V-CATH POLY PICC

K071875 · Hdc Corp. · LJS · Nov 5, 2007 · General Hospital

Device Facts

Record IDK071875
Device NameV-CATH POLY PICC
ApplicantHdc Corp.
Product CodeLJS · General Hospital
Decision DateNov 5, 2007
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5970
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Power-V PICC is indicated for patients that require repeated venous access for infusion or injection therapy. The Power-V PICC is indicated for peripheral access to the central venous system for intravenous therapy. The Power-V PICC is indicated for dwell times less than or greater than 30 days. The maximum recommended infusion rate is 5 ml/sec. The maximum pressure of power injectors used with the Power-V PICC catheter may not exceed 300 psi.

Device Story

Power-V PICC; peripherally inserted central venous catheter; radiopaque polyurethane tubing; kink-resistant, reverse-tapered design. Used for repeated venous access, infusion, or injection therapy; supports power injection of contrast media up to 5 cc/sec at 300 psi. Available in 4 Fr/5 Fr single lumen and 5 Fr/6 Fr dual lumen configurations; 60 cm length. Attached to injection-molded polyurethane hub with luer lock extension legs. Inserted by clinicians; used in clinical settings for intravenous therapy. Benefits include reliable central venous access for patients requiring long-term or repeated therapy; enables power injection for diagnostic imaging.

Clinical Evidence

Bench testing only. Device tested for functional requirements and specifications, including power injection capability (5 cc/sec at 300 psi) and biocompatibility per ISO 10993-1 and GLP Tripartite standards.

Technological Characteristics

Radiopaque polyurethane tubing; injection-molded polyurethane hub; luer lock fittings. Kink-resistant, reverse-tapered design. 4 Fr/5 Fr single lumen and 5 Fr/6 Fr dual lumen variants; 60 cm length. Sterile, non-pyrogenic. Biocompatibility per ISO 10993-1.

Indications for Use

Indicated for patients requiring repeated venous access for infusion or injection therapy; provides peripheral access to the central venous system for intravenous therapy; suitable for dwell times less than or greater than 30 days.

Regulatory Classification

Identification

A percutaneous, implanted, long-term intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings, such as luer hubs, and accessories that facilitate the placement of the device. The device allows for repeated access to the vascular system for long-term use of 30 days or more, and it is intended for administration of fluids, medications, and nutrients; the sampling of blood; and monitoring blood pressure and temperature. The device may be constructed of metal, rubber, plastic, composite materials, or any combination of these materials and may be of single or multiple lumen design.

Special Controls

*Classification.* Class II (special controls) Guidance Document: “Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters.”

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # 510(k) Summary #### General Information Classification: Class II Trade Name: V-Cath Poly PICC NOV - 5 2007 Submitter: HDC, Corporation. 628 Gibraltar Court Milpitas, CA 95035, USA > Earl Smart QA Tel # (408) 942-7340 #### Device Identification Contact: - A. Common Name: Power-V PICC - B. Classification Name: Percutaneous Implant, long-term intravascular catheter - C. Regulation Number: 880.5970 - D. Panel: General Hospital - E. Product Code: LJS - F. Class: Class II ### Predicate Devices V-Cath Poly PICC from HDC Corporation, K033853 OMNIPICC P.I. from RITA Medical Systems, Inc., K051102 OMNIPICC P.I. from RITA Medical Systems, Inc., K062579 PowerPICC Catheter, Bard Access Systems, Inc., K033389 #### Device Description The Power-V PICC is a family of peripherally inserted central venous catheters designed to allow for contrast media studies. The catheters, made of radiopaque polyurethane tubing, are inserted pcriphcrally. Each Power-V PICC has a kink resistant, reverse tapered catheter design. The Power-V PICC is indicated for dwell times greater than 30 days. The Power-V PICC catheter assemblies have been tested to withstand power injection of worst-case viscosity injection media at 5 cc/sec with a maximum power injection pressure of 300 psi (scc Appendix B). The catheter is supplied sterile and non-pyrogenic in a variety of kit configurations (scc Appendix C). The Power-V PICC is illustrated in the drawings at the end of this section. The Power-V PICC product line has catheters in 4 Fr and 5 Fr single lumen and 5 Fr and 6 Fr dual lumen. All catheters are 60 cm long. The catheters are attached to an injection-molded polyurethane hub that has extension legs with luer lock fittings for access attachment. {1}------------------------------------------------ ### Intended Use The Power-V PICC is indicated for patients that require repeated venous access for infusion or injection therapy. The Power-V PICC is indicated for peripheral access to the central venous system for intravenous therapy. The Power-V PICC is indicated for dwell times less than or greater than 30 days. The maximum recommended infusion rate is 5 ml/sec. The maximum pressure of power injectors used with the Power-V PICC catheter may not exceed 300 psi. #### Biocompatibility Biocompatibility testing on the device was in conformance with ISO 10993-1, and GLP TRIPARTIE . The device is identical to the previously approved device OMNIPICC P.I. from RITA Medical Systems and made by Health Line International Corporation. The device is now made for HDC Corporation by Health Line International Corporation. ### Device Performance/Product Testing The Power-V met all functional requirements and specifications. #### Technology Characteristics The Power-V is equivalent technologically to the devices mentioned under predicate devices above. #### Substantial Equivalence The design, methods of manufacturing, and materials used in the Power-V device are substantially equivalent to the predicate devices. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The logo also includes the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES . USA" arranged in a circular fashion around the caduceus symbol. The text is in all caps and appears to be in a sans-serif font. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 NOV - 5 2007 Mr. Earl Smart Quality Assurance HDC Corporation 628 Gibraltar Court Milpitas, California 95035 Re: K071875 Trade/Device Name: V-Cath (Polyurethane) Power PICC (Power -V) Regulation Number: 21 CFR 880.5970 Regulation Name: Percutaneous, Implanted, Long-Term Intravascular Catheter Regulatory Class: II Product Code: LJS Dated: October 3, 2007 Received: October 5, 2007 Dear Mr. Smart: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Page 2 - Mr. Smart Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Chiu Lin, Ph.D. fa Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ K071875 1 of 1 ## Indications For Use 510 (K) NUMBER: K071875 Device Name: Indications For Use: V-Cath (Polyurethane) Power PICC (Power-V) The Power-V PICC is indicated for patients that require repeated venous access for infusion or injection therapy. The Power-V PICC is indicated for peripheral access to the central venous system for intravenous therapy. The Power-V PICC is indicated for dwell times less than or greater than 30 days. The maximum recommended infusion rate is 5 ml/sec. The maximum pressure of power injectors used with the Power-V PICC catheter may not exceed 300 psi. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OROver-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) vision Sign-Off livision of Anestheslology, General fection Control, Dental Devices 510(k) Number: KD 71875 K071875 HDC Corporation Dated: July 12, 2007 Page 11-1
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...