K181718 · Becton, Dickinson and Company · FPA · Oct 18, 2018 · General Hospital
Device Facts
Record ID
K181718
Device Name
BD FlowSmart Set/MiniMed Pro-Set
Applicant
Becton, Dickinson and Company
Product Code
FPA · General Hospital
Decision Date
Oct 18, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The BD FlowSmart™ Set/MiniMed™ Pro-Set™ set is intended for the subcutaneous infusion of medication, including insulin, from an external infusion pump.
Device Story
Single-use subcutaneous infusion set; interfaces with external infusion pumps for medication delivery (e.g., insulin). Features flexible perforated catheter with proprietary side-port; perpendicular to adhesive patch for skin attachment. Detachable tubing connects pump reservoir to patient. Used in clinical or home settings; operated by patient or caregiver. Device facilitates continuous medication delivery; replaces every 48-72 hours. Modifications include changes to catheter length and introducer needle gauge. Bench testing confirms performance equivalence to predicate.
Clinical Evidence
Bench testing only. Studies included introducer needle integrity, skin penetration/drag force, needle-to-hub strength, catheter-to-base attachment strength (ISO 10555-1:1995), and tip radial strength. Biocompatibility evaluated per ISO 10993-1 (comparative chemical characterization, cytotoxicity). Insulin compatibility study performed. All tests met acceptance criteria.
Technological Characteristics
Single-use sterile infusion set. Flexible perforated catheter with side-port; 29G introducer needle; 6.6mm catheter length. Adhesive base for skin attachment. Tubing compatible with Paradigm® and Luer Lock connections. Materials evaluated for biocompatibility per ISO 10993-1. Mechanical performance tested per ISO 10555-1:1995.
Indications for Use
Indicated for subcutaneous infusion of medication, including insulin, from an external infusion pump in patients requiring such therapy.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 18, 2018
Becton, Dickinson and Company Meriam Youssef Staff Regulatory Affairs Specialist 1 Becton Drive Franklin Lakes, New Jersey 07417
Re: K181718
Trade/Device Name: BD FlowSmart Set/MiniMed Pro-Set Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: Class II Product Code: FPA, FPK Dated: September 9, 2018 Received: September 18, 2018
Dear Meriam Youssef:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarket safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Image /page/1/Picture/6 description: The image shows the name "Geeta K. Pamidimukkala -S" in a clear, sans-serif font. The text is stacked vertically, with "Geeta K." on the top line and "Pamidimukkala -S" on the second line. The text is black, and the background is white.
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| DEPARTMENT OF HEALTH AND HUMAN SERVICES |
|-----------------------------------------|
| Food and Drug Administration |
| <b>Indications for Use</b> |
| Form Approved: OMB No. 0910-0120 |
|----------------------------------|
| Expiration Date: 06/30/2020 |
| See PRA Statement below. |
| 510(k) Number (if known) | |
|--------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | BD FlowSmart ™ Set/MiniMed™ Pro-Set™ |
| Indications for Use (Describe) | The BD FlowSmart™ Set/MiniMed™ Pro-Set™ set is intended for the subcutaneous infusion of medication, including insulin, from an external infusion pump. |
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | <span style="display:inline-block; margin-right: 5px;"> <div style="display:inline-block; vertical-align:middle; width:15px; height:15px; border:1px solid black; text-align:center;"> <span><b>X</b></span> </div> </span> Prescription Use (Part 21 CFR 801 Subpart D) |
| | <span style="display:inline-block; margin-right: 5px;"> <div style="display:inline-block; vertical-align:middle; width:15px; height:15px; border:1px solid black; text-align:center;"> </div> </span> Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."*
| FORM FDA 3881 (7/17) | Page 1 of 1 |
|----------------------|-------------|
|----------------------|-------------|
။-9 ခုနှစ်များ
{3}------------------------------------------------
# 510(k) Summary
| Submitted By: | Meriam Youssef<br>Staff Regulatory Affairs Specialist, BD Medical<br>1 Becton Drive<br>Franklin Lakes, NJ 07417<br>Tel: 201 847 6557<br>Fax: 201 847 5307 | |
|----------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------|
| Date Prepared: | October 18, 2018 | |
| Device Name: | | |
| | Trade Name: | BD FlowSmart™ Set/MiniMed™ Pro-Set™ |
| | Common Name: | Intravascular adminstration set |
| | Classification: | Class II device; 21 CFR 880.5440,<br>(intravascular administration set) |
| | Product Code: | FPA, FPK |
## Legally marketed predicate devices to which substantial equivalence is being claimed: K160651: BD FlowSmart™ Set/MiniMedTM Pro-Set™
#### Device Description:
The subject infusion set is a single use infusion administration set intended to be used for 48-72 hours. The product is intended to interface with commercially available infusion pumps with suitable connections. The infusion set features a flexible perforated catheter perpendicular to an adhesive patch and detachable tubing is connected on one end to the medication reservoir of an infusion pump and on the other end to the patient, attached to the skin by an adhesive base. The plastic catheter of the device contains a proprietary side-port. It is a single-use sterile device.
#### Indications for Use:
BD FlowSmart™ Set/MiniMed™ Pro-Set™ is intended for the subcutaneous infusion of medication, including insulin, from an external infusion pump.
There are no changes to the Indications for Use as a result of the device modifications.
### Comparison with Predicate Devices:
The subject device has the same fundamental scientific technology as the predicated device. The purpose of this submission is to introduce design modifications to the catheter length and introducer needle gauge sizes, labeling and material modifications. BD has conducted bench studies confirming that the product performance is substantially equivalent to the predicate device. The table below provides a side by side comparison of the subject device compared to its predicate.
{4}------------------------------------------------
| Feature | Subject: BD Infusion Set | Predicate Device: BD<br>Infusion Set |
|-------------------------|--------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | Pending | K160651 |
| Manufacturer | BD | BD |
| Intended Use | Unchanged | The BD<br>FlowSmart™/MiniMed™<br>Pro-Set™ infusion set is<br>intended for the<br>subcutaneous infusion of<br>medication,<br>including insulin, from an<br>external infusion pump. |
| System components | Unchanged | Infusion set + tubing |
| Needle insertion method | Unchanged | Manual or Automatic via<br>Quickserter (K160860) |
| Angle of Insertion | Unchanged | Straight 90 degrees |
| Insertion Needle Gauge | 29G | 30G |
| Catheter Length (mm) | 6.6 | 6.0 |
| Cannula Gauge | Unchanged | 28G |
| Tubing Connection Type | Unchanged | Paradigm® and Luer Lock |
| Tubing Length (in) | Unchanged | 24, 42 |
| Connection Positions | Unchanged | Multiple |
| Replacement Frequency | Unchanged | Disposable, replaced<br>every 72 hours |
| Provided Sterile | Unchanged | YES |
### Testing:
The subject device has the same technological characteristics as the predicate devices cleared in K160651. BD has verified the modifications of the subject device through bench performance and biocompatibility studies.
Bench performance studies included the following test methods:
| Test Method | Acceptance Criteria | Results<br>(Pass/Fail) |
|--------------------------------------------------------|------------------------------------------------|------------------------|
| Introducer Needle Integrity during<br>Manual Insertion | Equivalent to predicate device | Pass |
| Skin Penetration and Drag Force | Equivalent to predicate device | Pass |
| Introducer Needle to Needle Hub<br>Strength | Equivalent to predicate device | Pass |
| Catheter to Base Attachment Strength | Testing in accordance with ISO<br>10555-1:1995 | Pass |
| Tip Radial Strength | Equivalent to predicate device | Pass |
The results of the bench studies successfully demonstrated the subject device has met the requirements.
Biocompatibility studies were conducted per ISO 10993-1 and included:
- Comparative chemical characterization
- Cytotoxicity evaluation ●
{5}------------------------------------------------
The results of the comparative chemical characterization demonstrated that there are no new compounds or compounds at higher levels in the subject device compared to the predicate device. An insulin compatibility study was performed to demonstrate insulin compatibility.
The testing demonstrates the BD FlowSmart™ Set/MiniMed™ Pro-Set™ device meet requirements for its intended use and is substantially equivalent to its predicate device.
#### Conclusion:
The testing performed demonstrates the BD FlowSmart™ Set/MiniMed™ Pro-Set™ device is substantially equivalent to its predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.