PARAGON INFUSION SYSTEM WITH BOLUS

K023827 · I-Flow Corp. · FRN · Dec 13, 2002 · General Hospital

Device Facts

Record IDK023827
Device NamePARAGON INFUSION SYSTEM WITH BOLUS
ApplicantI-Flow Corp.
Product CodeFRN · General Hospital
Decision DateDec 13, 2002
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 880.5725
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Paragon Infusion System with Bolus is intended for continuous and/or intermittent infusion of medications for general infusion use, including antibiotic delivery, chemotherapy and pain management. Routes of administration include the following: intravenous, subcutaneous, intramuscular and epidural. The Paragon Infusion System with Bolus is also intended for continuous and/or intermittent delivery of medication (such as local anesthetics or narcotics) to surgical wound sites for postoperative regional anesthesia and pain management. Routes of administration may be intraoperative or percutaneous.

Device Story

Paragon Infusion System with Bolus provides continuous or intermittent medication delivery; supports antibiotic, chemotherapy, and pain management therapies. Routes include intravenous, subcutaneous, intramuscular, epidural, intraoperative, and percutaneous. Device operates as an infusion pump for clinical use; enables precise medication administration to surgical wound sites or systemic circulation. Healthcare providers manage infusion parameters to support postoperative regional anesthesia and pain management. System benefits include controlled drug delivery, reducing manual administration requirements and supporting patient recovery through consistent medication dosing.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Infusion pump (Class II, Product Code FRN). Mechanical/electromechanical infusion delivery system. Designed for continuous and intermittent medication administration via multiple routes (IV, subcutaneous, intramuscular, epidural, percutaneous).

Indications for Use

Indicated for patients requiring continuous or intermittent infusion of medications including antibiotics, chemotherapy, and pain management agents. Routes include IV, subcutaneous, intramuscular, epidural, intraoperative, or percutaneous delivery for regional anesthesia and pain management.

Regulatory Classification

Identification

An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/0/Picture/1 description: The image shows a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three lines representing its wings. The eagle's head is facing left, and its beak is slightly open. The text "DEPARTMENT OF HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle. Food and Drug Administrat 9200 Corporate Boulevard Rockville MD 20850 DEC 1 3 2002 Mr. Shane Noehre Director, Regulatory Affairs I-Flow Corporation 20202 Windrow Drive Lake Forest, California 92630 Re: K023827 Trade/Device Name: Paragon Infusion System with Bolus Regulation Number: 880.5725 . Regulation Name: Infusion Pump Regulatory Class: II Product Code: FRN Dated: November 14, 2002 Received: November 18, 2002 Dear Mr. Noehre: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting vour device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ ## Page 2 - Mr. Noehre Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrb/dsma/dsmamain.html Sincerely yours, Timothy A. Ulatowski Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health {2}------------------------------------------------ Verl 3 - 4/24/96 Applicant: Flow Corporation 510(k) Number (if known): Device Name: Paragon Infusion System with Bolus Indications For Use: - 1. The Paragon Infusion System with Bolus is intended for continuous and/or intermittent infusion of medications for general infusion use, including antibiotic delivery, chemotherapy and pain management. Routes of administration include the following: intravenous, subcutaneous, intramuscular and epidural. - 2. The Paragon Infusion System with Bolus is also intended for continuous and/or intermittent delivery of medication (such as local anesthetics or narcotics) to surgical wound sites for postoperative regional anesthesia and pain management. Routes of administration may be intraoperative or percutaneous. Patricia Ciuente (Division Sign-Off) Division of Anesthesiology, General Hospital, Infection Control, Dental Devices 510(k) Number. K023827 (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) > Concurrence of CDRH, Office of Device Evaluation (ODE) (Per 21 CFR 801.109) > > (Optional Format 1-2-96)
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...