FEEL-FINE INSULIN PEN NEEDLE

K080904 · Feel Tech · FMI · Jun 4, 2008 · General Hospital

Device Facts

Record IDK080904
Device NameFEEL-FINE INSULIN PEN NEEDLE
ApplicantFeel Tech
Product CodeFMI · General Hospital
Decision DateJun 4, 2008
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5570
Device ClassClass 2
AttributesTherapeutic, Pediatric

Indications for Use

These disposable sterile insulin pen needles are intended for subcutaneous injection of insulin in the treatment of diabetes.

Device Story

Feel Fine Insulin Pen Needle is a sterile, single-use, disposable device designed for subcutaneous insulin delivery. It consists of a needle hub, needle, needle cap, and sterile cap with blister paper. The device attaches to an insulin pen to facilitate injection. It operates on the principle of a common piston syringe. The device is intended for use by patients or healthcare providers for diabetes management. It provides a sterile fluid path and is non-toxic and non-pyrogenic.

Clinical Evidence

Bench testing only. Testing included biocompatibility, mechanical performance, and sterility validation (including EO residues). No clinical data was required or provided.

Technological Characteristics

Sterile, single-use, disposable hypodermic needle. Components include needle hub, needle, needle cap, and sterile cap. Sterilized via ethylene oxide (EO). Operates as a common piston syringe. No software or electronic components.

Indications for Use

Indicated for subcutaneous insulin injection in patients with diabetes.

Regulatory Classification

Identification

A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ page 1 of 1 JUN - 4 2008 510(K) Summary for K080904 Feel Tech # 92-4 Haengjung-ri, Gwangdcok-myeon, Cheonan-si, Choong-nam Korea Tel. : 82-41-522-2446-8 Fax .: 82-41-522-2449 Contact person: Boo Sool Kim Date prepared: June 2, 2008 Trade Name: Feel Fine Insulin Pen Ncedle 1. - Common Name: Pen needle 2. - Classification Name: Syringe, piston, product code FMI, Regulation: 880.5570 3. - 4. Class of device: Class II. - The legally marketed device to which we are claiming equivalence [807.92(a)(3)) : 5. B.Braun "one.clickTM needle", K033575. - Description of device: Feel Fine Insulin Pen Needle consists of a sterile cap, needle cap 6. needle hub, which can be fixed with needle and blister paper. The sterile cap functions to maintain the sterility of the needle because sterile cap covers the needle hub and needle cap with blister paper sealed on the opening hole of sterile cap. The needle hub can be connected with pen. The necdle cap covers intended to provide physical protection to the needle tube. They are supplied with a sterile fluid path, (EO), nontoxic, and non pyrogenic, for single use only, disposable. The devices operate on the principles of common piston syringes. - Intended use: These disposable sterilc insulin pen needles are intended for 7. subcutaneous injection of insulin in the treatment of diabetes. - Technological characteristics: Feel Fine Insulin Pen Needles and the predicate devices 8. have identical technological characteristics and perform the same way as common piston syringes. These needles are EO sterilized. - Performance: Bench tests were performed. Bench testing included biocompatibility, 9. mechanical testing, sterility testing including EO residues. The tests demonstrated that the device is as safe, as effective, and performs in a substantially equivalent manner to the predicate device. {1}------------------------------------------------ Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circle around the eagle. The eagle is black, and the text is also black. Food and Drug Administration 9200 Corporate Bouievard Rockville MD 20850 JUN - 4 2008 Feel Tech C/O Mr. Daniel Kamm Principal Consultant Kamm & Associates P.O. Box 7007 Deerfield, Illinois 60015 Re: K080904 Trade/Device Name: Feel Fine Insulin Pen Needle Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: II Product Code: FMI Dated: March 28, 2008 Received: April 1, 2008 Dear Mr. Kamm: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register, {2}------------------------------------------------ Page 2 - Mr. Kamm Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Clare Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ K080904 ## Indications for Use 510(k) Number (if known): K080904 Feel Fine Insulin Pen Needle Device Name: Indications For Use: These disposable sterile insulin pen needles are intended for subcutaneous injection of insulin in the treatment of diabetes. V Prescription Us (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Anthony O. Natan (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices Page 1 of 1 510(k) Number: K4849994
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...