POWRSYRINGE MONITOR

K072345 · Pinyons Medical Technology, Inc. · FMF · Sep 21, 2007 · General Hospital

Device Facts

Record IDK072345
Device NamePOWRSYRINGE MONITOR
ApplicantPinyons Medical Technology, Inc.
Product CodeFMF · General Hospital
Decision DateSep 21, 2007
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5860
Device ClassClass 2
AttributesTherapeutic, 3rd-Party Reviewed

Indications for Use

The PowrSyringe Monitor is a piston syringe to inject fluids and monitor the pressure of that fluid including use in inflation and discography.

Device Story

PowrSyringe Monitor is a single-use, manual, hand-held piston syringe equipped with handles and a pressure gauge. Device enables clinicians to inject fluids and monitor pressure during inflation and discography procedures. Operation involves squeezing handles to advance the plunger into the barrel for injection or opening handles to aspirate fluid. Design features include a clear barrel for air bubble visualization, minimum dead space, and a handle mechanism to prevent plunger detachment during aspiration. Device provides real-time pressure monitoring via the integrated gauge, assisting clinicians in controlling fluid delivery during medical procedures.

Clinical Evidence

No clinical data. Bench and animal performance testing conducted to confirm functionality and substantial equivalence.

Technological Characteristics

Manual hand-held piston syringe; includes integrated pressure gauge; clear barrel for visualization; handle-actuated plunger mechanism; single-use; non-powered.

Indications for Use

Indicated for use as a piston syringe to inject fluids and monitor fluid pressure during procedures including inflation and discography.

Regulatory Classification

Identification

A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.

Submission Summary (Full Text)

{0}------------------------------------------------ # 5. 510(k) Summary In accordance with the Safe Medical Devices Act (SMDA) of 1990 and Title of the Code of Federal Regulations Part 807 (21 CFR \$807), and in particular \$807.92, the following summary of safety and effectiveness information is provided. ### Device Information | Trade Name: | PowrSyringe Monitor | |----------------------|---------------------| | Common Name: | Piston Syringe | | Classification Name: | FMF Syringe, Piston | #### Predicate Devices The PowrSyringe Monitor is substantially equivalent to multiple previous cleared piston. inflation, and pressure monitoring syringes. #### Device Description The PowrSyringe Monitor is a single use manual hand-held piston syringe with handles and a pressure gauge to inject fluids and monitor the pressure of that fluid including use in inflation and discography. The PowrSyringe Monitor handles allow the user to push the plunger into the barrel when the user squeezes the handles. Users may open the PowrSyringe Monitor's handles to aspirate fluid back into the barrel. The PowrSyringe Monitor safety features include: - 제 Clear barrel for visualization of air bubbles. - 트 Minimum dead space between the plunger and barrel with the handles are fully squeezed. - 다 Handle design to prevent the plunger from being pulled out of the barrel during aspiration. # Intended Use The PowrSyringe Monitor is a piston syringe to inject fluids and monitor the pressure of that fluid including use in inflation and discography. # Comparison to Predicate Devices Data is provided to demonstrate the PowrSyringe Monitor is substantially equivalent to previous cleared devices and does not introduce any new safety risks. Substantial equivalence is based on equivalence in indications for use, intended use, patient contact, materials, design, function, performance, sterilization, and safety. #### Non-Clinical Testing Bench and animal performance testing to confirm functionality in the intended use and equivalence to predicate device is included. # Clinical testing Not applicable. {1}------------------------------------------------ Image /page/1/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines forming its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. SEP 2 1 2007 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Pinyons Medical Technology, Incorporated C/O Mr. Mark Job Responsible Third Party Official Regulatory Technology Services, LLC 1394 25th Street North West Buffalo, Minnesota 55313 Re: K072345 Trade/Device Name: PowrSyringe Monitor Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: II Product Code: FMF Dated: September 5, 2007 Received: September 6, 2007 Dear Mr. Job: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {2}------------------------------------------------ Page 2 - Mr. Job Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market, If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Satic Y. Michael Dmd. Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # 4. Indication for Use Statement 510(k) Number: \\$\phi72345 Device Name: PowrSyringe Monitor Indications for Use: The PowrSyringe Monitor is a piston syringe to inject fluids and monitor the pressure of that fluid including use in inflation and discography. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) Anthony D. m (Division of (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices 510(k) Number: _ K473345
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...