K210150 · Hubei Wanli Protective Products Co., Ltd. · FXX · Jun 2, 2021 · General, Plastic Surgery
Device Facts
Record ID
K210150
Device Name
Disposable Medical mask
Applicant
Hubei Wanli Protective Products Co., Ltd.
Product Code
FXX · General, Plastic Surgery
Decision Date
Jun 2, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4040
Device Class
Class 2
Intended Use
The Disposable Medical Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.
Device Story
Disposable, single-use, three-layer flat-pleated medical mask; features ear loops and integrated nose clip. Constructed with outer/inner layers of non-woven spun-bond polypropylene and middle layer of melt-blown polypropylene filter. Worn by healthcare personnel and patients to provide barrier protection against microorganisms, body fluids, and particulates. Used in clinical settings for infection control. Device is non-sterile; provided in blue. Performance meets ASTM F2100 Level 2 requirements.
Clinical Evidence
No clinical data. Bench testing only. Performance verified against ASTM F1862 (synthetic blood penetration), ASTM F2299 (particulate filtration), ASTM F2101 (bacterial filtration), ASTM F2100/EN 14683 (differential pressure), and 16 CFR 1610 (flammability). Biocompatibility confirmed via ISO 10993-5 (cytotoxicity) and ISO 10993-10 (irritation/sensitization).
Technological Characteristics
Three-layer construction: non-woven spun-bond polypropylene (outer/inner), melt-blown polypropylene (middle). Spandex elastic ear loops. Dimensions: 175mm x 95mm. Non-sterile. ASTM F2100 Level 2 performance. Biocompatibility per ISO 10993.
Indications for Use
Indicated for use by patients and healthcare personnel to reduce exposure to microorganisms, body fluids, and particulate material during infection control practices.
Regulatory Classification
Identification
Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.
Special Controls
*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.
K221272 — Medical Face Mask · Shandong Xingyu Gloves Co., Ltd. · Jun 28, 2022
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 2, 2021
Hubei Wanli Protective Products Co., Ltd. % Eva Li Consultant Shanghai Sungo Management Consulting Company Limited Room 1309, Dongfang Building, 1500#Century Ave Shanghai, Shanghai 200122 China
Re: K210150
Trade/Device Name: Disposable Medical mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: January 21, 2021 Received: January 21, 2021
Dear Eva Li:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Ryan Ortega, PhD Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
K210150
Device Name Disposable Medical Mask
Indications for Use (Describe)
The Disposable Medical Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Image /page/2/Picture/19 description: The image shows the text "PSC Publishing Services (301) 443-6740". The text appears to be a company name and phone number. The area code is 301 and the phone number is 443-6740.
EF
{3}------------------------------------------------
Hubei Wanli Protective Products Co., Itd. Yuanshi, Ganhe, Xiantao, Hubei, China
### K210150
# 510(K) Summary
#### A. Applicant
Hubei Wanli Protective Products Co., Itd. Address: Yuanshi, Ganhe, Xiantao, Hubei, China Contact Person: Andy Wen Tel: 0086-728-3227299 Email: sale01@hbwanli.com Date Prepared: April 27, 2021
Submission Correspondent Primary contact: Ms. Eva Li Shanghai SUNGO Management Consulting Co., Ltd. Room 1309, Dongfang Building, 1500# Century Ave., Shanghai 200122, China Tel: +86-21-58817802 Email: eatereva@hotmail.com Secondary contact: Mr. Raymond Luo Room 1309, Dongfang Building, 1500# Century Ave., Shanghai 200122, China Tel: +86-21-68828050 Email: fda.sungo@gmail.com
#### B. Device
Trade Name: Disposable Medical Mask
| Model | Description |
|---------|-------------------------------------------------|
| WLM2002 | Ear Loops, Flat Pleated, 3 layers, 17.5cm*9.5cm |
Regulatory Information Classification Name: Surgical Face Mask Classification: Class II Product code: FXX Regulation Number: 878.4040 Review Panel: Surgical Apparel
### C. Predicate device:
K110455 Kimberly-Clark KC100 Mask Kimberly-Clark
#### D. Intended use of the device:
The Disposable Medical Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce
{4}------------------------------------------------
# Hubei Wanli Protective Products Co., Itd.
Yuanshi, Ganhe, Xiantao, Hubei, China
the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.
### E. Device Description:
The Disposable Medical Masks are single use, three-layer, flat –folded masks with ear loops and nose clip.
The Disposable Medical Masks are manufactured with three layers, the inner and outer layers are made of non-woven Spun-bond polypropylene, and the middle layer is made of melt blown polypropylene filter.
The ear loops are held in place over the users' mouth and nose by two strings welded to the facemask. The ear loops are made of 17cm spandex elastic strings(performance of the spandex elastic: Tensile strength≥100N; Elasticity≥2.8times; Breaking tension> 100N)
The nose clip in the layers of facemask is to allow the user to fit the facemask around their nose, which is not touch with the users' skin directly.
The Disposable Medical Masks will be provided in blue. The Disposable Medical Masks are sold non-sterile and are intended to be single use, disposable devices.
## F. Comparison with predicate device
Table 1 General Comparison
| Device | Proposed Device | Predicate Device | Comparison |
|----------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Manufacturer | Hubei Wanli Protective Products<br>Co., ltd. | Kimberly-Clark | — |
| 510(K) number | K210150 | K110455 | — |
| Model Name | Disposable Medical Mask | Kimberly-Clark KC100 Mask | — |
| Classification | Class II Device, FXX (21<br>CFR878.4040) | Class II Device, FXX (21<br>CFR878.4040) | same |
| Intend use | The Disposable Medical Masks are<br>intended to be worn to protect both<br>the patient and healthcare<br>personnel from transfer of<br>microorganisms, body fluids and<br>particulate material. These face<br>masks are intended for use in<br>infection control practices to reduce<br>the potential exposure to blood and<br>body fluids. This is a single use,<br>disposable device(s), provided non-<br>sterile. | The Kimberly-Clark KC100<br>Procedure Mask(s) is intended to<br>be worn to protect both the<br>patient and healthcare personnel<br>from transfer of microorganisms,<br>body fluids and particulate<br>material. These face masks are<br>intended for use in infection<br>control practices to reduce the<br>potential exposure to blood and<br>body fluids. The Kimberly-Clark<br>KC100 Procedure Mask(s) is a<br>single use, disposable devices,<br>provided non-sterile. | same |
| Model | Ear Loops, Flat Pleated, 3 layers | Ear Loops, Tie-On, Flat Pleated, 3<br>layers | same |
{5}------------------------------------------------
### Hubei Wanli Protective Products Co., Itd. Yuanshi, Ganhe, Xiantao, Hubei, China
| | Outer<br>facing<br>layer | | | |
|---------------------|--------------------------------------|--------------------------------------|---------------------------------|----------|
| Material | non-woven Spun-bond<br>polypropylene | non-woven Spun-bond<br>polypropylene | Spun-bond polypropylene | same |
| | Middle<br>layer | Melt blown polypropylene filter | Melt blown polypropylene filter | same |
| | Inner<br>facing<br>layer | non-woven Spun-bond<br>polypropylene | Spun-bond polypropylene | same |
| | Ear<br>loops/ties | Spandex elastic | Polyester/lycra knitted | similar |
| | Color | Blue | Variety (include blue) | same |
| | Dimension<br>(length) | $175mm \pm 5mm$ | $165 \pm 19mm$ | Similar |
| | Dimension<br>(width) | $95 mm \pm 5mm$ | $102 \pm 19mm$ | Similar |
| | OTC use | Yes | Yes | same |
| | Sterility | Non-Sterile | Non-Sterile | same |
| | Use | Single use, Disposable | Single use, Disposable | same |
| ASTM F2100<br>Level | | Level 2 | Level 1 | Similar* |
| Biocompatibility | | ISO10993 | ISO10993 | same |
## *Similar Discussion :
The proposed device conducted the test and the pass the Level 3 Acceptance Criteria, the predicate device pass the level 1 Acceptance Criteria. The test and the acceptance is following:
# G. Summary of Non-Clinical Performance Testing
Non-clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards and the requirements stated in the Guidance for Industry and FDA Staff: Surgical Masks - Premarket Notification [510(k)] Submission issued on March 5, 2004:
> ISO 10993-5: 2009 Biological Evaluation of Medical Devices -- Part 5: Tests For In Vitro Cytotoxicity
> ISO 10993-10: 2010 Biological Evaluation of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization
> ASTM F2100, Standard Specification for Performance of Materials Used in Medical Face Masks
> ASTM F1862, Standard Test Method for Resistance of Medical Face Masks to Penetration by Synthetic Blood (Horizontal Projection of Fixed Volume at A Known Velocity);
EN 14683, Medical Face Masks-Requirements and Test Methods; A
- A ASTM F2101, Standard Test Method for Evaluating the Bacterial Filtration Efficiency
{6}------------------------------------------------
# Hubei Wanli Protective Products Co., ltd.
Yuanshi, Ganhe, Xiantao, Hubei, China
(BFE) Of Medical Face Mask Materials, Using A Biological Aerosol of Staphylococcus Aureus;
> ASTM F2299, Standard test method for determining the initial efficiency of
materials used in medical face masks to penetration by particulates using latex spheres;
A 16 CFR 1610, Standard for the Flammability of clothing textiles;
### Table 2-performance Testing
| Item | Acceptance<br>Criteria (level<br>2) | Result of<br>Lot2020120501 | Result of<br>Lot2020120101 | Result of<br>Lot2020120801 |
|---------------------------------------------------------------------------------------------|----------------------------------------|-------------------------------------|-------------------------------------|-------------------------------------|
| Synthetic Blood<br>Performance<br>ASTM F1862 | 29 out of 32<br>pass at 120<br>mmHg | 32 out of 32<br>pass at 120<br>mmHg | 32 out of 32<br>pass at 120<br>mmHg | 32 out of 32<br>pass at 120<br>mmHg |
| Particulate Filtration<br>Efficiency<br>ASTM F2299 | ≥ 98% | ≥ 98% | ≥ 98% | ≥ 98% |
| Bacterial Filtration<br>Efficiency<br>ASTM F2101-19<br>EN 14683:2019 Annex B | ≥ 98% | ≥ 98% | ≥ 98% | ≥ 98% |
| Differential<br>Pressure(Delta P)<br>ASTM F2100-19<br>EN 14683:2019 Annex C<br>( mmH₂O/cm²) | < 6.0 | < 6.0 | < 6.0 | < 6.0 |
| Flammability<br>16 CFR 1610<br>(IBE=Test Article ignited,<br>but extinguished) | Class 1 (Burn<br>time ≥3.5<br>seconds) | Class I | Class I | Class I |
### Table3 Biocompatibility Comparison
| Item | Proposed device | Acceptance<br>Criteria | Result |
|---------------|------------------------------------------------------------------------|------------------------|----------------|
| Cytotoxicity | Under the conditions of<br>the study, the device is<br>non-cytotoxic. | Non-Cytotoxic | No Cytotoxic |
| Irritation | Under the conditions of<br>the study, the device is<br>non-irritating. | Non-Irritating | No irritating |
| Sensitization | Under the conditions of<br>the study, the device is<br>non-sensitizing | Non-Sensitizing | No Sensitizing |
{7}------------------------------------------------
## Hubei Wanli Protective Products Co., ltd. Yuanshi, Ganhe, Xiantao, Hubei, China
### H. Clinical Test Conclusion
No clinical study is included in this submission.
#### I. Conclusion
Based on the nonclinical tests performed, the subject device is as safe, as effective, and performs as well as the legally marketed predicate device, Kimberly-Clark KC100 Mask cleared under K110455.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.