K152606 · Jeesung Medical Co., Ltd. · MEG · Jun 28, 2016 · General Hospital
Device Facts
Record ID
K152606
Device Name
Jeesung Safety Syringe and Single Use Needles
Applicant
Jeesung Medical Co., Ltd.
Product Code
MEG · General Hospital
Decision Date
Jun 28, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Jeesung safety syringe and single use needles is a sterile, single-use, disposable and non-reusable, manual retractable safety syringe intended for injection of fluids into the body, while reducing the risk of sharps injuries and the potential for syringe reuse.
Device Story
Manual retractable safety syringe; 3cc capacity; 18-30G needles. Operates via internal barrel swell fixing needle hub; four legs on hub catch on barrel swell during retraction. Post-injection, user pulls plunger back; hub binds to plunger top; pushing plunger until audible click breaks hub off, retracting needle into barrel, rendering it unusable. Used in clinical settings by healthcare professionals. Prevents accidental needle sticks; ensures single-use compliance.
Clinical Evidence
Bench testing only. Biocompatibility testing (ISO 10993) included cytotoxicity, hemolysis, acute systemic toxicity, skin sensitization, pyrogenicity, and intracutaneous reactivity. Performance testing included ISO 7886-1 (syringe) and ISO 7864 (needle) compliance. Simulated clinical use study performed on 500 units; zero failures of the safety/protective feature reported.
Technological Characteristics
Materials: Polypropylene (barrel, plunger, sheath), Elastomer (piston/gasket), Polycarbonate (needle hub), Stainless steel (needle). Lubricant: Silicone oil. Energy source: Manual. Connectivity: None. Sterilization: E.O. gas (SAL 10^-6). Form factor: 3cc syringe with integrated manual retraction mechanism.
Indications for Use
Indicated for injection of fluids into the body for patients requiring parenteral administration, while reducing sharps injury risk and syringe reuse.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the circumference. Inside the circle is a stylized design featuring a series of overlapping human profiles, creating a sense of unity and connection.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 28, 2016
JeeSung Medical Co., Ltd. % Mr. Peter Chung Plus Global 300 Atwood Street, Pittsburgh, Pennsylvania 15213
Re: K152606
Trade/Device Name: Jeesung Safety Syringe and Single Use Needles Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: II Product Code: MEG Dated: May 9, 2016 Received: May 18, 2016
Dear Mr. Chung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Image /page/1/Picture/8 description: The image shows a signature followed by the name "Tina Kiang". Below the signature and name is the letter "S" preceded by a hyphen. The signature is illegible and appears to be a stylized scribble. The text is in a simple, sans-serif font.
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K152606
Device Name
Jeesung Safety Syringe and Single Use Needles
Indications for Use (Describe)
Jeesung safety syringe and single use needles is a sterile, single-use, disposable and non-reusable, manual retractable safety syringe intended for injection of fluids into the body, while reducing the risk of sharps injuries and the potential for syringe reuse.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K152606
# 510(k) Summary
[as required by 807.92]
### 1. Applicant
- 1) Company: JeeSung Medical Co.,Ltd.
- 2) Address: 54, Mujini 1-gil, Daedeok-gu, Daejeon, Korea
- 3) Tel: 82-42-932-6061
- 4) Fax: 82-42-932-6063
- 5) Contact Person: Peter Chung,
- 6) Tel: 412-687-3976
- 7) Email: peterchung210@gmail.com
- 8) Date of Preparation: June 22, 2016
#### 2. Device Information
- 1) Trade Name: Jeesung Safety Syringe and Single Use Needles
- 2) Common Name: Syringe, Antistick Piston Syringe
- 3) Classification Name: Piston Syringe
- 4) Product Code: MEG
- 5) Regulation Number: 880.5860
- 6) Class of device: Class II
- 7) Panel: General Hospital
#### 3. Predicate Device
K142435 3S Safety Syringe with/without Needle
#### 4. Device Description
Jeesung Safety Syringe and Single Use Needles is an integrated needle and piston syringe with an anti-needle-stick mechanism. There is a swell on the top of inside barrel, which can be used to fix the hub of needle to the top of inside barrel. Four legs on the bottom of hub are caught on the swell part on the top of inside barrel when the hub is pulled.
After using this syringe (such as injecting medicine into body etc.), the hub of needle is pulled back to the inside of the barrel. Because the four legs of hub is bound to the top of the plunger which has a smaller swell part than the top of inside barrel for being caught. Therefore, by pushing plunger until it makes a binding sound, the hub can follow with the plunger is broken off and the needle cannot be come out of the barrel. This renders the needle unusable and safe from accidental needle sticks.
Jeesung Safety Syringe and Single Use Needles has syringe capacity of 3cc attached with 18-30 gauge and the length of the needle (8mm, 10mm, 13mm, 19mm, 25mm, 32mm, 38mm).
{4}------------------------------------------------
#### 5. Indication for Use
Jeesung safety syringe and single use needles is a sterile, single-use, disposable and non-reusable, manual retractable safety syringe intended for injection of fluids into the body, while reducing the risk of sharps injuries and the potential for syringe reuse.
| Comparison table | | | |
|-------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------|
| Category | JeeSung Medical Co.,Ltd | Sincere Medical Device Co., Ltd. | |
| Element of<br>Comparison | Submission Device<br>(K152606) | Predicate Device<br>K142435 | |
| Regulation no. | 880.5860 | 880.5860 | |
| Product code | MEG | MEG | |
| Common name | Syringe, Antistick Piston Syringe | Syringe, Antistick Piston Syringe | |
| Class | Class II | Class II | |
| Indication for use | Jeesung safety syringe and single use<br>needles is a sterile, single-use,<br>disposable and non-reusable, manual<br>retractable safety syringe intended for<br>injection of fluids into the body, while<br>reducing the risk of sharps injuries and<br>the potential for syringe reuse. | The 3S Safety Syringe with/without<br>Needle is a sterile, single-use,<br>disposable and non-reusable, manual<br>retractable safety syringe intended<br>for injection of fluids into the body,<br>while reducing the risk of sharps<br>injuries and the potential for syringe<br>reuse. | |
| Principle of Operation | There is a swell on the top of inside<br>barrel, which can be used to fix the hub<br>of needle to the top of inside barrel.<br>Four legs on the bottom of hub are<br>caught on the swell part on the top of<br>inside barrel when the hub is pulled.<br>After using this syringe (such as<br>injecting medicine into body etc.), the<br>hub of needle is pulled back to the<br>inside of the barrel. Because the four<br>legs of hub is bound to the top of the<br>plunger which has a smaller swell part<br>than the top of inside barrel for being<br>caught. Therefore, by pushing plunger<br>until it makes a binding sound, the hub<br>can follow with the plunger. Then the<br>plunger is broken off and the needle<br>cannot be come out of the barrel. This<br>renders the needle unusable and safe<br>from accidental needle sticks. | The needle is contained within the<br>syringe barrel. After standard<br>techniques for injection, the plunger<br>is withdrawn completely into the<br>barrel and snapped off and the<br>needle is contained within the<br>puncture-resistant barrel. This<br>renders the needle unusable and safe<br>from accidental needle sticks. | |
| Syringe capacity | 3cc | Various Sizes<br>(smallest 0.5cc, largest 10cc) | |
| Lubricant for Barrel | Silicone Oil | Silicone Oil | |
| Barrel transparency | Transparent and Clear | Transparent and Clear | |
| Gradation Legibility | Legible | Legible | |
| Product configuration | Barrel<br>Plunger<br>Gasket<br>Needle Hub | Barrel<br>Plunger<br>Piston<br>Needle Hub | |
| Material | | Needle Protect cap | Needle Sheath |
| | Barrel | Polypropylene | Polypropylene |
| | Plunger | Polypropylene | Polypropylene |
| | Piston(Gasket) | Elastomer | Elastomer |
| | Needle hub | Polycarbonate | Polypropylene |
| | Needle | Stainless steel | Stainless Steel |
| | Needle sheath<br>(protect cap) | Polypropylene | Polypropylene |
| | Needle Gauge and<br>Length | Needle gauge (18, 19, 20, 21, 22, 23,<br>24, 25, 26, 27, 28, 29, 30G)<br>Needle length (8, 10, 13, 16, 19, 25, 32,<br>38mm) | Needle gauge (18, 19, 20, 21, 22, 23,<br>24, 25, 27, 28, 29, 30, 31G)<br>Needle length (8, 9, 11, 13, 16, 25,<br>32, 38mm) |
| | | | |
| Lubricant for Needle | | Silicone Oil | Silicone Oil |
| Sharp Injury<br>Prevention Features | | Manual Retractable | Manual Retractable |
| Performances | | Conforms to ISO7864 ISO7886 | Conforms to ISO7864 ISO7886 |
| Biocompatibility test | | Conforms to ISO10993<br>(ISO10993-4, ISO10993-5, ISO10993-10,<br>ISO10993-11) | Conforms to ISO10993 |
| Labeling | | Meet the requirements of 21 CFR Part<br>801 | Meet the requirements of 21 CFR<br>Part 801 |
| Sterilization information | | E.O gas sterilization<br>Assurance level : 10-6 | E.O gas sterilization<br>Assurance level : 10-6 |
#### 6. Predicate Device Comparison Table
{5}------------------------------------------------
The proposed and predicate devices are the same indications for use, principle of operation, Lubricant for Barrel, product configuration, lubricant for needle, sharp injury prevention feature. The proposed device syringe capacity includes 3cc which are within the range of the predicate syringe 0.5cc to 10cc. The subject device needle sizes are 18-30G and needle length 8, 10, 13, 16, 19, 25, 32, 38mm which are within the range of predicate needle gauge 18-31G and needle length 8-38mm. Therefore, these differences do not affect substantially equivalency between the proposed devices and predicate device.
#### 7. Performance Data
| No | Test item | Test method | |
|----|-----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------|
| | | Test criteria | Test result (pass / failure) |
| 1 | Cytotoxicity test | ISO10993-5:2009 Biological evaluation of medical devices -- Part 5: Tests for in vitro<br>cytotoxicity, Test on extracts method<br>Non-cytotoxicity | Non- cytotoxicity (Pass) |
| 2 | Hemolysis test | ISO10993-4: 2002(E) Biological evaluation of medical devices - Part 4: Selection for tests<br>for interactions with blood, Evaluation of hemolytic properties of medical and their<br>components<br>Non-hemolytic<br>ISO10993-4: 2002/Amd.1 : 2006 Biological evaluation of medical devices – Part 4:<br>Selection of tests for interactions with blood AMENDMENT 1 Annex C.6 Hemolysis<br>testing General considerations<br>Refer to test report | Non-hemolytic (Pass)<br>There was non-hemolytic. |
| 3 | Acute<br>systematic<br>toxicity<br>test | ISO10993-11: 2006(E) Biological evaluation of medical devices -- Part 11: Tests for<br>systemic toxicity, 5. Acute systemic toxicity<br>No dead animals<br>No strange reaction animals | Non-acute systemic toxicity (Pass) |
{6}------------------------------------------------
| 4 | Skin<br>sensitization<br>test | ISO10993-10: 2010(E), Biological evaluation of medical devices Part 10: Teat for Irritation<br>and sensitization. Guinea pig Maximization test (GPMT) | |
|---|-----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | | Non-skin hypersensitivity<br>reaction | There was non-skin hypersensitivity reaction. (Pass) |
| 5 | Pyrogen Test | ISO 10993-11 :2006, Biological evaluation of medical devices, Test for systemic, Annex(F)<br>Information on material-mediated pyrogens | |
| | | Non-pyrogenicity | Non-pyrogenicity (Pass) |
| | | USP 38 NF 33 <85> Bacterial Endotoxin test "Gel clot method" | |
| | | Concentration of the reaction<br>solution is determined as (-) | As a result of the reaction, all the concentration of<br>the test solution were determined as (-). Therefore,<br>endotoxin concentration of the test sample is less<br>than 4.0 EU/device. |
| 6 | Intracutaneous<br>reactivity test | ISO 10993-10 :2010(E), Biological evaluation of medical devices, Test for irritation and<br>skin sensitization, Animal Intracutaneous(intradermal) reactivity Test. | |
| | | Non-intracutaneous reactivity | There was non-intracutaneous reactivity. (Pass) |
| 7 | Particulate<br>matters | USP <788> Particulate Matter in Injections | |
| | | Particles does not greater than<br>10μm and greater than 25μm | Particles greater than 10μm - 0 particle<br>Particles greater than 25μm - 0 particle |
Based on the above testing results, the subject devices are biocompatible.
| | | Test items | Result summary |
|---|----------------------------------------|---------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1 | Syringe<br>(ISO7886-<br>1 : 1993) | Inner/outside and structure | Surface is smooth / no particle and foreign substance |
| | | Graduated scale | Dead space (0.06ml), capacity (2.19ml), Graduation lines were evenly spaces, overall length of scale (32.0mm) |
| | | Barrel | The length of the barrel is such that the syringe has a maximum usable capacity of at least 10% more than the nominal capacity |
| | | Piston(gasket)/plunger assembly | Design, fit of piston in barrel, fiducial line are suitable |
| | | Liquid leakage | Pressure test – no leaked water / aspirate test – no leaked air |
| | | Lubricant | The quantity of lubricant :0.18mg/cm³ |
| 2 | Needle<br>(ISO7864 :<br>1993) | Inner/outside and structure | No scratch, no crack, smooth, no foreign materials, sharped needle edge |
| | | Measurement | Outside diameter of needle : 0.645mm for 0.6000-0.673 criteria / length of needle : 24.7mm for 25(+1.5/-2.5) criteria |
| | | Elasticity test | Needle returned after remove weight. |
| | | Flexural rigidity | Needle was not broken. |
| | | Draw test | The union of the hub and needle tube was not broken. |
| 3 | Stimulated Clinical Use Testing report | | As a result of a simulated clinical use study using 500 subject devices, the number of failure of the subject devices safety feature is not founded and resulted to zero failure of the protective feature. |
Based on the result of the above syringe, needle testing and stimulated clinical use testing, the proposed devices are satisfied with all the acceptance criteria.
{7}------------------------------------------------
### 8. Conclusion
Both subject and predicate devices have the same Indications for Use. Based on the result of the device comparison and performance data, the proposed devices are determined to be Substantially Equivalent (SE) to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.