K220584 · Becton Dickinson Infusion Therapy Systems, Inc. · FOZ · Mar 31, 2022 · General Hospital
Device Facts
Record ID
K220584
Device Name
BD Cathena Safety IV Catheter
Applicant
Becton Dickinson Infusion Therapy Systems, Inc.
Product Code
FOZ · General Hospital
Decision Date
Mar 31, 2022
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.5200
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
BD Cathena™ Safety IV Catheters are intended to be inserted into a patient's peripheral vascular system for short term use to sample blood, monitor blood pressure, or administer fluids. These devices may be used for any patient population with consideration given to adequacy of vascular anatomy, procedure being performed, fluids being infused, and duration of therapy. These devices are suitable for use with power injectors set to a maximum pressure of 325 psi (2240 kPa).
Device Story
Over-the-needle peripheral IV catheter; includes radiopaque BD Vialon catheter, needle, grip, passive safety needle shield, and flash chamber with removable vent plug. Features BD Instaflash technology for immediate blood visualization; optional BD Multiguard technology prevents blood flow from hub until Luer connection. Needle tip passively protected upon removal to prevent needlestick injury. Used in clinical settings by healthcare providers for fluid administration, blood sampling, or pressure monitoring. Available in various gauges; color-coded for identification. Not made with natural rubber latex.
Clinical Evidence
Bench testing only. Performance testing conducted per ISO 10555-1:2013 for gravity flow rate. Where technological characteristics were identical to the predicate, predicate performance data was leveraged.
Technological Characteristics
Radiopaque BD Vialon catheter; passive safety needle shield; flash chamber with vent plug. Available with/without BD Multiguard blood-flow control. Color-coded by gauge (24G, 22G, 20G, 16G). Power injector compatible up to 325 psi (2240 kPa). Non-latex. Testing performed per ISO 10555-1:2013.
Indications for Use
Indicated for patients requiring peripheral vascular access for short-term blood sampling, blood pressure monitoring, or fluid administration. Suitable for all patient populations based on vascular anatomy, procedure, fluid type, and therapy duration. Compatible with power injectors up to 325 psi.
Regulatory Classification
Identification
An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.
K192493 — BD Cathena (TM) Safety IV Catheter · Becton Dickinson Infusion Therapy Systems, Inc. · Jan 24, 2020
K251155 — BD Cathena Safety IV Catheter · Becton Dickinson Infusion Therapy Systems, Inc. · Jul 11, 2025
K251654 — BD Insyte Autoguard Shielded IV Catheter, BD Insyte Autoguard BC Shielded IV Catheter and BD Insyte Autoguard BC Pro Shielded IV Catheter · Becton Dickinson Infusion Therapy Systems, Inc. · Jun 27, 2025
K170283 — BD Nexiva Diffusics Closed IV Catheter System · Becton, Dickinson and Company · Feb 28, 2017
K172506 — BD Cathena Safety IV Catheter · Becton, Dickinson and Company · Sep 17, 2017
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.
Becton Dickinson Infusion Therapy Systems Inc. Paul Holman Regulatory Affairs Specialist 9450 South State Street Sandy, Utah 84070
March 31, 2022
Re: K220584
Trade/Device Name: BD Cathena Safety IV Catheter Regulation Number: 21 CFR 880.5200 Regulation Name: Intravascular Catheter Regulatory Class: Class II Product Code: FOZ Dated: February 28, 2022 Received: March 1, 2022
Dear Paul Holman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Payal Patel Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K220584
Device Name BD Cathena™ Safety IV Catheter
#### Indications for Use (Describe)
BD Cathena™ Safety IV Catheters are intended to be inserted into a patient's peripheral vascular system for short term use to sample blood, monitor blood pressure, or administer fluids. These devices may be used for any patient population with consideration given to adequacy of vascular anatomy, procedure being performed, fluids being infused, and duration of therapy. These devices are suitable for use with power injectors set to a maximum pressure of 325 psi (2240 kPa).
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------------------|---------------------------------------------|
| <div> <span style="text-decoration: overline;">Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> | Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) Summarv Prepared on: 2022-02-28 Contact Details 21 CFR 807.92(a)(1) Becton Dickinson Infusion Therapy Systems Inc. Applicant Name Applicant Address 9450 South State Street Sandy UT 84070 United States 8015225132 Applicant Contact Telephone Applicant Contact Mr. Paul Holman Applicant Contact Email paul.holman@bd.com Device Name 21 CFR 807.92(a)(2) BD Cathena™ Safety IV Catheter Device Trade Name Common Name Intravascular catheter
Catheter, Intravascular, Therapeutic, Short-Term Less Than 30 Days
Classification Name
Requlation Number
Product Code
Predicate #
K201717
# Legally Marketed Predicate Devices
21 CFR 807.92(a)(3)
Predicate Trade Name (Primary Predicate is listed first)
BD Cathena™ Safety IV Catheter
880.5200
FOZ
Product Code
FOZ
21 CFR 807 92(a)(4)
# Device Description Summary
BD Cathena™ Safety IV Catheters are over-the-needle, intravascular (IV) catheters include a radiopaque BD Vialon™ catheter, needle, grip, passive safety needle shield, and flash chamber with removable vent plug. The needle and catheter are protected by a needle cover. These devices have BD Instaflash™ Needle Technology, allowing for immediate visualization of blood along the catheter. The flash chamber provides confirmation that the vessel. The needle tip is passively protected when the needle is removed, reducing the risk of accidental needlestick injury.
These devices are available with or without multi-access BD Multiguard™ Technology, which is designed to stop the flow of blood from the catheter hub until a Luer connection is made. Once a connection is made, fluids or blood can flow through the cather direction.
These devices are available with or without wings are color coded to indicate the catheter gauge size (24 GA (0.7 mm) = Yellow, 22 GA (0.9 mm) = Blue, 20 GA (1.3 mm) = Green, 16 GA (1.7 mm) = Green, 16 GA (1.7 mm) = Grey). These devices are not made with natural rubber latex.
# Intended Use/Indications for Use
## 21 CFR 807.92(a)(5)
BD Cathena™ Safety IV Catheters are intended to be inserted into a patient's peripheral vascular system for short term use to sample blood, monitor blood pressure, or administer fluids. These devices may be used for any patient population with consideration given to adequacy of vascular anatomy, procedure being performed, fluids being infused, and duration of therapy. These devices are suitable for use with power injectors set to a maximum pressure of 325 psi (2240 kPa).
{4}------------------------------------------------
## Indications for Use Comparison
21 CFR 807.92(a)(5)
10 Is the predicate device legally marketed? Yes. The predicate device, BD Cathena™ Safety IV Catheter, was cleared under K201717.
& Do the devices have the same intended use? Yes. The subject and predicate devices have the same intended use.
## Technological Comparison
@ Do the devices have the same technological characteristics? Yes. The subject and predicate devices have the same technological characteristics.
#### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)
A Risk Analysis in accordance with ISO 14971:2019 was conducted to assess the impact of the proposed modifications to the predicate device.
Performance tests completed on the subject device were limited to support a determination of substantial equivalence to the predicate device. When technological characteristics between the subject and predicate devices were found to be identical, results of performance testing conducted on the predicate device were applied to the subject device. The performance tests listed below were conducted to ensure that the subject device meets pre-determined design requirements:
· Testing per ISO 10555-1:2013 o Gravity Flow Rate
Based on the indications for use, technological characteristics, and results of performance testing, the subject BD Cathena™ Safety W Catheter is substantially equivalent to the predicate device.
### 21 CFR 807.92(a)(6)
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.