The Belmont Rapid Infuser RI-2 is designed to be used in general or alternative care environments to provide warmed blood and fluids to any patients ≥ 10 kg requiring warmed infusion from 2.5 mL/min to 1000 mL/min. - Infusion of crystalloid, colloid, or blood product, including packed red blood cells, as volume replacement for patients suffering from blood loss due to trauma or surgery. - Infusion of warmed fluid to rewarm patients after surgery or for hypothermia. - Infusion of warmed fluid for irrigation in urology procedures. The 3.0L reservoir is an optional accessory for use in adults only.
Device Story
High-speed infusion pump with integrated fluid warming; uses roller-type peristaltic pump and inductively heated stainless steel annular rings. Inputs: user-set flow rates (2.5-1000 mL/min), temperature, and line pressure. System monitors temperature, pressure, and air (ultrasonic detection); features hardware override for computer failure. Outputs: warmed infusate, real-time display of flow rate, volume, temperature, and pressure. Used in hospital/alternative care by clinicians. Battery backup allows transport (warming disabled, pump limited to 50 mL/min). Recirculation mode purges air. Output data informs clinical decisions regarding fluid resuscitation and patient temperature management.
Clinical Evidence
Bench testing only. No clinical data. Testing included verification of product requirements, IPX water intrusion (IEC 60529), electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), alarm systems (IEC 60601-1-8), software validation (IEC 62304), and biocompatibility (ISO 10993-1).
Indicated for patients ≥ 10 kg requiring warmed infusion (2.5-1000 mL/min) of crystalloid, colloid, or blood products for volume replacement (trauma/surgery), rewarming (post-surgery/hypothermia), or irrigation (urology). Contraindicated for platelets, cryoprecipitates, granulocyte suspensions, unprocessed/non-anti-coagulated blood, drug administration, and adding calcium-containing solutions, dextrose in water, or hypotonic sodium chloride to blood components.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.
December 23, 2024
Belmont Medical Technologies Lida Reed Director Quality Assurance and Regulatory Affairs 780 Boston Road Billerica, Massachusetts 01821
Re: K242735
Trade/Device Name: Belmont Rapid Infuser, RI-2 (Rapid Infuser RI-2, 1000 ml/min); Belmont Rapid Infuser, RI-2 (Rapid Infuser RI-2, 750 ml/min) Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: Class II Product Code: FRN, LGZ Dated: November 21, 2024 Received: November 21, 2024
Dear Lida Reed:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Jake K. Lindstrom -S
Jake Lindstrom, Ph.D Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K242735
#### Device Name
Belmont Rapid Infuser, RI-2 (Rapid Infuser RI-2, 1000 ml/min) Belmont Rapid Infuser, RI-2 (Rapid Infuser RI-2, 750 ml/min)
#### Indications for Use (Describe)
The Belmont Rapid Infuser RI-2 is designed to be used in general or alternative care environments to provide warmed blood and fluids to any patients ≥ 10 kg requiring warmed infusion from 2.5 mL/min to 1000 mL/min. - Infusion of crystalloid, colloid, or blood product, including packed red blood cells, as volume replacement for patients suffering from blood loss due to trauma or surgery.
- Infusion of warmed fluid to rewarm patients after surgery or for hypothermia.
- Infusion of warmed fluid for irrigation in urology procedures.
The 3.0L reservoir is an optional accessory for use in adults only.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
| Date: | December 17, 2024 | |
|----------------------------|--------------------------------------------------------------------------------------------------------------------------|--|
| Company: | Belmont Medical Technologies<br>780 Boston Road<br>Billerica, MA 01821 | |
| Official Contact: | Lida Reed<br>Director Quality Assurance and Regulatory Affairs<br>Phone: 978.696.9245<br>Email: Ireed@belmontmedtech.com | |
| Proprietary or Trade Name: | Belmont® Rapid Infuser, RI-2 | |
| Common/Usual Name: | Infusion Pump | |
| Classification Name: | 21 CFR 880.5725, Class II<br>Classification Product Code: FRN<br>Subsequent Product Code: LGZ | |
| Predicate Device: | K141654: The Belmont Rapid Infuser<br>This predicate has not been subject to a design-related recall | |
#### Device Description:
The Belmont® Rapid Infuser, RI-2 combines advanced microprocessor technology with an efficient mechanical system to provide a high speed, simple and safe system for rapid infusion of warmed fluids. The Rapid Infuses blood, replacement IV fluids or irrigation fluids warmed to physiologic temperature at user-set rates from 10 to 1000 milliliters per minute (mL/min). Low infusion rates of 2.5 mL/min (150 mL/hr) and 5.0 mL/min (300 mL/hr) are also available to keep the venous line open.
The system monitors temperature, line pressure, and air in the fluid path to ensure safe operation and alarms at all unsafe conditions. A hardware override circuit prevents unsafe operation in case of system computer failure. A touch screen displays flow rate, total fluid infused, temperature of the infusate, line pressure, alarm and status messages and proper procedures to proceed safely after an alarm situation.
A battery backup allows for mobile transport of the patient and system. During battery operation, fluid warming is disabled while pump operation limited to 50mL/min and safety monitoring remain active.
#### Indications for Use:
The Belmont Rapid Infuser RI-2 is designed to be used in general operation in hospital or alternative care environments to provide warmed blood and fluids to any patients ≥ 10 kg requiring warmed infusion from 2.5 mL/min to 1000 mL/min.
- Infusion of crystalloid, colloid, or blood product, including packed red blood cells, as volume replacement for patients suffering from blood loss due to trauma or surgery.
- . Infusion of warmed fluid to rewarm patients after surgery or for hypothermia.
- Infusion of warmed fluid for irrigation in urology procedures. .
The 3.0L reservoir is an optional accessory for use in adults only.
{5}------------------------------------------------
#### Contraindications:
- . The system should not be used to warm platelets, cryoprecipitates, granulocyte suspensions or unprocessed / non-anti-coagulated blood products.
- This system is not intended for drug administration.
- . Calcium containing solutions (ex. Lactated Ringer's solution), dextrose in water, and hypotonic sodium chloride solutions should not be added to blood components.
#### Substantial Equivalence:
The Belmont® Rapid Infuser RI-2 is substantially equivalent to the predicate device, the Belmont® Rapid Infuser RI-2 (510(k) K141654). The table below presents the similarities and differences between the products for substantial equivalence purposes. The differences between the subject device and the predicate device do not raise any new issues of safety and effectiveness. Performance data are available to support substantial equivalence.
| Element of Comparison | Predicate Device:<br>Belmont® Rapid<br>Infuser, RI-2<br>[510(k) K141654] | Subject Device:<br>Modified Belmont®<br>Rapid Infuser, RI-2 | Substantial<br>Equivalence |
|------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1.0 General: | | | |
| • Common/<br>Classification | Infusion Pump/<br>Pump, Infusion<br>Warmer Thermal Infusion<br>Fluid Warmer | Infusion Pump/<br>Pump, Infusion<br>Warmer Thermal<br>Infusion Fluid Warmer | SAME |
| • Regulation Number | 21 CFR 880.5725<br>FRN | 21 CFR 880.5725<br>FRN | SAME |
| • FDA Product Codes | LGZ | LGZ | SAME |
| • Trade name | Rapid Infuser | Rapid Infuser | SAME |
| • Model Number | RI2 | RI2 | SAME |
| Element of Comparison | Predicate Device:<br>Belmont® Rapid<br>Infuser, RI-2<br>[510(k) K141654] | Subject Device:<br>Modified Belmont®<br>Rapid Infuser, RI-2 | Substantial<br>Equivalence |
| • Indications for use | • Infusion of crystalloid, colloid, or blood product, including packed red blood cells, as volume replacement for patients suffering from blood loss due to trauma or surgery,<br>• Infusion of warmed fluid to rewarm patients after surgery or for hyperthermia, and<br>• Infusion of warmed fluid for irrigation in urology procedures. | The Belmont Rapid Infuser RI-2 is designed to be used in general operation in hospital or alternative care environments to provide warmed blood and fluids to any patients ≥ 10 kg requiring warmed infusion from 2.5 mL/min to 1000 mL/min.<br>• Infusion of crystalloid, colloid, or blood product, including packed red blood cells, as volume replacement for patients suffering from blood loss due to trauma or surgery,<br>• Infusion of warmed fluid to rewarm patients after surgery or for hyperthermia, and<br>• Infusion of warmed fluid for irrigation in urology procedures.<br>The 3.0L reservoir is an optional accessory for use in adults only. | SIMILAR<br>The subject device clarifies that the device is for use in patients ≥ 10 kg. This clarification does not impact the intended use which is rapid infusion of warmed blood, blood products and fluids. |
| • Device description | High speed infusion pump with fluid warming. | High speed infusion pump with fluid warming. | SAME |
| • Infusate | Blood, blood product, crystalloid and colloid solutions. | Blood, blood product, crystalloid and colloid solutions. | SAME |
| 2.0 Fluid Infusion (Pump) Characteristics: | | | |
| • Available flow rates | 2.5, 5.0, 10 to 1000 mL/min | 2.5, 5.0, 10 to 1000 mL/min | SAME |
| 510(k) Summary | | | |
| Element of Comparison | Predicate Device:<br>Belmont® Rapid<br>Infuser, RI-2<br>[510(k) K141654] | Subject Device:<br>Modified Belmont®<br>Rapid Infuser, RI-2 | Substantial<br>Equivalence |
| • Flow rate accuracy | ± 10% from 20 – 1000 mL/min<br>± 25% for 2.5, 5.0, 10 mL/min | ± 10% from 20 – 1000 mL/min<br>± 25% for 2.5, 5.0, 10 mL/min | SAME |
| • Type of Fluid Pump | Roller type peristaltic pump | Roller type peristaltic pump | SAME |
| • Digital readout of flow rate | Yes | Yes | SAME |
| • Automatic reduction of pump speed, if required, to maintain user-set line pressure limit | Yes | Yes | SAME |
| • Digital setting of speed by user from front panel | Yes | Yes | SAME |
| • Control and maintain pump speed with variable pressure head | Yes | Yes | SAME |
| • Maximum pressure at which set flow maintained (mmHg) | 300 mmHg standard.<br>Can be changed by the Operator to 100 or up to 300 mmHg, in 50 mmHg increments. | 300 mmHg standard.<br>Can be changed by the Operator to 100 or up to 300 mmHg, in 50 mmHg increments. | SAME |
| • Display total volume infused | Yes | Yes | SAME |
| • Continuous infuse and fixed bolus volume modes (fill until a fixed volume of fluid is delivered) | Yes | Yes | SAME |
| • Monitor pump speed and alarm if speed not equal to setting. | Yes | Yes | SAME |
| Summary-4 | | | |
| Element of Comparison | Predicate Device:<br>Belmont® Rapid<br>Infuser, RI-2<br>[510(k) K141654] | Subject Device:<br>Modified Belmont®<br>Rapid Infuser, RI-2 | Substantial<br>Equivalence |
| • Alternate fluid path | Yes, recirculate mode,<br>controlled by diversion<br>valve, routes fluid back to<br>system reservoir to purge<br>any air in the main fluid<br>circuit back into the<br>reservoir to be vented. | Yes, recirculate mode,<br>controlled by diversion<br>valve, routes fluid back<br>to system reservoir to<br>purge any air in the main<br>fluid circuit back into the<br>reservoir to be vented. | SAME |
| 3.0 Warming Characteristics: | | | |
| • Steady state fluid<br>output temperature | 37.5°C at high flow, 60<br>mL/min and higher; and<br>39°C at low flow, less<br>than 60 mL/min (Higher<br>temperature at low flow<br>to compensate for<br>cooling in patient line.) | 37.5°C at high flow, 60<br>mL/min and higher; and<br>39°C at low flow, less<br>than 60 mL/min (Higher<br>temperature at low flow<br>to compensate for<br>cooling in patient line.) | SAME |
| • High temperature<br>alarm condition<br>(Infusate Over<br>Temp) | Output temperature<br>>42°C for 20 mL of fluid<br>pumped or fluid<br>temperature >45°C for<br>0.25 seconds. | Output temperature<br>>42°C for 20 mL of fluid<br>pumped or fluid<br>temperature >45°C for<br>0.25 seconds. | SAME |
| • Maximum speed for<br>steady state 37°C | 1000 mL/min | 1000 mL/min | SAME |
| • Measure and<br>display output<br>temperature of<br>infusate | Yes | Yes | SAME |
| • Heat exchanger<br>type | Inductively heated<br>stainless steel annular<br>rings | Inductively heated<br>stainless steel annular<br>rings | SAME |
| • Method of<br>temperature control | Measure input and<br>output temperatures<br>from the temperature<br>sensors, compute<br>required power to the<br>heat exchanger to<br>maintain the targeted<br>temperature. Redundant<br>check using flow rate,<br>input power, input<br>temperature. | Measure input and<br>output temperatures<br>from the temperature<br>sensors, compute<br>required power to the<br>heat exchanger to<br>maintain the targeted<br>temperature. Redundant<br>check using flow rate,<br>input power, input<br>temperature. | SAME |
| 4.0 Alarms and System Surveillance: | | | |
| 510(k) Summary Summary-6 | | | |
| Element of Comparison | Predicate Device:<br>Belmont® Rapid<br>Infuser, RI-2<br>[510(k) K141654] | Subject Device:<br>Modified Belmont®<br>Rapid Infuser, RI-2 | Substantial<br>Equivalence |
| Air detection | | | |
| • Air detection at<br>input to machine to<br>detect empty<br>reservoir | Yes | Yes | SAME |
| • Redundant in-line<br>air bubble detection<br>near output | Yes | Yes | SAME |
| • Method to eliminate<br>air from system,<br>without disassembly | Yes | Yes | SAME |
| • Type of Air Bubble<br>detector | Ultrasonic | Ultrasonic | SAME |
| • Automatic Air<br>Elimination while<br>pumping | Yes, via "recirculation"<br>mode | Yes, via "recirculation"<br>mode | SAME |
| Element of Comparison | Predicate Device:<br>Belmont® Rapid<br>Infuser, RI-2<br>[510(k) K141654] | Subject Device:<br>Modified Belmont®<br>Rapid Infuser, RI-2 | Substantial<br>Equivalence |
| Line Pressure<br>monitoring | | | |
| • Monitor & Display<br>line pressure | Yes | Yes | SAME |
| • Pressure limit | 300 mmHg, standard.<br>User can adjust during<br>setup. | 300 mmHg, standard.<br>User can adjust during<br>setup. | SAME |
| • Automatic flow rate<br>control of pump to<br>maintain line<br>pressure below<br>pressure limit | Yes | Yes | SAME |
| • Message to<br>Operator when flow<br>rate is pressure<br>limited | Yes | Yes…
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.