K960963 · Osborn Medical Corp. · FMI · May 6, 1996 · General Hospital
Device Facts
Record ID
K960963
Device Name
CORD WRAP WITH NEEDLE RESHEATHER
Applicant
Osborn Medical Corp.
Product Code
FMI · General Hospital
Decision Date
May 6, 1996
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5570
Device Class
Class 2
Indications for Use
The Cord Wrap with Needle Resheather is an accessory to Infusion Stands, Classification (80 FOX).
Device Story
Device functions as infusion stand accessory; stores infusion pump electric cords via elongated spool; provides needle holder with five cavities for single-handed decoupling of needle covers from needle bodies. Operated by clinical staff in hospital/clinic settings. Clamp assembly with hand wheel and threaded hook arm secures device to infusion stand pole; anti-rotation protrusions prevent movement. Device is not a sharps receptacle. Benefits include organized cord management and simplified needle handling.
Clinical Evidence
No clinical data; bench testing only.
Technological Characteristics
Materials: Acrylonitrile-Butadiene-Styrene (ABS) resin for cord wrap and needle resheather; stainless steel for threaded hook arm. Adhesive: Loctite 494. Mechanical clamping mechanism with hand wheel. Non-powered, manual operation.
Indications for Use
Indicated for use as an accessory to infusion stands to store electric cords and provide a needle holder for infusion line needle covers.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
Predicate Devices
Ferguson Medical I.V. Hanger Bag or Holder (K883133)
Related Devices
K972705 — RESHEATHER DEVICE-METAL OR PLASTIC, RESHEATER/ NEEDLE HOLDER DEVICE-PLASTIC · Sage Products, Inc. · Apr 30, 1998
K980336 — IV STANDS · Clinton Industries, Inc. · Apr 6, 1998
K955748 — DOYLE EXTRACTOR · Safetech Intl., Inc. · Mar 14, 1996
K081449 — ULTIGUARD HOME INSULIN PEN NEEDLE DISPENSER AND SHARPS CONTAINER · Ultimed, Inc. · Jun 3, 2008
K043249 — ATRION NEEDLEVISE LARGE-BORE SHARPS SECURING DEVICE · Atrion Medical Products, Inc. · May 25, 2005
Submission Summary (Full Text)
{0}
K960963
MAY - 6 1996
## B. Summary
The Cord Wrap with Needle Resheather (Model CN-34) is an accessory to Infusion Stands, Classification (80 FOX).
As an accessory, it is substantially equivalent to the Ferguson Medical I.V. Hanger Bag or Holder. The 510(k) number for the Ferguson I.V. Hanger Bag or Holder is K883133.
Both the Ferguson Hanger Bag (K883133) and the Osborn Medical Corporation Cord Wrap with Needle Resheather are *accessories* to Infusion Stands.
## Submitter:
Osborn Medical Corporation
100 West Main
Utica, MN 55979
## Contact Person:
Bill D. Davis
## Trade Name:
Cord Wrap with Needle Resheather
## Common Name:
Infusion Stand Accessory
## Substantially Equivalent to:
Ferguson Medical I.V. Hanger Bag or Holder
## Description of Device:
The Cord Wrap with Needle Resheather comprises a reel for storing electric cords at a pole-type infusion stand. The stand provides a pole stanchion that extends a number of feet and supports an infusion pump and/or bags or bottles of liquid infusants.
-5-
{1}
Depending on the infusion pump, electric cords extend from the pole. Storage and containment can present difficulties during and between use of the equipment.
The Cord Wrap is constructed to conveniently store the electric cord of the infusion pump at the pole during and between use.
(Photographs in Section E, and drawings in Section F. depict the Cord Wrap with Needle Resheather.)
The cord is coiled about an elongated spool which is retained on the pole via a clamp assembly. The clamp assembly includes a hand wheel and threaded hook arm. The hook arm mounts about the pole, and upon tightening of the hand wheel the hook arm is drawn tight to the infusion stand.
Two protrusions are aligned to the hook arm and form a channel space where the infusion stand can nest. The Cord Wrap with Needle Resheather is thereby prevented from rotating about the pole.
The position of the Cord Wrap with Needle Resheather along the infusion stand can be varied by loosening the hand wheel and moving the Cord Wrap with Needle Resheather either up or down the infusion stand and retightening the hand wheel.
Protruding from the Cord Wrap with Needle Resheather is a needle holder which contains five cavities to receive and retain several cover pieces of needle storage containers.
The Cord Wrap with Needle Resheather is NOT a SHARPS RECEPTACLE. Each infusion line set typically requires an infusion needle which is stored in a two-piece container. Each container includes a cover and a needle body.
The five cavities are constructed to receive and hold the cover piece. This permits the staff member to de-couple the needle body from the cover with a single hand.
The Cord Wrap, hand wheel, and Needle Resheather are molded from a resin Acrylonitrile-Butadiene-Styrene (ABS).
The Needle Resheather is glued to the Cord Wrap with Loctite® 494 adhesive.
The threaded hook arm is made of stainless steel.
-6-
{2}
The Cord Wrap with Needle Resheather is packaged assembled in polyethylene bags.
Twenty-eight Cord Wrap with Needle Resheathers are shipped in standard shipping containers.
## Statement of Substantial Equivalence:
Osborn Medical Corporation considers the Cord Wrap with Needle Resheather to be substantially equivalent to the Ferguson Medical I.V. Hanger Bag or Holder.
-7-
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.