K142136 · Vascular Pathways, Inc. · FOZ · Jan 8, 2015 · General Hospital
Device Facts
Record ID
K142136
Device Name
ACCUCATH MIDLINE CATHETER SYSTEM
Applicant
Vascular Pathways, Inc.
Product Code
FOZ · General Hospital
Decision Date
Jan 8, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5200
Device Class
Class 2
Indications for Use
The AccuCath® Midline Catheter System is a midline catheter that is inserted into a patient's vascular system for short term use (< 30 days) to sample blood pressure, and administer fluids intravenously. This device may be used with consideration given to adequacy of vascular anatomy, appropriateness of the solution being infused, and duration of therapy. The AccuCath® Midline is suitable for use with low pressure power injectors having a maximum pressure setting of 300 psi and maximum flow rate of 7mL/second.
Device Story
AccuCath® Midline Catheter System is a single-use, sterile, radiopaque intravascular catheter; designed for short-term (<30 days) vascular access. Device features a notched needle for flashback visualization; guidewire with atraumatic coiled tip; and integrated safety handle to mitigate sharps injury. Includes a built-in blood control septum that automatically activates upon needle removal to restrict blood flow until a secure male luer connection is established. Used by clinicians in clinical settings for blood sampling, pressure monitoring, and fluid/power-injected contrast administration. Device provides a secure, closed-system flow path; benefits include reduced risk of blood exposure and sharps injury during insertion.
Clinical Evidence
No clinical data. Substantial equivalence established via biocompatibility and functional bench testing. Testing confirmed adherence to design inputs and standards, including ISO 10993-1 (biocompatibility), ISO 10555-1/5 (catheter performance), and ISO 594-1/2 (luer fittings).
Technological Characteristics
Radiopaque catheter; 3.1-inch usable length; multiple gauge sizes. Features: blood control septum, notched needle, guidewire with atraumatic coiled tip, integrated safety handle. Materials/Standards: ISO 594-1/2 (luer), ISO 7864 (needles), ISO 9626 (tubing), ISO 10555-1/5 (catheters), ISO 10993-1 (biocompatibility), ISO 11607-1 (packaging), ASTM D4169, ASTM F88, ASTM F1929, ASTM F1980. Sterile, single-use. No software/electronics.
Indications for Use
Indicated for patients requiring short-term (<30 days) vascular access for blood sampling, blood pressure monitoring, and intravenous fluid administration. Suitable for use with low-pressure power injectors (max 300 psi, 7mL/sec).
Regulatory Classification
Identification
An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.
K121073 — POWERGLIDE MIDLINE CATHETER · C.R. Bard, Inc. · Jun 1, 2012
K250292 — OSPREY Midline Closed IV Catheter System (OspreyEDC-F20) · Skydance Vascular, Inc. · May 2, 2025
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 8,2015
Vascular Pathways, Inc. Mr. Scott Pease Vice President Regulatory Affairs and Quality Assurance 1847 Trade Center Way Naples, Florida 34109
Re: K142136
Trade/Device Name: AccuCath® Midline Catheter System Regulation Number: 21 CFR 880.5200 Regulation Name: Intravascular Catheter Regulatory Class: II Product Code: FOZ Dated: November 20, 2014 Received: November 24, 2014
Dear Mr. Pease:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Mr. Pease
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Susan Runno, DDS, MA
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K142136
Device Name
AccuCath® Midline Catheter System
#### Indications for Use (Describe)
The AccuCath® Midline Catheter System is a midline catheter that is inserted into a patient's vascular system for short term use (< 30 days) to sample blood pressure, and administer fluids intravenously. This device may be used with consideration given to adequacy of vascular anatomy, appropriateness of the solution being infused, and duration of therapy. The AccuCath® Midline is suitable for use with low pressure power injectors having a maximum pressure setting of 300 psi and maximum flow rate of 7mL/second.
### Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY
# AccuCath® Midline Catheter System
Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared
> Vascular Pathways, Inc. 1847 Trade Center Way Naples, FL 34109
Cell: 678-689-8010 Office: 239-254-0391 Fax: 239-513-1533
Contact Person: Scott Pease, VP of Regulatory Affairs and Quality Assurance (spease@vascularpathways.com)
Date Prepared: July 31, 2014
| Name of Device | AccuCath® Midline Catheter System |
|----------------|-----------------------------------|
|----------------|-----------------------------------|
Classification Name Intravascular Catheter
Product Code / Regulation FOZ / 21 CFR § 880.5200
Classification Class 2
# Predicate Devices
Bard Access PowerGlide™ Device (K121073) Vascular Pathways, Inc., AccuCath® BC Intravascular Catheter System (K140504)
{4}------------------------------------------------
### Indications for Use
The AccuCath® Midline Catheter System is a midline catheter that is inserted into a patient's vascular system for short term use (< 30 days) to sample blood, monitor blood pressure, and administer fluids intravenously. This device may be used with consideration given to adequacy of vascular anatomy, appropriateness of the solution being infused, and duration of therapy. The AccuCath® Midline is suitable for use with low pressure power injectors having a maximum pressure setting of 300 psi and maximum flow rate of 7mL/second.
# Device Description / Technological Characteristics
The AccuCath® Midline Catheter System has a usable length catheter of 3.1 inches in multiple Gauge sizes. The device(s) are single use, sterile intravascular catheters designed to be inserted into a patient's vascular system to sample blood, monitor blood pressure, or administer fluids intravenously. This longer AccuCath® Midline is substantially equivalent functional characteristics and design to the predicates [Bard Access PowerGlide™ Device (K121073) and AccuCath® BC (K140504)]. The AccuCath® Midline catheter hub is identical in design to that of the AccuCath® BC's catheter's hub. including the built-in blood control septum. Additionally, the AccuCath® Midline, like the AccuCath® BC (K140504) predicate, consists of a radiopaque catheter with a blood control valve mechanism that is delivered over a guidewire with atraumatic coiled tip design; a notched needle to enhance flashback visualization, and a body / handle that serves as an integrated safety container to mitigate the risk of sharps injuries. The AccuCath® Midline is identical to the AccuCath BC® in that it features a septum that automatically activates to stop the blood flow in the catheter hub when the needle is removed from the catheter during initial insertion by the clinician. Blood flow from the catheter hub will be restricted immediately after the needle retraction until a secure male luer connection is made. The flow path will remain opened once a secure male luer connection has been made.
## Performance Data
Biocompatibility and functional bench testing performed by Vascular Pathways, Inc. demonstrates the AccuCath® Midline Cather is substantial equivalence, in terms of the safety and effectiveness, to the referenced predicate devices. In vitro bench testing included an assessment of all design input requirements and confirmation that the output of the design process met all design input requirements was completed, including those relating to appropriate standards, as follows:
- ASTM D4169-09 - Standard Practice for Performance Testing of Shipping Containers and Systems
- ASTM F88/F88M-09 - Standard Test Method for Seal Strength of Flexible Barrier Materials
- ASTM F1929 – Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration
- ASTM F1980-07 (2011) - Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices
{5}------------------------------------------------
- . Guidance Document No. 934 - Medical Devices with sharps injury prevention features (August 9, 2005)
- ISO 594-1:1986 Conical Fittings with a 6% (Luer) Taper for Syringes, Needles and ● Certain Other Equipment - Part 1: General Requirements
- o ISO 594-2:1998 – Conical Fittings with a 6% (Luer) taper for syringes, needles and certain other medical equipment - Part 2: Lock Fittings
- ISO 7864:1993 - Sterile Hypodermic Needles for Single Use
- . ISO 9626:1991 – Stainless Steel Needle Tubing for Manufacture of Medical Devices
- o ISO 10555-1:2013 – Sterile, Single Use Intravascular Catheters - Part 1: General Requirements
- o ISO 10555-5:2013 - Sterile, Single Use Intravascular Catheters - Part 5: Over-Needle Peripheral Catheters
- ISO 10993-1:2009 – Biological Evaluation of Medical Devices- Part I: Evaluation and Testing
- o ISO 11607-1:2009 – Packaging for terminally sterilized medical devices — Part 1: Requirements for materials, sterile barrier systems, and packaging systems
## Summary of Substantial Equivalence
Based upon the device description, technical characteristics and test data provided within this submission, the AccuCath® Midline Catheter System is substantially equivalent to the referenced predicate devices.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.