3D GraftRasp System

K201900 · Surgentec · FMF · Sep 3, 2020 · General Hospital

Device Facts

Record IDK201900
Device Name3D GraftRasp System
ApplicantSurgentec
Product CodeFMF · General Hospital
Decision DateSep 3, 2020
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5860
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The 3D GraftRasp System is intended to be used in orthopedic procedures to rasp or decorticate bone from the transverse processes and/or facets, and for the delivery of hydrated allograft, or synthetic bone graft material to an orthopedic surgical site.

Device Story

System comprises decortication rasps, graft pushers, bone funnel, tools, and dilators; used in orthopedic procedures. Physician dilates tissue for access; uses decortication rasp to manually scrape/remove cortical bone from transverse processes or facets. Once prepared, graft material is delivered via Graftgun Universal Graft Delivery System or bone funnel through the rasp lumen directly to the surgical site. System facilitates bone preparation and graft placement in a single workflow. Benefits include improved access to bony anatomy and efficient graft delivery. Used in operating room environments by orthopedic surgeons.

Clinical Evidence

Bench testing only. Included biocompatibility (ISO 10993-1), sterilization validation (ISO 17665-1), graft extrusion functionality, rasp decortication performance on bone blocks, and volume dispensing verification. Seven simulated use validation studies were conducted on cadavers and sawbones with orthopedic surgeons to evaluate usability, labeling, and technique, confirming the system's ability to perform intended decortication and graft delivery.

Technological Characteristics

Materials: Stainless Steel (316L, 304 per ASTM F899-12b), Nitinol Superelastic (ASTM F2063), Ti64 (ASTM F136-13). Passivation per ASTM A967-13. Manual mechanical operation. Components include rasps, dilators, pushers, and funnel. Sterilization: Steam (autoclave).

Indications for Use

Indicated for use in orthopedic procedures requiring bone decortication of transverse processes and/or facets and delivery of hydrated allograft, autograft, or synthetic bone graft material to the surgical site.

Regulatory Classification

Identification

A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. September 3, 2020 SurGenTec, LLC Andrew Shoup, COO 911 Clint Moore Road Boca Raton, Florida 33847 Re: K201900 Trade/Device Name: 3D GraftRasp System Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: Class II Product Code: FMF, HTR Dated: July 6, 2020 Received: July 8, 2020 Dear Mr. Shoup: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Laura C. Rose, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### 4.0 Indications for Use Statement | DEPARTMENT OF HEALTH AND HUMAN SERVICES | Form Approved: OMB No. 0910-0120 | |-----------------------------------------|----------------------------------| | Food and Drug Administration | Expiration Date: 06/30/2020 | | Indications for Use | See PRA Statement below. | 510(k) Number (if known) TBD K201900 Device Name 3D Graftrasp System ## Indications for Use (Describe) The 3D GraftRasp System is intended to be used in orthopedic procedures to rasp or decorticate bone from the transverse processes and/or facets, and for the delivery of hydrated allograft, or synthetic bone graft material to an orthopedic surgical site. Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) [ ] Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." FORM FDA 3881 (7/17) Page 1 of 1 {3}------------------------------------------------ #### 5.0 510(k) Summary In accordance with 21 CFR 807.87(h) and (21 CFR 807.92) the 510(k) Summary for the 3D GraftRasp System is provided below. #### 5.1 Submitter Information: | Submitter: | SurGenTec, LLC<br>911 Clint Moore Rd<br>Boca Raton, FL 33487<br>Telephone: (561) 990-7882 | |------------------------------------------------|-------------------------------------------------------------------------------------------| | Contact: | Andrew Shoup<br>COO<br>Phone 561-990-7882<br>Email: ashoup@surgentec.com | | Date Prepared: | July 6th, 2020 | | 5.2 Name of Device<br>Device Proprietary Name: | 3D GraftRasp System | | Device Common Name: | Piston Syringe and Bone Rasp | | Classification Regulation | 21 CFR 880.5860 Class II | | Classification name: | Piston Syringe | | Panel: | General & Plastic Surgery | | Product Code: | FMF, HTR | #### 5.3 Legally Marketed Predicate Device Predicate: 3D GraftRasp System Surgentec, LLC. Class II Device (K200431) #### 5.4 Device Description The 3D GraftRasp system contains various decortication rasps, graft pushers, a bone funnel, tools, and dilators. The 3D GraftRasp system is compatible with the Graftgun Universal Graft Delivery System for graft delivery to the surgical site. The decortication rasps can be used to rasp or decorticate bone from transverse processes and/or facets. After rasping, a 5cc or 7.5cc graft tube can be inserted into the rasp lumen to extrude graft from the tip of the rasp. A set of optional sequential dilators and rasp pushers may also be provided to aid in the use of the decortication rasp. A bone funnel may optionally be used to extrude graft through the rasps instead of the graftgun. {4}------------------------------------------------ # 5.5 Indication for Use The 3D GraftRasp System is intended to be used in orthopedic procedures to rasp or decorticate bone from transverse processes and/or facets, and for the delivery of hydrated allograft, autograft, or synthetic bone graft material to an orthopedic surgical site. | | Technological Characteristics and Substantial<br>Equivalence | 3D GraftRasp System | 3D GraftRasp System | |------------|--------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | Topic | | | | | 510(k) | TBD | K200431 | | | Regulation | 21 CFR 880.5860 | 21 CFR 880.5860 | | | Description | | | | | Device Name | 3D GraftRasp System | 3D GraftRasp System | | | Product Code | FMF, HTR | FMF, HTR | | | Classification | Class II | Class II | | | Indications for<br>Use | The 3D GraftRasp System is intended<br>to be used in orthopedic procedures to<br>rasp or decorticate bone from the<br>transverse processes and/or facets,<br>and for the delivery of hydrated<br>allograft, autograft, or synthetic bone<br>graft material to an orthopedic<br>surgical site. | The 3D GraftRasp System is intended<br>to be used in orthopedic procedures to<br>rasp or decorticate bone and for the<br>delivery of hydrated allograft,<br>autograft, or synthetic bone graft<br>material to an orthopedic surgical site. | | | Single Use | Yes / No* | Yes / No* | | | Sterility | Steam Sterilization | Steam Sterilization | | | Patient Contact<br>Material | Medical Grade:<br>Stainless Steel – 316L, 304<br>per ASTM F899-12b,<br>passivated per ASTM A967-<br>13Nitinol Superelastic per<br>ASTM F2063, passivated per<br>ASTM A967-13Ti64 per ASTM F136-13,<br>passivated per ASTM<br>A967-13. | Medical Grade:<br>Stainless Steel – 316L, 304<br>per ASTM F899-12b,<br>passivated per ASTM A967-<br>13Nitinol Superelastic per<br>ASTM F2063, passivated per<br>ASTM A967-13 | | | Packaging | Packaged in sterilization tray,<br>sterilized and cleaned prior to each<br>use via autoclave.<br>Replacement parts shipped in<br>nonsterile packaging including<br>labeling, IFU, and cleaning<br>instructions. Need to be assembled,<br>cleaned and sterilized prior to use. | Packaged in sterilization tray,<br>sterilized and cleaned prior to each<br>use via autoclave.<br>Replacement parts shipped in<br>nonsterile packaging including<br>labeling, IFU, and cleaning<br>instructions. Need to be assembled,<br>cleaned and sterilized prior to use. | | | Operating<br>Principle | The user can dilate tissue for easier<br>access to the surgical site. The<br>decortication rasp is then used to<br>remove cortical bone by manually<br>scraping the bony anatomy. Once the<br>bony anatomy is prepared, graft is<br>applied using the Groftoun Universal | The user can dilate tissue for easier<br>access to the surgical site. The<br>decortication rasp is then used to<br>remove cortical bone by manually<br>scraping the bony anatomy. Once the<br>bony anatomy is prepared, graft is<br>applied using the Groftoun Universal | | | Graft Delivery Device. It's inserted | Graft Delivery Device. It's inserted | | | | into the rasp and delivers graft | into the rasp and delivers graft | | | | through the rasp lumen directly to the | through the rasp lumen directly to the | | | | prepared surgical site. The shape of | prepared surgical site. The shape of | | | | the rasp allows to access to adjacent | the rasp allows to access to adjacent | | | | transverse processes and/or facets | vertebral bodies or other bony | | | | during decortication and graft | anatomy during decortication and | | | | delivery. | graft delivery. | | | Components | Orthopedic Rasp, Dilators, Pushers, Graft Funnel, Tools | Orthopedic Rasp, Dilators, Pushers, Graft Funnel | | Technological Characteristics and Substantial Equivalence 5.6 {5}------------------------------------------------ *Some components of the GraftRasp System are designed to be reusable while others are single use such as the replaceable rasp teeth and the flexible tip of the rasp pusher. #### 5.7 Performance Data The following non-clinical performance data were provided to demonstrate substantial equivalence of the subject device to the predicate. - Biocompatibility per ISO 10993-1:2018 ● - Sterilization validation per ISO 17665-1:2006/(R) 2013 ● - Bench Testing ● - o Functionality Common Material Test - l Various graft materials were tested to ensure the Graftgun can successfully extrude them through the various rasps of the 3D GraftRasp System. For each rasp, the Graftgun was able to successfully extrude the graft materials using an acceptable hand force. - Rasp Functionality Test o - The rasping surface of the rasp was tested on a sample bone ■ block to ensure the worst case scenario rasp of the 3D GraftRasp System is capable of bone decortication without failure. The rasp was able to successfully decorticate a portion of the bone block without failure of either the teeth of the rasping surface or of the main rasp body. - o Volume Dispensing Verification Test - The Graftgun Universal Graft Delivery system was tested to ensure an accurate amount of bone graft was dispensed with every squeeze of the trigger. It was justified that using the Graftgun along with either rasp and the accompanying graft pusher of the 3D GraftRasp System was adequate to maintain the acceptance criteria - Simulated Use Validation Testing O - Seven simulated use validation studies were conducted with ■ physician who are considered intended users. The validations were completed on cadavers and/or sawbones models to simulates typical use of the system in an operating room environment. These validations were designed to get feedback from a potential user and to validate the devices, labeling, usability, human factors, and technique. The user is supplied with a complete the appropriate labeling and Surgical {6}------------------------------------------------ Technique. The validations demonstrated that the GraftRasp System was able to be effectively used in orthopedic procedures to rasp or decorticate bone from transverse processes and/or facets, and for the delivery of hydrated allograft, autograft, or synthetic bone graft material to an orthopedic surgical site. It also evaluated multiple physicians' feedback on the safety and efficacy of the device, labeling and training. #### 5.8 Conclusion The design characteristics of the 3D GraftRasp System do not raise different questions of safety and effectiveness. This data supports that the 3D GraftRasp System is substantially equivalent to the predicate device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...