CROSSTEX ISOLITE AND ISOFLUID EARLOOP FACE MASKS AND ISOFLUID FOGFREE FACE MASKS WITH AND WITHOUT SPLASH VISOR

K082258 · Crosstex International · FXX · Oct 16, 2008 · General, Plastic Surgery

Device Facts

Record IDK082258
Device NameCROSSTEX ISOLITE AND ISOFLUID EARLOOP FACE MASKS AND ISOFLUID FOGFREE FACE MASKS WITH AND WITHOUT SPLASH VISOR
ApplicantCrosstex International
Product CodeFXX · General, Plastic Surgery
Decision DateOct 16, 2008
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4040
Device ClassClass 2

Intended Use

The following Crosstex® Surgical Masks are intended for use in infection control practices to minimize contamination caused by inhaled and exhaled microorganisms and reduce the potential exposure of the wearer to blood and body fluids. - Crosstex® Isolite® Earloop Face Masks Blue, Pink . - Crosstex® Isofluid® Earloop Face Masks Blue, Pink, White, Green ● - . Crosstex® Isofluid FogFree® Earloop Face Masks - Blue, Peach - . Crosstex® Isofluid FogFree® Face Masks with Splash Visor - Blue, Peach - . Crosstex® Procedural Earloop Face Masks - Blue, Pink, Yellow - Crosstex® Ultra Fluid Resistant No-Fog® Earloop Face Masks Blue - . Crosstex® Ultra Fluid Resistant No-Fog® Face Masks with Splash Visor -Blue

Device Story

Crosstex Surgical Masks are protective face masks designed for infection control. Constructed from spunbonded polypropylene layers and meltblown polypropylene filter media; feature 27-gauge aluminum nose pieces; some models include anti-fog strips or vapor barriers. Used in clinical settings by healthcare personnel to reduce exposure to microorganisms and bodily fluids. Masks are worn over the nose and mouth; secured via earloops. No electronic components or software.

Clinical Evidence

No clinical data. Bench testing only, including fluid resistance, bacterial filtration efficiency, differential pressure, flammability, and biocompatibility testing.

Technological Characteristics

Materials: Spunbonded polypropylene, meltblown polypropylene filter media, 27-gauge aluminum wire nose piece. Features: Elastic earloops, anti-fog strips, vapor barriers (specific models), splash visors (specific models). Biocompatibility: ISO 10993 compliant.

Indications for Use

Indicated for use in infection control practices to minimize contamination from inhaled/exhaled microorganisms and reduce wearer exposure to blood and body fluids. Applicable to general patient population and healthcare personnel.

Regulatory Classification

Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible. (ii) Analysis and nonclinical testing must: (A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and (B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device. (iii) NIOSH approved under its regulation. (2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # OCT 1 6 2008 Image /page/0/Picture/1 description: The image contains a handwritten word, "Korados". The letters are written in a cursive style, with some variations in thickness. The word appears to be written on a white background. ### 510(k) SUMMARY This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92. The assigned 510(k) number is: #### 1. Submitter's Identification: Crosstex International, Inc. 10 Ranick Road Hauppauge, New York 11788 Tel No .: 631-582-6777 Fax No.: 631-582-1726 ### Contact Person: Mr. Gary Steinberg President Crosstex International, Inc. 10 Ranick Road Hauppauge, New York 11788 Tel No .: 631-582-6777 Fax No .: 631-582-1726 Date Summary Prepared: July 1, 2008 #### 2. Name of Device: . - . Crosstex® Isolite® Earloop Face Masks - Blue, Pink - � Crosstex® Isofluid® Earloop Face Masks – Blue, Pink, White, Green - Crosstex® Isofluid FogFree® Earloop Face Masks Blue, Peach ● - . Crosstex® Isofluid FogFree® Face Masks with Splash Visor - Blue, Peach - . Crosstex® Procedural Earloop Face Masks - Blue, Pink, Yellow - Crosstex® Ultra Fluid Resistant No-Fog® Earloop Face Masks Blue - . Crosstex® Ultra Fluid Resistant No-Fog® Face Masks with Splash Visor -Bluc #### 3. Predicate Device Information: - a. American Threshold Industries, Surgical Masks, K# 801036, Asheville, North Carolina. - b. American Threshold Industries, Fluid Resistant Masks, K# 955556, Enka, North Carolina. {1}------------------------------------------------ #### 4. Device Description: The seven (7) Crosstex® Surgical Masks are constructed of either an inner/outer facing of tissue or 100% spunbonded polypropylene, a 100% meltblown polypropylene filter media, with white elastic loops and/or a fogfree strip. The nose piece for all seven (7) Crosstex® Surgical Masks is 27 gauge aluminum wire, some have an anti-fog strip, with the Crosstex® Ultra Fluid Resistant No Fog® having a vapor barrier. All of the materials used in the construction of the Crosstex® Surgical Masks are being used in currently marketed devices (see predicate information). #### 5. Intended Use: The following Crosstex® Surgical Masks are intended for use in infection control practices to minimize contamination caused by inhaled and exhaled microorganisms and reduce the potential exposure of the wearer to blood and body fluids. - Crosstex® Isolite® Earloop Face Masks Blue, Pink . - Crosstex® Isofluid® Earloop Face Masks Blue, Pink, White, Green ● - . Crosstex® Isofluid FogFree® Earloop Face Masks - Blue, Peach - . Crosstex® Isofluid FogFree® Face Masks with Splash Visor -- Blue, Peach - . Crosstex® Procedural Earloop Face Masks - Blue, Pink, Yellow - Crosstex® Ultra Fluid Resistant No-Fog® Earloop Face Masks Blue . - . Crosstex® Ultra Fluid Resistant No-Fog® Face Masks with Splash Visor -Blue #### 6. Comparison to Predicate Devices: See Attached #### 7. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as follows: - a. Fluid Resistance: - Liquid (Water) Resistance Test/Impact Penetration Test - b. Bacterial Filtration Efficiency (BFE) / Differential Pressure (△P) Tests - c. Flammability Testing - d. Latex Particle Challenge Test - Biocompatibility Testing Per ISO 10993 e. It was our conclusion that Performance Testing met all relevant requirements of the aforementioned test standards. {2}------------------------------------------------ #### 8. Discussion of Clinical Test Performed: Not Applicable #### 9. Conclusions: The Crosstex® Surgical Masks have the same intended use and technological characteristics as the predicate devices. Moreover, bench testing contained in this submission supplied demonstrates that the technological characteristics do no raise any new questions of safety or effectiveness. The Crosstex® Surgical Masks are substantially equivalent to the predicate device. {3}------------------------------------------------ Image /page/3/Picture/11 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. The logo is black and white. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 # OCT 1 6 2008 Crosstex International, Incorporated Mr. Richard M. Ormsbee Senior Regulatory Affairs Specialist Minntech Corporation 14605 28th Avenue North Minneapolis, Minnesota 55447-4822 Re: K082258 Trade/Device Name: Crosstex® Surgical Masks Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: II Product Code: FXX Dated: September 17, 2008 Received: September 18, 2008 Dear Mr. Ormsbee: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ ### Page 2 - Mr. Ormsbee Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, CBR Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # Indications for Use 510(k) Number (if known): K082258 Device Name: Crosstex® Surgical Masks Indications for Use: The following Crosstex® Surgical Masks are intended for use in infection control practices to minimize contamination caused by inhaled and exhaled microorganisms and reduce the potential exposure of the wearer to blood and body fluids. - Crosstex® Isolite® Earloop Face Masks Blue, Pink . - Crosstex® Isofluid® Earloop Face Masks Blue, Pink, White, Green . - Crosstex® Isofluid FogFree® Earloop Face Masks Blue, Peach . - Crosstex® Isofluid FogFree® Face Masks with Splash Visor Blue, Peach ● - . Crosstex® Procedural Earloop Face Masks - Blue, Pink, Yellow - Crosstex® Ultra Fluid Resistant No-Fog® Earloop Face Masks Blue . - Crosstex® Ultra Fluid Resistant No-Fog® Face Masks with Splash Visor Blue ● Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) ## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Shule A Murphy, MD (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices 510(k) Number: Page 1 of 1
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...