This is a single use sterile insulin pen-injector needle to be used by diabetic patients for the purpose of injecting insulin.
Device Story
Single-use sterile insulin pen-injector needle; consists of hub, needle, needle cap, sterile cap, and blister paper. Connects to various pen-type insulin syringes/injectors. Used by diabetic patients for self-administration of insulin. Needle cannula penetrates skin for subcutaneous delivery; withdrawn immediately after injection. Device ensures sterile delivery of medication; protects user via needle cap. Benefits include safe, effective insulin administration for diabetes management.
Clinical Evidence
Bench testing only. Biocompatibility testing performed per ISO 10993-1 (cytotoxicity, sensitization, intracutaneous reactivity, systemic toxicity, pyrogenicity, hemolysis). Performance testing per ISO 11608-2, ISO 7864, and ISO 9626 included dimension, draw, elasticity, flexural rigidity, flow rate, bond strength, and needle dose accuracy. Results demonstrate substantial equivalence to predicate.
Technological Characteristics
Single lumen hypodermic needle; stainless steel 304 cannula; polypropylene hub; polyethylene needle cap. EO gas sterilization. Available in 29G, 30G, 31G, and 32G. Complies with ISO 11608-2, ISO 7864, ISO 9626, and ISO 10993-1 standards.
Indications for Use
Indicated for diabetic patients requiring subcutaneous insulin injection via pen-injector systems.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around the perimeter. Inside the circle is an image of three human profiles facing to the right. The profiles are stacked on top of each other, with the first profile being the most prominent.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 21, 2016
CPL Co., Ltd. c/o Mr. Peter Chung President Plus Global 300 Atwood Street Pittsburgh, Pennsylvania 15213
Re: K151090
Trade/Device Name: CPL Insulin Pen Needle Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: II Product Code: FMI Dated: August 16, 2016 Received: August 23, 2016
Dear Mr. Chung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina Kiang
-S
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K151090
Device Name
CPL Insulin Pen Needle
Indications for Use (Describe)
This is a single use sterile insulin pen-injector need by diabetic patients for the purpose of injecting insulin.
Type of Use (Select one or both, as applicable)
الص Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120
Expiration Date: January 31, 2017
{3}------------------------------------------------
# 510(k) Summary
# [as required by 807.92(c)] K151090
## 1. Applicant
- 1) Company : CPL Co.,Ltd.
- 2) Address : 36, Yongteurim-gil, Danwon-gu, Ansan-si, Gyeonggi-do, Korea
- 3) Tel : 82-31-483-7301
- 4) Fax : 82-31-483-7351
- 5) Contact person : Peter Chung, 412-687-3976
- 6) Contact person address : 300, Atwood Street, Pittsburgh, PA, 15213, USA
- 7) Date prepared : September 16, 2016
## 2. Device Information
- 1) Trade Name : CPL Insulin Pen Needle
- 2) Common Name : Hypodermic single lumen needle
- 3) Classification Name : Needle, Hypodermic, Single Lumen
- 4) Product Code : FMI
- 5) Regulation Number : 880.5570
- 6) Class of device : Class II
- 7) Panel : General Hospital
## 3. The legally marketed device to which we are claiming equivalence
K080904 Feel Fine Insulin Pen Needle (Company : Feel Tech)
### 4. Device description
CPL insulin pen needle consist of a sterile cap, needle, hub and blister paper. The sterile cap and blister paper function to sustain sterilization of the product. The hub can be connected with pen type insulin syringe. The needle cap protects the needle. This device is single use.
| | Model | Needle length (mm) : | |
|----------------------------|--------------|------------------------|---------------------------|
| Patient-side tip | CPLPN-29 | 8mm, 10mm, 12mm | |
| | CPLPN-30 | 5mm, 8mm, 10mm, 12 mm | |
| | CPLPN-31 | 4mm, 5mm, 6mm, 8mm | |
| | CPLPN-32 | 4mm, 5mm, 6mm, 8mm | |
| Cartridge-side tip | CPLPN-29 | | |
| | CPLPN-30 | 5mm, | |
| | CPLPN-31 | | |
| | CPLPN-32 | | |
| Needle taper | | 3.5mm | |
| Wall type<br>(Thin-walled) | Needle gauge | Range of out diameters | Inside diameter of tubing |
| | CPLPN-29 | 29G(0.324mm~0.351mm) | 29G (0.190) |
| | CPLPN-30 | 30G (0.298mm~0.320mm) | 30G (0.165) |
| | CPLPN-31 | 31G (0.254mm~0.267mm) | 31G (0.125) |
| | CPLPN-32 | 32G (0.229mm~0.241mm) | 32G (0.105) |
{4}------------------------------------------------
### 5. Intended Use :
This is a single use sterile insulin pen-injector needle to be used by diabetic patients for the purpose of injecting insulin.
### 6. Performance data:
In accordance with ISO 10993-1, CPL Insulin Pen Needle is classified as : Externally Communicating Device, Blood Path Indirect, Shor Term (<24 hours) Use, as the cannula is immediately withdrawn after injection into the body. This classification was chosen as "worst case scenario."
PER ISO 10993-1, the following tests were performed for this classification:
- 1. Cytotoxicity
- 2. Sensitization
- 3. Intracutaneous reactivity (Acute)
- 4. Systemic toxicity (Acute)
- 5. Pyrogenicity
- 6. Hemolysis
Bench tests relating to the performance of the "CPL Insulin Pen Needle" were conducted. The principal device demonstrated equivalent performance to the predicate devices during bench testing. Bench testing consisted of:
| Inner/outside and structure | |
|---------------------------------------------------|---------------|
| Dimension | |
| Draw test | |
| Elasticity test | |
| Flexual rigidity | |
| Inner and outside of needle | |
| Size designation | |
| Elasticity of the needle tube | |
| Pull | Test standard |
| Lubricant | |
| Limits for acidity and alkalinity | ISO 11608-2 |
| Stiffness | |
| Resistance to breakage | ISO 7864 |
| Resistance to corrosion | |
| Determination of flow rate through the needle | ISO 9626 |
| Bond between hub and needle tube | |
| Freedom from defects | |
| Penetration resistance | |
| Cap-hub fitting strength | |
| Dislocation of measuring point at patient end | |
| Needle dose accuracy test | |
| Needle hub torque removal | |
| Ease of assembly and disassembly test | |
| Compatibility with needle-based injection systems | |
The performance tests demonstrated that CPL Insulin Pen Needle is performes in a substantially equivalent manner to the predicate device.
{5}------------------------------------------------
# Compatible pens
| Victoza Liraglutide injection | HumaLog KwikPen |
|--------------------------------------------------|---------------------------------------------|
| Apidra SoloStar | HumaLog Mix 75/25 KwikPen |
| Lantus SoloStar | HumaLog Mix 50/50 KwikPen |
| OptiClik for Lantus and Apidra | Humalog Pen |
| SymlinPen 120 (pramlintide acetate) pen injector | Humalog Mix 75/25 Pen |
| SymlinPen 60 (pramlintide acetate) pen injector | Humalog Mix 50/50 Pen |
| Byetta exenatide injection 10mog | Humulin N Pen |
| Byetta exenatide injection 5mog | Humulin 70/30 Pen |
| Levemir FlexPen | HumaPen LUXURA HD |
| Novolog FlexPen | HumaPen MEMOIR |
| Novolog Mix 70/30 FlexPen | AutoPen |
| NovoPen Junior | Forteo teriparatide (rDNA origin) injection |
| NovoPen 3 | |
# 7. Predicate device comparison table
| Manufacturer | CPL Co.,Ltd. | | Feel Tech | Results | | |
|----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|--------------------------------|----------------------------------------------------------------------------------------------------------------------------------------|-------------------------------|-------------------------|-----------------------|
| 510(k) No. | K151090 | | K080904 | N/A | | |
| Indication for use | This is a single use sterile<br>insulin pen-injector needle to be<br>used by diabetic patients for the<br>purpose of injecting insulin. | | These disposable sterile insulin pen<br>needles are intended for<br>subcutaneous injection of insulin in<br>the treatment of diabetes. | Same intended use | | |
| Product name | Hypodermic single lumen needle | | Hypodermic single lumen needle | Identical | | |
| Trade name | CPL Insulin Pen Needle | | Feel Fine Insulin Pen Needle | N/A | | |
| Model/type | CPLPN-29, CPLPN-30, CPLPN-31,<br>CPLPN-32 | | Feel Fine Insulin Pen Needle-29<br>Feel Fine Insulin Pen Needle-30<br>Feel Fine Insulin Pen Needle-31 | N/A | | |
| Appearance | Image: CPL Insulin Pen Needle | | Image: Feel Fine Insulin Pen Needle | Similar design | | |
| Product<br>configuration | Hub<br>Needle cap<br>Sterile cap<br>Needle<br>Sterile paper | | Hub<br>Needle cap<br>Sterile cap<br>Needle<br>Sterile paper | Similar device<br>componenets | | |
| Material | Sterile cap : Polypropylene<br>Needle cap : Polyethylene<br>Needle : Stainless steel 304<br>Hub : Polypropylene | | Sterile cap : Polypropylene<br>Needle cap : Polyethylene<br>Needle : Stainless steel 304<br>Hub : Polypropylene | Similar material | | |
| Length of parts (mm) | | | | | | |
| Total length of<br>assembly | 29.8 | | 28.8 | | | |
| Side length of<br>assembly(max) | 15.2 | | 15.5 | Similar dimensions | | |
| Inner diameter of<br>sterile cap | 11.2 | | 12.95 | | | |
| Gauge | 29G | 30G | 31G | 32G | 29, 30, 31G | Similar |
| Length of needle | 8 mm<br>10 mm<br>12 mm | 5 mm<br>8 mm<br>10 mm<br>12 mm | 4 mm<br>5 mm<br>6 mm<br>8 mm | 4 mm<br>5 mm<br>6 mm<br>8 mm | 5 mm<br>8 mm<br>12.7 mm | Similar needle length |
| Sterilization | EO Gas sterilization | | EO Gas sterilization | Same | | |
| Packagine | Sterile cap(PP)+sterile paper | | Sterile cap(PP)+sterile paper | Similar | | |
| Performance test | Accordance with ISO 11608-2 | | N/A | Suitable | | |
| Biocompatibility | Accordance with ISO 10993 series | | N/A | Suitable | | |
{6}------------------------------------------------
The device is investigated for function and effectiveness to compare the operation of function between CPL Co.,Ltd. and Feeltech (K080904). Comparison results demonstrate that the specifications and performance of the device are same as functional and effective as the legally marketed predicate device. Therefore, it is concluded that is Insulin pen needle of CPL Co.,Ltd. substantially equivalent to the legally marketed predicate device.
Intended use statement is similar with that of the predicate device. They are both single use only, used for insulin injection and for diabetes treatment. The difference in wording of these intended use does not change the intended use or raise questions of safety and effectiveness per 807.92(a)(5).
The bench tests of Needles with 32G demonstrated conformances to ISO 11608-2 and ISO 7864. Therefore the differences do not raise new concerns to establish substantial equivalence to the predicate.
### 9. Conclusion:
The subject device has been tested according to ISO 11608-2:2012 (Needle-based injection systems for medical use-Requirements and test methods-Part 2:Needles) as well as ISO 9626 (stainless steel tubing for manufacture of medical devices) and test results demonstrated it is substantial equivalent to as the cited predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.