K133100 · Facet Technologies · FMI · Jan 23, 2014 · General Hospital
Device Facts
Record ID
K133100
Device Name
CAREFINE PEN NEEDLE
Applicant
Facet Technologies
Product Code
FMI · General Hospital
Decision Date
Jan 23, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5570
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The CareFine™ Pen Needle is intended for use with a pen injector device for the subcutaneous injection of insulin.
Device Story
CareFine™ Pen Needles are single-use, sterile, double-ended cannulas designed for subcutaneous insulin delivery. The device consists of a cannula assembled into an injection-molded hub with internal threads for attachment to pen injectors. The hub allows the cartridge-end of the cannula to penetrate the pen's rubber septum. Both ends of the cannula are lubricated with silicone to facilitate injection and septum penetration. An inner shield protects the patient-end point, and an outer cover provides a sterility barrier. Used by consumers, caregivers, and healthcare professionals in home or clinical settings. The device facilitates insulin administration by connecting the insulin cartridge to the patient's subcutaneous tissue.
Clinical Evidence
No clinical data required; substantial equivalence supported by non-clinical performance testing, including biocompatibility, sterility, and comparative bench testing against predicate devices.
Technological Characteristics
Single-lumen, double-ended stainless steel cannula; injection-molded hub and shields; silicone lubricant; EtO terminal sterilization (SAL 10^-6); various gauges (29G-32G) and lengths (4mm-12.7mm).
Indications for Use
Indicated for use with pen injector devices for subcutaneous insulin injection in patients requiring insulin therapy.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
{0}------------------------------------------------
KI33100
### Facet Technologies, LLC Traditional 510(k) for the CareFine™ Pen Needles Submitted Under 21 CFR § 807.87
JAN 2 3 2014
# Confidential and Proprietary
9 510(k) Summary
55
{1}------------------------------------------------
# 510(k) Summary as Required by 21 CFR §807.92(c)
As required by the Safe Medical Devices Act (SMDA) of 1990 and in accordance with 21 CFR §807.92(a), the 510(k) summary provided below is of sufficient detailed to provide an understanding of the basis for a determination of substantial equivalence.
| Date<br>Prepared | Summary | September 25, 2013 |
|---------------------------|---------|----------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Sponsor<br>Address | | Facet Technologies, LLC<br>112 Town Park Drive, Suite 300<br>Kennesaw, Georgia 30144 |
| Contact Person | | Mary Ann Kinard<br>Senior Regulatory Affairs Manager<br>Phone: (770) 590-6462<br>FAX: (770) 590-6412<br>Email:MaryAnn_Kinard@facettechnologies.com |
| Trade Name | | CareFine™ Pen Needles |
| Common Name | | Insulin Pen Needle |
| Classification Name | | Needle, Hypodermic, Single Lumen |
| Regulation Number | | 21 CFR §880.5570 |
| Product Code | | FMI |
| Device Classification | | II |
| Review Panel | | 80 General Hospital |
### 510(k) Summary
CORPORATE HEADQUARTERS 112 Town Park Drive, Suite 300 Kennesaw, GA 30144 Phone: 770-590-6400 Fax: 770-590-6461
MANUFACTURING CENTER 101 Liberty Industrial Parkway McDonough, GA 30253 Phone: 770-898-3223 Fax: 770-898-1535
EUROPEAN OFFICE Unit 7 Mobbs Miller House Christchurch Road Northampton NN1 5LL U.K. Phone: +44 1604 638340 Fax: +44 1604 636 901
Confidential and Proprietary Copyright ® 2013 Facet Technologies, LLC
{2}------------------------------------------------
#### Predicate Marketed Devices:
| [807.92(a)(3)] | | |
|--------------------|--------------------------|---------------|
| Manufacturer | Trade Name | 510(k) Number |
| UltiMed, Inc. | UltiCare® Disposable Pen | K100812 |
| Novo Nordisk, Inc. | FlexPen® | K090111 |
| Becton Dickinson | BD Pen Needle | K100005 |
## [(807.92(a)(4)] Description of Device:
CareFine™ Pen Needles are single use, sterile medical devices designed to be used in conjunction with pen injectors and pen cartridges for the subcutaneous injection of insulin. Pen needles are used by consumers, caregivers and health care professionals. They are offered in various gauge sizes (29G, 30G, 31G and 32G) and lengths (4mm, 5mm, 6mm, 8mm, and 12.7mm). CareFine™ Pen Needles are sterile (EtO terminal sterilization), non-toxic and non-pyrogenic.
The pen needle assembly consists of a doubled-ended cannula that is assembled into an injection molded hub using adhesive. The hub has internal threads, which allow it to be screwed onto the pen injector device. This allows the cartridge end of the cannula to penetrate through the rubber septum of the cartridge. The patient-end and the cartridgeend of the cannula are lubricated using a silicone based lubricant for ease of injection and rubber septum penetration.
An injection-molded inner shield is assembled over the patient-end of the cannula to protect the point from damage and accidental needle-sticks. This needle assembly is inserted into a protective injection-molded outer cover and sealed with a peel-away label to provide a sterility barrier and tamper evidence seal. The peel-away label is pre-printed with information, which includes the lot number and needle gauge and length.
The individual needle assemblies are packaged in cartons, and placed into shippers with appropriate labeling. The shipper cases are palletized and sterilized to a SAL of 10° by EtO terminal sterilization.
# [(807.92(a)(5)] Indication for Use:
The CareFine™ Pen Needle is intended for use with a pen injector device for the subcutaneous injection of insulin.
{3}------------------------------------------------
# [(807.92(a)(6)] Technological Characteristics:
The CareFine™ Pen Needles are substantially equivalent to the UltiCare®, Flex Pen® and BD Pen Needles in terms of indications for use, compositions, material and design. Specifically, the following comparisons were made to determine equivalence of the CareFine™M Pen Needle to the referenced predicate pen needles:
- Needle sharpness, .
- . length, gauge,
- needle shield color(s), .
- needle shield dimensions, .
- . needle shield strength,
- hub/needle bond strength, .
- . biocompatibility, materials and
- . sterility
The CareFine™ Pen Needle product family is equivalent to the predicate devices, given that it:
- . Does not modify the indication for use or any labeling change that affects the intended use of the device,
- Does not alter the fundamental scientific technology, .
- Uses the same operating principles, .
- . Is manufactured from the same/similar materials,
- Is sterilized with a resulting SAL of 10-6, and .
- Is packaged using same/similar unit and case materials. .
Based on the comparisons described above to the predicate comparator devices, the CareFineTM Pen Needles do not raise any new issues of safety and effectiveness.
### [(807.92(b)(1)] Non-Clinical Performance Data:
The CareFine™ Pen Needle is manufactured in accordance with the requirements of the current Current Good Manufacturing Practices for Medical Devices and the testing performed which support a determination of substantial equivalence includes:
- . appropriate biocompatibility tests,
- sterility tests. . ●
- predicate product comparative tests and ●
Confidential and Proprietary Copyright " 2013 Facet Technologies, LLC
{4}------------------------------------------------
### Facet Technologies, LLC Traditional 510(k) for the CareFine™ Pen Needles Submitted Under 21 CFR § 807.87
. performance tests.
The test results are in compliance with existing domestic and international standards. Based on the above test parameter results, the CareFine™ Pen Needles do not raise any new issues of safety and effectiveness.
# · [(807.92(b)(2)] Clinical Performance Data:
Clinical data is not required.
[(807.92(b)(3)] Conclusion:
Based on the design equivalency and the functional testing performed, Facet Technologies has determined that CareFine™ Pen Needles are equivalent to the predicate comparator devices currently cleared for marketing in the United States. Differences between the devices do not raise any significant issues of safety and effectiveness.
# [(807.92(d)] Other Information:
None
End of 510(k) Summary
{5}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three stripes extending from its wing. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
#### January 23, 2014
Facet Technologies, LLC Ms. Mary Ann Kinard Senior Regulatory Affairs Manager 112 Town Park Drive, Suite 300 Kennesaw, GA 30144
Re: K133100
Trade/Device Name: CareFine™ Pen Needles Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: II Product Code: FMI Dated: December 19, 2013 Received: December 20, 2013
Dear Ms. Kinard:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{6}------------------------------------------------
Page 2 - Ms. Kinard
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Image /page/6/Picture/6 description: The image contains the text "Kwame Ulmers". The text is arranged vertically, with "Kwame" on top and "Ulmers" below. To the right of the text is a design that includes the letters "EDA" and a triangle shape. The design appears to be stylized or abstract.
for
Erin I. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
510(k) Number (if known) K133100
#### Device Name
CareFine™ Pen Needles
#### Indications for Use (Describe)
The CareFine™ Pen Needle is intended for use with a pen injector device for the subcuraneous injection of insulin.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
(2) Over-The-Counter Use (21 CFR 801 Subpart C)
# PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON A SEPARATE PAGE IF NEEDED.
THE THE THE PEORIFDAIUSE ONLY ONLY ONLY CLI , "게 대표 프 .
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Image /page/7/Picture/13 description: The image shows the name "Sajjad H. Syed" in large, bold font. To the right of the name is a digital signature. The signature includes the text "Digitally signed by Sajjad H. Syed -S" and other information such as "DN: c=US, o=U.S. Government, ou=HHS, ou=FDA, People, cn=Sajjad H. Syed -S, 2342\19200300.100.1.1=2000601742 0 Date: 2014.01.23 10:40:38 -05'00'".
FORM FDA 3881 (9/13)
PSC Publishing Services (131) 1434722 Fir
Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.
{8}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.