K100846 · International Medsurg Connection · FXX · Aug 16, 2010 · General, Plastic Surgery
Device Facts
Record ID
K100846
Device Name
IMC SURGICAL FACE MASK
Applicant
International Medsurg Connection
Product Code
FXX · General, Plastic Surgery
Decision Date
Aug 16, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4040
Device Class
Class 2
Intended Use
This device is intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material.
Device Story
IMC Surgical Face Mask is a flat, pleated, 3-ply surgical apparel device. It consists of inner and outer polypropylene layers sandwiching a middle melt-blown polypropylene filter layer. The device is available in yellow, with either tie-on or ear loop attachment mechanisms, and is provided in both sterile and non-sterile configurations. It is used by operating room personnel during surgical procedures. The mask acts as a physical barrier to reduce the transfer of microorganisms, body fluids, and particulate material between the wearer and the surgical environment. It is a single-use, disposable device.
Clinical Evidence
Bench testing only. Performance characteristics were evaluated against ASTM and military standards: Fluid Resistance (ASTM F1862-07), Particulate Filtration Efficiency (ASTM F2299-03, ≥98%), Bacterial Filtration Efficiency (ASTM F2101-07, ≥95-98%), Differential Pressure (MIL-M36945C, ≤4 mmH2O/cm²), and Flammability (16 CFR Part 1610, Class 1). Materials were tested for biocompatibility per ISO 10993-5, 10993-10, and 10993-11.
Technological Characteristics
3-ply flat pleated mask; inner/outer layers polypropylene; middle layer melt-blown polypropylene filter media. Features nose piece, ear loops or tie-on attachments. Dimensions: 7" x 3.5". Biocompatibility tested per ISO 10993-5, 10993-10, 10993-11. Sterile and non-sterile versions.
Indications for Use
Indicated for operating room personnel during surgical procedures to protect patients and personnel from transfer of microorganisms, body fluids, and particulate material.
Regulatory Classification
Identification
Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.
Special Controls
*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.
Predicate Devices
Medline Prohibit Series Surgical Mask With Fluid Shield (K991559)
Crosstex Isolite Earloop Face Masks-Blue, Pink, Crosstex Isofluid Earloop Face Masks-Blue, Pink, White, Green (K012602)
Related Devices
K210433 — Surgical Face Mask (Ear loops and Tie-on) · Wuhan Dymex Healthcare Co., Ltd. · Jun 7, 2021
K063043 — SURGICAL FACE MASK, TYPE: TIE-ON, EAR-LOOP · Modern Healthcare Corp. · Mar 21, 2007
K043199 — SURGICAL FACE MASK · Hong YA Non-Woven Products Co., Ltd. · Feb 16, 2005
K123115 — ACME SURGICAL FACE MASK WITH EAR-LOOP MODEL YN-501 AB/W/G/P · Acme Filter Mask, Inc. · May 20, 2013
K220777 — Surgical face mask · Hubei Qianjiang Kingphar Medical Material Co., Ltd. · Jul 19, 2022
Submission Summary (Full Text)
{0}------------------------------------------------
K100846
## AUG 1 6 2010
Trade Name: IMC Surgical Face Mask (non-sterile and sterile, yellow) Common Name: Surgical Face Mask 21 CFR 878.4040 (b): Surgical Apparel Classification Name: Submitter Information: International Medsurg Connection 935 N Plum Grove Rd, STE F Schaumburg, Illinois 60173 Summary Prepared By: Peter Kim Quality Manager International Medsurg Connection 935 N Plum Grove Rd, STE F Schaumburg, Illinois 60173 Telephone: 847-619-9926 Fax: 847-619-9927 e-mail: peterkim(@intlmedsurg.com Date Prepared: June 7, 2010 Predicate Devices: . Medline Prohibit Series Surgical Mask With Fluid Shield (K991559) Crosstex Isolite Earloop Face Masks-Blue, Pink, Crosstex Isofluid Earloop Face Masks-Blue, Pink, White, Green (K012602)
## 510(K) SUMMARY
## Device Name(s):
IMC Surgical Face Mask (non-sterile and sterile, yellow)
## Classification Panel:
General and Plastic Surgery
## Legally Marketed Device Under Which Substantial Equivalence is Being Claimed:
International Medsurg Connections, Inc is claiming substantial equivalence of the IMC Surgical Mask with the currently marketed:
| Description | 510(k) Number | Clearance Date |
|--------------------------------------------------------------------|---------------|----------------|
| Medline Prohibit Series Surgical Mask With Fluid<br>Shield | K991559 | 08/23/1999 |
| Crosstex Isolite Earloop, Crosstex Isofluid Earloop<br>Face Masks. | K012602 | 01/22/2002 |
#### Device Description
The IMC Surgical Face Mask devices are surgical apparel intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. The surgical masks feature either tie-on or ear loop attachment mechanisms
{1}------------------------------------------------
This device is a flat pleated 3 ply mask. The inner and outer layer of the mask is made of polypropylene, sandwiching a middle layer made of melt-blown polypropylene filter media. The outer layer of the mask is offered in a yellow color.
## Statement of Intended Use
Indications For Use: This device is intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material.
| Name/Description | Sterility |
|-----------------------------------------------|-------------|
| • IMC Surgical face mask Yellow with Ear loop | Sterile |
| • IMC Surgical face mask Yellow with Tie-on | Sterile |
| • IMC Surgical face mask Yellow with Ear loop | Non-Sterile |
| • IMC Surgical face mask Yellow with Tie-on | Non-Sterile |
The IMC Surgical Mask and the predicate devices (Medline Prohibit Series Surgical Mask With Fluid Shield and the Crosstex series: Isolite Earloop Face Masks and Isofluid Earloop Face Masks) are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material.
| Feature/<br>Characteristic | IMC Surgical Face Mask<br>K100846 | Medline<br>K991559<br>(Predicate) | Crosstex<br>K012602<br>(Predicate) |
|----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------|------------------------------------|
| Layer 1 - Outer | These materials were tested<br>in accordance with ISO<br>10993-5, 10993-10, and<br>10993-11 test methods and<br>were found to be acceptable<br>for the intended use | Same | Same |
| Layer 2 - Middle | These materials were tested<br>in accordance with ISO<br>10993-5, 10993-10, and<br>10993-11 test methods and<br>were found to be acceptable<br>for the intended use | Same | Same |
| Layer 3 - Inner | These materials were tested<br>in accordance with ISO<br>10993-5, 10993-10, and<br>10993-11 test methods and<br>were found to be acceptable<br>for the intended use | Same | Same |
| Ear loop | These materials were tested<br>in accordance with ISO<br>10993-5, 10993-10, and<br>10993-11 test methods and<br>were found to be acceptable<br>for the intended use | Same | Same |
{2}------------------------------------------------
| Feature/<br>Characteristic | IMC Surgical Face Mask<br>K100846 | Medline<br>K991559<br>(Predicate) | Crosstex<br>K012602<br>(Predicate) |
|-------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------|------------------------------------|
| Tie-on | These materials were tested<br>in accordance with ISO<br>10993-5, 10993-10, and<br>10993-11 test methods and<br>were found to be acceptable<br>for the intended use | Same | Same |
| Nose Piece | These materials were tested<br>in accordance with ISO<br>10993-5, 10993-10, and<br>10993-11 test methods and<br>were found to be acceptable<br>for the intended use | Same | Same |
| colors | | | |
| Yellow | These materials were tested<br>in accordance with ISO<br>10993-5, 10993-10, and<br>10993-11 test methods and<br>were found to be acceptable<br>for the intended use | Same | Same |
| Specification and Dimensions: | | | |
| Dimension | | | |
| • Length | 7 inches | Similar | Same |
| • Width | 3.5 inches | Similar | Same |
| Mask Style | Ear loop | Same | Same |
| Tie-on | Tie-on | Same | Same |
# Performance Data:
827
| Performance<br>Characteristics | Test Method | Acceptance<br>Criteria | IMC Surgical<br>Face Mask<br>K100846 | Medline<br>K991559<br>(Predicate) | Crosstex<br>K012602<br>(Predicate) |
|------------------------------------------------------|------------------------------------|------------------------------------------|--------------------------------------|-----------------------------------|------------------------------------|
| Fluid Resistance<br>Performance (mmHg) | ASTM F1862-07<br>(2007) | Low=80<br>Moderate = 120<br>High = 160 | Meets Standard<br>Criteria | Same | Same |
| Particulate Filtration<br>Efficiency Performance (%) | ASTM F2299-03<br>(2003) | ≥98% | Meets Standard<br>Criteria | Same | Same |
| Bacterial Filtration<br>Efficiency Performance (%) | ASTM F2101-07<br>(2007) | Low: ≥95<br>Moderate: ≥98%<br>High: ≥98% | Meets Standard<br>Criteria | Same | Same |
| Different Pressure (Delta-P)<br>(mmH2O/cm²) | MIL-M36945C<br>4.4.1.1.1<br>(1975) | ≤4 mmH2O/cm2 | Meets Standard<br>Criteria | Same | Same |
| Flammability | 16 CFR Part 1610<br>(1998) | Class 1 | Class 1 | Same | Same |
:
:
{3}------------------------------------------------
## Conclusions:
The indications for use, technology, specification, safety of the IMC Surgical Face Masks and the two predicate devices K991559 and K012602 are essentially the same. The differences between the face masks are minor and do not raise new issues of safety or effectiveness. Hence, the IMC Surgical Face Masks are substantially equivalent to the predicate devices.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized symbol featuring three overlapping, curved lines that resemble a human form or a flowing design.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
International MedSurg Connection C/O Mr. Peter Kim 935 North Plum Grove Road, Suite F Schaumburg, Illinois 60173
AUG 1 6 2010
Re: K100846
Trade/Device Name: IMC Surgical Face Mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: II Product Code: FXX Dated: July 28, 2010 Received: August 2, 2010
Dear Mr. Kim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
## Page 2- Mr. Kim
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the ·MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.html for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
th for
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure _
{6}------------------------------------------------
K100846
# Indications for Use
510(k) Number : K100846
Device Name: IMC Surgical Face Mask
Indications For Use: This device is intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material.
| Name/Description | Sterility |
|-----------------------------------------------|-------------|
| • IMC Surgical face mask Yellow with Ear loop | Sterile |
| • IMC Surgical face mask Yellow with Tie-on | Sterile |
| • IMC Surgical face mask Yellow with Ear loop | Non-Sterile |
| • IMC Surgical face mask Yellow with Tie-on | Non-Sterile |
AND/OR Over-The-Counter Use __________________________________________________________________________________________________________________________________________________ X Prescription Use (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Elaine S. Mayhall for Elizabeth F. Claverie
(Division Sign-Off) (Division of Anesthesiology, General Hospital Infection Control, Dental Devices
K 100846 510(k) Number: _______________________________________________________________________________________________________________________________________________________________
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.