The MiniMed Quick-serter is indicated to use as an aid for inserting compatible devices. It is for use by a single patient. It is not for use by multiple patients.
Device Story
MiniMed Quick-serter (MMT-305) is a hand-held, non-sterile, multi-use mechanical accessory; functions as an insertion aid for compatible infusion sets (MiniMed Quick-set and Pro-set). Device consists of plastic barrel, stainless steel spring, and handle. User loads infusion set into device; pulls handle to compress spring until locked; places device against insertion site; depresses side buttons to fire spring; drives needle/cannula into subcutaneous tissue; presses top release button to detach serter from infusion set. Used by patients or clinicians in home or clinical settings. Reduces technique dependency and insertion discomfort. Benefits patient through consistent, automated infusion set placement.
Clinical Evidence
Bench testing only. No clinical data. Verification testing included cocking force, trigger force, life cycle, drop testing, environmental storage, mechanical vibration, chemical resistance, and cleaning validation per AAMI TIR12/TIR30. Biocompatibility testing performed per ISO 10993-1, -5, -10, and -18. Usability evaluation confirmed successful insertion of Quick-set and Pro-set infusion sets.
Indicated for single-patient use as an aid for inserting compatible infusion sets into subcutaneous tissue.
Regulatory Classification
Identification
A syringe needle introducer is a device that uses a spring-loaded mechanism to drive a hypodermic needle into a patient to a predetermined depth below the skin surface.
Predicate Devices
MiniMed Quick-serter (MMT-395), infusion set insertion system (K992300)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of a human figure, represented by three overlapping profiles facing to the right. The profiles are black and are arranged in a way that suggests depth and connection. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 22, 2016
Medtronic MiniMed Noreen Bajwa Sr. Regulatory Affairs Specialist 18000 Devonshire St. Northridge. California 91325
Re: K160860
Trade/Device Name: MiniMed Quick-serter Regulation Number: 21 CFR 880.6920 Regulation Name: Syringe Needle Introducer Regulatory Class: II Product Code: KZH Dated: August 24, 2016 Received: August 25, 2016
Dear Noreen Bajwa:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Image /page/1/Picture/8 description: The image shows a signature and the name "Tina Kiang". The signature is a stylized, looping design to the left of the name. Below the name, there is a letter "S".
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K160860
Device Name
MiniMed Quick-serter
Indications for Use (Describe)
The MiniMed Quick-serter is indicated to use as an aid for inserting compatible devices. It is for use by a single patient. It is not for use by multiple patients.
Type of Use (Select one or both, as applicable)| | Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Page 1 of 1
{3}------------------------------------------------
### 510(K) SUMMARY (21 CFR 807.92)
### K160860
#### l. SUBMITTER [807.92(a)(1)]
Medtronic MiniMed, Inc. 18000 Devonshire Street Northridge, CA USA 91325
| Telephone: | (818) 576-5224 |
|------------|----------------|
| Fax: | (818) 576-6273 |
| Contact Person: | Noreen Bajwa |
|-----------------|----------------------------|
| Email: | noreen.bajwa@medtronic.com |
| Date Prepared: | September 21, 2016 |
#### II. DEVICE [807.92(a)(2)]
| Trade Name: | MiniMed Quick-serter |
|---------------------------|----------------------------------------|
| Common or Usual Name: | Syringe Needle Introducer |
| Regulation Name: | Introducer, Syringe Needle |
| Regulation Number: | 21 CFR §880.6920 |
| Product Code: | KZH |
| Regulatory Class: | II |
| Reviewing Product Branch: | General Hospital Devices Branch (GHDB) |
#### III. PREDICATE DEVICE [807.92(a)(3)]
MiniMed Quick-serter (MMT-395), infusion set insertion system (K992300)
### Reference Devices
- MiniMed Quick-set Infusion Set (K991759, K011071)
- . MiniMed Pro-set Infusion Set (K160651)
### Purpose of Submission
The scope of this 510(k) submission includes modifications to the device design and additional compatibility to MiniMed Pro-set infusion set.
#### IV. DEVICE DESCRIPTION [807.92(a)(4)]
The modified MiniMed Quick-serter (MMT-305) is a non-sterile, single patient, multi-use, hand held accessory product designed to be used as an insertion aid for compatible devices, including specific
{4}------------------------------------------------
infusion sets. It is used to insert the introducer needle and cannula through the skin and into the subcutaneous tissue. The MiniMed Quick-serter is intended to be used by a patient or clinician as a means to insert an infusion set with minimum discomfort and technique dependency.
The MiniMed Quick-serter consists of a plastic barrel containing a stainless steel spring and a handle. The device user places and securely seats the infusion set into the MiniMed Quick-serter is loaded and locked by pulling the handle (compressing the spring) until it clicks/locks into place. The serter is placed in contact with the insertion site and fired to release the infusion set. This is done by depressing the green side buttons to release the spring, which drives the infusion set forward and the insertion needle and cannula are inserted into the user's subcutaneous tissue. The release button on the top of the device is then pressed to ease the release of the infusion set from the serter.
Devices compatible to MiniMed Quick-serter (MMT-305) include:
- MiniMed Quick-set Infusion Set (K991759, K011071)
- . MiniMed Pro-set Infusion Set (K160651)
### V. INDICATIONS FOR USE [807.92(a)(5)]
The MiniMed Quick-serter is indicated to use as an aid for inserting compatible devices. It is for use by a single patient. It is not for use by multiple patients.
The Indications for Use statement for the MiniMed Quick-serter is not identical to the predicate device. However, the differences do not affect the intended use of the device, nor does it affect the safety and effectiveness of the device relative to the predicate.
The predicate MiniMed Quick-serter (MMT-395) is indicated for use with MiniMed Quick-set family of infusion sets. The subject MiniMed Quick-serter (MMT-305) is indicated for use with compatible infusion sets including both MiniMed Quick-set and MiniMed Pro-set families of infusion sets. Both the predicate and subject MiniMed Quick-serter are indicated for single patient, multi-use and have the same intended use for automatic insertion of infusion sets.
#### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS [807.92(a)(6)]
The MiniMed Quick-serter has similar technological characteristics as the predicate device.
The energy source for both the predicate (MMT-395) and subject (MMT-305) MiniMed Quick-serter devices is spring driven. The predicate and subject devices operational principle and there are no different questions of safety and effectiveness as the subject MiniMed Quick-serter (MMT-305) falls within the same classification regulation, has the same intended use and retains the same fundamental scientific technology of the predicate MiniMed Quick-serter (MMT-395).
{5}------------------------------------------------
A comparison of the technological characteristics of the predicate (MMT-395) and subject (MMT-305) MiniMed Quick-serter are provided below:
| Description | (Predicate) MiniMed Quick-serter<br>(MMT-395) | (Modified) MiniMed Quick-serter<br>(MMT-305) |
|------------------------------|--------------------------------------------------------------|----------------------------------------------------------------|
| Classification | Class II | Identical |
| Product Code | KZH | Identical |
| Type of Use | Over the Counter | Identical |
| Compatibility | Maersk Medical A/S Contour infusion<br>sets | MiniMed Quick-set Infusion Set<br>MiniMed Pro-set Infusion Set |
| Condition of use | Single patient, multi-use | Identical |
| Energy Source | Spring Driven | Identical |
| Mode of action | Manual operation | Identical |
| Cocking Force | <5 lbf | Identical |
| Trigger/release<br>Force | <7 lbf | 0.67 lbf -7lbf |
| Diameter | 1.89 inches | 1.96 inches |
| Width of Pull<br>Handle | 1.386 inches | 1.59 inches |
| Set release button | Set release button and pull handle<br>are the same component | Set release button and pull handle are<br>made distinct |
| Sterility | Non-sterile | Identical |
| Validated Cleaning<br>Method | Manual | Identical |
| Service Life | 3 years | Identical |
#### VII. PERFORMANCE DATA [807.92(b)]
The following performance data were provided in support of the substantial equivalence determinations:
## Biocompatibility
The biocompatibility evaluation was conducted in accordance with the FDA #G95-1 "Use of International Standard ISO-10993, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing"" and international standard EN ISO 10993-1 "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing Within A Risk Management Process" as recognized by FDA. The battery of testing included the following tests:
- . Chemical Characterization (EN ISO 10993-18)
- . Cytotoxicity (EN ISO 10993-5)
- Sensitization (EN ISO 10993-10)
- . Skin Irritation (EN ISO 10993-10)
The MiniMed Quick-serter has been evaluated for biocompatibility and is acceptable for its intended use by Biological Evaluation.
{6}------------------------------------------------
### Verification Testing
The design of MiniMed Quick-serter (MMT-305) was verified to provide objective evidence that the device retains its mechanical properties and functions with MiniMed Quick-set and MiniMed Pro-set infusion sets.
Performance testing performed to support the design modifications included:
- Cocking Force
- Trigger Force
- Visual/Functional
- Life Cycle testing
- Drop Test ●
- Warehouse Environment Storage ●
- Home Environment Storage
- Operating Temperature
- Operating Humidity ●
- Mechanical Vibration
- Chemical Resistance
- Reliability
- Cleaning ●
- Robust Cleaning Life Test ●
- Usability Evaluation of Quick-set and Pro-set Insertion with Quick-serter
The aforementioned tests were completed to internal standards:
- . Ship Test - ASTM D4169 "Distribution Cycle 12, Assurance Level 1-Standard Practice for Performance Testing of Shipping Containers and Systems
- . Cleaning studies- AAMI TIR12 "Designing, testing and labeling reusable medical devices for reprocessing in healthcare facilities. A guide for medical device manufacturers" and AAMI TIR30 "A compendium of processes, materials, test methods, and acceptance criteria for cleaning reusable medical devices.
Results from performance testing indicate that the product meets the established performance requirements.
#### VIII. CONCLUSION
Based on the 510(k) Summary and information provided herein, we conclude the subject device, MiniMed Quick-serter (MMT-305), is substantially equivalent to the predicate MiniMed Quick-serter (MMT-395).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.