K152334 · Sung Shim Medicare Co., Ltd. · FMI · Mar 17, 2016 · General Hospital
Device Facts
Record ID
K152334
Device Name
Sungshim Insulin Pen Needle
Applicant
Sung Shim Medicare Co., Ltd.
Product Code
FMI · General Hospital
Decision Date
Mar 17, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5570
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
This product is a single use sterile insulin pen needle intended for subcutaneous injection of insulin to patient with diabetes. It is to be used with compatible insulin injectors.
Device Story
Single-use sterile insulin pen needle; consists of hub, needle, needle cap, sterile cap, and sterile paper. Connects to pen-type insulin injectors for subcutaneous insulin delivery. Used by patients with diabetes or healthcare providers. Device maintains sterility via primary container and paper. Bench testing confirms mechanical integrity and compatibility with standard insulin injectors (e.g., Flex Pen, Lantus, Novo Let N).
Clinical Evidence
Bench testing only. Testing included biocompatibility, mechanical properties, sterility (ISO 11135:2007), and EO residuals (ISO 10993-7). Functional testing followed ISO 7864:1993. Compatibility with specific insulin injectors verified per ISO 11608-2.
Technological Characteristics
Materials: Polypropylene (sterile cap, hub), Polyethylene (needle cap), Stainless steel 304 (needle). EO gas sterilization. Dimensions: 29G-32G, lengths 4mm-13mm. Connectivity: Mechanical interface with insulin pens. Standards: ISO 7864:1993, ISO 10993-7, ISO 11135:2007, ISO 11608-2.
Indications for Use
Indicated for subcutaneous insulin injection in patients with diabetes. Compatible with insulin injectors.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features the department's name encircling a symbol of three stylized human profiles facing right, with flowing lines beneath them. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the circular seal.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 17, 2016
Sungshim Medicare Co., Ltd. c/o Mr. Peter Chung President Plus Global 300 Atwood Street Pittsburgh, Pennsylvania 15213
Re: K152334
Trade/Device Name: Sungshim Insulin Pen Needle Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: II Product Code: FMI Dated: February 4, 2016 Received: February 10, 2016
Dear Mr. Chung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina Kiang -S
for Erin I. Keith, M.S. Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K152334
Device Name Sungshim Insulin Pen Needle
Indications for Use (Describe)
This product to be used with insulin injector is a single use sterile insulin pen needle to be used for injecting insulin to patient with diabetes.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
PSC Publishing Services (301) 443-6740 EF
{3}------------------------------------------------
# 510(k) Summary
[as required by 807.92(c)]
# 1. Applicant
- 1) Company : Sungshim Medicare Co.,Ltd.
- 2) Address : 190, Maesil-ro, Sojeong-myeon, Sejong-si, Korea
- 3) Tel : 82-32-676-7066
- 4) Fax : 82-32-676-7063
- 5) Prepared date : Oct. 7, 2015
- 5) Contact person : Peter Chung, 412-687-3976
- 6) Contact person address : 300, Atwood Street, Pittsburgh, PA, 15213, USA
- 7) Submission date : Aug. 10, 2015
- 8) 510(k) number : K152334
### 2. Device Information
- 1) Trade name : Sungshim Insulin Pen Needle
- 2) Common name : Hypodermic single lumen needle
- 3) Classification name : Needle, Hypodermic, Single Lumen
- 4) Product code : FMI
- 5) Regulation number : 880.5570
- 6) Class of device : Class II
- 7) Panel : General Hospital
# 3. The legally marketed device to which we are claiming equivalence
K080904 Feel Fine Insulin Pen Needle
#### 4. Device description
Sungshim insulin pen needle consist of a Sterile cap, Needle, Hub and Sterile paper. The Primary container and Sterile paper function to sustain sterilization of the product. The Hub can be connected with pen type insulin syringe. The Needle cap protects the Needle. This device is single use.
#### 5. Intended Use :
This product is a single use sterile insulin pen needle intended for subcutaneous injection of insulin to patient with diabetes. It is to be used with compatible insulin injectors.
### 6. Performance data:
- (1) Bench test were performed. Bench testing included biocompatibility, mechanical testing, sterility testing including EO residues. The tests demonstrated that the device performs in a substantially equivalent manner to the predicate device. The following bench testing is performed to demonstrate the functionality is substantially equivalent.
Test criteria : ISO 7864:1993 Sterile hypodermic needles for single use
{4}------------------------------------------------
| Test item | Requirements | Results |
|----------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Inner/out side<br>and structure | When it was tested accordingly to the standard specifications for<br>medical device-"sterile hypodermic needles for single use" clause 2,<br>it should satisfy the requirements. | Pass |
| Measurement | When it was tested accordingly to the standard specifications for<br>medical device-"sterile hypodermic needles for single use" clause 3,<br>it should satisfy the requirements. | Pass |
| Elasticity test | When it was tested accordingly to the standard specifications for<br>medical device-"sterile hypodermic needles for single use" clause 4,<br>it should satisfy the requirements. | Pass |
| Draw test | When it was tested accordingly to the standard specifications for<br>medical device-"sterile hypodermic needles for single use" clause 6,<br>it should satisfy the requirements. | Pass |
| Package | When it was tested accordingly to the standard specifications for<br>medical device-"sterile hypodermic needles for single use" clause 11,<br>it should satisfy the requirements. | Pass |
| Ethylene oxide<br>sterilization<br>residuals | According to ISO 10993-7<br>ETO≤ 25ppm / ECH≤ 25ppm / EG≤250ppm | Pass |
| Sterility test | When it was tested accordingly to ISO 11135:2007 | Pass |
| Extraction test | | |
| Appearance | The sample solution shouled be colorless and transparent and<br>there should not be foreign materials. | Pass |
| pH | Difference in pH≤1.5 | |
| KMnO4 Reducing<br>adents | Difference in titiers≤ 2.0ml | |
| Evaporating residue | Difference in extractable≤ 1.0mg | |
| heavy Pb,Fe,Sn,Zn<br>metal Cd | Not greater than a comined total of 5mg/L of Pb, Fe, Sn and Zn<br>Shall be less than 0.1mg/L of Cd | |
| UV-vis spectrum | Difference in absorbance (250nm~350nm)≤ 0.1 | |
Compatibility testing was performed according to ISO 11608-2 and passed with the following injectors:
- Novo Nordisk A/S, Flex Pen / 3ml / Insulin sprat .
- Sanofi-Aventis Deutschland GmbH. LANTUS / 3ml / Insulin glargine ●
- Novo Nordisk A/S, Novo Let N / 3ml / Isophane insulin .
The performance tests demonstrated that Sungshim insulin pen needle performs in a substantially equivalent manner to the predicate device.
{5}------------------------------------------------
#### 7. Predicate device comparison table
| Manufacturer | | Sungshim Medicare Co.,Ltd. | | Feel Tech | | | | | | | | | | | | | | | | | | | |
|----------------------------------|--|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------|-------------------------------|--|--|--|--|--|--|--|--|-----------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|--|--|--|--|
| 510(k) No. | | | K080904 | | | | | | | | | | | | | | | | | | | | |
| Indication for use | | This product is a single use sterile insulin<br>pen needle intended for subcutaneous<br>injection of insulin to patient with<br>diabetes. It is to be used with compatible<br>insulin injectors. | These disposable sterile insulin pen<br>needles are intended for subcutaneous<br>injection of insulin in the treatment of<br>diabetes. | | | | | | | | | | | | | | | | | | | | |
| Product name | | Hypodermic single lumen needle | Hypodermic single lumen needle | | | | | | | | | | | | | | | | | | | | |
| Trade name | | Sungshim Insulin Pen Needle | Feel Fine Insulin Pen Needle | | | | | | | | | | | | | | | | | | | | |
| Model/type | | 20 model codes including 29Gx4mm | Feel Fine Insulin Pen Needle-29<br>Feel Fine Insulin Pen Needle-30<br>Feel Fine Insulin Pen Needle-31 | | | | | | | | | | | | | | | | | | | | |
| Appearance | | Image: Sungshim needle | Image: Feeltech needle | | | | | | | | | | | | | | | | | | | | |
| Product<br>configuration | | Hub<br>Needle cap<br>Primary container<br>Needle<br>Sterile paper | Hub<br>Needle cap<br>Sterile cap<br>Needle<br>Sterile paper | | | | | | | | | | | | | | | | | | | | |
| Material | | Part Material Sterile cap Polypropylene Needle cap Polyethylene Needle Stainless steel 304 Hub Polypropylene | | | | | | | | | | | Part Material Sterile cap Polypropylene Needle cap Polypropylene Needle Stainless steel 304 Hub Polypropylene | | | | | | | | | | |
| Length of parts (mm) | | | | | | | | | | | | | | | | | | | | | | | |
| Total length of<br>assembly | | 29.8 | 28.8 | | | | | | | | | | | | | | | | | | | | |
| Side length of<br>assembly(max) | | 15.2 | 15.5 | | | | | | | | | | | | | | | | | | | | |
| Inner diameter of<br>sterile cap | | 11.2 | 12.95 | | | | | | | | | | | | | | | | | | | | |
| Gauge | | 29G, 30G, 31G, 32G | 29, 30, 31G | | | | | | | | | | | | | | | | | | | | |
| Length of needle | | 4 mm, 5 mm, 6 mm, 8 mm, 13 mm | 5 mm, 8 mm, 12.7 mm | | | | | | | | | | | | | | | | | | | | |
| Sterilization | | EO Gas sterilization | EO Gas sterilization | | | | | | | | | | | | | | | | | | | | |
| Packagine | | Sterile cap(PP)+sterile paper | | Sterile cap(PP)+sterile paper | | | | | | | | | | | | | | | | | | | |
Although the Indications for Use is not identical to that of the predicate device, it does not change the intended use because both are single use sterile needles intended for subcutaneous injection of insulin for diabetes patients.
### 9. Conclusion:
Comparison results demonstrate that the specifications and performance of the device are substantially equivalent to the legally marketed predicate device.
Therefore, it is concluded that Sungshim insulin pen needle is substantially equivalent to the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.