Infusion of crystalloid, colloid, or blood product, including packed red blood cells, as volume replacement for patients suffering from blood loss due to trauma or surgery. Infusion of warmed fluid to rewarm patients after surgery or for hypothermia. Infusion of warmed fluid for irrigation in urology procedures.
Device Story
The Belmont® Rapid Infuser, RI2, is an infusion pump system designed for high-speed delivery of warmed fluids. It accepts blood, replacement IV fluids, or irrigation fluids as input. The device utilizes a microprocessor-controlled mechanical system to pump fluids at user-set rates from 10 to 1000 ml/min; a 2.5 ml/min rate is available without heating. It monitors temperature, line pressure, and air in the fluid path, triggering alarms for unsafe conditions. A hardware override circuit ensures safety during computer failure. The system features a touch screen for displaying flow rate, volume, temperature, and pressure. A battery backup supports mobile transport, though fluid warming is disabled during battery operation. Used in clinical settings like ORs or trauma centers, it is operated by healthcare professionals. The device provides clinicians with real-time monitoring and automated safety stops, facilitating rapid fluid resuscitation and temperature management, which helps stabilize patients experiencing significant blood loss or hypothermia.
Clinical Evidence
Bench testing only. Testing verified system performance across flow rates (10-1000 ml/min), varying input temperatures, back pressures, and fluid viscosities (crystalloids and packed red blood cells). Validation confirmed the system's ability to warm fluids to physiological temperatures and detect/alarm for over-temperature, high line pressure, air in line, and internal faults. Risk assessments were conducted for keypad operation and fluid spillage. Software was validated per IEC 62304. The device met all specifications for safety and electromagnetic compatibility.
Technological Characteristics
Microprocessor-controlled infusion pump. Features a monochromatic high-contrast LCD. Operates on AC power with battery backup for transport (warming disabled on battery). Monitors temperature, line pressure, and air in-line. Software developed per IEC 62304. Hardware includes a dedicated override circuit for safety. Sterilization/biocompatibility not specified; standard electrical safety and EMC compliance.
Indications for Use
Indicated for patients requiring rapid volume replacement due to trauma or surgery-related blood loss, patients requiring rewarming post-surgery or for hypothermia, and patients requiring warmed irrigation fluids during urology procedures.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for Belmont Instrument Corporation. The logo features a stylized letter "B" in a bold, flowing font. To the right of the "B" is the word "Belmont" in a bold, sans-serif font, and below that is the phrase "Instrument Corporation" in a smaller, sans-serif font.
was with the the one would but the restended
K141654
1 8 2014
## PREMARKET NOTIFICATION 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS [As Required By 21 CFR 807.92(c)]
| Submitter Name and Address: | Belmont Instrument Corporation<br>780 Boston Road<br>Billerica, MA 01821 |
|-------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Uraiwan P. Labadini<br>Quality Assurance/Regulatory Affairs Manager<br>(508) 663-0212 Ext. 128<br>Telephone:<br>(508) 663-0214<br>Fax:<br>Email address: ulabadini@belmontinstrument.com |
| Date Summary Prepared: | July 17, 2014 |
| Trade/Proprietary Name of Device: The Belmont® Rapid Infuser. R12 | |
| Common Name: | Infusion Pump |
| Classification Name: | Infusion Pump, 21 CFR 880.5725<br>Warmer, Thermal, Infusion Fluid |
| Legally Marketed Device Under<br>Which Substantial Equivalence is<br>Claimed: | The Belmont® Rapid Infuser cleared for market<br>entry originally under 510(k) K972284,<br>18 modification cleared under 510(k) K032674, and<br>200 modification cleared under 510(k) K091855. |
Device Description:
The Belmont® Rapid Infuser, RI2, (same as the existing Rapid Infuser) combines advanced microprocessor technology with an efficient mechanical system to provide a high speed, simple and safe system for rapid infusion of warmed fluid. . The Rapid Infuser infuses blood, replacement IV fluids or irrigation fluids warmed to physiologic temperature at user-set rates from 10 to 1000 milliliters per minute (ml/min). A low infusion rate at 2.5 ml/min (150 ml/hr) is also available without heating.
The system monitors temperature, line pressure, and air in the fluid path to ensure safe operation and alarms at all unsafe conditions. A hardware override circuit prevents unsafe operation in case
MADE IN USA
7-1
{1}------------------------------------------------
of system computer failure. A touch screen displays flow rate, total fluid infused, temperature, line pressure, alarm and status messages and proper procedures to proceed safely after an alarm situation.
A battery backup allows for mobile transport of the patient and system. During battery operation, fluid warming is disabled while pump operation and safety monitoring remain active.
The Belmont® Rapid Infuser, R12, is for use in high blood loss surgical procedures, trauma and any situation where rapid replacement of warmed blood or replacement fluid at 10 - 1000 ml/min is required. It can also be used to deliver irrigation fluids at rates up to 1000 ml/min.
Corporation is claiming Belmont Instrument substantial equivalence of the Belmont® Rapid Infuser, RI2, to the Belmont® Rapid Infuser which was originally cleared to market under 510(k) K972284 on November 13, 1997, and to the Belmont® Rapid Infuser 750 ml/min flow rate was cleared to market under 510(k) K032674 on September 16, 2003, and to the Belmont® Rapid Infuser 1000 ml/min flow rate was cleared to market under 510(k) K091855 on July 1, 2009.
The Belmont® Rapid Infuser has been in clinical use for more than 14 years; consequently a component used in the system had become obsolete and hard to obtain. Therefore, the system hardware (chip replacement), and software (to correspond with the chip changes) have been upgraded. At the same time, we are using a brighter monochromatic display with a higher contrast LCD to improve visibility. We have performed a rigorous validation to ensure that these updates are backward compatible with our existing system and no major functions have been altered.
Intended Use:
Comparison of Technological Characteristics of the Device vs. the Legally Marketed Device:
Specific Modifications being reviewed in this submission:
{2}------------------------------------------------
Brief Discussion of Nonclinical Tests and their Results Submitted in the Application:
In order to verify performance of the Belmont® . Rapid Infuser, R12, in support of substantial equivalence, the following tests were carried out:
- a. The ability of the system to pump fluids accurately over the full range of flow rate and operating conditions including different input fluid temperatures, different back pressure levels, change in ambient temperature, and changes in fluid viscosity including that of crystalloid solution and packed red blood cells.
- b. The ability of the system to warm cold fluids to physiological temperature over the full range of flow rate and operating conditions.
- c. The ability of the system to detect and alarm at unsafe or ineffective operating conditions including over temperature, the unsafe high line pressure condition, out of fluid, air in the line and at any of several internal fault conditions.
- d. We performed risk assessment regarding the keypad, i.e., key stick, key bounce/failure, operator error by pressing the wrong key, unintentional activation of the keypad, and fluid spillage on the keypad. We performed the verification and validation tests to ensure that these risks were mitigated and method of controls were implemented correctly.
- e. We generated a software life cycle in according to the IEC 62304. We validated each algorithm for each function element in the software specification using an approved procedure. We also performed extensive testing for the entire system with software in place to test both the hardware and software functions to ensure that all system requirements are met.
- Medical Electrical Equipment Part 1: General ﺖ Requirements for Basic Safety and Essential Performance.
- Equipment Collateral g. Medical Electrical Standard: Electromagnetic Compatibility
- The Belmont® Rapid Infuser, R12, performed within specifications in all of the above tests.
{3}------------------------------------------------
. :
Conclusion:
The Belmont® Rapid Infuser, RI2, is substantially equivalent to the Belmont® Rapid Infuser which received 510(k) approval at various times. Both systems are capable of high infusion flow rates, infusate at physiological while maintaining temperature, and both systems are suitable for use with blood products. Both systems monitor flow rate, temperature, line pressure, and the presence of air at the input line or within the system, and alarm and stop the system at all unsafe conditions.
The tests demonstrate that the Belmont® Rapid Infuser, R12, is substantial equivalent and performs as well as the legally marketed device.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized depiction of an eagle or bird with three curved lines forming its body and wings. The logo is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 0903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 18, 2014
Belmont Instrument Corporation Uraiwan Labadini Quality Assurance / Regulatory Affairs Manager 780 Boston Road Billerica, Massachusetts 01821
Re: K141654
Trade/Device Name: Belmont® Rapid Infuser, R12 Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: Class II Product Code: FRN, LGZ Dated: June 18, 2014 Received: June 20, 2014
Dear Ms. Labadini:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Ms. Labadini
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/Reportal?roblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tejashri Purohit-Sheth, M.D.
Tejashri Purohit Sheth, M.D. Clinical Deputy Director
DAGRID/ODE/CDRH FOR
Erin 1. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## Indications for Use
510(k) Number (if known): K141654
Device Name: The Belmont® Rapid Infuser, RI2
Indications for Use:
- Infusion of crystalloid, colloid, or blood product, including packed red A. blood cells, as volume replacement for patients suffering from blood loss due to trauma or surgery.
- B. Infusion of warmed fluid to rewarm patients after surgery or for hypothermia.
- ். Infusion of warmed fluid for irrigation in urology procedures.
Prescription Use _ X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Image /page/6/Picture/12 description: The image shows the FDA logo. The logo is made up of the letters FDA, with each letter being formed by three parallel lines. The letters are arranged in a slightly slanted fashion, with the F on the left, the D in the middle, and the A on the right.
Mary E. Brooks -A 2014.07.17 Page 1 of 1 15:27:47 -04'00'
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.