INOVEL HEALTH CARE N95 PARTICULATE RESPIRATORS & SURGICAL MASKS

K061859 · Inovel, LLC · MSH · Jul 12, 2006 · General, Plastic Surgery

Device Facts

Record IDK061859
Device NameINOVEL HEALTH CARE N95 PARTICULATE RESPIRATORS & SURGICAL MASKS
ApplicantInovel, LLC
Product CodeMSH · General, Plastic Surgery
Decision DateJul 12, 2006
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 878.4040
Device ClassClass 2
Attributes3rd-Party Reviewed

Intended Use

The various models of Inovel Type N95 Healthcare Particulate Respirators and Surgical Masks meet CDC Guidelines for TB Exposure Control within healthcare facilities. These devices are also intended to be worn by healthcare personnel during surgical procedures to protect both the patient and healthcare personnel from the transfer of microorganisms, body fluids and particulate material.

Device Story

Molded cup-style N95 particulate respirator and surgical mask; constructed from extruded plastic mesh outer cover, nonwoven spunbond inner/outer layers, and polypropylene melt-blown filter media; includes non-latex rubber head straps and closed-cell polyethylene foam nosepiece. Used by healthcare personnel in clinical/surgical settings. Provides barrier protection against microorganisms, body fluids, and particulates; meets NIOSH N95 certification (TC-84A-0013) for 95% filtration efficiency against 0.055-0.095 micron particles. Device is disposable; intended for single use.

Clinical Evidence

Bench testing only. Performance validated via NIOSH certification (42 CFR Part 84) for filtration efficiency (>95% average 98.58%) and breathing resistance. Fluid resistance tested per ASTM F 1862 (32/32 pass for most models). Bacterial filtration efficiency >99.9% per Modified Greene and Vesley Method. Biocompatibility (cytotoxicity, sensitization, irritation) verified per ISO 10993 standards.

Technological Characteristics

Molded cup design; materials: spunbond/meltblown polypropylene, extruded plastic mesh, polyethylene foam nosepiece, non-latex rubber head straps. NIOSH-certified N95 (TC-84A-0013). Filtration: >95% efficiency for 0.055-0.095 micron particles. Fluid resistance: ASTM F 1862. Flammability: 16 CFR 1610 Class I. Biocompatibility: ISO 10993-1 compliant.

Indications for Use

Indicated for healthcare personnel during surgical procedures and for TB exposure control in healthcare facilities to protect patients and personnel from microorganisms, body fluids, and particulates.

Regulatory Classification

Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible. (ii) Analysis and nonclinical testing must: (A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and (B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device. (iii) NIOSH approved under its regulation. (2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # K061859 JUL 1 2 2006 June 27, 2006 12 - I ### 510(k) Summary ### Submitter: lnovel LLC 10111 W. Jefferson Blvd. Culver City, CA 90232-3509 ### Contact: William Wawrzyniak Director Quality Assurance Telephone: 866-546-6835 Ext. 705 Fax: 310-837-0468 E-mail: billw@inovelmedical.com ### Trade Name: Inovel Health Care N95 Particulate Respirators and Surgical Masks, various models. ### Common Name: Health Care N95 Particulate Respirators and Surgical Masks. ### Classification: Name - Surgical Apparel, as described in 21 CFR 878.4040. Device Class - Class II Product Code - MSH CFR Section – 21 CFR 878.4040 ### Substantial Equivalency: Inovel Health Care N95 Particulate Respirators and Surgical Masks are found to be substantially equivalent to Inovel Health Care N95 Particulate Respirators model 3001N95-S, 3002N95-M, 3003N95-L and 3004N95-LP [(510(K) K051182] These products have also been tested and approved by NIOSH as N95 Respirators. {1}------------------------------------------------ ### 510(k) Summary (Continued) Description: The Inovel type N95 Healthcare Particulate Respirators and Surgical Masks are constructed from an extruded plastic mesh used in the outer cover and a nonwoven spunbond used in the inner and outer cover. The polypropylene melt blown filter media is layered between the inner and outer cover. The head strap is made of a non-latex rubber stapled to the mask. The inside nosepiece is a closed cell foam. The Inovel type N95 Healthcare Particulate Respirators and Surgical Masks are approved by NIOSH in accordance with 42 CFR Part 84. The certification numbers is TC-84A-0013. The type N95 must meet the prescribed test criteria which specifies the use of 0.055 to 0.095 micron diameter challenge and requiring a 95% efficiency or better. The masks are resistant to synthetic blood as per ASTM F 1862 Standard Test method for Resistance of Medical Face Mask to Penetration by Synthetic Blood. Breathing resistance was tested by NIOSH in accordance to 42 CFR Part 84. The devices have a Bacterial Filtration Efficiency greater than 99.9%. Testing was conducted by Nelson Laboratories using the Modified Green and Vesley Method for evaluation of bacterial filtration efficiency of surgical masks. #### Intended Use: The various models of Inovel Type N95 Healthcare Particulate Respirators and Surgical Masks meet CDC Guidelines for TB Exposure Control within healthcare facilities. These devices are also intended to be worn by healthcare personnel during surgical procedures to protect both the patient and healthcare personnel from the transfer of microorganisms, body fluids and particulate material. ### Limitations: These products do not eliminate the wearer from any risk of contracting any type of disease or infection. The mask should be changed immediately if contaminated with blood or body fluids. ### Comparison of Predicate Devices: The outside cover of the previously cleared devices incorporate a nonwoven polypropylene material with a layer of an extruded plastic mesh and the Inovel models 1511, 1512. 1513 and 1517 incorporate an extruded plastic mesh without the nonwoven material on the outer cover. The head strap color of the cleared device is the same as the Inovel device models for which clearance is being requested. The Inovel type N95 Healthcare Particulate Respirators and Surgical Masks incorporate a highly efficient filter media and is 95% efficiency or better against aerosols that have a count median diameter of 0.055 - 0.095 microns which was scientifically established as the most penetrating particle size. The legally marketed devices previously cleared 510(k) are manufactured from similar materials. {2}------------------------------------------------ # K 061859 ### 510(k) Summary (continued) ### Device and Predicate Devices Descriptions/ Comparisons | Description | Inovel Health Care N95 Particulate<br>Respirators and Surgical Masks, models<br>1511, 1512, 1513, 1517 | Inovel Health Care N95 Particulate<br>Respirators and Surgical Masks, models<br>3001N95-S, 3002N95-M, 3003N95-L,<br>3004N95-LP | |----------------------|--------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------| | Materials | | | | Outer Cover Fabrics | Spunbond polypropylene, Meltblown<br>polypropylene | Spunbond polypropylene, Meltblown<br>polypropylene | | Nosepiece | Polyethylene foam | Polyethylene foam | | Headband | Various colors elastic, latex free | Various colors elastic, latex free | | Specification & | Overall width: 5 - 5 5/8 inches | Overall width: 5 - 5 5/8 inches | | Dimensions | Overall height: $4 \frac{1}{4}$ - $5 \frac{1}{2}$ inches | Overall height: $4 \frac{1}{4}$ - $5 \frac{1}{2}$ inches | | Mask Style | Molded Cup | Molded Cup | | Design Features | Dual synthetic rubber | Dual synthetic rubber | | NIOSH Certification# | TC-84A-0013 | TC-84A-4102 | ### Risks to Health | Performance<br>Characteristics | Test Method | Acceptance criteria/ Results | Predicate Device Results | |-------------------------------------|------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Fluid Resistance<br>Performance | ASTM 1862 - 00a | Inovel Health Care N95 Particulate<br>Respirators and Surgical Masks<br>various models (4)<br>Models 1511, 1513 and 1517<br>32 of 32 pass<br>Model 1512<br>31 of 32 pass | Inovel Health Care N95 Particulate<br>Respirators and Surgical Masks<br>various models (4)<br>Models 3001N95-S, 3002N95-M,<br>3003N95-L, 3004N95-LP<br>32 of 32 pass | | Flammability Class | 16 CFR 1610 | Flame spread must be within upper<br>and lower limits/ No flame spread<br>on 10 of 10 samples, meets Class I | Flame spread must be within upper<br>and lower limits/ No flame spread<br>on 10 of 10 samples, meets Class I | | Filter Efficiency (%) | NIOSH, 42 CFR<br>Part 84 | > 95% Efficient/ average 98.58%<br>efficient of 17 samples | ≥95% Efficient/ average 99.11%<br>efficient of 20 samples | | Breathing<br>Resistance<br>(mm H2O) | NIOSH, 42 CFR<br>Part 84 | ≤ 35.0 mm H2O @ 85 lpm/ average<br>10.3 mm H2O @ 85 lpm of 3<br>samples | ≤ 35.0 mm H2O @ 85 lpm/ average<br>11.3 mm H2O @ 85 lpm of 3<br>samples | | Biocompatibility * | ISO 10993 - 1 | Cytotoxicity<br>Same as predicate device<br><br>Sensitization<br>Same as predicate device<br><br>Primary Skin Irritation<br>Same as predicate device | Cytotoxicity, score of 2 or less/<br>Score of 0<br><br>Sensitization, Grade 1 (no different<br>than control)/ Grade 1<br><br>Primary Skin Irritation,<br>Negligible/Negligible | | Bacterial Filtration<br>Efficeincy | Modified Greene<br>and Vesley<br>Method. J<br>Bacteriol 83:663-<br>667 | Test results show a bacterial<br>filtration efficiency greater than<br>99.9% | Greater than 99.9% | * Tests were conducted on Predicate Devices which are made from the same material as models identified in this 510(k) submission. 12 - 3 {3}------------------------------------------------ ## K061859 12 - 4 ### 510(k) Summary (Continued) Performance Tests: These products were tested and certified by NIOSH as an approved N95 Respirator. It meets all the requirements prescribed in 42 CFR Part 84 and is assigned TC-84A-0013 | Tests Performed | Laboratory | |--------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 1. Fluid Resistance - Resistance of Liquid<br>(Synthetic Blood Penetration Resistance)<br>ASTM F 1862. | Inovel LLC | | 2. Filtration Efficiency (Particulate and<br>Bacterial) 42 CFR Part 84 | NIOSH | | 3. Differential Pressure (Delta P) - Breathing<br>Resistance 42 CFR Part 84 | NIOSH | | 4. Flammability 16 CFR 1610 (Class 1) | Nelson Laboratories | | 5. Biocompatibility* (tested on<br>predicate devices)<br>Cytotoxicity (ISO 10993-5) Sensitization (ISO 10993-10) Irritation (ISO 10993-10) | Nelson Laboratories<br>Northview Pacific Laboratories, Inc<br>(Coordinated by Nelson Laboratories)<br>Northview Pacific Laboratories, Inc<br>(Coordinated by Nelson Laboratories) | | 6. Bacterial Filtration Efficiency<br>Modified Greene and Vesley<br>Method. J Bacteriol 83:663-667. | Nelson Laboratories | * Tests were conducted on Predicate Devices which are made from the same material as models identified in this 510(k) submission. ### Safety/ Effectiveness: The devices have a filtration equivalent to the previously cleared Inovel LLC N95 Particulate Respirator and Surgical mask models 3001N95-M, 3003N95-L and 3004N95-LP 510(k) number K051182. They are NIOSH approved and meet the CDC guidelines for TB and Avian Flu exposure control. ### Conclusion: The basic construction and material used in the cleared devices are the same as in the new devices. The cleared devices and the new devices are also approved by NIOSH, and meets all other required tests. The Inovel type N95 Healthcare Particulate Respirators and Surgical Masks are substantially equivalent to those listed on page 2 - 4. {4}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus, a symbol often associated with medicine and healthcare. The caduceus is depicted with three intertwined strands and a single wing. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" is arranged in a circular pattern around the caduceus. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JUL 1 2 2006 Inovel LLC C/O Mr. Neil E. Devine, Jr. Responsible Third Party Official Intertek Testing Services 2307 East Aurora Road Twinsburg, Ohio 44087 Rc: K061859 Trade/Device Name: Health Care N95 Respirators and Surgical Masks Inovel Models 1511,1512,1513, & 1517 Regulation Number: 878.4040 Regulation Name: Surgical Apparel Regulatory Class: II Product Code: MSH Dated: June 29, 2006 Received: June 30, 2006 Dear Mr. Devine: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {5}------------------------------------------------ ### Page 2 -Mr. Devine Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely vours. Quy Solom for Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ ## Indications for Use KO61859 510(k) Number (if known): Device Name: Health Care N95 Particulate Respirators and Surgical Masks Inovel Models 1511,1512,1513 & 1517 Indications for Use: > The various models of Inovel Type N95 Healthcare Particulate Respirators and Surgical Masks meet CDC Guidelines for TB Exposure Control within healthcare facilities. These devices are also intended to be worn by healthcare personnel during surgical procedures to protect both the patient and healthcare personnel from the transfer of microorganisms, body fluids and particulate material. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use X (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Page 1 of 1 Shade N. Murphy 1/2/04 ി ... വി-(Jin) of Anesthesiology, General Hospital, Jon Control. Dental Devices ) Number.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...