BD INTEGRA SYRINGE

K023752 · Bd · FMF · Feb 7, 2003 · General Hospital

Device Facts

Record IDK023752
Device NameBD INTEGRA SYRINGE
ApplicantBd
Product CodeFMF · General Hospital
Decision DateFeb 7, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5860
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The BD Integra™ Syringe is used for aspiration of fluids from vials and ampoules and a variety of fluid injections below the surface of the skin. It is not intended to be used for phlebotomy. The insulin syringe has scale lines in insulin units and is used for insulin injections. The tuberculin syringe can be used for any of the 3 types of common injections (intra-dermal, intra-muscular or subcutaneous). The BD Integra™ 1ml Syringe has a permanently attached needle. The BD Integra™ 1ml Syringe contains a tool used to cut through the hub and stopper allowing the needle to become retracted inside the plunger rod of the syringe after use. After activation the needle is fully contained inside the syringe guarding against accidental needle sticks during normal handling and disposal of the used needle/syringe combination.

Device Story

Single-use 1ml syringe with integrated needle-retraction safety mechanism; needle sizes 30G-25G. Device features internal tool to cut hub/stopper, allowing needle retraction into plunger rod post-injection. Operated by clinicians or patients for fluid aspiration and injection. Retraction mechanism guards against accidental needle sticks during disposal. Performance equivalent to standard syringes regarding aspiration, injection technique, and scale readability.

Clinical Evidence

No clinical trials performed. Evidence based on bench testing and a simulated use study comparing the device to conventional insulin/tuberculin syringes. Evaluated performance metrics included ease of injection, aseptic technique, aspiration ability, scale readability, and safety/disposal handling.

Technological Characteristics

Single-use piston syringe with integrated needle retraction mechanism. Permanently attached needle (30G-25G). Mechanical retraction via internal cutting tool. No electronic components or software.

Indications for Use

Indicated for aspiration of fluids from vials/ampoules and fluid injections below the skin surface (intra-dermal, intra-muscular, or subcutaneous). Insulin syringe variant indicated for insulin injections. Not for phlebotomy.

Regulatory Classification

Identification

A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ 1 Becton Drive Franklin Lakes, New Jersey 07417 tel: 201.847.6800 www.bd.com # Attachment 5 K023752 ## REVISED Summary of Safety & Effectiveness for the BD Integra™ 1ml Syringe ### 1. BD Contact Person: Pasquale Amato Regulatory Affairs Coordinator BD Medical Surgical BD 1 Becton Drive MC226 Franklin Lakes, NJ Phone: 201-847-4513 Fax: 201-847-4855 E-mail: pasquale amato@bd.com FEB 07 2003 - 2. Device Name: BD Integra™ 1ml Syringe - 3. Predicate Devices: BD Insulin Syringe, Latex Free K980580 BD Spring Based Syringe K011103 RTI VanishPoint® Syringe K980069 #### 4. Product Description/Function: 4.1 Description: Single-use syringe with a needle protection system. Needle sizes: 30G-25G, Syringe sizes: 1ml. - 5. Intended uses: The BD Integra™ Syringe is used for aspiration of fluids from vials and ampoules and a variety of fluid injections below the surface of the skin. It is not intended to be used for phlebotomy. The insulin syringe has scale lines in insulin units and is used for insulin injections. The tuberculin syringe can be used for any of the 3 types of common injections (intra-dermal, intra-muscular or subcutaneous). The BD Integra™ Iml Syringe has a permanently attached needle. The BD Integra™ Iml Syringe contains a tool used to cut through the hub and stopper allowing the needle to become retracted inside the plunger rod of the syringe after use. After activation the needle is fully contained inside the syringe guarding against accidental needle sticks during normal handling and disposal of the used needle/syringe combination. {1}------------------------------------------------ ### 6. Comparison of Modified & Predicate Devices: 6.1 Descriptive Comparison to a Legally Marketed Device The BD Integra™ Syringes can be used in the same fashion as standard BD Insulin/Tuberculin syringes. When the BD Integra™ Syringe is in the "ready-to-use" position it provides the same usable injection needle length and same barrel scale markings as a standard insulin / tuberculin syringes. Comparison has been made to the BD insulin/tuberculin Syringe. The BD Integra™ Syringe is used for aspiration of fluids from vials and ampoules and a variety of fluid injections below the surface of the skin. It is not intended to be used for phlebotomy. The BD Integra™ Syringe contains a tool used to cut through the stopper and hub allowing the needle to become retracted inside the plunger rod of the syringe after use. After activation the needle is fully contained inside the syringe guarding against accidental needle sticks during normal handling and disposal of the used needle/syringe combination. 6.2 Material Changes: There are no significant changes in the materials used for the BD Integra™ Syringe 6.3 Manufacturing Process Changes: The manufacturing processes are being modified to assemble the components associated with the BD Integra™ Syringe. 6.4 Manufacturing Site Changes: No new manufacturing site is being utilized. 6.5 Packaging Component Changes: The package design is being modified to accept the size of the BD Integra™ Syringe. ### 7. Equivalence determination: The BD Integra™ Syringe was compared to the predicate devices using the following criteria: shield removal, cannula removal, stopper breakout and sustaining force, hub and stopper leakage, dead space, activation forces, hub/barrel separation forces, cannula containment and splatter. The BD Integra™ Syringes performed in a similar manner to the predicate devices. A Simulated Use Study was performed on the BD Integra™ 1ml Syringe. The BD Integra™ 1 ml Syringe performed equivalently to the BD conventional insulin and tuberculin syringes with respect to the following characteristics: Overall performance - . Ease of maintaining standard injection technique - Ability to maintain aseptic technique ● - Ability to aspirate medication from a vial ● - Ability to read the syringe scale on the barrel - Ability to passively re-shield and transport - Ability to easily and safely dispose of the used device in a sharps container - Perceived safety of the device . {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of a bird or eagle with three curved lines representing its wings or feathers. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the bird symbol. The logo is black and white. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 ## FEB 0 7 2003 BD Mr. Pasquale Amato Regulatory Affairs Coordinator BD Medical Surgical 1 Becton Drive MC226 Franklin Lakes, New Jersey 07417 Re: K023752 Trade/Device Name: BD Integra™ 1 ml Syringe Regulation Number: 880.5860 Regulation Name: Piston Syringe Regulatory Class: II Product Code: MEG, FMF Dated: January 16, 2003 Received: January 17, 2003 Dear Ms. Amato: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting vour device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Page 2 - Mr. Amato Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4618. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Susan O'Kunner Susan Runner, DDS, MA Interim Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ #### Indications for Use Statement 510(k) Number (if known): K023752 Device Name: BD Integra™ 1 ml Syringe Indications for Use: The BD Integra™ Syringe is used for aspiration of fluids from vials and ampoules and a variety of fluid injections below the surface of the skin. It is not intended to be used for phlebotomy. The insulin syringe has scale lines in insulin units and is used for insulin injections. The tuberculin syringe can be used for any of the 3 types of common injections (intra-dermal, intra-muscular or subcutaneous). The BD Integra™ 1ml Syringe has a permanently attached needle. The BD Integra™ 1ml Syringe contains a tool used to cut through the hub and stopper allowing the needle to become retracted inside the plunger rod of the syringe after use. After activation the needle is fully contained inside the syringe guarding against accidental needle sticks during normal handling and disposal of the used needle/syringe combination. # (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) | Prescription Use | |----------------------| | (Per 21 CFR 801.109) | Over-The-Counter Use Patricia Lucente (Division Sign-Off) OR sthesiology, General Hospital 510(k) Number: K023752 K023752
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...