AccuCath Intravascular Catheter

K162894 · C.R. Bard, Inc. · FOZ · Nov 15, 2016 · General Hospital

Device Facts

Record IDK162894
Device NameAccuCath Intravascular Catheter
ApplicantC.R. Bard, Inc.
Product CodeFOZ · General Hospital
Decision DateNov 15, 2016
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 880.5200
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The AccuCath Intravascular Catheter is inserted into a patient's vascular system to sample blood, monitor blood pressure, or administer fluids intravenously. This device may be used with consideration given to adequacy of vascular anatomy, appropriateness of the solution being infused, and duration of therapy. The AccuCath IV Catheter is suitable for use with power injectors.

Device Story

AccuCath Intravascular Catheter is a single-use, sterile device for vascular access. Components include a catheter with luer lock fitting, needle with flashback chamber, safety container, and guidewire assembly (slider, spring, release button). During insertion, a built-in blood control septum automatically stops blood flow upon needle retraction until a secure luer connection is established. A safety mechanism shields the needle post-placement. Used by clinicians in clinical settings for short-term therapy. Device facilitates fluid administration, blood sampling, and pressure monitoring. Power injector compatibility allows for high-flow delivery.

Clinical Evidence

No clinical data. Bench testing only; risk analysis determined that labeling modifications did not necessitate additional verification or validation activities as the device design and performance remain identical to the predicate.

Technological Characteristics

Single-use, sterile intravascular catheter. Available in 18, 20, and 22 gauge sizes with 1.25 or 2.25-inch lengths. Features include a blood control septum, needle safety shielding mechanism, and guidewire-assisted insertion. Compatible with power injectors. No changes to materials or construction from predicate.

Indications for Use

Indicated for patients requiring vascular access for blood sampling, blood pressure monitoring, or intravenous fluid administration. Suitable for use with power injectors.

Regulatory Classification

Identification

An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 November 15, 2016 C.R. Bard, Inc. Jacob Lee Regulatory Affairs Specialist 605 North 5600 West Salt Lake City, Utah 84116 Re: K162894 Trade/Device Name: AccuCath™ Intravascular Catheter Regulation Number: 21 CFR 880.5200 Regulation Name: Intravascular Catheter Regulatory Class: II Product Code: FOZ Dated: October 13, 2016 Received: October 17, 2016 Dear Mr. Jacob Lee: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely, Image /page/1/Picture/8 description: The image shows a signature followed by the name "Tina Kiang". Below the name is the letter "S" preceded by a hyphen. The signature is a complex, looping design to the left of the name. Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K162894 Device Name AccuCath Intravascular Catheter ## Indications for Use (Describe) The AccuCath Intravascular Catheter is inserted into a patient's vascular system to sample blood, monitor blood pressure, or administer fluids intravenously. This device may be used with consideration given to adequacy of vascular anatomy, appropriateness of the solution being infused, and duration of therapy. The AccuCath IV Catheter is suitable for use with power injectors. Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW .* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below. {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for BARD ACCESS SYSTEMS. The word "BARD" is in a bold, sans-serif font, with each letter outlined in black. Below the word "BARD" is the text "ACCESS SYSTEMS", also in a bold, sans-serif font. K162894 | 510(k) Summary 21 CFR 807.92(a) | | | | |------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | General<br>Provisions | Submitter Name:<br>Submitter Address:<br>Contact Person:<br>Telephone Number:<br>Fax Number:<br>Date of Preparation: | Bard Access Systems, Inc.<br>605 North 5600 West<br>Salt Lake City, UT 84116<br>Jacob Lee<br>Regulatory Affairs Specialist<br>(801) 522-5823<br>(801) 522-5425<br>October 13, 2016 | | | Subject<br>Device | Trade Name:<br>Common Name:<br>Regulation Name:<br>Product Code:<br>Regulation:<br>Regulatory Class:<br>Classification Panel: | AccuCath™ Intravascular Catheter<br>Intravascular Catheter<br>Intravascular Catheter<br>FOZ<br>21 CFR §880.5200<br>Class II<br>General Hospital | | | Predicate<br>Devices | Predicate Trade Name:<br>Premarket Notification:<br>Manufacturer:<br>Common Name:<br>Regulation Name:<br>Regulation:<br>Regulatory Class:<br>Classification Panel: | AccuCath™ Intravascular Catheter<br>K153298 (cleared December 10, 2015)<br>Bard Access Systems, Inc.<br>Intravascular Catheter<br>Intravascular Catheter<br>21 CFR §880.5200<br>Class II<br>General Hospital | | | Device<br>Description | The AccuCath™ Intravascular Catheter has usable length catheters of 1.25<br>and 2.25 inches in 18, 20, and 22 gauge sizes. The devices are single use,<br>sterile intravascular catheters designed to be inserted into a patient's<br>vascular system to sample blood, monitor blood pressure, or administer<br>fluids intravenously. The AccuCath™ IV Catheter's hub has a built in blood<br>control septum. The blood control feature is a single-use septum that<br>automatically activates to stop the blood flow in the catheter hub when the<br>needle is removed from the catheter during initial insertion by the clinician.<br>Blood flow from the catheter hub will be restricted immediately after needle<br>retraction until a secure luer connection is made. The flow path is<br>permanently opened once a secure luer connection has been made. The<br>AccuCath™ IV Catheter is provided with a safety mechanism which allows<br>the needle to be shielded following placement of the catheter. All devices<br>have the basic structure of a protective cover, a catheter with a luer lock<br>fitting, a needle connected to a flashback chamber, a safety container, and a<br>guidewire within the lumen of the needle which is connected to a slider,<br>spring and release button. | | | | Intended Use | The AccuCath™ Intravascular Catheter is intended to be inserted in the<br>patient's vascular system for short term use to sample blood, monitor blood<br>pressure, or administer fluids intravenously. | | | | | Attribute | Subject Device - AccuCath™ Intravascular Catheter | Predicate Device – AccuCath™ Intravascular Catheter (K153298) | | | Owner | Same as predicate | Bard Access Systems, Inc. | | | Classification | Same as predicate | FOZ - 21 CFR 880.5200 -<br>Short-term - Intravascular<br>Catheter | | | 510(k) Status | Subject of this Premarket<br>Notification | K153298 - Concurrence date<br>December 10, 2015 | | | Indications for<br>Use | Proposed Indications for Use to<br>remove the pressure setting and<br>flow rate:<br>The AccuCath™ Intravascular<br>Catheter is inserted into a<br>patient's vascular system to<br>sample blood, monitor blood<br>pressure, or administer fluids<br>intravenously. This device may<br>be used with consideration given<br>to adequacy of vascular<br>anatomy, appropriateness of the<br>solution being infused, and<br>duration of therapy. The | The AccuCath™ Intravascular<br>Catheter is inserted into a<br>patient's vascular system to<br>sample blood, monitor blood<br>pressure, or administer fluids<br>intravenously. This device may<br>be used with consideration<br>given to adequacy of vascular<br>anatomy, appropriateness of<br>the solution being infused, and<br>duration of therapy. The<br>AccuCath™ IV Catheter is<br>suitable for use with low<br>pressure power injectors | | Technological<br>Characteristics | Name | Catheter | | | | Catheter<br>Dimensions | Same as predicate | | | | Duration of<br>Use | Same as predicate | | | | Primary Device<br>Components | Same as predicate | | | | Means of<br>Insertion | Same as predicate | | | | Insertion Site | Same as predicate | | | | Primary Device<br>Materials | Same as predicate | | | | Catheter<br>Proximal<br>Configuration | Same as predicate | | | | Catheter Distal<br>Configuration | Same as predicate | | | | Number of<br>Lumens | Same as predicate | | | | Power Injection<br>Maximum Flow<br>Rate | Same as predicate | | | | Sterility | Same as predicate | | | | Packaging<br>Configurations | Basic Configuration<br>Intermediate Configuration | | | Performance<br>Tests | As part of Bard Access Systems, Inc.'s design controls, a risk analysis was<br>conducted to assess the impact of the proposed subject device labeling<br>modifications. The results of the risk analysis determined that no verification<br>or validation activities were required because the subject device labeling<br>modifications to the Indications for use and instructions for use do not include<br>any changes to the design, materials, performance, or risk profile of the cited<br>predicate device. Therefore, it is not necessary to conduct additional<br>performance tests including verification and validation. | | | | Summary of<br>Substantial<br>Equivalence | The modification to the indications for use and product instructions for use<br>has no impact on the intended use, technological characteristics, or risk<br>profile of the subject device because there are no changes to the design or<br>performance of the predicate device. Therefore, the subject AccuCathTM<br>Intravascular Catheter is substantially equivalent to the cited predicate<br>device. | | | {4}------------------------------------------------ AccuCath™ IV Catheter is suitable for use with power Same as predicate injectors. Commercial setting of 300 psi and maximum flow rate of AccuCath™ Intravascular 6mL/second. {5}------------------------------------------------
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...