The Arrow Antimicrobial Pressure Injectable PICC is indicated for short-term or long-term peripheral access to the central venous system for intravenous therapy, blood sampling, infusion, pressure injection of contrast media, and allows for central venous pressure monitoring. The maximum pressure of pressure injector equipment used with the Arrow Antimicrobial Pressure Injectable PICC may not exceed 300 psi. Antimicrobial treatment on the external surface of the catheter body as well as the entire fluid pathway of the catheter has been shown to be effective in reducing microbial colonization. Antimicrobial effectiveness was evaluated using in vitro methods, and no correlation between in vitro and clinical outcome has currently been ascertained. It is not intended to be used for the treatment of existing infections.
Device Story
Arrow Antimicrobial Pressure Injectable PICC provides central venous access via peripheral insertion. Device consists of radiopaque polyurethane catheter with Blue Flex Tip; treated with chlorhexidine-based antimicrobial agent on external surface and internal fluid path to reduce microbial colonization. Available in 4.5 Fr single-lumen and 5.5 Fr double-lumen configurations. Used by clinicians in clinical settings for infusion, blood sampling, and pressure injection of contrast media (up to 5 mL/sec). Antimicrobial efficacy demonstrated in vitro and in vivo against gram-positive, gram-negative, and fungal pathogens for at least 7 days. Device includes securement features and distance markings. Output is clinical access for therapy; healthcare providers monitor patient status and catheter function. Benefits include reduced risk of microbial colonization during long-term venous access.
Clinical Evidence
No clinical data. Evidence based on bench testing and in vivo animal infection study. Bench testing verified mechanical performance (tensile strength, leakage, kink resistance, flow rates, pressure injection limits) per ISO 10555-1/3 and BS EN standards. Antimicrobial efficacy demonstrated via in vitro methods and in vivo animal model, showing minimum 4-log reduction of adherent biomass against gram-positive, gram-negative, and fungal organisms for at least 7 days.
Technological Characteristics
Polyurethane catheter body (blend of two durometers); Blue Flex Tip; chlorhexidine-based antimicrobial treatment on external surface and internal fluid path. Radiopaque. 4.5 Fr (single lumen) and 5.5 Fr (double lumen) configurations. Mechanical testing per ISO 10555-1, 10555-3, BS EN 10555-1:1997, BS EN 13868:2002, BS EN 20594-1:1994, and BS EN 1707:1997. Sterile, single-use.
Indications for Use
Indicated for patients requiring short-term or long-term peripheral access to the central venous system for IV therapy, blood sampling, infusion, contrast media pressure injection (max 300 psi), and central venous pressure monitoring. Not intended for treatment of existing infections.
Regulatory Classification
Identification
A percutaneous, implanted, long-term intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings, such as luer hubs, and accessories that facilitate the placement of the device. The device allows for repeated access to the vascular system for long-term use of 30 days or more, and it is intended for administration of fluids, medications, and nutrients; the sampling of blood; and monitoring blood pressure and temperature. The device may be constructed of metal, rubber, plastic, composite materials, or any combination of these materials and may be of single or multiple lumen design.
Special Controls
*Classification.* Class II (special controls) Guidance Document: “Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters.”
{1}------------------------------------------------
# KIDQ635
### 510(k) SUMMARY
#### SUMMARY OF SAFETY AND EFFECTIVENESS FOR ARROW ANTIMICROBIAL PRESSURE INJECTABLE PICC
AUG 27 2010
### 1. Submitter Information
| Name: | Arrow International, Inc (subsidiary of Teleflex Inc.) |
|-------------------|--------------------------------------------------------|
| Address: | 2400 Bernville Road |
| | Reading, PA 19605-9607 |
| Telephone Number: | (610) 378-0131 |
| Contact Person: | Tracy Maddock |
| | Regulatory Affairs Specialist |
| Telephone Number: | (610) 378-0131 Extension 3384 |
| Fax Number: | (610) 374-5360 |
| Email: | tracy.maddock@teleflexmedical.com |
| Date Prepared: | June 4, 2010 |
#### 2. Device Name
| Device Trade Name: | Arrow Antimicrobial Pressure Injectable Peripherally<br>Inserted Central Catheter (PICC) |
|----------------------|------------------------------------------------------------------------------------------|
| Common Name: | Peripherally Inserted Central Catheter |
| Classification Name: | Percutaneous, implanted, long-term intravascular catheter |
#### 3. Predicate Devices
| Predicate 1: | Pressure Injectable PICC (K061289) |
|--------------|----------------------------------------------------------|
| Predicate 2: | 6 French Triple Lumen Pressure Injectable PICC (K080604) |
| Predicate 3: | ARROWg+ard Blue PLUS® Multi-Lumen CVC (K993691) |
#### 4. Device Description
The Arrow Antimicrobial Pressure Injectable PICC is a short-term or long-term, single use catheter designed to provide access to the central venous system. It consists of a non-tapered, radiopaque polyurethane extruded catheter body with a softer, contoured Blue Flex Tip. The catheter is available in 4.5 Fr. single lumen and 5.5 Fr. double lumen configurations with usable lengths of 40 - 55 cm. The catheters can be used for the injection of contrast media. The maximum recommended infusion rate is 5 mL/sec. The external catheter body and the internal fluid path of the device are treated with Chlorhexidine based antimicrobial technology.
The catheters will be packaged sterile in both nursing and radiology The canoters will be passfigurations will include components to facilitate insertion. .
{2}------------------------------------------------
#### 5. Indications for Use
The Arrow Antimicrobial Pressure Injectable PICC is indicated for short-term or long-term peripheral access to the central venous system for intravenous therapy, blood sampling, infusion, pressure injection of contrast media, and allows for central venous pressure monitoring. The maximum pressure of pressure injector equipment used with the Arrow Antimicrobial Pressure Injectable PICC may not exceed 300 psi. Antimicrobial treatment on the external surface of the catheter body as well as the entire fluid pathway of the catheter has been shown to be effective in reducing microbial colonization. Antimicrobial effectiveness was evaluated using in vitro methods, and no correlation between in vitro and clinical outcome has currently been ascertained. It is not intended to be used for the treatment of existing infections.
## 6. Summary Comparing Technological Modifications
Modifications to existing Arrow products include:
- Inclusion of 4.5 Fr. single lumen and 5.5 Fr double lumen catheters to ● those ARROW PICCs already marketed. The increase in OD of the catheters allow for the subject devices to achieve 5 mL/sec pressure injection and to accommodate external / internal treatment.
- The catheter body material for the subject devices consist of a blending of . two durometers of polyurethane resin. A blue colorant was added to the catheter body resins to differentiate from non-antimicrobial PIC catheters.
- A blue colorant was added to the juncture hub material of the subject . devices to further differentiate the antimicrobial catheters.
- The catheter tip material of the subject devices includes a different . radiopacifier for enhanced radiopacity.
- Antimicrobial treatment has been applied to the external catheter body ● surface and the entire fluid path of the device.
- The antimicrobial treatment present on the external catheter body of the . subject devices consist of chlorhexidine only as opposed to chlorhexidine and silver sulfadiazine present on currently marketed Arrowgard Blue Plus central venous catheters.
#### 7. Nonclinical Testing
Bench testing was performed on the Arrow Antimicrobial Pressure Injectable PICC in accordance with ISO 10555-1, 10555-3 and FDA Guidance on Premarket Notification [510(k)] Submission for Short- Term and Long- Term Intravascular Catheters. In vitro and in vivo testing was performed to assess the safety and efficacy of the proposed device. Testing included biocompatibility, in vitro antimicrobial efficacy, and in vivo animal infection study.
{3}------------------------------------------------
## 8. Summary of Verification Activities
| Test | Acceptance Criteria | Results | | | | | | | | | | | | |
|---------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------|--|--|--|--|--|--|--|--|--|--|
| Air Leakage<br>during<br>aspiration | There shall be no air leakage in the form of an air bubble in the<br>syringe connected to the PICC after the first 5 seconds when<br>tested per BS EN ISO 10555-1:1997 Annex D.<br><br>All catheters must pass to achieve a 5% LTPD with 95%<br>confidence | Pass | | | | | | | | | | | | |
| Collapse<br>Resistance | The catheter shall not collapse during aspiration as evidenced by<br>water being able to be pulled out of the catheter when vacuum is<br>applied by a minimum of a 10 cc syringe. The extension line<br>clamps, if present, shall be in the fully constrained position.<br><br>All catheters must pass to achieve a 5% LTPD with 95%<br>confidence | Pass | | | | | | | | | | | | |
| Liquid<br>Leakage under<br>pressure | There shall be no liquid leakage in the form of a falling drop of<br>water at 300-320 kPa (43.5 -46.4) for 30 sec when tested per BS<br>EN ISO 10555-1:1997 Annex C.<br><br>All catheters must pass to achieve a 5% LTPD with 95%<br>confidence | Pass | | | | | | | | | | | | |
| Force at break<br>-Tensile<br>Testing and<br>Catheter<br>Elongation | There must be a 95% confidence level that 95% of the population<br>meets the specification. | Pass | | | | | | | | | | | | |
| | Tensile attribute | | | | | | | | | | | | | |
| | Requirement per BS EN ISO 10555-1 and 10555-3 Catheter Body Force at Break ≥ 10N Blue Flex Tip to Catheter Body Force at Break ≥ 4N Catheter Body to Juncture Hub Force at Break ≥ 10N Extension Line to Juncture Hub Force at Break ≥ 15N Extension Line to Luer Hub Force at Break ≥ 15N Catheter Body Elongation > 100% | | | | | | | | | | | | | |
| | Radio-<br>Detectability | | The optical density contrast must be at least 0.1. | Pass | | | | | | | | | | |
| | Catheter Body<br>Kink | | Does not kink at a radius greater than 0.5 inch when tested per BS<br>EN 13868:2002 Annex A under simulated <i>in vivo</i> conditions. This<br>requirement shall be met with 95% assurance. | Pass | | | | | | | | | | |
:
.
.
{4}------------------------------------------------
| Test | Acceptance Criteria | Results |
|--------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Central<br>Venous<br>Pressure<br>Monitoring | The average amplitude difference between input and output<br>signals shall be less than or equal to 1 mmHg when tested using a<br>1 Hz sinusoidal input signal. This requirement shall be met with<br>95% assurance. | Pass |
| Column<br>Strength and<br>Tip Stiffness | For catheters having a tip of different construction to the catheter<br>body, the tip shall be constructed in accordance with the<br>requirement 5.1.6 and shall be made of lower durometer material<br>than that of the catheter body.<br><br>Design of tip shall ensure that the average force required to<br>deflect or compress the tip is no greater than the average force<br>required to deflect or compress the catheter body.<br><br>NOTE: Requirement 5.1.6 referenced above is taken from an<br>internal Arrow requirements document. Requirement 5.1.6 is<br>identical to the requirement found in ISO 10555-3 Section 4.3. | Pass |
| Static Burst<br>Pressure | The maximum internal static pressure during pressure injection<br>shall not exceed the static burst pressure. | Pass |
| Static Burst<br>Pressure | The maximum internal static pressure during pressure injection<br>shall not exceed the static burst pressure. | Pass |
| Rate Limited<br>Injection<br>Testing | Each pressure injectable lumen shall withstand at least 5 repeat<br>injections without rupture or visually evident yielding of the<br>catheter when injected at the maximum indicated flow rate using<br>125 mL of contrast media or equivalent (maximum viscosity of<br>$11.8 \pm 0.2$ cP) at $37 \pm 2$ °C. | Pass |
| Pressure<br>Limited<br>Injection<br>Testing | The average flow rate of each catheter lumen shall be at least 90%<br>of the maximum indicated flow rate. | Pass |
| Ink Adhesion<br>Testing | The catheter shall remain legible when examined without<br>magnification with exposure to ChloroPrep and Iodine for 1<br>minute each, then application and removal of semi-permeable<br>adhesive dressing and Biopatch after 7 days. The acceptance<br>criteria for meeting this requirement will be a legible marking. | Pass |
| Step Stress<br>Testing | The catheters shall pass the first 10 injections at the maximum<br>flow rate without visually evident yielding or rupture. | Pass |
| Trim Tool | After trimming with the provided trimming tool and visualized<br>under 2.5X magnification, the indwelling catheter shall terminate<br>at the distal end with a square tip that:<br>Has no pointsProduces a clean, smooth surfaceWith a sample size of n=60, zero failures are required to show a | Pass |
| Test | Acceptance Criteria | Result |
| Luer Hub Slip | Luer Hub Slip<br>The hub shall meet the following Luer slip requirements with 95% confidence and a LTPD of 10% when tested per BS EN 20594-1:1994, ISO 594-1:1986 Clauses 5.1 through 5.5:<br>Gauging: The plane of the maximum diameter at the opening of the female conical fitting shall lie between the two limit planes of the gauge. Rocking shall not be evident between the gauge and the fitting made of rigid material undergoing test.<br>Liquid Leakage: No liquid leakage shall occur in the form of one or more falling drops of water.<br>Air Leakage: There shall be no signs of continued formation of air bubbles.<br>Separation force: The Luer hub shall remain attached to the reference fitting.<br>Stress cracking: There shall be no evidence of stress cracking of the fitting. | Pass |
| Luer Hub Lock | Luer Hub Lock<br>The hub shall meet the following Luer lock requirements with 95% confidence and a LTPD of 10% when tested per BS EN 1707:1997 Clauses 5.2 through 5.8:<br>Gauging: When tested with the appropriate gauge, the conical part of the lock fitting shall have the plane of the maximum diameter at the opening of the female conical fitting shall lie between the two limit planes of the gauge. Rocking shall not be evident between the gauge and the fitting made of rigid material undergoing test.<br>Liquid Leakage: No liquid leakage shall occur in the form of one or more falling drops of water.<br>Air Leakage: There shall be no signs of continued formation of air bubbles.<br>Separation force: The Luer hub shall remain attached to the reference fitting.<br>Unscrewing torque: The Luer hub shall remain attached to the reference fitting.<br>Ease of Assembly to Male Fitting: No resistance shall be observed until the taper of the fitting under test and the reference fitting fit together securely.<br>Resistance to Overriding Male to Female Luer Connection: The reference fitting shall not override the threads or lugs of the fitting under test.<br>Stress cracking: There shall be no evidence of stress cracking of the fitting | Pass |
| Catheter Securement | The catheter shall include a feature that enables the catheter to be secured to the patient's skin.<br>Demonstrate a 95% confidence level and LTPD=5% by having the suture holes for all catheters fit over the Securement posts with zero failures and the retainer wings from all catheters lock | Pass |
| Test | Acceptance Criteria | Results |
| First Article<br>Inspection | If the catheter is provided with distance markings, the marking<br>system shall indicate distance from the distal end. From the first<br>mark, the distance between marks shall not exceed 5cm. (BS EN ISO<br>10555-3: 1997 Section 4.4 and JIS T 3218:2005, Section 5.7)<br>For multilumen catheters, identification of each lumen shall be<br>apparent to the user (BS EN ISO 10555-3:1997, Item 4.5 and JIS T<br>3218:2005, Item 5.8)<br>The French size of the catheter shall be printed on the integral<br>juncture hub or in a location that can be seen after the catheter has<br>been inserted.<br>The tradename and/or name of the manufacturer of the catheter shall<br>be printed on the integral juncture hub or in a location that can be<br>seen after the catheter has been inserted. | Pass |
| Clamp<br>Closure<br>Efficacy | The clamp closure capability shall be such that when the clamps<br>are in the fully constrained position, there shall be no flow<br>through the lumen being tested when tested in accordance with<br>BS EN ISO 10555-3 Annex A or JIS T 3218 Annex C. | Pass |
| Flow<br>restriction<br>after clamping | The extension lines shall not be permanently deformed from the<br>use of extension line clamps during the maximum expected clamp<br>duration of the catheter to the point where a restriction in the<br>extension line decreases the gravity flow through the catheter<br>below the minimum gravity flow rate requirement (i.e. 90 mL/hr) | Pass |
| In vitro<br>efficacy testing<br>- external<br>antimicrobial<br>treatment | The antimicrobial agent release rate will be sufficiently slow to<br>provide efficacy against gram (+), gram (-) and fungi for a<br>minimum of 7 days.<br>Note: Efficacy will be based upon a minimum 4 log reduction of<br>adherent biomass (microbial colonization) when compared to the<br>initial inoculum concentration. | Pass |
| In vitro<br>efficacy testing<br>- internal<br>antimicrobial<br>treatment | The antimicrobial agent release rate will be sufficiently slow to<br>provide efficacy against gram (+), gram (-) and fungi for a<br>minimum of 7 days.<br>Note: Efficacy will be based upon a minimum 4 log reduction of<br>adherent biomass (microbial colonization) when compared to the<br>initial inoculum concentration. | Pass |
| In vivo animal<br>infection study | The product shall exhibit efficacy against Staphylococcus aureus<br>at minimum 7 days for in-vivo studies. Efficacy will be based<br>upon a minimum 4 log reduction of adherent biomass (microbial<br>colonization) when compared to the initial inoculum<br>concentration. | Pass |
,
.
and the comments of the comments of the comments of
and the comments of the comments of the comments of
{5}------------------------------------------------
and the comments of the comments of
{6}------------------------------------------------
#### 9. Conclusions
The Arrow Antimicrobial Pressure Injectable PICCs are substantially equivalent to the Arrow Pressure Injectable PICCs (K061289) and the Arrow 6 French Triple to the Pressure Injectable PICCs (K080604). The subject devices have the same intended use, principles of operation and technological characteristics as the predicates. The indications for use, for the proposed catheters, are the same as the
{7}------------------------------------------------
Arrow PICC predicate device K080604 with the addition of the proposed catheter's effectiveness in reducing microbial colonization.
The antimicrobial agent for the proposed device is a similar Chlorhexidine-based solution used for the ARROWg ard Blue PLUS® Multi-Lumen CVC (K993691). The process of application of the antimicrobial agent is also similar to that of the predicate device.
The results of the testing performed have demonstrated that the Arrow Antimicrobial Pressure Injectable PICC devices are safe and perform as intended. The differences, between subject devices and predicate devices, do not raise any new issues of safety and effectiveness. Thus, the Arrow Antimicrobial Pressure Injectable PICCs are substantially equivalent to the predicate devices.
{8}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the department's name around the perimeter. In the center of the seal is a stylized representation of a human figure, with three lines forming the body and head. The figure is facing to the right.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Ms. Tracy Maddock Regulatory Affairs Specialist Arrow International, Incorporated 2400 Bernville Road Reading, Pennsylvania 19605
AUG 2 7 2010
Re: K100635
Trade/Device Name: Arrow Antimicrobial Pressure Injectable Peripherally Inserted Central Catheter (PICC) Regulation Number: 21 CFR 880.5970 Regulation Name: Percutaneous, Implanted, Long-Term Intravascular Catheter Regulatory Class: II Product Code: LJS Dated: August 23, 2010 Received: August 24, 2010
Dear Ms. Maddock:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{9}------------------------------------------------
Page 2- Ms. Maddock
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Sisa Tuone
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and
Radiological Health
Enclosure
{10}------------------------------------------------
K100635
## Indications for Use
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: Arrow Antimicrobial Pressure Injectable Peripherally Inserted Central Catheter (PICC)
#### Indications for Use:
The Arrow Antimicrobial Pressure Injectable PICC is indicated for short-term or longterm peripheral access to the central venous system for intravenous therapy, blood sampling, infusion, pressure injection of contrast media, and allows for central venous pressure monitoring. The maximum pressure of pressure injector equipment used with the Arrow Antimicrobial Pressure Injectable PICC may not exceed 300 psi. Antimicrobial treatment on the external surface of the catheter body as well as the entire fluid pathway of the catheter has been shown to be effective in reducing microbial colonization. Antimicrobial effectiveness was evaluated using in vitro methods, and no correlation between in vitro and clinical outcome has currently been ascertained. It is not intended to be used for the treatment of existing infections.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Anesthesiology, General HospitalPage *_ of _*
Infection Control, Dental Devices
510(k) Number: K100635
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.