JIE+HONG MEDICAL SURGICAL GOWNS - STERILE, JIE+HONG MEDICAL SURGICAL GOWNS - NON STERILE
K111067 · Kunshan Jiehong Nonwoven Product Co., Ltd. · FYA · Aug 2, 2011 · General, Plastic Surgery
Device Facts
Record ID
K111067
Device Name
JIE+HONG MEDICAL SURGICAL GOWNS - STERILE, JIE+HONG MEDICAL SURGICAL GOWNS - NON STERILE
Applicant
Kunshan Jiehong Nonwoven Product Co., Ltd.
Product Code
FYA · General, Plastic Surgery
Decision Date
Aug 2, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4040
Device Class
Class 2
Intended Use
Disposable gowns are worn by operating room personnel during surgical procedures to protect both the surgical patient and OR personnel from the transfer of body fluids and particulate material. Gowns are provided as sterile and non-sterile. Non-sterile surgical gowns are to be sold to OEMs, which require EtO sterilization according to the ISO 11135 standard. Sterile surgical gowns are to be sold directly to users and must be sterilized by an EtO cycle validated to ISO 11135 standard.
Device Story
Jiehong Medical Surgical Gown is a disposable protective garment for operating room personnel. Constructed from SMS (spunbond-meltblown-spunbond) material; designed to prevent transfer of body fluids and particulate matter between patient and staff. Available in reinforced styles; provided in sterile or non-sterile formats. Non-sterile versions intended for OEM distribution for subsequent EtO sterilization; sterile versions sold directly to end-users. Used in surgical environments to maintain sterile fields and protect personnel.
Clinical Evidence
No clinical data. Bench testing only.
Technological Characteristics
Material: SMS (spunbond-meltblown-spunbond) nonwoven fabric. Form factor: Reinforced surgical gown. Sterilization: Ethylene Oxide (EtO) per ISO 11135.
Indications for Use
Indicated for operating room personnel during surgical procedures to provide a barrier against body fluids and particulate material transfer. No specific age or gender restrictions.
Regulatory Classification
Identification
Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.
Special Controls
*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.
K061308 — TRIMAX, MODEL 9515N, BREATHABLE SURGICAL GOWN, MODELS 3247AN AND 9015N AND OPTIMA OR SURGICAL GOWN, MODEL 9506N · Cardinal Health 200, Inc. · Jun 5, 2006
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" around the perimeter. Inside the circle is a stylized symbol that resembles three abstract human figures or waves, stacked on top of each other.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 1 0903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring. MD 20993-0002
Kunshan Jichong Nonwoven Product Company, Limited C/O Mr. Glen Feye President Accurate Consultants Incorporated 1340 West Pennsylvania Avenue San Diego, California 92103
AUG - 2 2011
Re: K111067
Trade/Device Name: Jiehong Medical Surgical Gown Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: II Product Code: FYA Dated: June 28, 2011 Received: June 29, 2011
Dear Mr. Feve:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Mr. Feye
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/C DRHOffice s/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Susan Runner
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known):
Device Name: Jiehong Medical Surgical Gown
Indications for Use:
Disposable gowns are worn by operating room personnel during surgical procedures to protect both the surgical patient and OR personnel from the transfer of body fluids and particulate material.
Gowns are provided as sterile and non-sterile.
Non-sterile surgical gowns are to be sold to OEMs, which require EtO sterilization according to the ISO 11135 standard. Sterile surgical gowns are to be sold directly to users and must be sterilized by an EtO cycle validated to ISO 11135 standard.
The following disposable surgical gown types and model numbers are included in this submission
| MATERIAL | GOWN STYLE | STERILE/NON | SIZES | Model number |
|----------|------------|-------------|-------------------------|--------------|
| SMS | Reinforced | Sterile | Medium (M) | JHSTG 65022 |
| | | | Large (L) | JHSTG 65023 |
| | | | Extra Large (XL) | JHSTG 65024 |
| | | | Extra-Extra Large (XXL) | JHSTG 65025 |
| | Reinforced | Non Sterile | Medium (M) | JHSTG 35022 |
| | | | Large (L) | JHSTG 35023 |
| | | | Extra Large (XL) | JHSTG 35024 |
| | | | Extra-Extra Large (XXL) | JHSTG 35025 |
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use X (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Elaine S. Marshall for EFC
Division Sign Off
Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K111067
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.