HM SURGICAL GOWN, NON REINFORCED HM SURGICAL GOWN, FABRIC REINFORCED HM SURGICAL GOWN, POLYERINFORCED

K121998 · Changzhou Holymed Products Co., Ltd. · FYA · Dec 20, 2012 · General, Plastic Surgery

Device Facts

Record IDK121998
Device NameHM SURGICAL GOWN, NON REINFORCED HM SURGICAL GOWN, FABRIC REINFORCED HM SURGICAL GOWN, POLYERINFORCED
ApplicantChangzhou Holymed Products Co., Ltd.
Product CodeFYA · General, Plastic Surgery
Decision DateDec 20, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4040
Device ClassClass 2

Intended Use

HM Surgical gowns are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from the transfer of micro-organisms, body fluids, and particulate matter. HM Surgical Gowns are disposable, single use, sterile or nonsterile surgical gowns manufactured from nonwoven Spunbond-Meltblown-Spunbond (SMS) polypropylene fabrics. Non-sterile HM surgical gowns are to be sold to repackager/relabeler establishments for EtO sterilization according to ISO 11135-1 and Sterile surgical gowns are to be sold directly to users after EtO sterlization validation to ISO 11135-1.

Device Story

HM Surgical Gowns are disposable, single-use protective apparel for operating room personnel. Constructed from nonwoven Spunbond-Meltblown-Spunbond (SMS) polypropylene fabrics; available in non-reinforced, fabric-reinforced, and poly-reinforced configurations. Designed to prevent transfer of micro-organisms, body fluids, and particulate matter between patient and staff. Provided in sterile or non-sterile formats; non-sterile versions intended for further processing by repackagers/relabelers. Sterilization via EtO per ISO 11135-1.

Clinical Evidence

No clinical data; bench testing only.

Technological Characteristics

Materials: Spunbond-Meltblown-Spunbond (SMS) polypropylene. Configurations: Non-reinforced, fabric-reinforced, poly-reinforced. Sterilization: Ethylene Oxide (EtO) per ISO 11135-1. Form factor: Disposable surgical gowns in various sizes (M, L, XL, XXL).

Indications for Use

Indicated for operating room personnel during surgical procedures to provide barrier protection against micro-organisms, body fluids, and particulate matter.

Regulatory Classification

Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible. (ii) Analysis and nonclinical testing must: (A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and (B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device. (iii) NIOSH approved under its regulation. (2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized graphic of an eagle with its wings spread, positioned to the right of the department's name. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the left side of the eagle graphic. The logo is presented in black and white. ## DEPARTMENT OF HEALTH & HUMAN SERVICES #### December 20, 2012 Mr. Alan Yu Manager, Quality Assurance/Regulatory Affairs Changzhou Holymed Products Company, Limited 528 Changwu Road South Wujin District Changzhou, Jiangsu, China 213167 Re: K121998 Trade/Device Name: HM Surgical Gowns Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: II Product Code: FYA Dated: November 27, 2012 Received: November 29, 2012 Dear Mr. Yu: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely vours. Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # INDICATIONS FOR USE K121998 510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________ Device Name: HM Surgical Gowns ### Indications for Use: HM Surgical gowns are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from the transfer of micro-organisms, body fluids, and particulate matter. HM Surgical Gowns are disposable, single use, sterile or nonsterile surgical gowns manufactured from nonwoven Spunbond-Meltblown-Spunbond (SMS) polypropylene fabrics. Non-sterile HM surgical gowns are to be sold to repackager/relabeler establishments for EtO sterilization according to ISO 11135-1 and Sterile surgical gowns are to be sold directly to users after EtO sterlization validation to ISO 11135-1. Gowns are provided as below: ### Gown Family A: 960100/860100 | PROD. CODE | STERILE/NONSTERILE | DESC. | SIZE | COLOR | |------------|--------------------|-------------------------------|------|-------| | 960102 | STERILE | Surgical Gown, Non Reinforced | M | BLUE | | 960103 | STERILE | Surgical Gown, Non Reinforced | L | BLUE | | 960104 | STERILE | Surgical Gown, Non Reinforced | XL | BLUE | | 960105 | STERILE | Surgical Gown, Non Reinforced | XXL | BLUE | | 860102 | NONSTERILE | Surgical Gown, Non Reinforced | M | BLUE | | 860103 | NONSTERILE | Surgical Gown, Non Reinforced | L | BLUE | | 860104 | NONSTERILE | Surgical Gown, Non Reinforced | XL | BLUE | | 860105 | NONSTERILE | Surgical Gown, Non Reinforced | XXL | BLUE | Page 1 of 2 {3}------------------------------------------------ #### Gown Family B: 960200/860200 | PROD. CODE | STERILE/NONSTERILE | DESC. | SIZE | COLOR | |------------|--------------------|----------------------------------|------|-------| | 960202 | STERILE | Surgical Gown, Fabric Reinforced | M | BLUE | | 960203 | STERILE | Surgical Gown, Fabric Reinforced | L | BLUE | | 960204 | STERILE | Surgical Gown, Fabric Reinforced | XL | BLUE | | 960205 | STERILE | Surgical Gown, Fabric Reinforced | XXL | BLUE | | 860202 | NONSTERILE | Surgical Gown, Fabric Reinforced | M | BLUE | | 860203 | NONSTERILE | Surgical Gown, Fabric Reinforced | L | BLUE | | 860204 | NONSTERILE | Surgical Gown, Fabric Reinforced | XL | BLUE | | 860205 | NONSTERILE | Surgical Gown, Fabric Reinforced | XXL | BLUE | ### Gown Family C: 960300/86000 | PROD. CODE | STERILE/NONSTERILE | DESC. | SIZE | COLOR | |------------|--------------------|--------------------------------|------------|-------| | 960302 | STERILE | Surgical Gown, Poly Reinforced | M | BLUE | | 960303 | STERILE | Surgical Gown, Poly Reinforced | L | BLUE | | 960304 | STERILE | Surgical Gown, Poly Reinforced | XL | BLUE | | 960306 | STERILE | Surgical Gown, Poly Reinforced | XL, X-Long | BLUE | | 860302 | NONSTERILE | Surgical Gown, Poly Reinforced | M | BLUE | | 860303 | NONSTERILE | Surgical Gown, Poly Reinforced | L | BLUE | | 860304 | NONSTERILE | Surgical Gown, Poly Reinforced | XL | BLUE | | 860306 | NONSTERILE | Surgical Gown, Poly Reinforced | XL, X-Long | BLUE | Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use X (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Page 2 of 2 Elizabeth F. Claverie 2012.12.19 21:51:29 -05'00' (Division Sign-Off) Division of Anesthesiology, General Hospital Infaction Control, Dental Devices 510(k) Number;
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...