K160575 · Yidobio, Inc. · FMI · Sep 7, 2016 · General Hospital
Device Facts
Record ID
K160575
Device Name
CareSens Pen Needle, Softip Pen Needle
Applicant
Yidobio, Inc.
Product Code
FMI · General Hospital
Decision Date
Sep 7, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5570
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The CareSens Pen Needle and Softip Pen Needle are intended for use with pen injector devices for subcutaneous injection of fluids, including insulin and exenatide.
Device Story
CareSens Pen Needle and Softip Pen Needle are sterile, single-use hypodermic needles designed for attachment to pen injectors. The device consists of a cannula, needle hub, and protective caps. It facilitates the subcutaneous delivery of medications such as insulin and exenatide. The device is intended for use by patients or clinicians in various settings. By providing a sterile, single-use interface between the medication pen and the patient, the needle enables accurate and safe subcutaneous administration of prescribed fluids, supporting effective disease management.
Clinical Evidence
Bench testing only. Performance evaluated per ISO 11608-2, including dimensions, flow rate, bond strength, needle point defects, lubrication, and functional compatibility. Biocompatibility testing conducted per ISO 10993 (cytotoxicity, intracutaneous reactivity, sensitization, acute systemic toxicity, pyrogenicity, and hemolysis). Sterility validated per ISO 11135 with SAL of 10^-6.
Technological Characteristics
Materials: 304 stainless steel needle, polypropylene hub/outer cap, polyethylene inner cap, polydimethylsiloxane lubricant. Dimensions: 30G-33G gauges, 4mm-8mm lengths. Energy: None (manual). Connectivity: None. Sterilization: EO gas (ISO 11135). Standards: ISO 11608-2 (performance), ISO 10993 (biocompatibility).
Indications for Use
Indicated for subcutaneous injection of fluids, including insulin and exenatide, using pen injector devices.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
K110105 — BD PEN NEEDLE · Becton, Dickinson and Company (BD) · Apr 28, 2011
K160199 — Advocate Insulin Pen Needles · Diabetic Supply of Suncoast, Inc. · Oct 19, 2016
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized graphic of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 7, 2016
Yidobio Inc. Kyung Hee Yoon Excutive Director #3-1, 59-12, Dong-gil, Hyangnam-eup, Hwaseong-si Gyeonggi-do 18624 KOREA
Re: K160575
Trade/Device Name: CareSens Pen Needle and Softip Pen Needle Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: II Product Code: FMI Dated: July 29, 2016 Received: August 10, 2016
Dear Kyung Hee Yoon:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Tina
Kiang -S
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K160575
Device Name
CareSens Pen Needle and Softip Pen Needle
Indications for Use (Describe)
The CareSens Pen Needle and Softip Pen Needle are intended for use with pen injector devices for subcutaneous injection of fluids, including insulin and exenatide.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <svg height="16" width="16"> <path d="M2 2h12v12H2z" fill="none" stroke="black" stroke-width="2"></path> <path d="M4 8l4 4l6-6" stroke="black" stroke-width="2"></path> </svg> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span> <svg height="16" width="16"> <path d="M2 2h12v12H2z" fill="none" stroke="black" stroke-width="2"></path> <path d="M4 8l4 4l6-6" stroke="black" stroke-width="2"></path> </svg> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
| > Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows a logo for YIDOBIO. The logo consists of two stylized letters, 'O' and 'B', intertwined together in a calligraphic style. Below the letters, the word "YIDOBIO" is written in a simple, sans-serif font. The color of the logo and the text is a muted gray.
# 510(k) SUMMARY – K160575
A summary of 510(k) information in accordance with the requirements of 21 CFR 807.92.
| 1. Prepared by: | Joon Ho Jung<br>Regulatory Affairs<br>i-SENS, Inc.<br>Tel. +82-02-910-0453<br>Fax. +82-02-942-2514 |
|-----------------------|----------------------------------------------------------------------------------------------------------------------|
| 2. Prepared for: | Owner/Operator<br>YidoBio Inc.<br>59-12, Dongo4-gil, Hyangnam-eup<br>Hwaseong-si, Gyeonggi-do |
| 3. Contact person: | Kyung Hee Yoon<br>Executive Director<br>YidoBio Inc.<br>59-12, Dongo4-gil, Hyangnam-eup<br>Hwaseong-si, Hyangnam-eup |
| 4. Date Prepared: | February 1, 2016 |
| 5. Subject Device: | |
| Proprietary Name: | CareSens Pen Needle and Softip Pen Needle |
| Common Name: | Insulin Pen Needle |
| Regulation Name: | Needle, Hypodermic, Single Lumen |
| Regulation Number: | 21 CFR 880.5570 |
| Product Code: | FMI |
| Regulatory Class: | Class II |
| 6. Device Description | |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows a stylized monogram of the letters 'O' and 'B' in a flowing, intertwined design. The letters are rendered in a muted gray color, giving them a soft and elegant appearance. The design is intricate, with the curves of the letters overlapping and connecting to create a unified and visually appealing emblem. The overall impression is one of sophistication and classic style.
YIDOBIO
CareSens Pen Needle and Softip Pen Needle are sterile, non-pyrogenic, single-use injection needles intended for use with injection pens. This needle is a device for subcutaneous injection of medicines such as insulin and exenatide used with pen injectors for subcutaneous injection. CareSens Pen Needle and Softip Pen Needle are the same product in exception to the brand name.
#### 7. Predicate Device
The legally marketed device(s) to which substantial equivalence is claimed is/are:
- K063466 TopFine® Insulin Pen, ● Manufactured by Daejin Tech Medical Manufacturing Co., Ltd., Korea
| Regulation Name: | Needle, Hypodermic, Single Lumen |
|--------------------|----------------------------------|
| Regulation Number: | 21 CFR 880.5570 |
| Product Code: | FMI |
| Regulatory Class: | Class II |
#### 7.1 Reference device
- K 131358 BD 31G and 32G Extra Thin Wall (XTM) Pen Needles with PentaPoint, manufactured by Becton, Dickinson and Company, New Jersey.
| Regulation Name: | Needle, Hypodermic, Single Lumen |
|--------------------|----------------------------------|
| Regulation Number: | 21 CFR 880.5570 |
| Product Code: | FMI |
| Regulatory Class: | Class II |
#### 8. Reason for 510(K) Submission
This 510(k) is submitted to demonstrate substantial equivalence to the predicate device. CareSens Pen Needle and Softip Pen Needle. CareSens Pen Needle and Softip Pen Needle are the same product in exception to the brand name.
The specifications of the product are as follows:
| No. | Gauge (Outer diameter of Needle) | Length of needle | Weight |
|-----|----------------------------------|------------------|--------|
|-----|----------------------------------|------------------|--------|
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows a stylized monogram consisting of the letters 'O' and 'B'. The letters are intertwined and feature elegant, flowing lines with flourishes, giving the monogram a sophisticated and decorative appearance. The color of the monogram is a muted gray, which adds to its subtle and refined aesthetic. The overall design suggests a logo or emblem, possibly for a brand or personal use.
| | | (mm) l | (g) |
|---|-------------------------|------------|------|
| 1 | 30G (0.298mm ~ 0.320mm) | 8mm(5/16") | |
| 2 | 31G (0.254mm ~ 0.267mm) | 5mm(3/16") | |
| 3 | | 6mm(1/4") | |
| 4 | | 8mm(5/16") | 1.22 |
| 5 | 32G (0.229mm ~ 0.241mm) | 4mm(5/32") | |
| 6 | | 5mm(3/16") | |
| 7 | | 6mm(1/4") | |
| 8 | 33G (0.204mm ~ 0.215mm) | 4mm(5/32") | |
#### 9. Indication For Use
The CareSens Pen Needle and Softip Pen Needle are intended for use with pen injector devices for subcutaneous injection of fluids, including insulin and exenatide.
# 10. Technological Characteristics
#### Comparison with Predicate Device (K063466 TopFine Insulin Pen Needle)
The subject device, CareSens Pen Needle and Softip Pen Needle are similar in its intended use, principle of operation, design construction, sizes, material, and sterilization method. The subject device and predicate device both completed biocompatibility and sterility testing with similar results.
The subject device and the predicate device are classified under 21 CFR 880.5570. which states: "A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin." The proposed device and the predicate device use statements similar in that they both are used with pen injector devices for subcutaneous injection of fluids, including insulin. The subject device is also used to inject exenatide.
| Similarities between the Subject Devices and the Predicate Device | | |
|-------------------------------------------------------------------|------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------|
| Item | Topfine Insulin Pen Needle<br>(k063466, Predicate device) | CareSens Pen Needle and<br>Softip Pen Needle<br>(subject device) |
| Intended use | Intended for subcutaneous<br>injection of insulin in the<br>treatment of diabetes. | Intended for use with pen<br>injector devices for<br>subcutaneous injection of fluids,<br>including insulin and exenatide. |
| Operation Principle | To be used with pen<br>injectors/Single use | Same |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows a stylized monogram of the letters 'O' and 'B' intertwined. The letters are rendered in a soft gray color and feature flowing, calligraphic lines. The 'O' is positioned to the left and slightly above the 'B', with the curves of the letters overlapping to create a unified design. The overall effect is elegant and decorative.
#### YIDOBIO
| Design/Construction | • Needle assembly (cannula,<br>needle hub, protector cap) | Same |
|-----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|------|
| Material | • Outer cap - Polypropylene<br>• Inner Cap - Polyethylene<br>• Needle – 304 Stainless<br>• Needle Silicone -<br>Polydimethylsiloxane<br>• Needle Hub – Polypropylene | Same |
| Gauges/<br>length of needle | 30G 8mm, 31G 5mm, 31G 6mm,<br>31G 8mm, 32G 4mm, 32G 5mm,<br>32G 6mm, 33G 4mm | Same |
| Sterilization | EO Gas sterilization | Same |
| Biocompatibility | In accordance to ISO 10993 | Same |
# Comparison with the Reference Device (K131358 BD 31G and 32G Extra Thin Wall (XTM) Pen Needles with PentaPoint)
The reference device is used in comparison for performance bench testing in accordance to ISO 11608-2. The reference device is also indicated for use of the injection of both insulin and exenatide.
| Similarities between the Subject Devices and the Reference Device | | |
|-------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|
| Item | CareSens Pen Needle and<br>Softip Pen Needle<br>(subject device) | BD 31G and 32G Extra<br>Thin Wall (XTM) Pen<br>Needles with PentaPoint<br>(K131358, Reference Device) |
| Intended use | CareSens Pen Needle and Softip<br>Pen Needle are Intended for use<br>with pen injector devices for<br>subcutaneous injection of fluids,<br>including insulin and exenatide. | BD Pen Needle is intended<br>for use with pen injector<br>device for subcutaneous<br>injection of drugs, including<br>insulin and exenatide. |
| Operation Principle | To be used with pen<br>injectors/Single use | Same |
| Design/Construction | Needle assembly (cannula,<br>needle hub, protector cap) | Same |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows a stylized logo with intertwined letters, likely a monogram. The letters appear to be 'C' and 'B', rendered in a flowing, calligraphic style. Below the monogram, there is some text, but it is too blurry to read. The logo is presented in a light gray color against a white background.
| Material | • Outer cap - Polypropylene<br>• Inner Cap - Polyethylene<br>• Needle – 304 Stainless<br>• Needle Silicone –<br>Polydimethylsiloxane<br>• Needle Hub – Polypropylene | • Outer cap – Unknown plastic<br>• Inner Cap – Unknown plastic<br>• Needle - Stainless steel<br>• Needle Silicone – Unknown<br>• Needle Hub – Unknown<br>plastic |
|-------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sterilization | EO Gas sterilization | Gamma irradiation |
| Performance Bench | ISO 11608-2 | ISO 11608-2 |
Bench tests relating to the performance of CareSens Pen Needle and Softip Pen Needle were conducted. Testing to voluntary standards ISO 11608-2 provides additional evidence that the performance of CareSens Pen Needle and Softip Pen Needle demonstrated equivalent performance to the Reference Device.
#### 11. Summary of performance tests
| Dimension testing | ISO11608-2:2012 (4.2 Dimensions / 4.2.1 General / 4.2.2 Dimensions for needles), meets requirements |
|---------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Flow rate testing | ISO11608-2:2012 (4.3 Determination of flow rate through the needle), meets requirements |
| Bond between hub and needle tube testing | ISO11608-2:2012 (4.4 Bond between hub and needle tube), meets requirements |
| Needle point freedom from defects lubrication test | ISO11608-2:2012 (4.5 Needle point / 4.6 Freedom from defects / 4.7 Lubrication), meets requirements |
| Dislocation of measuring point at patient end | ISO11608-2:2012 (4.8 Dislocation of measuring point at patient end), meets requirements |
| Compatibility of needles and injector system test / Ease of assembly and disassembly. | ISO11608-2:2012 (4.9 Determination of functional compatibility with needle-based injection systems / 4.10 Ease of assembly and disassembly), meets requirements |
# Biocompatibility
Biocompatibility testing of the subject device, Softip Pen Needle and CareSens Pen Needle, were conducted so ensure no physical harm on the human skin. Results of the testing demonstrate that the device is biocompatible. Screening tests were performed on accelerated aged devices to show that the biocompatibility is maintained. Results of the testing demonstrate that the device is biocompatible.
| No. | Test Items | Test Method | Result |
|-----|-------------------|-----------------------------------------------|-----------|
| 1 | Cytotoxicity test | ISO 10993-5, Tests for in vitro Cytotoxicity, | Non- |
| | | Test on extracts | Cytotoxic |
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows a stylized monogram consisting of the letters 'O' and 'B' intertwined. The letters are rendered in a flowing, calligraphic style with delicate curves and flourishes. The color of the monogram is a muted gray, giving it a soft and elegant appearance.
#### YIDOBIO
| 2 | Intracutaneous<br>(Intradermal)<br>Reactivity Test | ISO 10993-10,<br>Intracutaneous(Intradermal) reactivity test | Negligible |
|---|--------------------------------------------------------|--------------------------------------------------------------|--------------------|
| 3 | Sensitization<br>Test(Guinea pig<br>maximization test) | ISO 10993-10, Sensitization(Guinea pig<br>maximization test) | Weak<br>sensitizer |
| 4 | Acute systemic<br>toxicity test | ISO 10993-11, Acute systemic toxicity | Non-toxic |
| 5 | Pyrogen test | ISO 10993-11, USP Rabbit Pyrogen Test | Non-<br>pyrogenic |
| 6 | Haemolysis test | ISO 10993-4, Haemolysis testing-general<br>considerations | Non-<br>hemolytic |
#### Sterilization
EO gas sterilization method is utilized for assuring sterility of the subject device, Softip Pen Needle and CareSens Pen Needle. The sterility of the device is assured using a sterilization method validated in accordance with ISO 11135. The sterilization validation has provided a Sterility Assurance Level (SAL) of 10°.
| No. | Test items | Test Method | Result |
|-----|-------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| 1 | Packaging | The case of injection or outer package should keep clean<br>from microorganism and should not be broken before<br>use. Also needle should be packed one by one. | PASS |
| 2 | Sterility Test-Direct<br>inoculation | The Korean pharmacopoeia Tenth edition<br>(same as the FDA recognized consensus standard<br>"USP 25<71> sterility test" are the same in test<br>protocol.) | PASS |
| 3 | Ethylene oxide<br>sterilization residuals | If the product is sterilized by EO gas, it should be tested<br>in accordance with ISO 10993-7, Ethylene oxide<br>sterilization residuals. The Sterilant Residual Limits per<br>ISO 10993-7 for a limited exposure (<24 hours) device<br>is as follows: EO ≤4 mg first 24 hrs and ECH ≤9 mg<br>first 24 hrs. | PASS |
# 12. Conclusion:
Based on the submitted information in this premarket notification, the subject devices are substantially equivalent to the predicate device.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.