O-Genesis Graft Delivery System
K200606 · Orthofix, Inc. · FMF · May 1, 2020 · General Hospital
Device Facts
| Record ID | K200606 |
| Device Name | O-Genesis Graft Delivery System |
| Applicant | Orthofix, Inc. |
| Product Code | FMF · General Hospital |
| Decision Date | May 1, 2020 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 880.5860 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
The O-GENESIS Graft Delivery System is intended to be used for the delivery of allograft, autograft or synthetic bone graft material to an orthopedic surgical site.
Device Story
Manual graft delivery system; used in orthopedic surgery to deliver bone graft materials (allograft, autograft, synthetic). System components: loading syringe, plunger, funnel, delivery cannula, delivery gun with ratchet-actuated trigger handle. Operation: clinician loads graft material into cannula; trigger handle advances plunger to express material through cannula directly to surgical site. Single-use; sterile. Benefits: controlled, manual delivery of graft material to surgical site.
Clinical Evidence
Bench testing only. Included simulated use testing (loading and delivery of bone graft material), functional verification (ratchet plunger force, T-handle strength, threaded component interface), and biocompatibility evaluation (cytotoxicity, sensitization, acute systemic toxicity, material-mediated pyrogenicity per ISO 10993-1:2009). All acceptance criteria met.
Technological Characteristics
Materials: HDPE DMDA-8904 HEALTH+, HEPE DMDA-8907 NT7. Operating principle: ratchet-actuated plunger. Form factor: handheld delivery gun with cannula. Sterilization: Gamma irradiation (SAL 10^-6). Single-use.
Indications for Use
Indicated for patients requiring delivery of allograft, autograft, or synthetic bone graft material to an orthopedic surgical site.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
Predicate Devices
- GraftGun Universal Graft Delivery System (K170675)
Related Devices
- K160978 — LITe BIO Delivery System · Stryker Corporation · Jul 8, 2016
- K180937 — Graftgun Universal Graft Delivery System · SurGenTec, LLC · May 10, 2018
- K230492 — Avitus® Precision Autograft Delivery System · Avitus Orthopaedics, Inc. · Jun 26, 2023
- K210997 — PreferX Delivery System · Md3, LLC · Sep 7, 2021
- K243580 — GraftGun Universal Graft Delivery System (GDS) · SurGenTec, LLC · Feb 5, 2025
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left side of the image is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 1, 2020
Orthofix Troy Brooks, RAC Director, Regulatory Affairs 3451 Plano Parkway Lewisville, Texas 75056
Re: K200606
Trade/Device Name: O-GENESIS Graft Delivery System Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: Class II Product Code: FMF Dated: March 6, 2020 Received: March 9, 2020
Dear Troy Brooks:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K200606
Device Name O-GENESIS™ Graft Delivery System
Indications for Use (Describe)
The O-GENESIS Graft Delivery System is intended to be used for the delivery of allograft, autograft or synthetic bone graft material to an orthopedic surgical site.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Orthofix. The logo consists of a blue abstract shape on the left, followed by the word "ORTHOFIX" in black, bold letters. A small trademark symbol is located to the right of the "X" in "ORTHOFIX".
Page 1/3
# 510(k) Summary
# O-GENESIS™ Graft Delivery System
#### 510(k) Owner Information:
| Name: | Orthofix |
|-----------------------------------|-----------------------------------------------------------------------|
| Address: | 3451 Plano Parkway<br>Lewisville, TX 75056 |
| Phone Number: | 214-937-2047 |
| Fax Number: | 214-937-3322 |
| Email: | TroyBrooks@Orthofix.com |
| FDA Registration Number: | 2183449 |
| Contact Person: | Troy Brooks, RAC<br>Director, Regulatory Affairs<br>Orthofix |
| Date Prepared: | April 24, 2020 |
| Name of Device: | |
| Trade Name /<br>Proprietary Name: | O-GENESIS Graft Delivery System |
| Common Name: | Graft Delivery Device |
| Classification Name: | Piston Syringe |
| Product Code: | FMF |
| Regulatory Class: | Class II – 21 CFR 880.5860 |
| Review Panel: | General Hospital |
| Predicate Device: | K170675 - GraftGun Universal Graft Delivery System<br>(SurGenTec LLC) |
## Reason for 510(k) Submission:
Orthofix is submitting this Traditional 510(k) premarket notification for the new O-GENESIS Graft Delivery System.
#### Device Description:
The O-GENESIS Graft Delivery System is designed to deliver allograft, autograft or synthetic bone graft material to an orthopedic surgical site. The system consists of a loading syringe, a loading plunger, a loading funnel, a delivery cannula and a delivery gun with an actuating trigger handle. The system is provided sterile and is for single-use only.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the Orthofix logo. The logo consists of a blue abstract shape on the left and the word "ORTHOFIX" in black letters on the right. The abstract shape is a stylized representation of a circle with a gap in the middle. The word "ORTHOFIX" is written in a bold, sans-serif font, and there is a registered trademark symbol next to the "X".
Page 2/3
## Indications for Use:
The O-GENESIS Graft Delivery System is intended to be used for the delivery of allograft, autograft or synthetic bone graft material to an orthopedic surgical site.
#### Technological Characteristics of the Device Compared to the Predicate Device:
The technological characteristics of the O-GENESIS Graft Delivery System are similar to the predicate device in terms of intended use, indications for use, and fundamental technology, including basic design, sterility and operating principle. There are no significant differences between the subject device and the predicate that raise new issues regarding safety or effectiveness. The table below compares the key technological characteristics of the subject device to the predicate device.
| Comparator | Subject Device<br>O-GENESIS<br>Graft Delivery System<br>(K200606) | Predicate Device<br>GraftGun Universal<br>Graft Delivery System<br>(K170675) |
|------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code | FMF | FMF |
| Classification | Class II - 21 CFR 880.5860 | Class II - 21 CFR 880.5860 |
| Intended Use | Delivery of graft material to an<br>orthopedic surgical site. | Delivery of graft material to an<br>orthopedic surgical site. |
| Operating<br>Principle | Graft material expressed from a<br>cannula via a plunger, operated by<br>a ratchet-actuated handle.<br>Graft material expressed from the<br>cannula directly to the surgical site. | Graft material expressed from a graft<br>tube via a plunger, operated by a<br>ratchet-actuated handle.<br>Graft material expressed from a graft<br>tube directly to the surgical site. |
| Volume<br>Capacity | Up to 6.6ml in cannula | Up to 7.5ml in graft tube (cannula) |
| Direct Patient<br>Contacting<br>Material | HDPE DMDA-8904 HEALTH+<br>HEPE DMDA-8907 NT7<br>(High Density Polyethylene) | Polypropylene<br>Polycarbonate |
| Single-Use | Yes | Yes |
| Sterility | Provided Sterile<br>Gamma Irradiation<br>SAL 10-6 | Provided Sterile<br>Gamma Irradiation<br>SAL 10-6 |
## Performance Data:
Testing was conducted for the O-GENESIS Graft Delivery System to confirm the device performs as intended. Testing was performed using test units representative of the finished devices. Testing included simulated use, functional verification, and biocompatibility evaluation.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Orthofix. The logo consists of a blue abstract shape on the left and the word "ORTHOFIX" in black letters on the right. The abstract shape is a stylized representation of a circle or sphere, with three curved lines that converge at a point. The word "ORTHOFIX" is written in a bold, sans-serif font.
Page 3/3
Simulated use testing included loading of bone graft material into the cannula and delivery of bone graft material from the cannula using the graft delivery gun. Testing was successful with all acceptance criteria met.
Functional verification testing based on design requirements and risk analysis was performed and included:
- Delivery Gun Ratchet Plunger force
- Plunger T-Handle strength
- Syringe Body/Stopper force - Threaded Components interface
-
Functional verification testing was successful with all acceptance criteria met.
A biocompatibility evaluation was performed in accordance with ISO 10993-1:2009, Biological Evaluation of Medical Devices - Part 1, and included testing for the following biocompatibility endpoints; Cytotoxicity, Sensitization, Acute Systemic Toxicity and Material-Mediated Pyrogenicity. Based on the evaluation and testing results the O-GENESIS Graft Delivery System is biocompatible and meets the requirements of ISO 10993-1:2009.
## Conclusion:
Based upon similarities in intended use, indications for use, and fundamental technology, including basic design, sterility and operating principle, the O-GENESIS Graft Delivery System is substantially equivalent to the predicate device.