CROSSTEX ISOLITE EARLOOP FACE MASKS-BLUE, PINK, CROSSTEX ISOFLUID EARLOOP FACE MASKS-BLUE, PINK, WHITE, GREEN. CROSSTEX
K012602 · Crosstex Intl. · FXX · Jan 22, 2002 · General, Plastic Surgery
Device Facts
Record ID
K012602
Device Name
CROSSTEX ISOLITE EARLOOP FACE MASKS-BLUE, PINK, CROSSTEX ISOFLUID EARLOOP FACE MASKS-BLUE, PINK, WHITE, GREEN. CROSSTEX
Applicant
Crosstex Intl.
Product Code
FXX · General, Plastic Surgery
Decision Date
Jan 22, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4040
Device Class
Class 2
Indications for Use
The following Crosstex® Surgical Masks are intended for use in infection control practices to minimize contamination caused by inhaled and exhaled microorganisms and reduce the potential exposure of the wearer to blood and body fluids. - Crosstex® Isolite® Earloop Face Masks Blue, Pink . - Crosstex® Isofluid® Earloop Face Masks -- Blue, Pink, White, Green . - Crosstex® Isofluid FogFree® Earloop Face Masks Blue, Peach . - Crosstex® Isofluid FogFree® Face Masks with Splash Visor Blue, Peach ● - Crosstex® Procedural Earloop Face Masks Blue, Pink, Yellow ● - Crosstex® Ultra Fluid Resistant No-Fog® Earloop Face Masks Blue . - Crosstex® Ultra Fluid Resistant No-Fog® Face Masks with Splash Visor -. Blue
Device Story
Surgical masks designed for infection control; constructed from medical grade tissue or 100% spunbonded polypropylene; features include elastic earloops, fog-free strips, and 27-gauge aluminum nose wire; specific models include vapor barriers and splash visors. Used by healthcare personnel in clinical settings to provide a physical barrier against microorganisms and fluid splashes. Masks are disposable; worn over nose and mouth. Benefit includes reduced risk of cross-contamination and protection from bodily fluid exposure.
Clinical Evidence
No clinical data. Bench testing only: fluid resistance (impact penetration), bacterial filtration efficiency (BFE), differential pressure (ΔP), flammability, latex particle challenge, and biocompatibility (ISO 10993).
Technological Characteristics
Materials: medical grade tissue or 100% spunbonded polypropylene. Components: 27-gauge aluminum wire, elastic loops, anti-fog strips, vapor barrier (select models). Biocompatibility: ISO 10993. Form factor: earloop face masks with optional splash visors.
Indications for Use
Indicated for use in infection control practices to minimize contamination from inhaled/exhaled microorganisms and reduce wearer exposure to blood and body fluids.
Regulatory Classification
Identification
Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.
Special Controls
*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.
Predicate Devices
American Threshold Industries, Surgical Masks (K801036)
American Threshold Industries, Fluid Resistant Masks (K955556)
Submission Summary (Full Text)
{0}------------------------------------------------
## Attachment #2
# 510(K) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
س2012 The assigned 510(k) number is:_KO124
### Submitter's Identification: 1.
Crosstex International, Inc. 10 Ranick Road Hauppauge, New York 11788 Tel No.: 631-582-6777 Fax No.: 631-582-1726
## Contact Person:
Mr. Richard Allen, Sr. President Crosstex International, Inc. 10 Ranick Road Hauppauge, New York 11788 Tel No.: 631-582-6777 Fax No.: 631-582-1726
Date Summary Prepared: July 27, 2001
#### Name of the Device: 2.
- Crosstex® Isolite® Earloop Face Masks Blue, Pink ●
- Crosstex® Isofluid® Earloop Face Masks -- Blue, Pink, White, Green .
- Crosstex® Isofluid FogFree® Earloop Face Masks Blue, Peach ●
- Crosstex® Isofluid FogFree® Face Masks with Splash Visor Blue, Peach ●
- Crosstex® Procedural Earloop Face Masks Blue, Pink, Yellow ●
- Crosstex® Ultra Fluid Resistant No-Fog® Earloop Face Masks Blue .
- Crosstex® Ultra Fluid Resistant No-Fog® Face Masks with Splash Visor -. Blue
#### Predicate Device Information: 3.
- American Threshold Industries, Surgical Masks, K# 801036, Asheville, a. North Carolina
{1}------------------------------------------------
- American Threshold Industries, Fluid Resistant Masks, K# 955556, Enka, b. North Carolina
## Device Description: 4.
The seven (7) Crosstex® Surgical Masks are constructed of either an inner/outer facing of medical grade tissue or 100% spunbonded polypropylene, a 100% facing of mediour grade tiesas of Tea, with white elastic loops and/or a fogfree strip. monown performs most (7) Crosstex® Surgical Masks is 27 gauge aluminum wire, some have an anti-fog strip, with the Crosstex® Ultra Fluid Resistant No Fog® having a vapor barrier. All of the materials used in the construction of the Crosstex® Surgical Masks are being used in currently marketed devices (see predicate information).
### Intended Use: ട.
The following Crosstex® Surgical Masks are intended for use in infection control practices to minimize contamination caused by inhaled and exhaled microorganisms and reduce the potential exposure of the wearer to blood and body fluids.
- Crosstex® Isolite® Earloop Face Masks Blue, Pink .
- Crosstex® Isofluid® Earloop Face Masks -- Blue, Pink, White, Green .
- Crosstex® Isofluid FogFree® Earloop Face Masks Blue, Peach .
- Crosstex® Isofluid FogFree® Face Masks with Splash Visor Blue, Peach ●
- Crosstex® Procedural Earloop Face Masks Blue, Pink, Yellow ●
- Crosstex® Ultra Fluid Resistant No-Fog® Earloop Face Masks Blue ●
- Crosstex® Ultra Fluid Resistant No-Fog® Face Masks with Splash Visor -. Blue
### Comparison to Predicate Devices: 6.
See Attached
## Discussion of Non-Clinical Tests Performed for Determination of 7. Substantial Equivalence are as follows:
- a. Fluid Resistance:
- Liquid (Water) Resistance Test/Impact Penetration Test
- b. Bacterial Filtration Efficiency (BFE) / Differential Pressure (ΔΡ) Tests
- c. Flammability Testing
- d. Latex Particle Challenge Test
- e. Biocompatibility Testing Per ISO 10993
{2}------------------------------------------------
It was our conclusion that Performance Testing met all relevant requirements of the aforementioned test standards.
## 8. Discussion of Clinical Tests Performed:
Not Applicable
## 9. Conclusions:
The Crosstex® Surgical Masks have the same intended use and technological characteristics as the predicate devices. Moreover, bench testing contained in this submission supplied demonstrates that the technological characteristics do not raise any new questions of safety or effectiveness. The Crosstex® Surgical Masks are substantially equivalent to the predicate device.
{3}------------------------------------------------
Public Health Service
Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three curved lines.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN 2 2 2002
Crosstex, International C/O Ms. Susan D. Goldstein-Falk Official Correspondent MDI Consultants, Incorporated 55 Northern Boulevard, Suite 200 Great Neck, New York 11021
Re: K012602
Trade/Device Name: Crosstex® Surgical Masks Regulation Number: 878.4040 Regulation Name: Surgical Mask Regulatory Class: II Product Code: FXX Dated: November 20, 2001 Received: November 26, 2001
Dear Ms. Falk:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements
{4}------------------------------------------------
Page 2 - Ms. Falk
of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours.
Timothy A. Ulatowski
Timothy A. Ulatowski Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health
## Enclosure
{5}------------------------------------------------
## Attachment 1
of 1 ہے Page
## KO12602 510(k) Number (if known): __
Device Name: Crosstex® Surgical Masks
## Indications For Use:
The following Crosstex® Surgical Masks are intended for use in infection control practices to minimize contamination caused by inhaled and exhaled practices to minimize ochtan the potential exposure of the wearer to blood and body fluids.
- Crosstex® Isolite® Earloop Face Masks Blue, Pink .
- Crosstex® Isofluid® Earloop Face Masks Blue, Pink, White, Green ●
- Crosstex® Isofluid FogFree® Earloop Face Masks Blue, Peach .
- Crosstex® Isofluid FogFree® Face Masks with Splash Visor Blue, Peach .
- Crosstex® Procedural Earloop Face Masks Blue, Pink, Yellow .
- Crosstex® Ultra Fluid Resistant No-Fog® Earloop Face Masks Blue .
- Crosstex® Ultra Fluid Resistant No-Fog® Face Masks with Splash Visor -● Blue
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Dental, Infection Control, and General Hospital Devices
| 510(k) Number | K012602 |
|---------------|---------|
|---------------|---------|
OR
Prescription Use (Per 21 CFR 801.109) Over-The-Counter Use (Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.