K251733 · The Surgical Company International BV (As Tsc Life) · LGZ · Sep 4, 2025 · General Hospital
Device Facts
Record ID
K251733
Device Name
Fluido® AirGuard System
Applicant
The Surgical Company International BV (As Tsc Life)
Product Code
LGZ · General Hospital
Decision Date
Sep 4, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5725
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Fluido® AirGuard System is for use in high blood loss surgical procedures, trauma, and any situations where replacement of warmed blood or replacement fluid is required.
Device Story
Fluido® AirGuard System (FAS) is a pressurized infusion system for warming blood and fluids to prevent perioperative hypothermia. System components include pressure chambers, fluid warmer, air guard, compressor, IV pole, and sterile single-use disposables (standard/trauma sets). Operation involves placing fluid bags into pressure chambers; compressor applies pressure to drive fluid through disposable tubing. Infrared heating technology warms fluid as it passes through the disposable set. AirGuard monitors fluid for air to prevent embolism. Used in hospital/alternate care environments by trained healthcare professionals. Output includes warmed fluid delivery; user interface displays operation/error codes, temperature, and estimated flow/volume. Benefits include rapid volume replacement and hypothermia prevention.
Clinical Evidence
No clinical data. Substantial equivalence established via non-clinical bench testing, including biocompatibility (ISO 10993), aluminum leaching, sterilization validation (ISO 11135), EMC (IEC 60601-1-2), electrical safety (ANSI/AAMI ES60601-1), software lifecycle (IEC 62304), human factors validation, hemocompatibility, and performance testing per ASTM F2172-02 and ISO 8536-4.
Technological Characteristics
Infrared heating technology; non-contact dry heating. Components: pressure chambers, fluid warmer, air guard, compressor, IV pole, single-use disposables. Max pressure 300 mmHg (+/-10%). Max flow rate ~400-900 mL/min. Power: AC source with sealed battery for indicators/alarms. Sterilization: Ethylene Oxide. Software-controlled temperature monitoring via infrared sensors at cassette entrance/exit.
Indications for Use
Indicated for infusion of crystalloid, colloid or blood product, including packed red blood cells, as volume replacement for patients suffering from blood loss due to trauma or surgery; and infusion of warmed fluid to patients with the ability to prevent hypothermia. Adult patients only.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
K252636 — Fluido Compact System · The Surgical Company International B.V. · Nov 18, 2025
K232107 — °M Warmer System · Mequ A/S · May 31, 2024
K180154 — QiF Blood and Fluid Warmer · Quality IN Flow , Ltd. · Apr 27, 2018
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
The Surgical Company International BV (as TSC Life)
Gianpaolo Soares
Regulatory Affairs Project Manager
Paalbergweg 3
Amsterdam, North Hollans 1105 AG
Netherlands
Re: K251733
Trade/Device Name: Fluido® AirGuard System
Regulation Number: 21 CFR 880.5725
Regulation Name: Infusion Pump
Regulatory Class: Class II
Product Code: LGZ, FRN
Dated: June 4, 2025
Received: June 6, 2025
Dear Gianpaolo Soares:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251733 - Gianpaolo Soares
Page 2
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K251733 - Gianpaolo Soares
Page 3
Sincerely,
Jake K. Lindstrom -S
Jake Lindstrom, Ph.D.
Assistant Director
DHT3C: Division of Drug Delivery and
General Hospital Devices and Human Factors
OHT3: Office of Gastrorenal, ObGyn,
General Hospital and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
Fluido® AirGuard System
Page 11 of 51
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This
textbox will be left blank for original applications/submissions. | | ? |
| Please provide the device trade name(s). | | ? |
| Fluido® AirGuard System | | |
| Please provide your Indications for Use below. | | ? |
| • Infusion of crystalloid, colloid or blood product, including packed red blood cells, as volume replacement
for patients suffering from blood loss due to trauma or surgery.
• Infusion of warmed fluid to patients with the ability to prevent hypothermia. | | |
| Please select the types of uses (select one or both, as
applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
TSCLIFE
Innovation with purpose
# 510(k) Summary
(in accordance with 21 CFR 807.92)
# Fluido® AirGuard System
June 2025
The Surgical Company International B.V. | Paalbergweg 3 | 1105 AG Amsterdam | The Netherlands | T +31 20 7900 2100 | letsconnect@tsc-life.com
www.tsc-life.com | CC no. 31029240 | VAT no. NL 8111.49.948.B03 | BANK: ING Commercial Banking | Bijlmerplein 888 | 1102 MG Amsterdam | The Netherlands
IBAN NL15 INGB 0653 7522 02 | BIC: INGBNL2A
{5}
TSC LIFE
Innovation with purpose
Submitter Name and Address
The Surgical Company International B.V., doing business as TSC Life
Bedrijvenpark Twente Noord 48, 7602 KR Almelo, Netherlands
Contact Person
Name: Gianpaolo Soares
Title: Regulatory Affairs Project Manager
Telephone: +31 642771951
Email: fda.tsclife-us@tsc-life.com
Device Information
Name of Device: Fluido® AirGuard System
Common or Usual Name: Infusion Pump
Classification Name: Infusion Pump, 21 CFR 880.5725
Regulatory Class: II
Product Code: LGZ
Predicate Information
Name of Predicate: The Belmont® Rapid Infuser, RI2
510k Number: K141654
This predicate has not been subject to a design-related recall.
Name of Reference Device: Ranger Irrigation Fluid Warming System
510k Number: K060939
Device Description
The Fluido® AirGuard System (FAS) is a pressurized infusion system intended for warming blood and fluids prior to administration to help prevent perioperative hypothermia. The FAS uses infrared heating technology to safely warm blood and fluids; the system detects air to prevent it from entering the patient.
The FAS consists of the Fluido Pressure Chambers, Fluido Blood and Fluid Warmer, Fluido AirGuard, Fluido Compressor, Fluido IV Pole, and Fluido Single-Use (Disposables): Standard Set and Trauma Set.
| Component | Functionality |
| --- | --- |
| Pressure Chambers | Hold the IV fluid or blood bags and control the flow of the fluid by increasing or decreasing the pressure. |
| Blood and Fluid Warmer | Warms the blood or fluid to the set temperature. |
| AirGuard | Monitors the blood or fluid for the presence of air and prevents it from reaching the patient. |
| IV Pole | Ensures a stable operating platform for the system. |
| Compressor | Supplies pressurized air to the Pressure Chambers. |
| Disposables | Define a sterile pathway for the blood and fluids to be administered to the patient, with adjustable flow rates. Allows for manual air removal. |
Indications for Use
- Infusion of crystalloid, colloid or blood product, including packed red blood cells, as volume replacement for patients suffering from blood loss due to trauma or surgery.
- Infusion of warmed fluid to patients with the ability to prevent hypothermia.
The Surgical Company International B.V. | Paalbergweg 3 | 1105 AG Amsterdam | The Netherlands | T +31 20 7900 2100 | letsconnect@tsc-life.com
www.tsc-life.com | CC no. 31029240 | VAT no. NL 8111.49.948.B03 | BANK: ING Commercial Banking | Bijlmerplein 888 | 1102 MG Amsterdam | The Netherlands
IBAN NL15 INGB 0653 7522 02 | BIC: INGBNL2A
{6}
TSCLIFE
Innovation with purpose
## Table of Contents
1. Comparison of Technological Characteristics with the Predicate Device ... 4
2. Performance Data ... 11
3. Conclusion ... 11
## Table of Tables
Table 1 Comparison of Fluido AirGuard System to Belmont Rapid Infuser RI2 ... 4
The Surgical Company International B.V. | Paalbergweg 3 | 1105 AG Amsterdam | The Netherlands | T +31 20 7900 2100 | letsconnect@tsc-life.com www.tsc-life.com | CC no. 31029240 | VAT no. NL 8111.49.948.B03 | BANK: ING Commercial Banking | Bijlmerplein 888 | 1102 MG Amsterdam | The Netherlands IBAN NL15 INGB 0653 7522 02 | BIC: INGBNL2A
{7}
TSCLIFE
Innovation with purpose
# 1. Comparison of Technological Characteristics with the Predicate Device
Table 1 Comparison of Fluido AirGuard System to Belmont Rapid Infuser RI2
| Features | Subject Device | Predicate Device | Reference Device | Summary |
| --- | --- | --- | --- | --- |
| | Fluido AirGuard System (FAS) | Belmont Rapid Infuser RI2 (K141654) | Ranger Irrigation Fluid Warming System (K060939) | |
| Manufacturer | The Surgical Company International B.V. (TSC Life) | Belmont Instrument Corporation | Formerly: Arizant Healthcare, Incorporated Currently: 3M Health Care | Not applicable |
| Regulation Name | Infusion Pump | Infusion Pump | Infusion Pump | Not applicable |
| Regulation Number | 21 CFR 880.5725 | 21 CFR 880.5725 | 21 CFR 880.5725 | Not applicable |
| Product Code | LGZ | FRN, LGZ | LGZ | Not applicable |
| Indications for Use | • Infusion of crystalloid, colloid or blood product, including packed red blood cells, as volume replacement for patients suffering from blood loss due to trauma or surgery.
• Infusion of warmed fluid to patients with the ability to prevent hypothermia. | • Infusion of crystalloid, colloid or blood product, including packed red blood cells, as volume replacement for patients suffering from blood loss due to trauma or surgery.
• Infusion of warmed fluid to rewarm patients after surgery or for hypothermia. | The Ranger Irrigation Fluid Warming System is intended to warm irrigation fluids. | Substantially Equivalent and is a subset of the predicate devices Indication for Use. |
The Surgical Company International B.V. | Paalbergweg 3 | 1105 AG Amsterdam | The Netherlands | T +31 20 7900 2100 | letsconnect@tsc-life.com www.tsc-life.com | CC no. 31029240 | VAT no. NL 8111.49.948.B03 | BANK: ING
Commercial Banking | Bijlmerplein 888 | 1102 MG Amsterdam | The Netherlands
IBAN NL15 INGB 0653 7522 02 | BIC: INGBNL2A
{8}
TSCLIFE
Innovation with purpose
| Features | Subject Device | Predicate Device | Reference Device | Summary |
| --- | --- | --- | --- | --- |
| | Fluido AirGuard System (FAS) | Belmont Rapid Infuser RI2 (K141654) | Ranger Irrigation Fluid Warming System (K060939) | |
| Intended Use | The Fluido® AirGuard System is for use in high blood loss surgical procedures, trauma, and any situations where replacement of warmed blood or replacement fluid is required. | The Belmont Rapid Infuser RI2 is for use in high blood loss surgical procedures, trauma, and any situations where rapid replacement of warmed blood or replacement fluid 10 - 1000 ml/min is required. It can also be used to deliver irrigation fluids at up to 1000 ml/min. | The Ranger Irrigation Fluid Warming System is intended to warm irrigation fluids. | Substantially Equivalent. |
| Intended User | The Fluido AirGuard System is intended to be used by trained healthcare professionals. | The Belmont Rapid Infuser RI2 is intended to be used by trained healthcare professionals. | The Ranger Irrigation Fluid Warming System is intended to be used by trained healthcare professionals. | Substantially Equivalent. |
| Patient Population | Adult Patients Only | Adult Patients Only | Adult and Pediatric Use | Substantially Equivalent. |
| Intended Use Environment | The Fluido AirGuard System is intended for general operation in a hospital or alternate care environment. | The operating environment for the Belmont Rapid Infuser RI2 is general operation in hospital or alternate care environments. The Belmont Rapid Infuser RI2 will be subject to the temperature, humidity, and pressure typical of a health care environment. Sources of shock, drop, and vibration are | Healthcare facilities. | Substantially Equivalent |
The Surgical Company International B.V. | Paalbergweg 3 | 1105 AG Amsterdam | The Netherlands | T +31 20 7900 2100 | letsconnect@tsc-life.com www.tsc-life.com | CC no. 31029240 | VAT no. NL 8111.49.948.B03 | BANK: ING
Commercial Banking | Bijlmerplein 888 | 1102 MG Amsterdam | The Netherlands
IBAN NL15 INGB 0653 7522 02 | BIC: INGBNL2A
{9}
TSCLIFE
Innovation with purpose
| Features | Subject Device | Predicate Device | Reference Device | Summary |
| --- | --- | --- | --- | --- |
| | Fluido AirGuard System (FAS) | Belmont Rapid Infuser RI2 (K141654) | Ranger Irrigation Fluid Warming System (K060939) | |
| | | also those typically found in a health care environment. | | |
| Principle of Operation | Non-contact dry heating | Non-contact dry heating | Contact dry heating | Substantially Equivalent |
| Intended Fluid(s) to be Warmed | Crystalloid, colloid, or blood product, including packed red blood cells. | Crystalloid, colloid, or blood product, including packed red blood cells. | Irrigation fluids (Not applicable to SE discussion) | Substantially Equivalent |
| Delivery Mechanism/Intended Route of Administration | Intravenous Delivery | Intravenous Delivery | Surgical sites (Not applicable to SE discussion) | Substantially Equivalent |
| Components | Control System | Control System | Control System | Substantially Equivalent. |
| | IV Pole | IV Pole (optional) | IV Pole | |
| | Single use plastic disposables | Single use plastic disposables | Single use plastic disposables | |
| Monitoring and alarm system? | Yes | Yes | Yes | Substantially Equivalent |
| Power Source | AC Power Source, sealed battery (only operates indicators and alarms for AirGuard) | AC Power Source, sealed battery (only operates as a backup battery for mobile transport of the patient) | AC Power Source, no battery used | Substantially Equivalent. |
| User Interface Controls | Temperature | Flow, volume, temperature | None – can view temperature, but no changing from 41°C set-point. | Substantially Equivalent. For FAS, the flow and volume control is not managed through the |
The Surgical Company International B.V. | Paalbergweg 3 | 1105 AG Amsterdam | The Netherlands | T +31 20 7900 2100 | letsconnect@tsc-life.com www.tsc-life.com | CC no. 31029240 | VAT no. NL 8111.49.948.B03 | BANK: ING
Commercial Banking | Bijlmerplein 888 | 1102 MG Amsterdam | The Netherlands
IBAN NL15 INGB 0653 7522 02 | BIC: INGBNL2A
{10}
TSCLIFE
Innovation with purpose
| Features | Subject Device | Predicate Device | Reference Device | Summary |
| --- | --- | --- | --- | --- |
| | Fluido AirGuard System (FAS) | Belmont Rapid Infuser RI2 (K141654) | Ranger Irrigation Fluid Warming System (K060939) | |
| | | | (Not applicable to SE discussion) | user interface. Instead, it is controlled by pressurizing the fluid bag and using the rolling clamp within the disposable set. The user interface, however, does display the estimated flow and volume. |
| User Displays | Operation codes, error codes, warning codes, and maintenance codes | Operation codes, error codes, warning codes, and maintenance codes | Error codes, warning codes | Substantially Equivalent |
| Maximum In-Line Pressure | 300 mmHg | 300 mmHg | None, no pressure system | Substantially Equivalent |
| Prime Volume | Standard Disposable: 90ml
Trauma Set: 145 ml | 100 mL | 308 mL
(Not applicable to SE discussion) | Substantially Equivalent, |
| Max Flow Rate | Standard set ≈ 400 mL per minute
Trauma Set ≈ 900 mL per minute | 1,000 mL per minute | 867 mL per minute
(Not applicable to SE discussion) | Substantially equivalent |
| Warmer Type | In-Line | In-Line | In-Line | Substantially Equivalent |
The Surgical Company International B.V. | Paalbergweg 3 | 1105 AG Amsterdam | The Netherlands | T +31 20 7900 2100 | letsconnect@tsc-life.com www.tsc-life.com | CC no. 31029240 | VAT no. NL 8111.49.948.B03 | BANK: ING
Commercial Banking | Bijlmerplein 888 | 1102 MG Amsterdam | The Netherlands
IBAN NL15 INGB 0653 7522 02 | BIC: INGBNL2A
{11}
TSCLIFE
Innovation with purpose
| Features | Subject Device | Predicate Device | Reference Device | Summary |
| --- | --- | --- | --- | --- |
| | Fluido AirGuard System (FAS) | Belmont Rapid Infuser RI2 (K141654) | Ranger Irrigation Fluid Warming System (K060939) | |
| | | | (Not applicable to SE discussion) | |
| Warming Technology | Infrared Heating | Electromagnetic Inductive Heater | Metal plate heated by electrical resistance (Not applicable to SE discussion) | Substantially equivalent. The subject device shares substantially equivalent in-line warming of blood and fluids through the use of an indirect heat source, with the Belmont Rapid Infuser RI2 predicate device. |
| Heating Methodology | The hardware device radiates infrared heat directly to the blood or IV fluid circulating through the single use plastic disposables. | The hardware device heats the heat exchanger, which transfers the heat to the blood or IV fluid circulating through the single use plastic disposables. | A metal plate is heated by electrical resistance, the disposable cassette contacts the plates and the fluid is warmed. (Not applicable to SE discussion) | Substantially Equivalent |
| Heating Control | Input/Output Infrared Temperature Sensors, controlled by software | Input/Output Infrared Sensor controlled by software | Temperature is measured by temperature sensor and controlled by software. (Not applicable to SE discussion) | Substantially Equivalent |
The Surgical Company International B.V. | Paalbergweg 3 | 1105 AG Amsterdam | The Netherlands | T +31 20 7900 2100 | letsconnect@tsc-life.com www.tsc-life.com | CC no. 31029240 | VAT no. NL 8111.49.948.B03 | BANK: ING
Commercial Banking | Bijlmerplein 888 | 1102 MG Amsterdam | The Netherlands
IBAN NL15 INGB 0653 7522 02 | BIC: INGBNL2A
{12}
TSCLIFE
Innovation with purpose
| Features | Subject Device | Predicate Device | Reference Device | Summary |
| --- | --- | --- | --- | --- |
| | Fluido AirGuard System (FAS) | Belmont Rapid Infuser RI2 (K141654) | Ranger Irrigation Fluid Warming System (K060939) | |
| Temperature Monitoring | The temperature sensors measure the temperature at the entrance and the exit of the cassette, located in the single use disposable component. | Infrared temperature sensors which are mounted at the entrance and the exit of the heat exchanger, monitor the temperature of fluid as it enters and exits the heat exchanger. | Sensor measures temperature before it leaves the cassette. | Substantially Equivalent. |
| Temperature viewing (on screen)? | Yes | Yes | Yes (Not applicable to SE discussion) | Substantially Equivalent |
| Occlusion Detection | No occlusion detection system | Integrated occlusion detection system | No occlusion detection system | Substantially Equivalent |
| Mechanism of Pressurizing Fluid | Compression of fluid bag | Roller compression of IV tubing | No pressure | Substantially Equivalent |
| Contains software? | Yes | Yes | Yes | Substantially Equivalent |
| Alarms | Audible and visual alarms | Audible and visual alarms | Audible and visual alarms | Substantially Equivalent |
| Protection from air embolism | Yes | Yes | No (Not applicable to SE discussion) | Substantially Equivalent |
| Pressure Regulator (max) | 300 mmHg (+/-10%) | 300 mmHg (+/-50mm Hg) | N/A – no pressure applied (Not applicable to SE discussion) | Substantially Equivalent |
| Maximum Heating Power (watts) | 1200W | 1440W | 900W | Substantially Equivalent |
The Surgical Company International B.V. | Paalbergweg 3 | 1105 AG Amsterdam | The Netherlands | T +31 20 7900 2100 | letsconnect@tsc-life.com www.tsc-life.com | CC no. 31029240 | VAT no. NL 8111.49.948.B03 | BANK: ING
Commercial Banking | Bijlmerplein 888 | 1102 MG Amsterdam | The Netherlands
IBAN NL15 INGB 0653 7522 02 | BIC: INGBNL2A
{13}
TSCLIFE
Innovation with purpose
| Features | Subject Device | Predicate Device | Reference Device | Summary |
| --- | --- | --- | --- | --- |
| | Fluido AirGuard System (FAS) | Belmont Rapid Infuser RI2 (K141654) | Ranger Irrigation Fluid Warming System (K060939) | |
| | | | (Not applicable to SE discussion) | |
| Cardiac Floating Capability | Yes | Yes | No (Not applicable to SE discussion) | Substantially Equivalent |
| Sterilization Method for single use disposables | Ethylene Oxide | Ethylene Oxide | Ethylene Oxide (Not applicable to SE discussion) | Substantially Equivalent |
The Surgical Company International B.V. | Paalbergweg 3 | 1105 AG Amsterdam | The Netherlands | T +31 20 7900 2100 | letsconnect@tsc-life.com www.tsc-life.com | CC no. 31029240 | VAT no. NL 8111.49.948.B03 | BANK: ING Commercial Banking | Bijlmerplein 888 | 1102 MG Amsterdam | The Netherlands IBAN NL15 INGB 0653 7522 02 | BIC: INGBNL2A
{14}
TSCLIFE
Innovation with purpose
## 2. Performance Data
**Nonclinical safety testing:**
Nonclinical testing was conducted to support the safety and performance of the Fluido AirGuard System. This testing covered a range of subjects, including but not limited to:
1. Biocompatibility Testing per ISO 10993 and USP
- Cytotoxicity
- Irritation
- Sensitization
- Material Mediated Pyrogenicity
- Bacterial Mediated Pyrogenicity
- Acute Systemic Toxicity
2. Aluminum Leaching Testing on disposable sets, where aluminum levels were in accordance with the limits stated in 21 CFR 201.323
3. Transport testing per ISTA 3A
4. Sterilization validation per ISO 11135
5. EMC Testing per IEC 60601-1-2
6. UL Electrical Safety Testing per ANSI/AAMI ES60601-1
7. Cybersecurity Testing
**Nonclinical performance testing:**
1. Software Lifecycle Testing per IEC 62304
2. Human Factors Validation Testing, in accordance with FDA Guidance “Applying Human Factors and Usability Engineering to Medical Devices” issued February 2016 and IEC 62366
3. Hemocompatibility (hemolysis, platelet aggregation, S-protein, C-protein)
4. Non-Clinical Bench Performance Testing per ASTM F2172-02
5. Flow rate testing per ISO 8536-4
All non-clinical test results were found acceptable.
**Clinical testing:**
Clinical testing was unnecessary to establish substantial equivalence.
**Nonclinical test comparison to predicate:**
Substantial equivalence between the FAS and the Belmont Rapid Infuser RI2 was demonstrated through the following tests: warm-up time, output temperature, maximum flow speed, overtemperature alarm and alarm suppression, air removal, pressure comparison, minimum/maximum operating temperatures.
## 3. Conclusion
Based on the performance testing conducted according to FDA recognized standards and guidance documents, along with the comparative analysis provided in the substantial equivalence table, it can be concluded that the FAS is substantially equivalent to the Belmont Rapid Infuser RI2 cleared under K141654.
The Surgical Company International B.V. | Paalbergweg 3 | 1105 AG Amsterdam | The Netherlands | T +31 20 7900 2100 | letsconnect@tsc-life.com www.tsc-life.com | CC no. 31029240 | VAT no. NL 8111.49.948.B03 | BANK: ING Commercial Banking | Bijlmerplein 888 | 1102 MG Amsterdam | The Netherlands IBAN NL15 INGB 0653 7522 02 | BIC: INGBNL2A
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.