K073451 · Arrow International, Inc. · LJS · Jan 15, 2008 · General Hospital
Device Facts
Record ID
K073451
Device Name
PRESSURE INJECTABLE PICC
Applicant
Arrow International, Inc.
Product Code
LJS · General Hospital
Decision Date
Jan 15, 2008
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.5970
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Arrow International Pressure Injectable PICC is intended for short-term or long-term peripheral access to the central venous system for intravenous therapy and blood sampling.
Device Story
Radiopaque polyurethane peripherally inserted central catheter (PICC); available in 4 Fr and 5 Fr sizes; includes BlueFlex and non-BlueFlex tip configurations. Device provides peripheral access to central venous system for IV therapy, blood sampling, and power injection of contrast media. Luer hubs labeled for specific pressure injection rates (4ml/sec or 5ml/sec). Used by clinicians in clinical settings. Enables safe delivery of contrast media via power injectors up to 300 psi. Bench testing confirms structural integrity under repetitive bending, cleaning, and high-pressure injection conditions.
Clinical Evidence
Bench testing only. Testing included flow rate qualification, repeat injection, and catheter static burst pressure (post-sterilization, simulated shipping, cleaning, repetitive bending, and vesicant fluid exposure). Results demonstrate performance criteria met based on FDA-recognized standards.
Technological Characteristics
Radiopaque polyurethane catheter; 4 Fr and 5 Fr diameters; 40cm to 60cm lengths. Features BlueFlex and non-BlueFlex tip designs. Sterile kit configuration. Pressure injection rated up to 300 psi. No software or electronic components.
Indications for Use
Indicated for short or long-term peripheral access to the central venous system for intravenous therapy, blood sampling, infusion, and power injection of contrast media. Maximum power injector pressure must not exceed 300 psi.
Regulatory Classification
Identification
A percutaneous, implanted, long-term intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings, such as luer hubs, and accessories that facilitate the placement of the device. The device allows for repeated access to the vascular system for long-term use of 30 days or more, and it is intended for administration of fluids, medications, and nutrients; the sampling of blood; and monitoring blood pressure and temperature. The device may be constructed of metal, rubber, plastic, composite materials, or any combination of these materials and may be of single or multiple lumen design.
Special Controls
*Classification.* Class II (special controls) Guidance Document: “Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters.”
Predicate Devices
Arrow International's Peripherally Inserted Central Catheter (K061289)
Submission Summary (Full Text)
{1}------------------------------------------------
Kø73451 (P.1 of A2)
:
## 510(k) Summary
.
.
| | Arrow International, Inc.<br>2400 Bernville Road<br>Reading, PA 19605-9607 USA<br>JAN 15 2008 |
|------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact person: | Kevin Lentz<br>Regulatory Affairs Project Manager<br>Phone: 610-378-0131<br>Fax: 610-478-3172<br>Email: kevin.lentz@teleflexmedical.com |
| Date summary prepared: | December 7, 2007 |
| Device trade name: | Pressure Injectable Peripherally Inserted Central Catheter<br>(21 CFR 880.5970, Product Code LJS) |
| Device common name: | PICC |
| Device classification name: | Catheter, intravascular, therapeutic, long-term greater than<br>30 days |
| Legally marketed devices<br>which the device is<br>substantially equivalent: | Arrow International's Peripherally Inserted Central<br>Catheter (K061289) |
| Description of the device: | The Pressure Injectable Peripherally Inserted Central Catheters<br>have the following characteristics:<br>Radiopaque polyurethane catheters4 Fr single-lumen BlueFlex® Tip and non-BlueFlex® Tip5 Fr double lumen BlueFlex® Tip and non-BlueFlex® TipUsable length of 4 Fr catheters are 40cm to 60cmUsable length of 5 Fr catheters are 40cm to 60cmCatheters are provided in sterile kit configurationsThe 4 Fr BlueFlex® Tip, 4Fr non-BlueFlex® Tip and 5 Fr<br>BlueFlex® Tip catheters are labeled for "4ml/sec, pressure<br>injectable" on the Luer hub(s) to facilitate the proper use of<br>the deviceThe 5 Fr non-BlueFlex® Tip catheter is labeled for<br>"5ml/sec, pressure injectable" on the Luer hubs to facilitate<br>the proper use of the device |
. .
{2}------------------------------------------------
## K473451 (0.2002)
Intended use of the device: The Arrow International Pressure Injectable PICC is intended for short-term or long-term peripheral access to the central venous system for intravenous therapy and blood sampling.
## Performance tests: The following tests were performed to demonstrate substantial equivalence:
- . Flow Rate Qualification Test
- . Repeat Injection Test
- Catheter Static Burst Pressure (after sterilization, simulated . shipping conditions, cleaning solutions, repetitive bending of catheter, and use of high vesicant fluid on catheter)
The Arrow International Pressure Injectable PICC met performance data: performance criteria of the safety and effectiveness tests performed based on FDA recognized standards and guidance. The results of bench tests demonstrate that Arrow's PICC's are as safe and effective as compared to the predicate Arrow International PICC (K061289).
Summary: Arrow International's Pressure Injectable Peripherally Inserted Central Catheter (PICC) has the same intended use as the predicate device, Arrow International's PICC, (K061289). Based on the assessment of non-clinical performance data, Arrow International's Pressure Injectable Peripherally Inserted Central Catheter (PICC) is substantially equivalent to the legally marketed predicate device.
> The Arrow International predicate device (K061289) was used as a reference device to show the primary design has not changed significantly and that proper verification was done to prove that the catheter can be used safely for its intended use.
Assessment of non-clinical
{3}------------------------------------------------
Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three distinct head profiles, suggesting a sense of community or collaboration. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES. USA" is arranged in a circular pattern around the emblem.
JAN 15 2008
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Kevin Lentz Regulatory Affairs Project Manager Arrow International, Incorporated 2400 Bernville Road Reading, Pennsylvania 19605-9607
Re: K073451
Trade/Device Name: Pressure Injectable PICC Regulation Number: 880.5970 Regulation Name: Percutaneous, Implanted, Long-Term Intravascular Catheter Regulatory Class: II Product Code: LJS Dated: January 4, 2008 Received: January 7, 2008
Dear Mr. Lentz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Lentz
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Chiu Lin, Ph.D.
Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use Statement
510(k) Number : K073451
Device Name: Pressure Injectable PICC
Indications For Use: The Pressure Injectable PICC is indicated for short or long term peripheral access to the central venous system for intravenous therapy, blood sampling, infusion and power injection of contrast media. The maximum pressure of power injector equipment used with the pressure injectable PICC may not exceed 300 psi.
X Prescription Use AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED Concurrence of CDRH, Office of Device Evaluation (ODE)
signature
on of Ancementology, General Houpes), caon Clumbl, Chrisal Devices
15473451
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.