AN1036 Dosimeter

K200334 · Andersen Sterilizers, Inc. · JOJ · Nov 12, 2020 · General Hospital

Device Facts

Record IDK200334
Device NameAN1036 Dosimeter
ApplicantAndersen Sterilizers, Inc.
Product CodeJOJ · General Hospital
Decision DateNov 12, 2020
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.2800
Device ClassClass 2

Indications for Use

AN1036 Dosimeter is a single-use color change chemical indicator. It is calibrated for sterilization temperature and used to verify adequate cumulative ethylene oxide exposure in a 3-hour or 6-hour gas exposure at 50°C in the EOGas 4 Ethylene Oxide Gas Sterilizer manufactured by Andersen Sterilizers, Inc.

Device Story

AN1036 Dosimeter is a single-use chemical indicator accessory for EOGas 4 Ethylene Oxide Gas Sterilizer. Device consists of pH indicator within glass capillary tube mounted on plastic tray. Upon exposure to ethylene oxide, indicator changes color from yellow-orange to dark blue; color progression is proportional to cumulative ethylene oxide exposure. Device features calibration marks corresponding to 3-hour and 6-hour sterilization cycles. Healthcare providers place dosimeter in sterilizer; color change provides visual verification of adequate exposure to inactivate 6-Log Bacillus atrophaeus biological indicator. Device is not a replacement for biological indicators. It provides immediate visual feedback to confirm sterilization cycle parameters (temperature, time, gas concentration) were met, assisting in quality assurance of sterilized loads.

Clinical Evidence

Bench testing only. Performance characterized using Andersen Chemical Indicator Evaluator Resistometer and EOGas 4 Sterilizer. Testing validated response to critical parameters (temperature, time, EO concentration) per ANSI/AAMI/ISO 11140-1 and FDA guidance. Results confirmed correlation between color change and cumulative EO exposure, stability of endpoint color, and 3-year shelf life. Biocompatibility testing confirmed no hazardous ingredients and compliance with ISO 10993-7 for residuals.

Technological Characteristics

Single-use chemical indicator; pH-sensitive indicator ink in glass capillary tube on plastic tray. Responds to ethylene oxide concentration, time, and temperature (50°C ± 3°C). Calibrated for 3-hour and 6-hour cycles in EOGas 4 system. Shelf life: 3 years. Non-patient contacting.

Indications for Use

Indicated for use as a single-use chemical indicator to verify adequate cumulative ethylene oxide exposure during 3-hour or 6-hour sterilization cycles at 50°C in the EOGas 4 Ethylene Oxide Gas Sterilizer.

Regulatory Classification

Identification

Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. November 12, 2020 Andersen Sterilizers, Inc. William Andersen President 3154 Caroline Drive Haw River, North Carolina 27258 Re: K200334 Trade/Device Name: AN1036 Dosimeter Regulation Number: 21 CFR 880.2800 Regulation Name: Sterilization Process Indicator Regulatory Class: Class II Product Code: JOJ Dated: October 26, 2020 Received: October 29, 2020 #### Dear William Andersen: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part {1}------------------------------------------------ 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, CAPT Elizabeth Claverie, MS Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below. 510(k) Number (if known) K200334 Device Name AN1036 Dosimeter #### Indications for Use (Describe) AN 1036 Dosimeter is a single-use color change chemical indicator. It is calibrated for sterilization temperature and used to verify adequate cumulative ethylene oxide exposure in a 3-hour gas exposure at 50°C in the EOGas 4 Ethylene Oxide Gas Sterilizer manufactured by Andersen Sterilizers, Inc. Critical process parameters for the cycles are summarized in Table 1. Table 1. Critical sterilization cycle parameters in the EOGas 4 Ethylene Oxide Gas Sterilizer | Ethylene Oxide | Temperature | Relative Humidity | EO Exposure Time | Total Cycle Time | |----------------|-------------|-------------------|------------------|------------------| | 17.6 g ± 5% | 50°C ± 3°C | 35-70% | 3 hours | 3.5 hours | | | | | 6 hours | 7 hours | Prescription Use (Part 21 CFR 801 Subpart D) > Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995, #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary K200334 # Applicant's Name and Address Andersen Sterilizers, Inc. Establishment Registration Number 3004634710 3154 Caroline Drive Haw River, NC 27258 Contact Person William K. Andersen, BE, MD, FAAOS President Phone: 336-376-8622; Fax: 336-376-5428 ## Date of Preparation October 26, 2020 ## Device Proprietary Name AN1036 Dosimeter Common Name Indicator, Physical/Chemical Sterilization Process Classification Class II (21 CFR 880.2800) Chemical Indicator Product Code 101 The refill kits for the EOGas 4 Ethylene Oxide Gas Sterilizer, including the accessories (sterilization bags, EOGas 4 cartridges, Dosimeters, and Humidichips), are registered with the US Environmental Protection Agency (EPA #69340-7). ## Predicate Device | Device Name | AN1087 Dosimeter | |---------------|----------------------------| | 510(k) number | K150645 | | Manufacturer | Andersen Sterilizers, Inc. | The predicate AN1087 Dosimeters were cleared for use with 3-hour gas exposures in the EOGas 4 Ethylene Oxide Gas Sterilization system manufactured by Andersen Sterilizers, Inc. The sterilization time claim of the predicate device was modified in order to indicate the AN1087 Dosimeters for use in 6-hour gas exposures in the EOGas 4 Ethylene Oxide Gas Sterilization System. No modifications were made to the manufacturing method, technology, or intended use. {4}------------------------------------------------ # Device Description The AN1036 Dosimeter is a single-use chemical indicator for cumulative ethylene oxide exposure. It is an accessory for the EOGas 4 Ethylene Oxide Gas Sterilizer. The AN1036 Dosimeter contains a proprietary pH indicator in a glass capillary tube that is sealed on one end and mounted on a plastic tray. It is calibrated for a 50℃ sterilization temperature and responds to ethylene oxide concentration and sterilization time. With exposure to ethylene oxide, the indicator turns from yellow-orange to a dark blue color from the open end toward the closed end. The extent of the color change is proportional to cumulative ethylene oxide exposure. The cycle-specific calibration marks represent adequate cumulative ethylene oxide exposure to inactivate a 6-Log Bacillus atrophaeus biological indicator at the location of the AN1036 Dosimeter after 3-hour gas exposures or at the center of the endoscope lumens after 6-hour gas exposures. The AN1036 Dosimeter is not a replacement for a biological indicator. # Indications for Use AN1036 Dosimeter is a single-use color change chemical indicator. It is calibrated for sterilization temperature and used to verify adequate cumulative ethylene oxide exposure in a 3-hour or 6-hour gas exposure at 50°C in the EOGas 4 Ethylene Oxide Gas Sterilizer manufactured by Andersen Sterilizers, Inc. Critical process parameters for the cycles are summarized in Table 5-1. | Ethylene Oxide | Temperature | Relative<br>Humidity | Ethylene Oxide<br>Exposure Time | Total Cycle Time | |----------------|-------------|----------------------|---------------------------------|------------------| | 17.6 g ± 5% | 50°C ± 3°C | 35-90% | 3 hours | 3.5 hours | | | | | 6 hours | 7 hours | Table 5-1. Critical sterilization cycle parameters in the EOGas 4 Ethylene Oxide Gas Sterilizer # Technological Characteristics Comparison The subject AN1036 Dosimeter is compared to the predicate AN1087 Dosimeter (K150645): there is no difference in intended use, design principles, technical characteristics, or performance between the devices. The difference between the predicate AN1087 and subject AN1036 Dosimeters is the addition of a calibration mark suitable for a 6-hour gas exposure. A comparison between the indicators is listed in Table 5-2. {5}------------------------------------------------ | | Predicate AN1087 Dosimeter<br>(K150645) | Subject AN1036 Dosimeter<br>(K200334) | Comparison | |----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------| | Intended Use | Chemical indicator for<br>cumulative EO exposure in the<br>EOGas 4 Ethylene Oxide Gas<br>Sterilizer | Chemical indicator for<br>cumulative EO exposure in the<br>EOGas 4 Ethylene Oxide Gas<br>Sterilizer | Identical | | Design | A pH indicator changes color<br>when exposed to EO;<br>The extent of the color changes<br>is proportional to cumulative<br>EO exposure;<br>Calibrated for use at 50±3°C;<br>One calibration mark for use in<br>3-hour gas exposures | A pH indicator changes color<br>when exposed to EO;<br>The extent of the color<br>changes is proportional to<br>cumulative EO exposure;<br>Calibrated for use at 50±3°C;<br>Two calibration marks for use<br>in both 3-hour and 6-hour gas<br>exposures | Identical | | Technology | Chemical reactions with EO<br>change the pH, and therefore<br>the color, of the indicator ink | Chemical reactions with EO<br>change the pH, and therefore<br>the color, of the indicator ink | Identical | | Performance | Indicates cumulative EO<br>exposure | Indicates cumulative EO<br>exposure | Identical | | Shelf Life | 3 years | 3 years | Identical | | Endpoint<br>Specifications | Endpoint blue color is stable<br>for 28 days at 20-25°C and 3<br>days at 50°C. | Endpoint blue color is stable<br>for 28 days at 20-25°C and 4<br>days at 50°C. | Similar | Table 5-2. Comparison between the subject AN1036 Dosimeter and the predicate AN1087 Dosimeter (K150645) # Summary of Non-clinical Testing: The AN1036 Dosimeters were validated using applicable tests in: - 1) FDA 2003 guideline, "Premarket Notification [510(k)] Submissions for Chemical Indicators: Guidance for Industry and FDA Staff"; and - 2) ANSI AAMI ISO 11140-1, "Sterilization of health care products Chemical Indicators -Part 1: General requirements" (FDA Recognition Number 14-460). The performance of the AN1036 Dosimeter was characterized in an Andersen Chemical Indicator Evaluator Resistometer as well as in the EOGas 4 Ethylene Oxide Gas Sterilizer using 3- and 6-hour gas exposures at 50°C. The critical parameters to which the AN1036 responds include temperature, time, and ethylene oxide gas concentration in a relative humidity-controlled environment. The AN1036 indicates adequate cumulative EO exposure in the cycle for all validated loads in the EOGas 4 Ethylene Oxide Gas Sterilizer. Using the AN1036 Dosimeter at various stages of shelf life, the distance the blue line travels is stable for a minimum of 28 days after the sterilization cycle when AN1036 Dosimeters are stored at room temperature (20-25°C) and 4 days when stored at 50°C. Real-time shelf life testing with the AN1036 Dosimeter supports a shelf life of 3 years. {6}------------------------------------------------ Performance testing is summarized in Table 5-3. | Test | Purpose | Acceptance Criteria | Results | |-----------------------------|--------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Chemical<br>Indicator | To validate AN1036<br>Dosimeter as a<br>chemical indicator per<br>ANSI/AAMI/ISO<br>11140-1 and FDA<br>Chemical Indicator<br>guidance | Critical parameters<br>with pass/fail criteria;<br>Able to correctly<br>indicate pass/fail<br>results from the EOGas<br>4 Sterilizer | AN1036 Dosimeters respond to<br>EO concentration, sterilization<br>time, and temperature, at a<br>relative humidity of 35-90%.<br>AN1036 Dosimeters correctly<br>indicate pass/fail in various<br>endoscope loads.<br>A correlation exists between the<br>extent the blue color travels and<br>cumulative EO exposure. | | Biocompatibility | To demonstrate the<br>AN1036 Dosimeters<br>are safe to use | Provides reasonable<br>assurance for safety | AN1036 Dosimeter is not a direct<br>or indirect patient-contacting<br>device. | | | | | None of the major ingredients is<br>hazardous. | | | | | No chemicals leach out of the<br>AN1036 Dosimeters during the<br>sterilization cycle. | | | | | Without any additional aeration,<br>the residual EO detected in<br>Dosimeters met ISO 10993-7<br>requirements. | | Endpoint Color<br>Stability | To evaluate the<br>stability of the<br>endpoint color | Stability demonstrates<br>reasonable assurance<br>for effectiveness | Endpoint blue color is stable for<br>28 days at 20-25°C or 4 days at<br>50°C after 6-hour gas exposures. | | Shelf Life | Evaluate the ability to<br>correctly indicate<br>pass/fail in cycles<br>through the shelf life | Maintains performance<br>specifications<br>throughout the stated<br>shelf life | Data supports claimed shelf life<br>of 3 years | Table 5-3. Summary of bench tests performed to demonstrate safety and effectiveness of the AN1036 Dosimeter # Conclusions: The conclusions drawn from the nonclinical tests demonstrate that the AN1036 Dosimeter device is as safe, as effective, and performs as well as or better than the legally marketed AN1087 Dosimeter (K150645).
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...