K955180 · Sims Deltec, Inc. · FRN · May 23, 1996 · General Hospital
Device Facts
Record ID
K955180
Device Name
CADD-PCA, -MICRO, -PRIZM AMBULATORY INFUSION PUMP
Applicant
Sims Deltec, Inc.
Product Code
FRN · General Hospital
Decision Date
May 23, 1996
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5725
Device Class
Class 2
Attributes
Therapeutic
Intended Use
CADD-PCA® Model 5800 or 5800R Pump: The CADD-PCA pump is indicated for intravenous, subcutaneous, epidural space or subarachnoid space infusions. CADD-Micro® Model 5900 Pump: The CADD-Micro pump is indicated for intravenous, intra-arterial, subcutaneous, intraperitoneal, epidural space or subarachnoid space microinfusions. CADD-Prizm™ Model 6100 Pump: The CADD-Prizm pump is indicated for intravenous, intra-arterial, subcutaneous, intraperitoneal, epidural space or subarachnoid space infusions.
Device Story
Microcomputer-controlled ambulatory infusion pumps deliver medication via parenteral routes (intravenous, intra-arterial, subcutaneous, intraperitoneal, epidural, subarachnoid). Devices operate via programmed infusion parameters to deliver medication through compatible catheters. Used in clinical or ambulatory settings by healthcare providers or patients. Output consists of controlled fluid delivery; informs clinical pain management or therapeutic infusion regimens. Benefit includes expanded site-specific delivery options for existing pump platforms.
Clinical Evidence
No clinical studies or functional testing were performed. The submission relies on the existing clearance of the pump hardware and the established clinical indication for subarachnoid use associated with compatible catheter systems.
Technological Characteristics
Microcomputer-controlled ambulatory infusion pumps. No hardware or software modifications were made for this labeling expansion. Connectivity and materials are consistent with previously cleared versions of the CADD-PCA, CADD-Micro, and CADD-Prizm pump lines.
Indications for Use
Indicated for patients requiring medication delivery via intravenous, intra-arterial, subcutaneous, intraperitoneal, epidural, or subarachnoid routes using microcomputer-controlled ambulatory infusion pumps.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
Reference Devices
IPI AlgoLine™ Catheter Kits (PS Medical Corporation)
Related Devices
K982836 — CADD-LEGACY PLUS AMBULATORY INFUSION SYSTEM · Sims Deltec, Inc. · Nov 2, 1998
K982839 — CADD-LEGACY PCA AMBULATORY INFUSION SYSTEM, MODEL 6300 · Sims Deltec, Inc. · Nov 2, 1998
K960826 — CADD-PRIZM MODEL 6100 PUMP · Sims Deltec, Inc. · Aug 23, 1996
K971844 — MCKINLEY OUTBOUND-2 SYSTEM · Mckinley, Inc. · Jan 23, 1998
{0}
K955180
510(k) Summary
Date Prepared: November 8, 1995
## I. GENERAL INFORMATION
Applicant's name and address:
SIMS Deltec, Inc.
1265 Grey Fox Road
St. Paul, MN 55112
Contact:
David H. Short
Manager, Regulatory Affairs
(612)628-7214
## II. PRODUCT INFORMATION
Common/Usual Name:
Ambulatory Infusion Pumps
Classification Name:
Infusion Pump
Proprietary Name:
CADD-PCA® Model 5800 Pump
CADD-PCA® Model 5800R Pump
CADD-Micro® Model 5900 Pump
CADD-Prizm™ Model 6100 Pump
## III. EQUIVALENCE DEVICE COMPARISON
IPI AlgoLine™ Catheter Kits manufactured by PS Medical Coporation, Goleta, CA.
## IV. DEVICE DESCRIPTION
The subject devices are microcomputer controlled ambulatory infusion pumps intended for the delivery of medication via various parenteral routes, i.e., intravenous, intra-arterial, subcutaneous, intraperitoneal, or epidural space infusion. The purpose of this 510(k) Notification is to add subarachnoid space infusion to the labeling and thereby expanding the intended use of each pump. The subject devices, as a matter of course, will be used for the delivery of medication through catheters that have this indication.
Page 1 of 3
241
{1}
510(k) Summary
Date Prepared: November 8, 1995
## IV. DEVICE COMPARISON
Provided in Table 1 is a comparison of devices and their respective indications for use.
Table 1. Comparison of indications for use of various devices.
| | CADD-Prizm™ Pump | CADD-PCA® Pump
(Model 5800 & 5800R) | CADD-Micro® Pump | AngoLine™ Catheter Kits |
| --- | --- | --- | --- | --- |
| Manufacturer | SIMS Deltec, Inc. | SIMS Deltec, Inc. | SIMS Deltec, Inc. | IPI PS Medical Corporation |
| SYSTEM | | | | |
| Microcomputer controlled pump | YES | YES | YES | n/a |
| INDICATIONS FOR USE | | | | |
| Intravenous | YES | YES | YES | n/a |
| Intra-arterial | YES | NO | YES | n/a |
| Subcutaneous | YES | YES | YES | n/a |
| Intraperitoneal | YES | NO | YES | n/a |
| Epidural space | YES | YES | YES | YES |
| Subarachnoid space | NO | NO | NO | YES |
## V. INTENDED USE
CADD-PCA® Model 5800 or 5800R Pump:
The CADD-PCA pump is indicated for intravenous, subcutaneous, epidural space or subarachnoid space infusions.
CADD-Micro® Model 5900 Pump:
The CADD-Micro pump is indicated for intravenous, intra-arterial, subcutaneous, intraperitoneal, epidural space or subarachnoid space microinfusions.
CADD-Prizm™ Model 6100 Pump:
The CADD-Prizm pump is indicated for intravenous, intra-arterial, subcutaneous, intraperitoneal, epidural space or subarachnoid space infusions.
Page 2 of 3
{2}
510(k) Summary
Date Prepared: November 8, 1995
## VI. SUMMARY OF STUDIES
### A. Functional Testing
No software and/or functional modifications were required as it relates to the addition of the subarachnoid space infusion indication to the labeling of each subject device, therefore, no functional testing was deemed necessary.
### B. Clinical Studies
Clinical studies were not deemed necessary.
## VII. CONCLUSION
Based upon the information provided above, the addition of subarachnoid space infusion to the labeling of each subject device is substantially equivalent to the labeling of commercially available catheters, such as the IPI AlgoLine™ Catheter Kits manufactured by PS Medical Coporation, with this indication.
Page 3 of 3
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.