BASIXCONPAK ANALOG INFLATION SYRINGE

K122321 · Merit Medical Ireland, Ltd. · MAV · Dec 19, 2012 · Cardiovascular

Device Facts

Record IDK122321
Device NameBASIXCONPAK ANALOG INFLATION SYRINGE
ApplicantMerit Medical Ireland, Ltd.
Product CodeMAV · Cardiovascular
Decision DateDec 19, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1650
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Merit analog inflation syringe is used to inflate and deflate an angioplasty balloon or other interventional device, and to measure the pressure within the balloon

Device Story

Single-use disposable inflation syringe; generates and monitors pressure for angioplasty or interventional devices. Features threaded plunger assembly with lock/release bar, flexible high-pressure extension tube, and three-way stopcock. Includes analog pressure gauge for real-time monitoring. Operated by clinicians in interventional settings. Provides manual control of balloon inflation/deflation; gauge allows visual pressure verification to ensure appropriate device deployment. Benefits include precise pressure management during interventional procedures.

Clinical Evidence

Bench testing only. No clinical data. Testing included functional (fluid/air), torqueability, tensile, thermal shock/leak, transportation simulation, gauge responsiveness, volumetric comparison, greening effect, vacuum capability, and gauge accuracy/precision tests. All tests met acceptance criteria.

Technological Characteristics

Single-use disposable syringe with threaded plunger, lock/release bar, high-pressure extension tube, and 3-way stopcock. Analog pressure gauge. Materials evaluated per ISO 10993-1. Sterilization via ethylene oxide (ISO 11135-1). Luer fittings per ISO 594-1/594-2. No software or electronic components.

Indications for Use

Indicated for inflation and deflation of angioplasty balloons or other interventional devices and measurement of balloon pressure.

Regulatory Classification

Special Controls

*Classification.* Class II (special controls). The device, when it is a non-patient contacting balloon inflation syringe intended only to inflate/deflate balloon catheters and monitor pressure within the balloon, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K122321 (pg. 1 of 2 Merit Medical Systems, Inc. Analog Inflation Syringe Premarket Notification 510(k) Section 5 510(k) Summary ## DEC 1 9 2012 ಿ ## Section 5 ## 510(k) Summary | General<br>Provisions | Submitter Name:<br>Address:<br>Telephone Number:<br>Fax Number:<br>Contact Person:<br>Registration Number: | Merit Medical Systems, Inc.<br>1600 West Merit Parkway,<br>South Jordan, UT 84095<br>(801) 208-4349<br>(801) 253-6967<br>Stephanie Erskine<br>1721504 | |-----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | Correspondent Name:<br>Address:<br>Telephone Number:<br>Fax Number:<br>Contact Person:<br>Date of Preparation:<br>Registration Number: | Merit Medical Ireland Ltd.<br>Parkmore Business Park,<br>Galway, Ireland<br>(353) 91 703 761<br>(353) 91 771 888<br>Mark Mullaney<br>December 19th, 2012<br>9616662 | | Subject<br>Device | Trade Name:<br>Common/Usual Name:<br>Classification Name: | BasixCOMPAK<br>Inflation Syringe<br>74 MAV Balloon Inflation Syringe | | Predicate<br>Device | Trade Name:<br>Classification Name:<br>Premarket Notification:<br>Manufacturer: | Merit Monarch COMPAK Inflation Syringe &<br>Universal Fluid Dispensing<br>74 MAV Balloon Inflation Syringe<br>K083523<br>Merit Medical Systems, Inc. | | Classification | Class II<br>21 CFR § 870.1650, 74 MAV<br>Division of Cardiovascular Devices | | | Intended Use | The Merit analog inflation syringe is used to inflate and deflate an<br>angioplasty balloon or other interventional device, and to measure the<br>pressure within the balloon | | | Device<br>Description | The Merit analog inflation syringe is a single use disposable device capable<br>of generating and monitoring pressure in angioplasty or other similar<br>interventional devices. It is fitted with a threaded plunger assembly with<br>lock/release bar, flexible high pressure extension tube. It is provided with a<br>510(k)-cleared three way medium pressure stopcock. | | | | Comparison to<br>Predicate | The Technological characteristics of the subject Merit analog inflation<br>device are substantially equivalent to those of the predicate, the Merit<br>Monarch COMPAK Inflation Syringe & Universal Fluid Dispensing Device.<br>The significant difference between the devices relate to the pressure gauge<br>display. The Merit analog inflation device has an analog gauge display. The<br>predicate device has a digital display. | | | Safety &<br>Performance<br>Tests | No applicable mandatory performance standards or special controls have<br>been established under Section 514 of the Food, Drug and Cosmetic Act for<br>this device. However, a battery of tests was performed according to<br>protocols based on the requirements of industry standards and guidance<br>and the device met the acceptance criteria necessary to demonstrate the<br>safety and efficacy of the device. | | | | Where appropriate, the tests were based on the requirements of the<br>following documents:<br>ISO 11135-1:2007 Sterilization of health care products - Ethylene<br>oxide - Part 1: Requirements for the development, validation, and<br>routine control of a sterilization process for medical devices. ISO 10993-7:2008 - Biological Evaluation of Medical Devices - Part<br>7: Ethylene Oxide Sterilization Residuals ASTM F1980-07 Standard Guide for Accelerated Aging of Sterile<br>Barrier Systems for Medical Devices. ASTM D4169 - 09 Standard Practice for Performance Testing of<br>Shipping Containers and Systems ISO 10993-1: 2009 Biological Evaluation of Medical Devices Part 1:<br>Evaluation and Testing within a risk management process, and the<br>FDA Modified ISO 10993 Test Profile FDA Memo G95-1. ISO 594-1:1996 Conical Fittings with a 6% (Luer) taper for syringes,<br>needles and certain other medical equipment - general<br>requirements ISO 594-2:1998 Conical fittings with 6% (Luer) taper for syringes,<br>needles and certain other medical equipment - Part 2: Lock fittings | | | Safety &<br>Performance<br>Tests | The following is a list of all significant testing that has been successfully<br>completed for this device:<br>Functional Test (Fluid and Air) Torqueability Test Tensile Test Thermal Shock - Leak Test Transportation Simulation Gauge Responsiveness Test Volumetric Comparison Test Greening Effect Vacuum Capability Test Gauge Accuracy and Precision | | | Summary of<br>Substantial<br>Equivalence | Based on the indications for use, design, safety, and performance testing,<br>the subject Merit analog inflation device meets the requirements that are<br>considered essential for its intended use and is substantially equivalent to<br>the predicate device, the Merit Monarch Inflation Syringe (K083523)<br>manufactured by Merit Medical Systems, Inc. Differences between the<br>devices do not raise any different questions of safety or effectiveness. | {1}------------------------------------------------ K122321 (pg. 282) Merit Medical Systems, Inc. Analog Inflation Syringe Premarket Notification 510(k) Section 5 510(k) Summary {2}------------------------------------------------ ## DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized human figure with three arms reaching upwards, symbolizing growth and development. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the figure. #### Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 Merit Medical Systems, Inc. c/o Stephanie Erskine 1600 West Merit Parkway South Jordan, UT 84095 DEC 1 9 2012 Re: K122321 Trade/Device Name: BasixCOMPAK Regulation Number: 21 CFR 870.1650 Regulation Name: Angiographic Injector and Syringe Regulatory Class: Class II Product Code: MAV Dated: November 29, 2012 Received: November 30, 2012 Dear Ms. Erskine: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {3}------------------------------------------------ forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely vours. # Matthew G. Hillebrenner for Bram Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Merit Medical Systems, Inc. Analog Inflation Syringe Premarket Notification 510(k) ### Section 4 ## Indications for Use Statement 510(k) Number (if known): K12232 Device Name: BasixCOMPAK Indications for Use: The Merit analog inflation syringe is used to inflate and deflate an angioplasty balloon or other interventional device, and to measure the pressure within the balloon. × Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE—CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) M.Y. Hillel n Sign-C (Division Sign-Oh) Division of Cardiovascular Devices 510(k) Numbe
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...