BAYLAB 3-Ply Surgical Mask (BEACON I)

K212302 · Baylab USA, LLC · FXX · Dec 7, 2021 · General, Plastic Surgery

Device Facts

Record IDK212302
Device NameBAYLAB 3-Ply Surgical Mask (BEACON I)
ApplicantBaylab USA, LLC
Product CodeFXX · General, Plastic Surgery
Decision DateDec 7, 2021
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4040
Device ClassClass 2

Intended Use

BAYLAB 3-Ply Surgical Masks (BEACON I) are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These surgical masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device provided nonsterile.

Device Story

Nonwoven, pleated, 3-ply, single-use, disposable surgical mask; manufactured in blue; secured via elastic ear loops. Composed of spunbond polypropylene (inner/outer layers) and meltblown polypropylene (middle filter layer); includes plastic-clad aluminum nose piece. Used in clinical settings by healthcare personnel and patients to provide a physical barrier against microorganisms, body fluids, and particulates. Device is non-sterile and intended for single use. Performance meets ASTM F2100 Level 3 requirements for bacterial filtration efficiency, sub-micron particulate filtration, fluid resistance, and breathability.

Clinical Evidence

Bench testing only. Performance evaluated against ASTM F2100-19 standards: Bacterial Filtration Efficiency (>98%), Sub-micron Particulate Filtration (>99%), Differential Pressure (<6mm H2O/cm2), Synthetic Blood Penetration (160 mmHg), and Flammability (Class 1). Biocompatibility testing performed per ISO 10993-5 (cytotoxicity) and ISO 10993-10 (irritation and sensitization) confirmed the device is non-cytotoxic, non-irritating, and non-sensitizing.

Technological Characteristics

3-ply nonwoven construction: spunbond polypropylene (inner/outer), meltblown polypropylene (filter). Ear loops: spandex, nylon, polyester. Nose piece: plastic-clad aluminum. Dimensions: 17.78 cm x 9.525 cm. Non-sterile, single-use. Meets ASTM F2100 Level 3. Biocompatibility per ISO 10993-5 and ISO 10993-10.

Indications for Use

Indicated for use by healthcare personnel and patients to reduce the transfer of microorganisms, body fluids, and particulate material, and to reduce potential exposure to blood and body fluids in infection control practices.

Regulatory Classification

Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible. (ii) Analysis and nonclinical testing must: (A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and (B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device. (iii) NIOSH approved under its regulation. (2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. December 7, 2021 Baylab USA, LLC % Vardhini Kirthivas Vice President - Regulatory Services Freyr Solutions Level 4, Building No. H-08, Phoenix SEZ, Phase 2, Gachibowli Hyderabad, Telangana 500081 India Re: K212302 Trade/Device Name: BAYLAB 3-Ply Surgical Mask (BEACON I) Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical apparel Regulatory Class: Class II Product Code: FXX Dated: November 9, 2021 Received: November 10, 2021 Dear Vardhini Kirthivas: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, For Clarence W. Murray III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K212302 Device Name BAYLAB 3-Ply Surgical Masks (BEACON I) #### Indications for Use (Describe) BAYLAB 3-Ply Surgical Masks (BEACON I) are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device provided nonsterile. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| |-------------------------------------------------|--| | | Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Phone: E-mail: ### 510(k) Summary K212302 ### 1. Submitter Information: Application Correspondent (US Agent): Vardhini Kirthivas Freyr Global Regulatory Solutions and Services Level 4, Building No. H-08, Phoenix SEZ, Phase 2, Gachibowli, Hyderabad, Telangana, India, 500081 +91 - 9940445965 vardhini.kirthivas@FreyrSolutions.com | Legal Manufacturer: | BAYLAB USA LLC | |-----------------------------------|---------------------------| | Phone: | 214-907-2527 | | Contact Person: | Ashley Park | | E-mail: | Ashley.park@baylabusa.com | | Fax Number (including area code): | 469-372-0414 | | Date Prepared: | 22-June-2021 | ### 2. Device Identification: | Device Trade Name: | BAYLAB 3-Ply Surgical Masks (BEACON I) | |-------------------------------|----------------------------------------| | Device: | Mask, Surgical | | Regulation Description: | Surgical Apparel | | Regulation Medical Specialty: | General & Plastic Surgery | | Review Panel: | General Hospital | | Device Class: | II | | Regulation Number: | 878.4040 | | Product Code: | FXX | ### 3. Predicate Devices: Table 1 - List of Predicate Devices | Device Name | 510(k) Number | |-------------|---------------| | DemeMASK | K201479 | ### 4. Device Description BAYLAB 3-Ply Surgical Masks (BEACON I) are nonwoven, pleated, 3-ply, single use, disposable flat surgical masks, manufactured in Bay Blue color. The surgical masks are to be secured on users by means of elastic ear loops. {4}------------------------------------------------ # 5. Intended Use & Indications for Use BAYLAB 3-Ply Surgical Masks (BEACON I) are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These surgical masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device provided nonsterile. {5}------------------------------------------------ # 6. Comparison of Technological Characteristics Table 2 – Comparison Table | S.No | Parameters | | Predicate Device | Subject device | Comments | |------|-----------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------| | 1. | Name | | DemeMASK | BAYLAB 3-Ply Surgical Masks (BEACON I) | N/A | | 2. | 510(k) Number | | K201479 | K212302 | N/A | | 3. | Manufacturer | | DemeTECH Corporation | BAYLAB USA LLC | N/A | | 4. | Product Code | | FXX | FXX | Same | | 5. | Regulation Number | | 878.4040 | 878.4040 | Same | | 6. | Intended Use/Indications of Use | | The Disposable Surgical Face Masks are<br>intended to be worn to protect both the patient<br>and healthcare personnel from transfer of<br>microorganisms, body fluids and particulate<br>material. These face masks are intended for use<br>in infection control practices to reduce the<br>potential exposure to blood and body fluids.<br>This is a single use, disposable device provided<br>nonsterile. | BAYLAB 3-Ply Surgical Masks (BEACON I) are<br>intended to be worn to protect both the patient and<br>healthcare personnel from transfer of<br>microorganisms, body fluids and particulate material.<br>These surgical masks are intended for use in<br>infection control practices to reduce the potential<br>exposure to blood and body fluids. This is a single<br>use, disposable device provided nonsterile. | Same | | 7. | Material of<br>Construction | Inner Layer | Spun-bond Polypropylene | Spunbond polypropylene | Same | | | | Outer Layer | Spun-bond Polypropylene | Spunbond polypropylene | Same | | | | Middle Layer | Meltblown polypropylene filter | Meltblown polypropylene filter | Same | | | | Ear loops | Spandex and Nylon- Not made with natural<br>rubber latex | Spandex, Nylon, Polyester - Not made with natural<br>rubber latex | Different,<br>Note 1 | | | | Nose Piece | Galvanized wire coated with polyethylene | Plastic (polyethylene) clad aluminum | Different,<br>Note 2 | | 8. | Dimensions | | Length: 17.5 cm±1 cm<br>Width: 9.5 cm±1 cm | Mask size: 7" x 3 ¾" (17.78 cm x 9.525 cm) | Same | | 9. | Color | | Blue | Blue | Same | | 10. | Mask Style | | Flat - pleated | Flat pleated, 3-ply | Same | | 11. | Design Features | | Ear loop | Ear loop | Same | | 12. | Sterility | | Non - Sterile | Non- Sterile | Same | | 13. | Use | | Single Use | Single Use | Same | | S.No | Parameters | Predicate Device | Subject device | Comments | | | 14. | Latex | Not made with Natural Rubber Latex | Not made with Natural Rubber Latex | Same | | | 15. | ASTM F2100 Level | Level 3 | Level 3 | Same | | | 16. | Bacterial filtration efficiency | Pass at ≥ 99%<br>(ASTM F2101) | Pass at >98%<br>(ASTM F2101) | Different,<br>Note 3 | | | 17. | Differential pressure (Delta-P) | Average 3.6 mmH2O/cm²<br>(MIL-M-36954C) | ΔP <6mm H2O/cm²<br>(EN 14683) | Different,<br>Note 4 | | | 18. | Sub-micron particulate filtration efficiency | Pass at ≥99%<br>(ASTM F2299) | Pass at >99%<br>(ASTM F2299) | Similar,<br>Note 5 | | | 19. | Resistance to penetration by<br>synthetic blood, minimum pressure<br>in mm Hg for pass result | Pass at 160 mm Hg (ASTM F1862) | Pass at 160 mm Hg (ASTM F1862) | Same | | | 20. | Flammability | Class 1 (16 CFR Part 1610) | Class 1 (16 CFR Part 1610) | Same | | | 21. | Biocompatibility Cytotoxicity,<br>ISO 10993<br>5:2009 | Pass ISO 10993-5:2009/ under the conditions of<br>study the subject device was non-cytotoxic. | Pass ISO 10993-5:2009/ under the conditions of<br>study the subject device was non-cytotoxic. | Same | | | | Biocompatibility Irritation,<br>ISO 10993-<br>10:2010 | Pass ISO 10993-10:2010/ under the conditions<br>of the study, the subject device was non-irritating. | Pass ISO 10993-10:2010/ under the conditions of the<br>study, the subject device was non-irritating. | Same | | | | Biocompatibility Sensitization,<br>ISO 10993-<br>10:2010 | Pass ISO 10993-10:2010/ under the conditions<br>of study, the subject device was non-sensitizing. | Pass ISO 10993-10:2010/ under the conditions of<br>study, the subject device was non-sensitizing. | Same | | {6}------------------------------------------------ Note 1 & Note 2: The differences in the materials do not raisety and effectiveness. Performance testing including biocompatibility evaluation has been performed on the final finished device which includes all construction materials. Note 3: The performance requirement for BFE as per ASTM F2100 is ≥98%. The performance of the requirements of ASTM F2100 version 2020 which is the revised version of FDA recognized consensus standard [Rec# 6-425] ASTM F2100-19. The subject device meets the criteria. Note 4: We understand that MIL-M-36954C is the FDA recommended standard for differential pressure, However, the conformance to FDA Recognized consensus standard [Rec# 6-425], ASTM F2100-19 requires that Differential pressure be performed as per EN 14683:2019, Anex C. We also understand that as per ASTM F2100-19, passing criteria for an ASTM level 3 mask with respect to differential pressure is <6.0mm HJVcm² when tested in accordance with EN 14683:2019. Annex C. The subject device meets the criteria. Note 5: The performance requirement for PFE as per ASTM F2100 is ≥98% when tested in accordance with ASTM F2299. The performance of the subject device meets the requirements of ASTM F2100 version 2020 which is the revised consensus standard [Rec# 6-425] ASTM F2100-19. The subject device meets the criteria. {7}------------------------------------------------ Traditional 510(k) BAYLAB 3-ply Surgical Masks (BEACON I) ## 7. Non-Clinical Testing The following performance tests, in accordance with ASTM F2100 were conducted for the subject device: - . ASTM F1862 Standard test method for resistance of medical face masks to penetration by synthetic blood (Horizontal projection of fixed volume at a known velocity) - 16 CFR Part 1610 Standard for The Flammability of Clothing Textiles - EN 14683:2019+AC:2019 Medical face masks Requirements and test methods ● Annex C - Method for determination of breathability (differential pressure) - ASTM F2299- Standard Test Method for Determining the Initial Efficiency of Materials Used in Medical Face Masks to Penetration by Particulates Using Latex Spheres - ASTM F2101-19 Standard Test Method for Evaluating the Bacterial Filtration ● Efficiency (BFE) of surgical masks using a Biological Aerosol of Staphylococcus aureus. | Sr. No. | Test Method/<br>Standard | Purpose | Acceptance<br>criteria | Results | |---------|--------------------------|-------------------------------------------------|--------------------------|------------------| | 1. | ASTM F 2101 | Bacterial filtration efficiency | Level 3: ≥98% | Pass at >98% | | 2. | EN 14683 | Differential Pressure (Delta-<br>P) | Level 3: ΔP <6mm H2O/cm2 | ΔP <6mm H2O/cm2 | | 3. | ASTM F2299 | Sub-micron particulate<br>filtration efficiency | Level 3: ≥98% | Pass at >99% | | 4. | ASTM F1862 | Resistance to penetration by<br>synthetic blood | Level 3: 160mm<br>Hg | Pass at 160mm Hg | | 5. | 16 CFR Part 1610 | Flammability | Class I | Class I | | Table 3 - Performance Testing - summary | | |-----------------------------------------|--| | | | FDA's guidance, "Surgical Masks - Premarket Notification [510(k)] Submissions ", recommends evaluating the biocompatibility as described in the standard ISO10993, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing" for limited contact devices, contacting intact skin. The following Biocompatibility End points have been identified and tested in accordance appropriate biocompatibility standards. - Cytotoxicity (ISO 10993-5) - - Irritation or intracutaneous reactivity (ISO 10993-10) - - -Sensitization (ISO 10993-10) | Biological<br>endpoint | Test<br>Method | Purpose | Acceptance<br>criteria | Test Result | |---------------------------------|---------------------|--------------------------------------------------------------|-------------------------------------------|----------------------------------------------------------------------------------------------------| | Cytotoxicity | ISO<br>10993-5 | Verify<br>Cytotoxicity<br>potential of the<br>subject device | Non-cytotoxic | Pass - ISO 10993-5:2009/ under the<br>conditions of study the subject<br>device was non-cytotoxic. | | Irritation and<br>Sensitization | ISO<br>10993-<br>10 | Verify irritation<br>and sensitization | Non-irritating<br>and non-<br>sensitizing | Pass ISO 10993-10:2010/ under the<br>conditions of the study, the subject | | Table 4 - Biocompatibility Testing - summary | | |----------------------------------------------|--| | | | {8}------------------------------------------------ # Traditional 510(k) BAYLAB 3-Ply Surgical Masks (BEACON I) | | potential of the | device was non-irritating and non- | |--|------------------|------------------------------------| | | subject device | sensitizing. | # 8. Conclusion The conclusions drawn from the nonclinical tests demonstrate that the proposed device is as safe, as effective, and performs as well as or better than the legally marketed predicate device, DemeMASK cleared under K201479.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...