under GLP controlled conditions. The results of the testing showed that the fluid path and Pod base adhesive pose no concerns for human use due to problems associated with biocompatibility, toxicity or dermal sensitization. {3}------------------------------------------------ #### Insulet Corporation OmniPod® Insulin Management System 510(k) Premarket Notification Submission - Insulin compatibility; the OmniPod Insulin Management System has been tested and . is considered to be compatible with NovoLog®, HumaLog®, and Apidra® U-100 insulin. - . Sterilization; the proposed device (Pod) and disposable set have been tested and will be sterilized by 100% ethylene oxide. - . Software; documentation was prepared and submitted for a MAJOR level of concern device in accordance with FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical devices. - . Electrical safety; the OmniPod Insulin Management System has been tested and successfully passed all of the relevant sections of IEC 60601-1 Medical electrical equipment, General requirements for Safety. - . RF wireless safety and performance; the OmniPod Insulin Management System has been tested and verified to ensure proper wireless communication between the Pod and PDM. - . Electromagnetic interference; the OmniPod Insulin Management System has been tested and successfully met all of the relevant sections (Radiated emissions, Electrostatic discharge immunity test, radiated radio frequency, electromagnetic field immunity, and Power frequency magnetic field immunity test) to satisfy compliance. #### F. Conclusion Insulet believes that the information and data provided clearly describes the OmniPod Insulin Management System and demonstrates that the device is adequately designed for the labeled indication for use. Performance, verification and validation testing was conducted to characterize performance of the proposed device and the predetermined acceptance criteria was met. Results of this testing have documented that the proposed device is substantially equivalent to the predicate device and is suitable for the labeled indication for use. Therefore, the proposed OmniPod Insulin Management System is considered to be substantially equivalent to the identified predicate. {4}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it. Above the caduceus, the text "HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion. Below the caduceus, the text "DEPARTMENT OF HEALTH" is arranged in a circular fashion. #### Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002 December 7, 2012 Mr. Michael J. Doyle Director Regulatory & Clinical Affairs Insulet Corporation 9 Oak Park Drive Bedford, Massachusetts 01730 Re: K122953 Trade/Device Name: OmniPod Insulin Management System Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: II Product Code: LZG, NBW Dated: November 26, 2012 Received: November 28, 2012 Dear Mr. Doyle: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {5}------------------------------------------------ Page 2 – Mr. Doyle Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Anthony De min Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health #### Enclosure {6}------------------------------------------------ ## Indications for Use # 510(k) Number (if known): K122 953 Device Name: OmniPod Insulin Management System #### Indications for Use: The OmniPod Insulin Management System is intended for subcutaneous delivery of insulin at set and variable rates for the management of diabetes mellitus in persons requiring insulin and for the quantitative measurement of glucose in fresh whole capillary blood (in vitro) from the finger. The glucose measurements should not be used for the diagnosis of or screening for diabetes. The PDM glucose meter is intended for single-patient use and should not be shared. Abbott FreeStyle test strips are used with the built-in FreeStyle meter for the quantitative measurement of blood glucose in fresh whole capillary blood from the finger, upper arm and palm. Abbott Freestyle Control Solutions are used to verify that the meter and test strips are working together properly and that the test is performed correctly. X Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Digitally signed by Richard C. Chapman Date: 2012.12.07 11:33:25 -05'00' | (Division Sign-Off) | |----------------------------------------------| | Division of Anesthesiology, General Hospital | | Infection Control, Dental Devices | | 510(k) Number: | K122953 | |----------------|---------| |----------------|---------| Page 1 of 1
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...